APEX SPINE SYSTEM 5.50 MM TITANIUM ROD & POLYAXIAL SCREW WASHER
K092825 · Spinecraft, Inc. · KWP · Mar 16, 2010 · Orthopedic
Device Facts
Record ID
K092825
Device Name
APEX SPINE SYSTEM 5.50 MM TITANIUM ROD & POLYAXIAL SCREW WASHER
Applicant
Spinecraft, Inc.
Product Code
KWP · Orthopedic
Decision Date
Mar 16, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The APEX Spine System 5.5mm Rod and Polyaxial Screw Washers intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; spinal tumor; and failed previous fusion (pseudoarthrosis). The APEX Spine System 5.5mm Rod and Polyaxial Screw Washers is also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. The APEX Spine System 5.5mm Rod and Polyaxial Screw Washers is also a hook and sacral/liac screw fixation system of the non-cervical spine indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudo-arthrosis). The APEX Spine System 5.5mm Rod and Polyaxial Screw Washers when used with pedicle screws are indicated for degenerative discogenic back pain with degeneration of the disc. confirmed by history and radiographic studies). Levels of fixation are for the thoracic, lumbar and sacral spine.
Device Story
Rod-based spinal fixation system; includes 5.50mm titanium rods and polyaxial screw washers. Designed to interface with previously cleared APEX system components (monoaxial/polyaxial screws, hooks, connectors). Washers available in two sizes to accommodate screw shaft diameters ranging from 4.75mm to 8.50mm. Used by surgeons in clinical settings to provide immobilization and stabilization of spinal segments during fusion procedures. Mechanical performance verified via bench testing to ensure comparability to predicate systems. Benefits patients by stabilizing spinal segments to facilitate fusion in cases of deformity, trauma, or degenerative disease.
Clinical Evidence
Bench testing only. Mechanical and dynamic testing performed to compare performance against predicate devices; results demonstrate comparable or superior mechanical performance.
Technological Characteristics
Implant grade titanium alloy (Ti-6Al-4V) per ASTM F-136 and ISO 5832-3. Rod-based system with polyaxial screw washers. Mechanical fixation device; no software or energy source.
Indications for Use
Indicated for skeletally mature patients requiring spinal immobilization/stabilization as an adjunct to fusion. Treats thoracic, lumbar, and sacral spine conditions: severe spondylolisthesis (Grades 3-4, L5-S1), degenerative spondylolisthesis with neurologic impairment, fracture, dislocation, scoliosis, kyphosis, lordosis, spinal stenosis, spinal tumor, and pseudoarthrosis. Used with autogenous bone graft for pedicle screw fixation (L3-sacrum).
Regulatory Classification
Identification
A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.
{0}------------------------------------------------
SpineCraft
510(k) Premarket Notification APEX Spine System 5.5mm Rod & Polyaxial Screw Washer
K092825
## 510(k) SUMMARY
# MAR 1 6 2010
| SUBMMITTER: | SpineCraft, LLC<br>2215 Enterprise Drive<br>Westchester, IL 60154-5819 |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ESTABLISHMENT<br>REGISTRATION No: | 3004717358 |
| CONTACT PERSON: | Ami Akallal-Asaad, Director of Regulatory Affairs |
| DATE PREPARED: | September 8, 2009 |
| CLASSIFICATION NAME: | Spinal Interlaminal Fixation Orthosis - 888.3050<br>Spinal Pedicle Fixation - 888.3070(b) (1) |
| PROPRIETARY NAME: | APEX Spine System 5.50mm Titanium Rod and Polyaxial<br>Screw Washers |
| COMMON NAME: | Spinal Fixation System |
| PRODUCT CODE: | KWP, MNH, MNI |
| CLASSIFICATION PANEL: | 87 |
| PRODUCT DESCRIPTION: | The APEX Spine System 5.50mm Titanium Rod and Polyaxial<br>Screw Washers is a rod-based system designed to interface with<br>various spinal anatomies. The 5.50mm Rods are available in<br>various lengths and are designed for use with the previously<br>cleared titanium alloy components of the APEX Spine System<br>which can accept a 5.50mm spinal rod, including monoaxial,<br>polyaxial screws, hooks, and connectors. The polyaxial screw<br>washers are available in 2 sizes. The smaller washer size is for<br>use with the APEX Polyaxial Screws with shaft diameter from<br>4.75mm to 7.00mm and the larger washer size to be used with<br>the APEX Polyaxial Screws with shaft diameters 7.75mm and<br>8.50mm. |
| MATERIALS: | The APEX Spine System 5.50mm Titanium Rod and Polyaxial<br>Screw Washers is manufactured from implant grade titanium<br>Alloy (Ti-6Al-4V) conforming to ASTM standard F-136 and ISO<br>5832-3. |
| INDICATIONS FOR USE: | The APEX Spine System 5.50mm Titanium Rod and Polyaxial<br>Screw Washers is intended to provide immobilization and<br>stabilization of spinal segments in skeletally mature patients as<br>an adjunct to fusion in the treatment of the following acute and<br>chronic instabilities or deformities of the thoracic, lumbar, and<br>sacral spine: severe spondylolisthesis (grades 3 and 4) of the<br>L5-S1 vertebra; degenerative spondylolisthesis with objective |
{1}------------------------------------------------
evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudoarthrosis).
The APEX Spine System 5.50mm Titanium Rod and Polyaxial Screw Washers is also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.
The APEX Spine System 5.50mm Titanium Rod and Polyaxial Screw Washers is also a hook and sacral/iliac screw fixation system of the non-cervical spine indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the confirmed by by history and radiographic disc studies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudo-arthrosis).
The APEX Spine System 5.50mm Titanium Rod and Polyaxial Screw Washers when used with pedicle screws are indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies). Levels of fixation are for the thoracic, lumbar and sacral spine.
The APEX Spine System is substantially equivalent to the following predicate devices:
- I APEX Spine System: K062513 - SpineCraft
- 트 MONARCH Spine System: K010576, K024348 - DePuy AcroMed, Inc.
- . MOSS MIAMI Spinal System: K013296 - DePuy AcroMed, Inc.
The substantial equivalence of the APEX Spine System 5.50mm Titanium Rod and Polyaxial Screw Washers to the above mentioned predicate devices is based upon equivalence in design, material, manufacturing standards, intended use, as well as indications and contraindications. The fundamental scientific technology of this system is identical to previously cleared svstems.
Mechanical and dynamic testing of the APEX Spine System 5.50mm Titanium Rod and Polyaxial Screw Washers was performed. The test results demonstrate that the mechanical performance of the APEX Spine System 5.50mm Titanium Rod and Polyaxial Screw Washers is at least comparable to, if not better than, those of the predicate devices.
#### SUBSTANTIAL EQUIVALENCE:
PERFORMANCE DATA:
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with three overlapping wing-like shapes.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
SpineCraft, LLC % Ami Akallal-Asaad Director of Regulatory Affairs 2215 Enterprise Drive Westchester, Illinois 60154-5819
# MAR 1 5 2010
Re: K092825
Trade/Device Name: APEX Spine System 5.50mm Titanium Rod and Polyaxial Screw Washers Regulation Number: 21 CFR 888.3050, 888.3070 Regulation Name: Spinal interlaminal fixation orthosis Pedicle screw spinal system Regulatory Class: Class II Product Code: KWP, MNI, MNH Dated: February 17, 2010
Received: February 18, 2010
Dear Ami Akallal-Asaad:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
## Page 2 - Ami Akallal-Asaad
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please ) go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Barbara Brimm.
Mark N. Melkers Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indication for Use Statement
510(k) Number (if known): ______KO92825
### Device Name: APEX Spine System 5.5mm Rod and Polyaxial Screw Washers
### Indication for Use:
The APEX Spine System 5.5mm Rod and Polyaxial Screw Washers intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; spinal tumor; and failed previous fusion (pseudoarthrosis).
The APEX Spine System 5.5mm Rod and Polyaxial Screw Washers is also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.
The APEX Spine System 5.5mm Rod and Polyaxial Screw Washers is also a hook and sacral/liac screw fixation system of the non-cervical spine indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudo-arthrosis).
The APEX Spine System 5.5mm Rod and Polyaxial Screw Washers when used with pedicle screws are indicated for degenerative discogenic back pain with degeneration of the disc. confirmed by history and radiographic studies). Levels of fixation are for the thoracic, lumbar and sacral spine.
Prescription Use: (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use:
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
(Division Sigh-Off) (Division Sign of Surgical, Orthopedic, And Restorative Devices
K092825 510(k) Number_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.