← Product Code [DSH](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DSH) · K092785

# COMBINED VASOMEDICAL-BIOX AMBULATORY ECG & BLOOD PRESSURE RECORDER & ANALYSIS SOFTWARE, 2301 (K092785)

_Vasomedical, Inc. · DSH · Apr 2, 2010 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DSH/K092785

## Device Facts

- **Applicant:** Vasomedical, Inc.
- **Product Code:** [DSH](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DSH.md)
- **Decision Date:** Apr 2, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2800
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Software as a Medical Device, Pediatric

## Indications for Use

Vasomedical-Biox Combined Ambulatory ECG and Blood Pressure Recorder is a Non-Invasive oscillometric device intended to acquire Ambulatory 3 Channel ECG signals and non- invasive Blood Pressure signals from the upper body surfaces. Cardiac rhythm is acquired via 3 Channel ECG signals. The Recorders are intended for adults and children Who are over the age of six years. Vasomedical-Biox Ambulatory ECG CB Series Analysis System Software allows transfer of ECG and Blood Pressure data from the Recorder to a Windows based PC-based computer program via a removable and large capacity storage card (SD) for the purpose of creating reports and printouts. The Software does not perform diagnostics. Physicians carry out diagnostic evaluations of this data. The system is only for measurement, recording and display. It makes no diagnosis.

## Device Story

Portable, microprocessor-based recorder worn by patient; acquires 3-channel ECG signals and oscillometric blood pressure measurements via upper body sensors. Data stored on SD memory card; transferred to Windows-based PC via infrared adaptor or card reader. Analysis software provides full disclosure for arrhythmia and ST events, synchronous ECG/ABP pulses, summary tables, statistics, and trends. Used by trained medical personnel to assist in diagnosis; software does not perform automated diagnostics. Benefits include continuous cardiac and blood pressure monitoring for up to 24 hours to aid clinical decision-making.

## Clinical Evidence

No clinical data provided; device relies on substantial equivalence to predicate devices.

## Technological Characteristics

Portable microprocessor-based recorder; 3-channel ECG; oscillometric BP measurement; 5 or 7 electrodes; 256 Hz/channel sample rate; 2048 Hz pacemaker detection; 10-bit resolution; 512 MB SD card storage; 4 AA alkaline batteries; LCD display; PC-based analysis software.

## Regulatory Identification

A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.

## Predicate Devices

- SunTech Medical Instruments Inc., Automatic Blood Pressure Measurement System, Oscar 2 ([K003004](/device/K003004.md))
- Vasomedical-Biox 1305, 3 Channel ECG Holter Monitor and CB Series Analysis Software ([K083820](/device/K083820.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 22 March, 2010

K092785 1/3

## 510(k) Summary

### Vasomedical, Incorporated

Vasomedical-Biox Combined Ambulatory ECG and Blood Pressure Recorder 2301 and Ambulatory ECG and Blood Pressure Recorder Analysis Software

- 1. Date Prepared: 22 March, 2010 2. Submitter's Name: Vasomedical, Inc. 180 Linden Ave. and Address Westbury, NY 11590 Richard Gordon 3. Contact Person: Director, Regulatory and Quality Affairs Vasomedical, Incorporated Telephone: (516) 997-4600 Facsimile: (516) 997-2299 E-mail: rgordon@vasomedical.com Combined Ambulatory ECG and Blood Pressure Recorder 4. Device Name: Proprietary Names: Vasomedical-Biox Combined Ambulatory ECG and Blood Pressure Recorder 2301 and Analysis Software Common Name: Combined Ambulatory ECG and Blood Pressure Recorder 2301 Classification Name: 870.2800 Medical Magnetic Tape Recorder 870.1130 Non invasive Blood Pressure Measurement System 5. Predicate Device: The AECG/BP Recorder listed above is substantially equivalent to the SunTech Medical Instruments Inc., Automatic Blood Pressure Measurement System, Oscar 2 and the Vasomedical-Biox 1305, 3 Channel ECG Holter Monitor and CB Series Analysis Software.
Vasomedical's Ambulatory ECG and Blood Pressure Recorder is 6. Device Description: intended to be used as a combined Holter Ambulatory Electrocardiograph device and a non-invasive Ambulatory Blood Pressure Monitor for the purpose of screening ECG rhythms and blood pressure measurements for periods up to 24 hours. Blood Pressure measurements are obtained using oscillometric signals at intervals set by the physician or on demand. Cardiac rhythm is acquired by 3 channel ECG signals. The Recorders are intended for adults and children over the age of six years old.

25, 2000 (K003004), and April 1, 2009 (K083820).

FDA granted 510(k) clearances for these predicate devices on October

APR - 2 2010

11

{1}------------------------------------------------

K092785 2/3

The CB series AECG/BP Analysis system is a PC based diagnostic application running on a Microsoft Windows operating system. It is designed in conjunction with the CB series Holter Recorder and analyzes prerecorded patient's ECG and/or ABP data that has been stored by CB series Recorder, or other compatible Holter Recorders. The system provides multi-channel full disclosure for arrhythmia and ST events and synchronous three channel ECG and ABP pulses, and creates summary tables, statistics, trends, and final report regarding a variety of cardiac data indices. The cardiac data provided by the system is used by trained medical personnel to assist in the diagnosis of patients with various rhythm patterns.

The Recorders are portable, microprocessor based devices that are worn by a patient with the use of a carrying case and strap.

| Lead                    | 3               | Dimensions          | 4.88x2.68x1.22"                              |
|-------------------------|-----------------|---------------------|----------------------------------------------|
| Electrode               | 5 or 7          | Weight              | 6.35 oz. w/o batteries                       |
| Lead wire type          | Lead-7C/5C      | Storage             | SD Memory Card                               |
| Display                 | LCD             | Memory              | 512 MB or4 more                              |
| Batteries               | 4 "AA" alkaline | Sample Rate         | 256 Hz/Channel<br>2048Hz Pacemaker Detection |
| BP Cuff                 | Medium          | Resolution          | 10 Bits                                      |
| Carrying Case/<br>Strap | Supplied        | Infrared<br>Adaptor | Supplied                                     |
| Card Reader             | Supplied        |                     |                                              |

Model 2301 specifications are listed in Table 1 below:

#### Table 1: Model CB2301 Specifications

7. Intended Use:

Vasomedical-Biox Combined Ambulatory ECG and Blood Pressure Recorder, #2301, is a Non-Invasive device intended to acquire Ambulatory 3 Channel ECG signals and non- invasive Oscillometric Blood Pressure signals from the upper body surfaces. Cardiac rhythm is acquired via 3 Channel ECG Signals. The Recorders are intended for adults and children who are over the age of six years.

Vasomedical-Biox Ambulatory ECG CB Series Analysis System Software allows transfer of ECG and Blood Pressure data from the Recorder to a Windows based PC program via a removable and

{2}------------------------------------------------

13

K092785 3/3

large capacity storage card (SD) for the purpose of creating reports

and printouts. The Software does not perform diagnostics. Physicians carry out diagnostic evaluations of this data.

The system is only for measurement, recording and display. It makes no diagnosis.

Refer to Attachment I, Vasomedical-Biox Model 2301 AECGBP Operator's Manual, Sections 2.2, Indications for Use, 2.2.1 Intended Use and 2.2.2 for Contraindications.

- Technological and functional characteristics of the 8. Comparison of device listed in this 510(k) Premarket Notification is Technological essentially the same as those of the predicate devices. The device Characteristics: listed in this 510(k) Premarket Notification is therefore substantially equivalent to the predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its head facing left and its wings forming three curved lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002

APR - 2 2010

Vasomedical, Inc. c/o Mr. Richard Gordon Manager, Regulatory and Quality Affairs 180 Linden Ave Westbury, NY 11590

Re: K092785

Vasomedical-Biox 2301 Combined Ambulatory ECG and Blood Pressure Recorder Regulation Number: 21 CFR 870.2800 Regulation Name: Medical Magnetic Tape Recorder Regulatory Class: Class II (two) Product Code: DSH and DXN Dated: March 22, 2010 Received: March 24, 2010

Dear Mr. Gordon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Richard Gordon

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

| 510(k) Number:                                                              | k092785                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                    |
|-----------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|
| Device Name:                                                                | Vasomedical-Biox Combined Ambulatory ECG and Blood<br>Pressure Recorder 2301 and Analysis Software                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                    |
| Indications for Use:                                                        | Vasomedical-Biox Combined Ambulatory ECG and Blood<br>Pressure Recorder is a Non-Invasive oscillometric device<br>intended to acquire Ambulatory 3 Channel ECG signals and<br>non- invasive Blood Pressure signals from the upper body<br>surfaces. Cardiac rhythm is acquired via 3 Channel ECG<br>signals. The Recorders are intended for adults and children<br>Who are over the age of six years.<br><br>Vasomedical-Biox Ambulatory ECG CB Series Analysis<br>System Software allows transfer of ECG and Blood Pressure<br>data from the Recorder to a Windows based PC-based<br>computer program via a removable and large capacity<br>storage card (SD) for the purpose of creating reports and<br>printouts. The Software does not perform diagnostics.<br>Physicians carry out diagnostic evaluations of this data.<br>The system is only for measurement, recording and display.<br>It makes no diagnosis. |                                                    |
| Prescription Use: YES<br>(Part 21 CFR 801 Subpart D)                        | AND/OR                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Over-The-Counter Use: NO<br>(21 CFR 801 Subpart C) |
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF<br>NEEDED) |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                    |

Concurrence of CDRH, Office of Device Evaluation (ODE)

Signature

Page 1 of 1

.

9

(Division Sign-Off) Division of Cardiovascular Devices

510(k) Number_

.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DSH/K092785](https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DSH/K092785)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
