← Product Code [GEI](/productcode/GEI) · K092761

# PROPLATE, MODELS 9641C, 9641F, P9641C, P9641F, 9642C, 9642F AND P9642C (K092761)

_Bio Protech, Inc. · GEI · Nov 24, 2009 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K092761

## Device Facts

- **Applicant:** Bio Protech, Inc.
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Nov 24, 2009
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The PROPLATE is a disposable, single-use, neutral electrode, which provides a return path for high frequency electrical current to the ESU device.

## Device Story

PROPLATE is a disposable, single-use neutral electrode (grounding plate) used during electrosurgical procedures. It functions as a return path for high-frequency electrical current from the patient back to the electrosurgical unit (ESU). The device is applied to the patient's skin to safely dissipate current, preventing thermal injury at the site. It is used in clinical settings (e.g., OR) by surgeons or surgical staff. The device ensures the integrity of the electrosurgical circuit, allowing for safe cutting and coagulation. It benefits the patient by minimizing the risk of electrical burns during surgery.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

Disposable, single-use neutral electrode. Conductive material interface for electrical current return. Designed for compatibility with standard electrosurgical units (ESU). Passive device; no energy source or software.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized bird or human figure, with flowing lines representing movement or connection.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

NOV 2 4 2009

Bio Protech, Inc. % Advena, Ltd. Ms. Judy Burton Director of US Operations 2626 Valley View Lane, Suite 4 Dallas, Texas 75234-6274

Re: K092761

Trade/Device Name: PROPLATE Electrosurgical Grounding Plate Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: October 21, 2009 Received: October 27, 2009

Dear Ms. Burton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{1}------------------------------------------------

Page 2 - Ms. Judy Burton

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

http://www.fda.gov/MedicalDevices/ResourceforYou/Industry/default.htm.

Sincerely yours.

Sincerely yours,

Mark A. Milken

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

ATTACHMENT 2

INDICATIONS FOR USE STATEMENT

510(k) Number: K092761

Device Name: PROPLATE Electrosurgical Grounding Plate

Indications for Use: The PROPLATE is a disposable, single-use, neutral electrode, which provides a return path for high frequency electrical current to the ESU device.

FOR M. MELKERSON

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K092761

Prescription Use: X (per 21 cfr 801.809) AND/OR

Over the Counter Use _

PLEASE DO NOT WRITE BELOW THIS LINE

Concurrence of CDRH, Office of Device Evaluation (ODE)

---

**Source:** [https://fda.innolitics.com/device/K092761](https://fda.innolitics.com/device/K092761)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
