TUMARK PROFESSIONAL, MODEL 271560, MR-TUMARK PROFESSIONAL, MODEL 601560

K093064 · Somatex Medical Technologies GmbH · NEU · Feb 17, 2010 · General, Plastic Surgery

Device Facts

Record IDK093064
Device NameTUMARK PROFESSIONAL, MODEL 271560, MR-TUMARK PROFESSIONAL, MODEL 601560
ApplicantSomatex Medical Technologies GmbH
Product CodeNEU · General, Plastic Surgery
Decision DateFeb 17, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4300
Device ClassClass 2

Indications for Use

The Tumark® Professional and MRI Tumark® Professional are intended to attach a marker to soft tissue at the surgical site during an open or a percutaneous procedure. The devices are indicated for use to radiographically and radiologically mark the surgical location in breasts following an open or percutaneous procedure. Only the MRI Tumark® Professional is indicated to be used in Magnetic Resonance Imaging (MRI) procedures.

Device Story

Tissue site marking system; consists of non-absorbable nickel-titanium marker, introducer cannula, and handheld plastic applier. Cannula features beveled tip, 1 cm depth marks, and ultrasound enhancement. Applier includes slide-button for one-handed placement and safety catch to prevent premature deployment. Used by clinicians during open or percutaneous breast procedures to mark surgical sites for subsequent identification. MRI Tumark® Professional version is MRI safe; Tumark® Professional version is not. Marker remains in soft tissue to provide radiographic/radiological reference for follow-up imaging.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including biocompatibility, sterility, and packaging validation.

Technological Characteristics

Non-absorbable nickel-titanium marker. Introducer cannula: stainless steel (Tumark® Professional) or cobalt-chrome alloy (MRI Tumark® Professional). Features: beveled tip, 1 cm depth marks, ultrasound enhancement. Handheld plastic applier with mechanical slide-button deployment and safety catch. Sterile, single-use.

Indications for Use

Indicated for radiographic and radiological marking of surgical locations in breasts following open or percutaneous procedures. MRI Tumark® Professional is indicated for use in MRI procedures.

Regulatory Classification

Identification

An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for SOMATEX MEDICAL TECHNOLOGIES GMBH. Above the logo is the text "K093064 Page 1/4" in handwritten font. The SOMATEX logo is in a bold, sans-serif font, with the registered trademark symbol next to the "X". — MADE IN GERMANY Rheinstraße 7d | D-14513 Teltow Tel +49 (0) 3328 30 76-0 | Fax +49 (0) 3328 30 76-99 info@somatex.com | www.somatex.com FEB 1 7 2010 ## 510(k) SUMMARY # Tumark® Professional / MRI Tumark® Professional | Submitter: | Somatex Medical Technologies' GmbH<br>Rheinstrasse 7d<br>14513 Teltow<br>Germany<br>Phone: +49 3328 3076 13<br>Fax: +49 3328 3076 99 | |----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Official<br>Correspondent: | Susanne Raab<br>Regulatory Affairs Consultant<br>1480 Cambridge Street<br>Cambridge, MA 02139<br>Phone: 617 547 0628<br>Fax: 617 520 2136<br>e-mail: sbraab@comcast.net | | Trade Name: | Tumark® Professional / MRI Tumark® Professional | | Common Name: | Tissue Site Marking System | | Classification Name: | Radiographic Implantable Marker, 21 C.F.R. 878.4300 | | Regulatory Class: | II | | Product Code: | NEU | | Device Description: | The Tumark® Professional and MRI Tumark® Professional are<br>sterile, single use, preloaded tissue site marking systems<br>consisting of a non-absorbable nickel-titanium marker, an<br>introducer cannula and a plastic handheld applier with<br>deployment mechanism. The Tumark® Professional is not | Page 1 of 4 {2}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the logo for SOMATEX MEDICAL TECHNOLOGIES GMBH. Above the logo is the text "K093064 Page 2/4". The logo is in black and white and is a simple design. MADE IN GERMANY Rheinstraße 7d | D-14513 Teltow Tel +49 (0) 3328 30 76-0 | Fax +49 (0) 3328 30 76-99 info@somatex.com | www.somatex.com indicated to be used in Magnetic Resonance Imaging (MRI) procedures. The MRI Tumark® Professional is MRI Safe and indicated to be used during MRI procedures. The introducer cannula which will be available in stainless steel (Tumark® Professional) or cobalt-chrome alloy (MRI Tumark® Professional) for use with magnetic resonance imaging techniques is designed with 1 cm depth marks, a beveled tip and an ultrasound enhancement on the distal end. The cannulas will be available in different lengths. The handle is equipped with a slide-button which allows for a one handed placement. A safety catch system prevents the slide-button from inadvertently moving forward and therefore prevents a premature deployment of the marker. The Tumark® Professional will be available with a U-shaped or X-shaped marker. The MRI Tumark® Professional will only be available with a U-shaped marker. The Tumark® Professional and MRI Tumark® Professional Intended Use: are intended to attach a marker to soft tissue at the surgical site during an open or a percutaneous procedure. > The devices are indicated for use to radiographically and radiologically mark the surgical location in breasts following an open or percutaneous procedure. Only the MRI Tumark® Professional is indicated to be used in Magnetic Resonance Imaging (MRI) procedures. {3}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for SOMATEX MEDICAL TECHNOLOGIES GMBH. Above the logo is the text "K093064 Page 3/4". The logo is in black and white and is a simple design. MADE IN GERMANY Rheinstraße 7d | D-14513 Teltow Tel +49 (0) 3328 30 76-0 | Fax +49 (0) 3328 30 76-99 info@somatex.com | www.somatex.com #### Substantial Equivalence: The Tumark® Professional and MRI Tumark® Professional are substantial equivalent to the Tumark® Professional, also manufactured by Somatex Medical Technologies GmbH, that has been cleared by FDA on March 19, 2008 (K073095), the ClipLoc Soft Tissue marker, manufactured by MRI Devices Corporation (K033447) and the UltraClip Tissue Marker, manufactured by Inrad Inc. The Tumark® Professional and MRI Tumark® Professional and each of their predicate devices are intended to attach a marker to soft tissue at the surgical site during an open or a percutaneous procedure. All devices are indicated for use to radiographically and radiologically mark the surgical location in breasts following an open or percutaneous procedure. In addition, both the proposed devices and the predicate devices are identical or similar in technology, design and material. In particular, the proposed devices and the predicate devices consist of the same primary components and the component materials of the proposed devices and the predicate devices are substantially equivalent. Furthermore, biocompatibility, sterility and packaging testing of the proposed devices demonstrate the safety and effectiveness of the system. {4}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for SOMATEX Medical Technologies GmbH. Above the logo is the text "K093064 Page 4/4". The SOMATEX logo is in a bold, sans-serif font, with the registered trademark symbol to the right of the X. MADE IN GERMANY Rheinstraße 7d | D-14513 Teltow Tel +49 (0) 3328 30 76-0 | Fax +49 (0) 3328:30 76-99 info@somatex.com | www.somatex.com Based on the same intended use and the similiarities in technology, design and materials the Turnark® Professional and MRI Tumark® Professional are substantially equivalent to their predicate devices. The minor technological differences between the proposed devices and the predicate devices do not raise new questions of safety and effectiveness. #### Date Prepared: January 18, 2010 {5}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and three wavy lines below it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 FEB 1 7 2010 Somatex Medical Technologies, GmbH % Ms. Susanne Raab 1480 Cambridge Street Cambridge Massachusetts 02139 Re: K093064 Trade/Device Name: Tumark® Professional/MRI Tumark® Professional Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable clip Regulatory Class: II Product Code: NEU Dated: January 20, 2010 Received: January 25, 2010 Dear Ms. Raab: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {6}------------------------------------------------ Page 2 - Ms. Susanne Raab forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 100-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm1115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notfication" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its the fire number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours incerely yours, Peter Romano Mark N. Melkerson Director . Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ ### Indication for Use 510(k) Number (if known): Device Name: Tumark® Professional / MRI Tumark® Proféssional Indications for Use: The Tumark® Professional and MRI Tumark® Professional are intended to attach a marker to soft tissue at the surgical site during an open or a perculancous procedure. The devices are indicated for use to radiographically and radiologically mark the surgical location in breasts following an open or percutaneous procedure. Only the MRI Tumark® Professional is indicated to be used in Magnetic Resonance Imaging (MRI) procedures. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Daniel Kime for MXM Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K093064 Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...