← Product Code [JJX](/productcode/JJX) · K092848

# ELECSYS TROPONIN T CALCHECK 5 (K092848)

_Roche Diagnostics Corp. · JJX · Mar 19, 2010 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K092848

## Device Facts

- **Applicant:** Roche Diagnostics Corp.
- **Product Code:** [JJX](/productcode/JJX.md)
- **Decision Date:** Mar 19, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1660
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

The Elecsys Troponin T CalCheck 5 is an assayed control for use in calibration verification and for use in the verification of the assay range established by the Elecsys Troponin T reagent on the indicated Elecsys and cobas e immunoassay analyzers. For in vitro diagnostic use. For prescription use only. The Elecsys Troponin T CalCheck 5 is not intended to be used as a primary calibrator or routine control material.

## Device Story

Elecsys Troponin T CalCheck 5 is an assayed quality control material used to verify calibration and assay range for Troponin T reagents on Elecsys and cobas e immunoassay analyzers. It functions as a reference standard to ensure analytical performance within established ranges. Used by laboratory professionals in clinical settings to validate system accuracy. Does not provide diagnostic results directly; serves as a quality assurance tool to confirm instrument performance before patient sample testing.

## Clinical Evidence

No clinical data. Bench testing only. Performance established via traceability to internal reference material and recovery analysis on reference analyzers (Elecsys 2010 and E170) to confirm target values and stability.

## Technological Characteristics

Lyophilized recombinant human cardiac Troponin T in human serum matrix. Five concentration levels (Level 1: <0.01 ng/mL to Level 5: 25.0 ng/mL). Reconstitution required (1.0 mL distilled/deionized water). Stability: 18 months unopened at 2-8°C; 4 hours reconstituted at 20-25°C. Analyte traceable to internal reference material.

## Regulatory Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

## Predicate Devices

- Elecsys Troponin T CalCheck 5 ([K040373](/device/K040373.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE

A. 510(k) Number:
k092848

B. Purpose for Submission:
New device

C. Measurand:
Calibration verification and assay range verification material for cardiac troponin T

D. Type of Test:
Not applicable

E. Applicant:
Roche Diagnostics

F. Proprietary and Established Names:
Elecsys Troponin T CalCheck 5

G. Regulatory Information:
1. Regulation section:
21 CFR §862.1660, Quality Control Material (Assayed and Unassayed)
2. Classification:
Class I, reserved
3. Product code:
JJX - Single (Specified) Analyte Controls (Assayed and Unassayed)
4. Panel:
75 Clinical Chemistry

{1}

# H. Intended Use:

1. Intended use(s):

Refer to indications for use below

2. Indication(s) for use:

The Elecsys Troponin T CalCheck 5 is an assayed control for use in calibration verification and for use in the verification of the assay range established by the Elecsys Troponin T reagent on the indicated Elecsys and cobas e immunoassay analyzers.

3. Special conditions for use statement(s):

For in vitro diagnostic use. For prescription use only. The Elecsys Troponin T CalCheck 5 is not intended to be used as a primary calibrator or routine control material.

4. Special instrument requirements:

The labeling states that the CalChecks are for use with the Elecsys Troponin T reagent on the Elecsys 2010/cobas e411 and Modular Analytics E170/cobas e601 test systems.

# I. Device Description:

The Elecsys Troponin T CalCheck 5 is a lyophilized product consisting of recombinant human cardiac Troponin T in human serum matrix. During manufacture, the analyte is spiked into the matrix at the desired concentration levels. CalCheck Level 1 contains no analyte.

|  Elecsys Troponin T CalCheck | Target Value (ng/mL)  |
| --- | --- |
|  Level 1 | <0.01  |
|  Level 2 | 0.5  |
|  Level 3 | 10.0  |
|  Level 4 | 20.0  |
|  Level 5 | 25.0  |

The human source materials were prepared exclusively from the blood of donors tested individually and shown to be free from HBsAg and antibodies to HCV and HIV. The testing methods applied were FDA-approved or cleared in compliance with the European Directive 98/79/EC, Annex II, List A.

{2}

3

J. Substantial Equivalence Information:

1. Predicate device name(s):
Elecsys CalCheck Troponin T

2. Predicate 510(k) number(s):
k984372

3. Comparison with predicate:
|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device (k092848) | Predicate (k984372)  |
|  Format | Lyophilized | Same  |
|  Reactive Component | Recombinant human cardiac Troponin T | Same  |
|  Matrix | Human serum | Same  |
|  Handling | Reconstitute the contents of each vial with exactly 1.0 mL distilled or deionized water. Allow the bottle to stand closed for 15 minutes. Mix gently by inversion to ensure homogeneity. | Same  |
|  Stability | Unopened
Store at 2-8°C until expiration date.
Reconstituted
20-25°C for 4 hours. | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device (k092848) | Predicate (k984372)  |
|  Intended Use | For use in calibration verification and for use in the verification of the assay range established by the Elecsys Troponin T reagent on the indicated Elecsys and cobas e immunoassay analyzers. | For use in the verification of the calibration established by the Elecsys Troponin T reagent on Elecsys 1010 or 2010 immunoassay analyzers.  |
|  Levels | 5 | 3  |

K. Standard/Guidance Document Referenced (if applicable):
Guidance for Industry and FDA Staff - Assayed and Unassayed Quality Control Material

{3}

L. Test Principle:

Not applicable

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

Not applicable

b. Linearity/assay reportable range:

Not applicable

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Traceability

The Troponin T used in this quality control material is traceable to an internal reference material assayed with the Elecsys Troponin T 4th Generation assay. The control material is analyzed on the sponsor’s reference analyzer and recovered values must fall within pre-determined acceptance criteria.

Expected Values and Value Ranges

Expected values for the Elecsys Troponin T CalCheck 5 are determined by duplicate analyses of levels 1 to 5 using N ≥ 3 analyzers (i.e. Elecsys 2010 and E170). Pre-determined acceptance criteria for Troponin T recovery must be met for each control lot. The target value for each CalCheck is the median of the observed values. Troponin T expected values and value ranges are lot dependent and are listed in the lot-specific value sheet. Test results must fall within the range printed on the lot-specific value sheet. The labeling states that laboratories should establish appropriate acceptance criteria when using this product for its intended use.

Stability

The manufacturer claims CalCheck 5 is stable until the expiration date printed on the vial when stored unopened at 2 – 8°C. Accelerated stability testing supports the target shelf life claim of 18 months. The reconstituted vials are stable for four hours at 20-25°C.

4

{4}

d. Detection limit:
Not applicable

e. Analytical specificity:
Not applicable

f. Assay cut-off:
Not applicable

2. Comparison studies:
a. Method comparison with predicate device:
Not applicable
b. Matrix comparison:
Not applicable

3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable

4. Clinical cut-off:
Not applicable

5. Expected values/Reference range:
The expected values are provided in the labeling for each specific lot.

5

{5}

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

6

---

**Source:** [https://fda.innolitics.com/device/K092848](https://fda.innolitics.com/device/K092848)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
