K093080 · Treymed, Inc. · BZL · Jan 22, 2010 · Anesthesiology
Device Facts
Record ID
K093080
Device Name
METAPHOR METABOLIC MONITOR
Applicant
Treymed, Inc.
Product Code
BZL · Anesthesiology
Decision Date
Jan 22, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1730
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Metaphor uses the direct measurement of Oxygen uptake (VO2) and Carbon Dioxide production (VCO2) to objectively and noninvasively assess breath by breath metabolic function during rest or exercise. The Metaphor can assess energy expenditure to support nutritional assessment. Optional measurement of Nitrous Oxide (N2O) is available for enhanced accuracy of the CO2 measurement in the presence of N2O. The Metaphor is a comprehensive pulmonary mechanics monitor which can display graphic and numeric data in patients who are mechanically ventilated or breathing spontaneously. The Metaphor pulse oximetry function provides continuous SpO2 and pulse rate measurements in neonatal through adult populations with appropriate probes. The Metaphor provides adjustable visible and audible alarms. The device is not intended for use as a transport monitor. Caution: Federal law (U.S.A.) restricts this device to sale by or on the order of a physician or other licensed practitioner.
Device Story
Metaphor Metabolic Monitor integrates sidestream respiratory gas monitoring (CO2, O2, N2O) and pulmonary mechanics monitoring (pressure, flow, volume). System consists of a PC-based host application and BXB DAS (breath-by-breath data acquisition subsystem). Inputs include gas samples via sidestream sampling, pressure/flow via EZ-Flow sensors, and SpO2 via integrated third-party pulse oximetry module. Device computes VO2, VCO2, and metabolic parameters breath-by-breath. Used in clinical settings for metabolic and pulmonary assessment; operated by clinicians. Output displayed as graphic/numeric data on touch-screen terminal to assist in nutritional assessment and respiratory management. Provides adjustable alarms. Not for transport use.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including electrical safety, electromagnetic compatibility, environmental testing, and biocompatibility. System-level performance testing verified respiratory gas monitoring, pressure/flow measurements, and metabolic measurement accuracy. Pulse oximeter integration was validated via simulation testing. Software verification and validation were performed, and risk/hazard analysis confirmed acceptable residual risk.
Technological Characteristics
PC-based monitor with BXB DAS subsystem. Gas sensing: infrared (CO2, N2O), electrochemical (O2). Flow/pressure sensing: differential pressure via EZ-Flow sensors. Pulse oximetry: integrated third-party module. Connectivity: PC-based control program. Biocompatible patient-contact materials. Electrical safety/EMC tested per standards.
Indications for Use
Indicated for neonatal through adult patients requiring metabolic assessment (rest or exercise), nutritional assessment, pulmonary mechanics monitoring (mechanically ventilated or spontaneously breathing), or continuous SpO2 and pulse rate monitoring.
Regulatory Classification
Identification
An oxygen uptake computer is a device intended to compute the amount of oxygen consumed by a patient and may include components for determining expired gas volume and composition.
Predicate Devices
Datex-Ohmeda S/5 Compact Airway Module (Model Family E-CAiOVX) (K051092)
GE HealthCare MedGraphics CPX/D with Breeze EX software (K955432)
Meteor Hand Held Respiratory Mechanics Monitor (K982487)
Shape-HF Cardiopulmonary Exercise Testing System (K090722)
Submission Summary (Full Text)
{1}------------------------------------------------
JAN 22 2010
K093080
Page 1 of 4
510(k) Summary of Safety and Effectiveness in accordance with 21 CFR 807.92
| (a) (1) | Submitted by: | TreyMed, Inc. |
|---------|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | N56 W24790 N Corporate Circle<br>Sussex, WI 53089<br>262-820-1294 phone<br>262-820-1295 fax<br>www.treymed.com |
| | | E-mail: rhr@treymed.com |
| | Contact Person: | Robert H. Ricciardelli |
| | Position/Title: | President |
| | Date of Preparation: | September 25, 2009 |
| | Trade Name: | Metaphor Metabolic Monitor |
| | Common/Classification Name: | Computer, Oxygen Uptake; Pulse<br>Oximeter; Analyzer, Gas, Carbon-<br>Dioxide, Gaseous-phase; Analyzer, Gas,<br>Oxygen, Gaseous-phase; Analyzer, Gas,<br>Nitrous-Oxide, Gaseous-phase;<br>Spirometer, Monitoring; |
| | Product Code(s): | The following Class II classifications<br>appear to apply:<br>BZL 21 CFR § 868.1730<br>DQA 21 CFR § 870.2700<br>CCK 21 CFR § 868.1400<br>CCL 21 CFR § 868.1720<br>CBR 21 CFR § 868.1700<br>BZK 21 CFR § 868.1850 |
| | Class: | Class II |
| (3) | Predicate Device(s): | K051092 Datex-Ohmeda S/5 Compact Airway Module (Model Family |
- E-CAiOVX), GE HealthCare MedGraphics CPX/D with Breeze EX software; Medical K955432 Graphics Corporation
{2}------------------------------------------------
- K982487 Meteor Hand Held Respiratory Mechanics Monitor, TreyMed, Inc. (formerly CardioPulmonary Technologies, Inc.) BCI WW1020 SPECTRO2 Pulse Oximeter, Smiths Medical PM, K083705 Inc. (formerly BCI) K090722 Shape-HF Cardiopulmonary Exercise Testing System, Shape Medical Systems, Inc.
Reason for Submission: New Device
### (4) Description of Device:
The Metaphor Metabolic Monitor integrates the functions of a sidestream respiratory gas monitor (CO2, O2, N2O), and pulmonary mechanics monitor (pressure, flow, and volume). These features allow the Metaphor to measure Oxygen uptake (VO2) and Carbon Dioxide production (VCO2) to objectively and noninvasively assess breath by breath metabolic function during rest or exercise.
The system is comprised of two main elements:
- Metaphor PC software host application running on a dedicated touch . screen computer terminal, which presents computed physiological monitoring parameters, including metabolic measurements, as graphic and numeric data. This application manages the user interface, including alarms, and it interfaces to the BXB_DAS.
- The BXB DAS embedded measurement platform this subsystem . contains the sensor technology to perform the gas, pressure, and flow measurements, and integrates a 510(k) cleared third party pulse oximeter module. The BXB DAS subsystem interfaces with single patient use EZ-Flow sensors and SpO2 sensors.
### (2) Intended use:
Respiratory gas measurements are useful in critical care monitoring scenarios. Pulmonary mechanics measurements are useful in ventilated patients. Metabolic measurements are useful to determine the rate of energy expenditure in a variety of clinical scenarios, as described below.
# Indications for Use:
The Metaphor uses the direct measurement of Oxygen uptake (VO)) and Carbon Dioxide production (VCO2) to objectively and noninvasively assess breath by breath metabolic function during rest or exercise. The Metaphor can assess energy expenditure to support nutritional assessment. Optional
{3}------------------------------------------------
measurement of Nitrous Oxide (N>O) is available for enhanced accuracy of the CO2 measurement in the presence of N2O.
The Metaphor is a comprehensive pulmonary mechanics monitor which can display graphic and numeric data in patients who are mechanically ventilated or breathing spontaneously.
The Metaphor pulse oximetry function provides continuous SpO2 and pulse rate measurements in neonatal through adult populations with appropriate probes.
The Metaphor provides adjustable visible and audible alarms. The device is not intended for use as a transport monitor.
Prescription Device.
### (6) Technological Characteristics:
The Metaphor utilizes similar measurement methods as the referenced predicate devices:
- Flow detection based on differential pressure measurement .
- o Sidestream qas sampling
- � CO2 and N₂O gas monitoring via infrared optical measurement
- O> monitoring using electrochemical cell 0
- Integration of a listed subsystem for pulse oximetry .
- WIN PC-based control program .
As described above, the Metaphor implements a PC based monitor application with the TreyMed BXB DAS (breath by breath data acquisition subsystem) to perform and display the measurements.
Metaphor Metabolic Monitor has the same The technological characteristics as the referenced combination of 510(k) listed devices:
- The Metaphor integrates the same flow measurement technology and o EZ-Flow sensor technology as two predicate devices (Meteor, Shape-HF), and the same measurement principle (differential pressure). The Metaphor uses the EZ-Flow sensor.
- The Metaphor. Shape-HF and BCI WW1020 devices use identical . oximetry technology, the BCI technology.
- Metaphor reports metabolic parameters similar to CPX/D and Shape-. HF, including VO2 and RQ, employing similar technical means.
- Metaphor provides respiratory gas monitoring parameters (CO2, O2, . optional N2O) similar to the Datex CAiOVX using similar means.
{4}------------------------------------------------
#### (b) (1) Non-Clinical Tests Submitted:
The Metaphor Metabolic Monitor was tested in accordance with applicable standards for medical device electrical safety, electromagnetic compatibility, shock and vibration, and environment (temperature and humidity). The device passed the tests.
Materials utilized in patient contact surfaces were reviewed for conformance with biocompatibility requirements. The materials met the requirements.
System level tests of the Metaphor Metabolic Monitor included testing in support of requirements for respiratory gas monitors, pressure/flow measurement testing, and extended range bench testing of metabolic measurements, including, where possible, comparison testing. Pulse oximeter simulation tests were performed to demonstrate successful integration. Alarms testing was performed. The device satisfied performance criteria.
The Metaphor Metabolic Monitor PC software and BXB_DAS subsystem embedded software were verified to requirements and validated to meet intended use. Risk, hazard and failure mode analysis was performed and residual risks were determined to be acceptable.
### (2) Clinical Tests Submitted:
(none)
### Conclusions from Tests: (3)
As described in (b)(1) and (b)(2) above, the testing demonstrates that the Metaphor Metabolic Monitor is as safe and effective as, and functions in a manner equivalent to the predicate devices.
ર-4
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -W066-G609 Silver Spring, MD 20993-0002
Mr. Robert H. Ricciardelli President TreyMed, Incorporated N56 W24790 North Corporate Circle Sussex, Wisconsin 53089
JAN 22 2010
Re: K093080
Trade/Device Name: Metaphor Metabolic Monitor Regulation Number: 21 CFR 868.1730 Regulation Name: Oxygen Uptake Computer Regulatory Class: II Product Code: BZL, DQA, CCK, CBR, CCL, BZC Dated: January 18, 2010 Received: January 19, 2010
Dear Mr. Ricciardelli:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
### Page 2 - Mr. Ricciardelli
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR. Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.goy/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
### Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: TreyMed, Inc. Metaphor Metabolic monitor
### Indications for use:
The Metaphor uses the direct measurement of Oxygen uptake (VO2) and Carbon Dioxide production (VCO2) to objectively and noninvasively assess breath by breath metabolic function during rest or exercise. The Metaphor can assess energy expenditure to support nutritional assessment. Optional measurement of Nitrous Oxide (N₂O) is available for enhanced accuracy of the CO2 measurement in the presence of N2O.
The Metaphor is a comprehensive pulmonary mechanics monitor which can display graphic and numeric data in patients who are mechanically ventilated or breathing spontaneously.
The Metaphor pulse oximetry function provides continuous SpO2 and pulse rate measurements in neonatal through adult populations with appropriate probes.
The Metaphor provides adjustable visible and audible alarms. The device is not intended for use as a transport monitor.
Caution: Federal law (U.S.A.) restricts this device to sale by or on the order of a physician or other licensed practitioner.
Prescription Use × (Part 21 CFR 801 Subpart D) AND / OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Schulthe
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of 1
510(k) Number: K093080
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.