K092937 · Diagnostica Stago, Inc. · JIT · Aug 2, 2010 · Clinical Chemistry
Device Facts
Record ID
K092937
Device Name
STA-HYBRID HEP CALIBRATOR
Applicant
Diagnostica Stago, Inc.
Product Code
JIT · Clinical Chemistry
Decision Date
Aug 2, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1150
Device Class
Class 2
Intended Use
The STA® - Hybrid Hep Calibrator is a set of calibrator plasmas intended for use with analyzers of the STA® line suitable to these reagents, for the calibration of heparin (UFH and LMWH) activity assay by measuring the anti-Xa activity using the chromogenic method, STA® - Rotachrom® Heparin.
Device Story
STA® - Hybrid Hep Calibrator is a bundled kit of lyophilized human plasmas used to generate calibration curves for heparin (UFH and LMWH) activity assays. The kit contains five reagents: one heparin-free plasma, two plasmas with defined UFH concentrations, and two plasmas with defined LMWH concentrations. It is intended for use with the STA® line of IVD coagulation analyzers. The device operates by providing known reference values to calibrate the analyzer's chromogenic anti-Xa activity measurements. By establishing these curves, the analyzer can accurately quantify heparin levels in patient samples, assisting clinicians in monitoring anticoagulant therapy. The device is used in clinical laboratory settings by trained laboratory personnel.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Set of lyophilized human plasmas. Reagents include heparin-free plasma and plasmas with defined UFH or LMWH concentrations. Designed for use with STA® line IVD coagulation analyzers. Operates via chromogenic anti-Xa method. No software or electronic components.
Indications for Use
Indicated for use as a calibration set for heparin (UFH and LMWH) activity assays on STA® line analyzers using the chromogenic anti-Xa method. No specific patient population or contraindications are defined.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the word "Stago" in a serif font, with the letters evenly spaced and of uniform size. Above the word, there is a stylized graphic element that appears to be composed of three curved shapes arranged in a circular fashion, resembling a stylized flower or abstract design. The graphic is solid black, contrasting with the white background.
### 510(k) Summary
Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared:
Umberto V. Parrotta, Jr. Diagnostica Stago, Inc. Five Century Drive Parsippany, New Jersey 07054
K092937
AUG 0 2 2010
Phone: (973) 631-1200, x-2044 Facsimile: (973) 695-0095
Contact Person: Umberto V. Parrotta, Jr.
Date Prepared: September 23, 2009
#### Name of Device and Name/Address of Sponsor:
STA® - Hybrid Hep Calibrator
Diagnostica Stago, Inc. Five Century Drive Parsippany, New Jersey 07054
#### Common or Usual Name:
- IVD Coagulation Calibration Device. ●
- IVD Calibrator. .
#### Classification Name:
- Name: Calibrator, Secondary. .
- Description: Calibrator .
#### Predicate Device:
- STA® Calibrator HBPM/LMWH Kit (K010350). ●
- STA® Hepanorm® H (K854762) { formerly cleared as Hepanorm® . Calibration Plasma Set}.
### Purpose of the Traditional 510(k) Notice:
STA® - Hybrid Hep Calibrator is a new device that bundles two of the Company's previously FDA cleared devices, STA® - Hepanorm® H and STA® - Calibrator
DSI-STA-HHC-Attachment-04v1-510(k)Summary.doc
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the word "Stago" in a serif font. Above the word is a graphic consisting of two curved shapes that resemble stylized leaves or petals. The shapes are arranged in a circular fashion, creating a sense of movement or rotation above the word.
# 510(k) Summary
HBPM/LMWH into a single kit. Thus, the primary difference between the subject product and predicate devices is that it contains a set of calibration plasmas for calibration of assays of unfractionated heparin (UFH) and Low Molecular Weight Heparin (LMWH) bundled together in a single kit.
#### Indication/Intended Use:
The STA® - Hybrid Hep Calibrator is a set of calibrator plasmas intended for use with analyzers of the STA® line suitable to these reagents, for the calibration of heparin (UFH and LMWH) activity assay by measuring the anti-Xa activity using the chromogenic method, STA® - Rotachrom® Heparin.
#### Technological Characteristics:
- The STA® Hybrid Hep Calibrator is a set of lyophilized human plasmas . used to create the calibration curve on the STA® line of IVD instruments performing the chromogenic method for heparin (UFH and LMWH) assays.
- Each STA® Hybrid Hep Calibrator available contains: .
- 0 4 x 1-ml vials of Reagent 1: STA® Hybrid Hep Calibrator O lyophilized human plasma free of heparin.
- 4 x 1-ml vials of Reagent 2: STA® Hybrid Hep Calibrator 3 o lyophilized human plasma containing a well-defined quantity of UFH.
- 4 x 1-ml vials of Reagent 3: STA® Hybrid Hep Calibrator 6 o lyophilized human plasma containing a well-defined quantity of UFH that is greater than that of Reagent 2.
- (9) 4 x 1-ml vials of Reagent 4: STA® - Hybrid Hep Calibrator o lyophilized human plasma containing a well-defined quantity of LMWH.
- 4 x 1-ml vials of Reagent 5: STA® Hybrid Hep Calibrator (18 o lyophilized human plasma containing a well-defined quantity of LMWH that is greater than that of Reagent 4.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for Stago. The logo consists of the word "Stago" in a serif font, with a stylized graphic above it. The graphic appears to be an abstract shape, possibly representing a stylized eye or a leaf. The logo is simple and modern in design.
# 510(k) Summary
#### Principles of Operation
The STA® - Hybrid Hep Calibrator is designed to operate utilizing the STA® product line of IVD coagulation analyzers for the purpose of creating calibration curves for assays of heparin by measuring the anti-Xa activity. While the subject device performs calibration of assays based on chromogenic method, the predicate devices perform calibration of assays based on chromogenic method and on clotting method, both being anti-Xa method.
#### Substantial Equivalence:
The STA® - Hybrid Hep Calibrator, the subject product of this submission, and the predicate devices, STA® - Calibrator HBPM/LMWH Kit and STA® - Hepanorm® H are identical products regarding indication/intended use, formulation or materials of construction and design, technology, and safety. The primary difference between the subject product and predicate devices is that the product contains a set of calibration plasmas for assays of both unfractionated heparin (UFH) and Low Molecular Weight Heparin (LMWH) bundled in a single kit whereas the predicate devices are marketed as two (2) separate and distinct products - one kit for UFH and the other kit for LMWH. In summary, the predicate devices are bundled together as a single kit. The STA® - Hybrid Hep Calibrator like the predicate devices, is a calibrator used in creating the calibration curve on IVD devices for the purpose of Heparin activity assay by measuring the anti-Xa activity.
The product, STA® - Hybrid Hep Calibrator and the predicate devices. STA® -Hepanorm® H and STA®- Calibrator HBPM/LMWH are similar in indication/intended use, technology, principles of operation, and application of use (with IVD medical devices) thus yielding no new questions in safety, effectiveness, or technology. Therefore, this concludes the product STA® - Hybrid Hep Calibrator is substantially equivalent to the predicate devices, STA® - Calibrator HBPM/LMWH Kit (K010350) and STA® - Hepanorm® H {previously cleared as Hepanorm® Calibration Plasma Set} (K854762).
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle with three curved lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring. MD 20993-0002
Diagnostica Stago, Inc. c/o Mr. Umberto V. Parrotta Director of Regulatory Affairs and Quality Assurance Five Century Drive Parsippany, New Jersey 07054
AUG 0 2 2010
Re: k092937
Trade/Device Name: STA® - Hybrid Hep Calibrator Regulation Number: 21 CFR 864.7525 Regulation Name: Heparin Assay Regulatory Class: Class II Product Code: KFF, JIT Dated: July 21, 2010 Received: July 22, 2010
Dear Mr. Parrotta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket
{4}------------------------------------------------
Page 2 - Mr. Umberto Parrotta
notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours.
m. chan
Maria M. Chan, Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
#### Enclosure
{5}------------------------------------------------
#### Indications for Use
AUG 0 2 2010
510(k) Number (if known): 《092937
Device Name: STA®- Hybrid Hep Calibrator.
Indication for Use:
The STA® - Hybrid Hep Calibrator is a set of calibrator plasmas intended for use with analyzers of the STA® line suitable to these reagents, for the calibration of heparin (UFH and LMWH) assay by measuring the anti-Xa activity using the chromogenic method, STA® - Rotachrom® Heparin.
Prescription Use X (21 CFR Part 801 Subpart D)
And/Or
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Lenck R. Si.
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K092937
DSI-STA-HHC-Attachment-01-IndicationsForUseStatementForm.doc
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.