K092904 · Custom Spine, Inc. · ODP · Dec 22, 2009 · Orthopedic
Device Facts
Record ID
K092904
Device Name
PATHWAY IVB/VBR
Applicant
Custom Spine, Inc.
Product Code
ODP · Orthopedic
Decision Date
Dec 22, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
When used as an intervertebral body fusion device, the Custom Spine PATHWAY ACIF is intended for spinal fusion procedures at one level (C2-T1) in skeletally mature patients with degenerative disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Implants are to be implanted via an open, anterior approach and packed with autogenous bone. Patients should receive 6 weeks of non-operative treatment prior to treatment with the Custom Spine ACIF. When used as a vertebral body replacement, the Custom Spine PATHWAY ACIF is intended for use in the thoracic and/or throcolumbar spine (TI-L5) to replace a collapsed, or unstable vertebral body resected or excised (i.e. partial or total vertebrectomy procedures) due to tumor or trauma (i.e. fracture). The PATHWAY ACIF, when used as a vertebral body replacement, can be packed with autograft. For all indications, the device is intended to be used with supplemental fixation systems that have been cleared for use in cervical, thoracic, or lumbar spine (posterior pedicle screw systems, anterior plate systems, and anterior screw and rod systems).
Device Story
PATHWAY ACIF is a trapezoidal intervertebral body fusion/vertebral body replacement device; manufactured from PEEK OPTIMA LT-1; includes tantalum markers for radiographic visualization. Device features central cavity for autograft packing; superior/inferior surfaces include serrated teeth to resist migration. Implanted via open anterior approach; provides mechanical support until biologic fusion occurs. Used in conjunction with cleared supplemental fixation systems (e.g., anterior plates, pedicle screws).
Clinical Evidence
Bench testing only. Testing performed per ASTM F2267-04 (load-induced subsidence) and ASTM F2077-03 (intervertebral body fusion device testing).
Technological Characteristics
Material: PEEK OPTIMA LT-1; Tantalum markers (ASTM F560). Form factor: Trapezoidal, 5-12mm height, 15mm width, 13mm length. Features: Central cavity for autograft, serrated teeth for fixation. Non-sterile. Mechanical support device; no software or energy source.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (C2-T1) requiring spinal fusion, or patients with thoracic/thoracolumbar (T1-L5) vertebral body collapse/instability due to tumor or trauma requiring replacement. Requires 6 weeks prior non-operative treatment for DDD. Must be used with supplemental fixation.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
# 510(K) SUMMARY
#### Name of Firm
Custom Spine, Incorporated 1140 Parsippany Blvd, Suite 201 Parsippany, NJ 07054 Phone: (973) 808-0019 Fax: (973) 808-0707
DEC 2 2 2009
# Official Correspondent
Saad Attiyah Suad Attiven
Manager of Regulatory Affairs and Quality Assurance 1140 Parsippany Blvd, Suite 201 Parsippany, NJ 07054 Phone: (973) 265-5036 Fax: (973) 808-0707 E-mail: saad@customspine.com
# Establishment Number
3005129649
#### Device Name
Legally Marketed Trade Name: PATHWAY ACIF Legally Marketed Trade Name: PATHWAT ACT
Common Name: Intervertebral fusion device with bone graft, cervical/Vertebral Body Replacement Device Classification: Class II Device Clussmeatier: 21 CFR 888.3060, 21 CFR 888.3080 Device Product Codes: MQP, ODP
### Predicate Devices
r r eurcate De vices
LDR Spine MC+ (K043479), Zimmer Spine BAK/C Cervical Interbody Fusion Device (P980048), and X-Spine Calix ™Spinal Implant (K083637)
#### Device Description
DCVICC DODGE = p = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = The PATHWAY ACIF is a cervical interbody fusion active as identification markers for the (Polyetheretherketone). The device continues of the implant both intra-operatively and postoperatively.
The device is trapezoidal in nature and is provided in various heights ranging from 5 mm to 12 mm and a The device is trapezondan in hatale and is pro naiss a central cavity which is to be packed with width of 15 mm, and 13 mm in length This abouts of Indusis (0 Degree and 7 Degree and 7 Degree). The superior
autograft. This device is provided in various degrees of lordos autograft. This device is provided in the resist implant migration/displacement.
and inferior surfaces have serrated teeth to help resist implant migration/displacement.
{1}------------------------------------------------
The device is intended to provide mechanical support to the implanted level until biologic fusion is achieved.
The product is provided clean and "non-sterile".
#### Indications for Use
The PATHWAY ACIF, when used as an intervertebral body fusion device, is intended is for spinal fusion procedures at one level (C2-T1) in skeletally mature patients with degenerative disc disease (defined as procedures as and religion with degeneration of the disc confirmed by history and radiographic neel pain of also be and and and and and and open, anterior approach and packed states, or the use bone. Patients should receive 6 weeks of non-operative treatment with the PATHWAY ACIF intervertebral body fusion device.
When used as a vertebral body replacement, the PATHWAY ACIF is intended for use in the thoracic vincen atos aoumbar spine (TI-L5) to replace a collapsed, damaged, or unstable vertebral body resected ana/of will be partial or total vertebrectomy procedures) due to tumor or trauma (i.e. fracture). The PATHWAY ACIF, when used as a vertebral body replacement, can be packed with autograft.
For all indications, the device is intended to be used with supplemental fixation systems that have been , or an franciese in cervical, thoracic, or lumbar spine (posterior pedicle screw systems, anterior plate systems, and anterior screw and rod systems).
#### Materials
The PATHWAY ACIF is made from PEEK OPTIMA LT-1 (Polyetheretherketone) and contains tantalum markers (ASTM F560).
#### Performance Data
The PATHWAY ACIF device was tested in accordance with ASTM F2267-04 "Standard Test Method for Measuring Load Induced Subsidence of Intervertebral Body Fusion Devices Under Static Axial Websited by Jan ASTM F2077-03 "Test Method For Intervertebral Body Fusion Devices". The data are located in Section 19 of this submission, "Bench Testing". The data demonstrate that the device is capable of performing in accordance with its intended use.
### Substantial Equivalence Statement
The PATHWAY ACIF is equivalent to the previously cleared systems, as they utilize the same principle of operation, and has similar indications for use as the predicates, and materials.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle, indicating the department's name and national affiliation.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
## OEC 22 2009
Custom Spine, Inc. % Mr. Saad Attiyah Manager of Regulatory Affairs and Quality Assurance 1140 Parsippany Boulevard, Suite 201 Parsippany, New Jersey 07054
Re: K092904
Trade Name: Pathway ACIF Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: ODP, MQP Dated: November 24, 2009 Received: November 27, 2009
Dear Mr. Attiyah:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above).into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
Page 2 - Mr. Saad Attiyah
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
lttp://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation . Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
KO92904 510(k) Number:
When used as an intervertebral body fusion device, the Custom Spine PATHWAY ACIF is intended for spinal fusion procedures at one level (C2-T1) in skeletally mature patients with degenerative disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Implants are to be implanted via an open, anterior approach and packed with autogenous bone. Patients should receive 6 weeks of non-operative treatment prior to treatment with the Custom Spine ACIF.
When used as a vertebral body replacement, the Custom Spine PATHWAY ACIF is intended for use in the thoracic and/or throcolumbar spine (TI-L5) to replace a collapsed, or unstable vertebral body resected or excised (i.e. partial or total vertebrectomy procedures) due to tumor or trauma (i.e. fracture). The PATHWAY ACIF, when used as a vertebral body replacement, can be packed with autograft.
For all indications, the device is intended to be used with supplemental fixation systems that have been cleared for use in cervical, thoracic, or lumbar spine (posterior pedicle screw systems, anterior plate systems, and anterior screw and rod systems).
Prescription Use _____________________________________________________________________________________________________________________________________________________________
AND/OR
Over-The-Counter Use
(Part 21 C.F.R. 801 Subpart D)
(21 C.F.R. 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign Off
Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number k092204
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.