K092997 · Edan Instruments, Inc. · KNG · Nov 10, 2009 · Obstetrics/Gynecology
Device Facts
Record ID
K092997
Device Name
ULTRASONIC TABLETOP DOPPLER, MODELS SD5 AND SD6
Applicant
Edan Instruments, Inc.
Product Code
KNG · Obstetrics/Gynecology
Decision Date
Nov 10, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.2660
Device Class
Class 2
Indications for Use
The Ultrasonic TableTop Doppler is intended to be used by health care professionals including registered nurses, practical nurses, midwives, ultrasound technicians, and physician assistants, by prescription from licensed physicians in hospitals, clinics and private offices. The 2 MHz and/or 3 MHz probes are indicated for the detection of fetal heart rate from early gestation thru delivery and as a general indication of fetal well being. They can also be used to verify fetal heart viability following patient trauma.
Device Story
Ultrasonic TableTop Doppler (Models SD5, SD6) uses continuous wave (CW) Doppler transducers (2MHz or 3MHz) to detect fetal heart rate (FHR); device records and plays back 240 seconds of fetal heart sound. Operated by healthcare professionals (nurses, midwives, technicians, PAs) in hospitals, clinics, and private offices; not for ICU, OR, or home use. SD6 model features wireless probe with Li-ion battery and infrared communication. Output provides FHR data to clinicians for assessment of fetal well-being and viability. Benefits include non-invasive, real-time fetal monitoring.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including software, hardware, safety, and environmental testing, alongside risk analysis and final validation.
Technological Characteristics
Continuous wave (CW) Doppler fetal monitor. Models SD5 (wired) and SD6 (wireless probe). 2MHz and 3MHz transducers. Ni-MH battery (main unit), Li-ion battery (SD6 probe). Infrared communication (SD6). Class II device. Software-based signal processing for FHR detection.
Indications for Use
Indicated for detection of fetal heart rate from early gestation through delivery, general indication of fetal well-being, and verification of fetal heart viability following trauma.
Regulatory Classification
Identification
A fetal ultrasonic monitor is a device designed to transmit and receive ultrasonic energy into and from the pregnant woman, usually by means of continuous wave (doppler) echoscopy. The device is used to represent some physiological condition or characteristic in a measured value over a period of time (e.g., perinatal monitoring during labor) or in an immediately perceptible form (e.g., use of the ultrasonic stethoscope). This generic type of device may include the following accessories: signal analysis and display equipment, electronic interfaces for other equipment, patient and equipment supports, and component parts. This generic type of device does not include devices used to image some relatively unchanging physiological structure or interpret a physiological condition, but does include devices which may be set to alarm automatically at a predetermined threshold value.
{1}------------------------------------------------
C092997
Section 1
### 510(k) Summary of Safety and Effectiveness
This summary of 510k safety and effectiveness is being submitted in according with 21CFR part 807.92
| Submitter: | Edan Instruments, Inc<br>3/F - B, Nanshan Medical<br>Equipments Park, Nanhai Rd 1019#,<br>shekou, Nanshan Shenzhen.<br>518067 P.R. China<br>Tel: 86-755-26882220<br>Fax:86-755-26882223<br>Contact person: Jiang Yucai | NOV 10 2009 |
|------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
|------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
William Stern Official correspondent: Multigon Industries,Inc. l Odell Plaza Yonkers, N.Y. 10701 Phone: 914.376 5200 X27 914 376 6111 Fax:
2009-9-23
KNG
Date of Preparation: Proprietary Name: Classification Name: Product code: Predicate Devices:
| Predicate devices | IMEXDOP CT+ | Sonotrax series pocket<br>doppler |
|-------------------|------------------------------|-----------------------------------|
| Manufacturer | Imex Medical<br>Systems, Inc | Edan Instruments, Inc |
| K # | K942441 | K080087 |
Device Description:
Ultrasonic TableTop Doppler provides the following primary features:
Ultrasonic TableTop Doppler (Models SD5, SD6)
21 CFR 884.2660 Fetal ultrasonic monitor and accessories
- · Basic parameters: FHR
- · 240 seconds fetal heart sound record and playback
- · Infrared communication(for SD6 only)
- · Ni-MH battery for 20 hours continuous working of main unit
- · Li-ion battery for 2.5 hours continuous working of SD6 probe
- · Charge the SD6 probe battery by main unit
- · Continuous wave Doppler transducer for FHR detection
#### Comparison with predicate device
The Ultrasonic TableTop Doppler models including SD5 and SD6
3
{2}------------------------------------------------
. '.'
have the same device characteristics as the predicate devices mentioned above, the 2MHz, 3MHz probes for heart rate detection uses the same technology and circuitry as the cleared Sonotrax series pocket Doppler under K080087. The Ultrasonic TableTop Doppler is also similar and comparable to the IMEXDOP CT+ cleared under K024197. Hence Ultrasonic TableTop Doppler above is substantially equivalent to the predicate devices cited.
The Ultrasonic TableTop Doppler is intended to be used by health Intended Use: care professionals including registered nurses, practical nurses, midwives, ultrasound technicians, and physician assistants, by prescription from licensed physicians in hospitals, clinics and private offices.
> The 2 MHz and/or 3 MHz probes are indicated for the detection of fetal heart rate from early gestation thru delivery and as a general indication of fetal well being. They can also be used to verify fetal heart viability following patient trauma.
It is not intended for use in intensive care units, operating rooms or Contraindications: for home use.
Test Summary:
The following quality assurance measures were applied to the development of the Ultrasonic Doppler
- · Software testing
- · Hardware testing
- · Safety testing
- · Environment test
- · Risk analysis
- Final validation
Conclusion:
Verification and validation testing was done on the Ultrasonic TableTop Doppler. This premarket notification submission demonstrates that Ultrasonic TableTop Doppler is substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/2/Picture/13 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with outstretched wings. The bird is positioned to the right of the text, which is arranged in a circular pattern around the bird. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
Edan Instruments, Inc. % Mr. William Stern Official Correspondent Multigon Industries, Inc. 1 Odell Plaza YONKERS NY 10701
## NOV 1 0 2009
Re: K092997
Trade/Device Name: Ultrasonic TableTop Doppler (Models SD5 and SD6) Regulation Number: 21 CFR 884.2660 Regulation Name: Fetal ultrasonic monitor and accessories Regulatory Class: II Product Code: KNG Dated: September 25, 2009 Received: September 28, 2009
Dear Mr. Stern:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Ultrasonic TableTop Doppler (Models SD5 and SD6) as described in your premarket notification:
Transducer Model Number
2MHz CW fetal probe - model: SD5 3MHz CW fetal probe - SD5 2MHz CW wireless fetal probe - model: SD6 3MHz CW wireless fetal probe - model: SD6
{4}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fdagov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Paul Hardy at (301) 796-6542.
Sincerely vours.
Janine M. Morris
to Janine M. Morris Acting Director. Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{5}------------------------------------------------
# Diagnostic Ultrasound indications for Use Form Fill out one form for each ultrasound system and each transducer. 2MHz CW fetal probe- model: SD5
Intended use : Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows
| Clinical Application | | Mode Operation | | | | | | | | |
|------------------------------|---|----------------|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | P | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative(specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ(specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheraln Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal | | | | | | | | | | |
| Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other | | | | | | | | | | |
N=new indication; P=previously cleared by FDA; e=ADDED UNDER appendix E The above is a 2MHz CW transducer for the fetal heart rate detection. Additional Comments:
> PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE I FNEEDED CONCURRENCE OF cdrh, Office of Device Evaluation (ODE)
Jose M. Perez
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Device 510(k) Number
6
Prescription Use
(Per 21 CFR 801.109)
(Per 21 CFR 801.109)
{6}------------------------------------------------
## Diagnostic Ultrasound indications for Use Form Fill out one form for each ultrasound system and each transducer. 3MHz CW fetal probe- model: SD5
Intended use : Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows
| Clinical Application | | Mode Operation | | | | | | | | |
|------------------------------|---|----------------|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | P | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative(specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ(specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal | | | | | | | | | | |
| Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other | | | | | | | | | | |
N=new indication; P=previously cleared by FDA; e=ADDED UNDER appendix E N=1iew multanon, - I =previously vicured of FDT, OW transducer for the fetal heart rate detection.
> PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE I FNEEDED CONCURRENCE OF cdrb, Office of Device Evaluation (ODE)
Aom Mhz
7
(Division Sign-Off) (Division of Reproductive, Abdominal, and Radiological Devic 510(k) Number
**Prescription Use**
(Per 21 CFR 801.109)
{7}------------------------------------------------
## Diagnostic Ultrasound indications for Use Form Fill out one form for each ultrasound system and each transducer. 2MHz CW wireless fetal probe- model: SD6
Intended use : Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows
| Clinical Application | | Mode Operation | | | | | | | | |
|------------------------------|---|----------------|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | N | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative(specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ(specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal | | | | | | | | | | |
| Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other | | | | | | | | | | |
N=new indication; P=previously cleared by FDA; e=ADDED UNDER appendix E The above is a 2MHz CW transducer for the fetal heart rate detection. Additional Comments: __
> PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE I FNEEDED CONCURRENCE OF cdrb, Office of Device Evaluation (ODE)
Arzu M. Whang
(Division Sign-Off)
(Division Sign-Off) Division of Reproductive, Abdominal and Radiogical Devic 510(k) Number
8
Prescription Use
(Per 21 CFR 801.109)
{8}------------------------------------------------
## Diagnostic Ultrasound indications for Use Form Fill out one form for each ultrasound system and each transducer. 3MHz CW wireless fetal probe- model: SD6
Intended use : Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows
| Clinical Application | | Mode Operation | | | | | | | | |
|------------------------------|---|----------------|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | N | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative(specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ(specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal | | | | | | | | | | |
| Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other | | | | | | | | | | |
N=new indication; P=previously cleared by FDA; e=ADDED UNDER appendix E The above is a 3MHz CW transducer for the fetal heart rate detection. Additional Comments: __
> PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE I FNEEDED CONCURRENCE OF cdrh, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division a Reproductive, Abdominal, and Radio gical Devices 510(k) Number.
9
Prescription Use. (Per 21 CFR 801.109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.