← Product Code [ORC](/productcode/ORC) · K092801

# ABS BLOWER, ABS NOZZLE, MODEL 1001 (K092801)

_Nimbic Systems, LLC · ORC · May 20, 2011 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K092801

## Device Facts

- **Applicant:** Nimbic Systems, LLC
- **Product Code:** [ORC](/productcode/ORC.md)
- **Decision Date:** May 20, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.5070
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The Air Barrier System is a portable device for use in a surgical operating room that produces a directed, non-turbulent flow of air to the surgical site. The air flow from the device is HEPA-filtered to reduce the presence of particulate matter and microorganisms at the surgical site during hip arthroplasty. The ABS Nozzle is intended to be used only where: (1) it can be placed on an anatomical surface with no gap between the bottom of the nozzle and the surface, (2) the incision plane is parallel to the direction of airflow, and (3) the incision dimensions are within: 6″ (15.2 cm) in width and 20" (50.8 cm) in length. Device effectiveness may not be reliably detectable at a distance of 20 inches from the Nozzle, and effectiveness depreciates beyond this specified area.

## Device Story

Air Barrier System (ABS) provides localized air purification in surgical operating rooms. System consists of non-sterile, reusable Filter/Blower and sterile, single-use Air Delivery System (hose and nozzle). Filter/Blower draws ambient OR air through HEPA filter; filtered air travels through hose to nozzle. Nozzle placed directly on surgical drape adjacent to incision; delivers non-turbulent, HEPA-filtered air to surgical site. Used by surgical staff during hip arthroplasty. Reduces airborne particulate and microorganism counts at incision site; potentially lowers risk of surgical site contamination. Effectiveness limited to 20-inch range from nozzle.

## Clinical Evidence

Prospective clinical study (n=29) during total hip arthroplasty compared three groups: active ABS, sham (present but inactive), and control (no device). Airborne particulate and microorganism samples collected at 10-minute intervals. Results: Active ABS group showed significantly lower (P<0.001) particulate and microorganism counts compared to sham and control. Mean microorganism density: 1.60 CFU/m³ (active) vs 10.73 CFU/m³ (outside area). Mean particulate: 524 particles/ft³ (active) vs 3853 and 4092 particles/ft³ (sham/control).

## Technological Characteristics

System comprises HEPA-filtered blower and sterile, single-use air delivery hose/nozzle. Operates via directed, non-turbulent airflow. Materials, biocompatibility, and sterilization methods are equivalent to predicate. Bench testing included electrical safety, EMC, and microbial precipitation. No software algorithm or connectivity specified.

## Regulatory Identification

Air-handling apparatus for a surgical operating room is a device intended to produce a directed, nonturbulent flow of air that has been filtered to remove particulate matter and microorganisms to provide an area free of contaminants to reduce the possibility of infection in the patient.

## Special Controls

*Classification.* Class II (special controls). The device, when it is an air handling bench apparatus, an air handling room apparatus, or an air handling enclosure apparatus, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.

## Predicate Devices

- SIBS ([K831531](/device/K831531.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

### APPENDIX B

### 510(k) SUMMARY - K092801

### The 510(k) Summary is submitted per 21 CFR Section 807.92(c)

Submitter's Name: Nimbic Systems, Inc.

Submitter's Address: 4910 Wright Road, Suite 170, Stafford, Texas 77477

125 Telephone: 281-565-5715

Fax: 281-565-5712

Contact Person: Sean D. Self

Date Prepared: 1 April 2011

Device Trade Name: Air Barrier System

130 Device Common Name: Air Barrier System or ABS

Device Classification Name: Air Handling Apparatus for Surgical Operating Room

Device Classification: Class II

Summary of Substantial Equivalence: The design, materials, mode of operation and intended use features and performance of the ABS System is substantially equivalent with regard to features and performance of the predicate device, SIBS (K831531).

Device Description: The Air Barrier System (ABS) is used in the surgical operating room. The Air Barrier System (ABS) has two components: a Filter/Blower and an Air Delivery System. The Filter/Blower is nonsterile and reusable. The Air Delivery System is sterile, single-use. The ABS Filter/Blower captures and filters ambient air found in a typical surgical operating room through a High Efficiency Particle Arresting (HEPA)

- The HEPA filtered air leaves the Filter/Blower via the exit port where it is connected to the Hose 140 filter. portion of the sterile, single-use Air Delivery System. The Hose ends in the Nozzle portion of the Air Delivery System; the Nozzle is gently applied to the incision drape adjacent to a surgical incision. The Nozzle delivers HEPA filtered air directly to the surgical site area to reduce the presence of airborne warticulate and microorganisms.
#### 145 Indications for Use:

150

135

The Air Barrier System is a portable device for use in a surgical operating room that produces a directed, non-turbulent flow of air to the surgical site. The air flow from the device is HEPA-fittered to reduce the presence of particulate matter and microorganisms at the surgical site during hip arthroplasty. The ABS Nozzle is intended to be used only where: (1) it can be placed on anatomical surface with no gap between the bottom of the nozzle and the surface, (2) the incision plane is parallel to the direction of airflow, and (3) the incision dimensions are within: 6″ (15.2 cm) in width and 20" (50.8 cm) in length. Device effectiveness may not be reliably detectable at a distance of 20 inches from the Nozzle, and effectiveness depreciates beyond this specified area.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows a handwritten string of characters. The string starts with the letter 'K', followed by the numbers '092801'. The characters are written in a cursive style, with the numbers being slightly slanted. The image has a white background.

- Technological Characteristics: Comparisons of the ABS System and predicate device show that technological 155 characteristics such as materials, biocompatibility, mode of operation, performance properties, sterilization and packaging are substantially equivalent to the predicate device.
Non-clinical Performance Data: Bench testing included: medical electrical safety, electromagnetic compatibility, sterilization, biocompatibility, microbial precipitation testing, simulated use performance 160 testing, simulated worst-case use performance testing, cadaver surgical procedure testing, maximum air velocity testing in simulated and actual OR settings, turbulence testing at distances from simulated incision sites, and turbulence testing in clinical environment. Results of all bench and cadaver testing demonstrate the safety and effectiveness of the Air Barrier System and its substantial equivalence to the predicate device.

- Clinical Performance Data: A clinical study at a single investigator was conducted to demonstrate the performance characteristics of the Air Barrier System (ABS) during total hip arthroplasty 165 surgery. Twenty-nine (29) patients were randomized into one of three groups: with the ABS active (experiment), with the ABS present but not active (sham), and with no ABS device present (control). Airborne particulate and microorganism samples were collected simultaneously at a discrete single location at the surgical incision within the ABS area of effect and from a discrete single location within the sterile field,
- but not within the ABS area of effect. The airborne particulate and microorganism samples were the nt 170 ten-minute increments throughout the duration of the surgical procedures. The airborne particulate and microorganism counts observed in the experiment group (with the ABS device active) were significantly lower (P<0.001) than that observed in the sham and control groups. In the experiment group, with the ABS device active, the mean microorganism density at the discrete single sampling location within the ABS near S
- effect was 1.60 colony-forming units per cubic meter compared to 10.73 colony-forming units per cubic : 175 meter measured at a discrete single location outside the ABS area of effect. For particulate of size Sum sampled from single location points throughout each study group, the mean observed particulated for in the experiment group was 524 particles per cubic foot compared with 3853 and 4092 particles per cultic foot in the sham and control groups, respectively.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular fashion around the symbol. The caduceus is a common symbol associated with healthcare and medicine.

Food and Drug Administration 1 0903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Mr. Sean Self President Nimbic Systems, LLC 4910 Wright Road, Suite 170 Stafford, Texas 77477

# MAY 2 0 2011

Re: K092801

Trade/Device Name: Air Barrier System (ABS) Regulation Number: 21 CFR 878.5070 Regulation Name: Air-handing Apparatus for a Surgical Operating Room Regulatory Class: II Product Code: ORC Dated: March 25, 2010 Received: March 26, 2010

Dear Mr. Self:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Self

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Antony D. Natin

Anthony Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## APPENDIX A

Indications for Use

510(k) Number (if known): K092801 તેર

Device Name: _________________________________________________________________________________________________________________________________________________________________

Indications For Use:

100

The Air Barrier System is a portable device for use in a surgical operating room that produces a directed, non-turbulent flow of air to the surgical site. The air flow from the device is HEPA-filtered to reduce the presence of particulate matter and microorganisms at the surgical site during hip arthroplasty. The ABS Nozzle is intended to be used only where: (1) it can be placed on an anatomical surface with no gap between the bottom of the nozzle and the surface, (2) the incision plane is parallel to the direction of airflow, and (3) the incision dimensions are within: 6″ (15.2 cm) in width and 20" (50.8 cm) in length. Device effectiveness may not be reliably detectable at a distance of 20 inches from the Nozzle, and effectiveness depreciates beyond this specified area.

110

ਹੈ ਹਵ

115 Prescription Use _

Over-The-Counter Use _

(Per 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Elizabeth & Clavie-Williams

(Division Sigh-Off)
Division of Anesthesiology, General Hospital
Infection Control and Dental Devices
\$10(k) Number: K 092801

Page 6 of 8

# 000007

---

**Source:** [https://fda.innolitics.com/device/K092801](https://fda.innolitics.com/device/K092801)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
