← Product Code [HSB](/productcode/HSB) · K092748

# SURESHOT TAN NAILS AND ACCESSORIES (K092748)

_Smith & Nephew, Inc. · HSB · Apr 26, 2010 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K092748

## Device Facts

- **Applicant:** Smith & Nephew, Inc.
- **Product Code:** [HSB](/productcode/HSB.md)
- **Decision Date:** Apr 26, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3020
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

SURESHOT™ TAN Nails are indicated for fractures of the femur including simple long bone fractures, severely comminuted, spiral, large oblique and segmental fractures; nonunions and malunions; polytrauma and multiple fractures; prophylactic nailing of impending pathologic fractures; reconstruction, following tumor resection and grafting; supracondylar fractures; bone lengthening and shortening; and for fixation of fractures that occur in and between the proximal third and distal fourth of the femur. In addition, SURESHOT™ TAN Nails contain holes/slots proximally to accept screws that thread into the femoral head for compression and rotational stability and are indicated for the following: subtrochanteric fractures; intertrochanteric fractures; ipsilateral femoral shaft/neck fractures; and intracapsular fractures. SURESHOT™ TAN Nails and Accessories are for single use only.

## Device Story

Intramedullary nail system for femoral fracture fixation; includes nails, lag screws, compression screws, and set screws. Titanium components implanted surgically to provide structural support, compression, and rotational stability to fractured bone. Used by orthopedic surgeons in clinical/OR settings. Provides mechanical stabilization to facilitate bone healing; prevents further displacement of fractures; supports reconstruction post-tumor resection.

## Clinical Evidence

Bench testing only. Evidence includes finite element analysis, construct fatigue testing of nail constructs with accessories, and axial pullout testing of lag screws from test media. No clinical data provided.

## Technological Characteristics

Intramedullary nails and accessories (lag, compression, set screws) manufactured from titanium. Dimensions: nails 9-13mm diameter, 32-42cm length; screws vary by type. Mechanical fixation device; non-powered. Single-use.

## Regulatory Identification

An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.

## Predicate Devices

- TriGen™ Titanium Intramedullary Nail System ([K981529](/device/K981529.md))
- TriGen™ InterTAN Nail ([K040212](/device/K040212.md))
- TriGen™ Trochanteric Antegrade Nail ([K040462](/device/K040462.md))
- TriGen™ Meta-Nails Retrograde Femoral and Tibia Nails ([K061019](/device/K061019.md))
- Asian Intramedullary Hip Screw (IMHS) ([K050226](/device/K050226.md))
- T2 Femoral Nail ([K010801](/device/K010801.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K092748

## 510(k) Summary of Safety and Effectiveness SURESHOT™ TAN Nails and Accessories

| Submitted By:                       | Smith & Nephew, Inc.<br>Orthopaedics<br>1450 Brooks Road<br>Memphis, TN 38116                            | APR 8 3 2010 |
|-------------------------------------|----------------------------------------------------------------------------------------------------------|--------------|
| Date:                               | September 4, 2009                                                                                        |              |
| Contact Person:                     | David Henley, Regulatory Affairs Project Manager<br>Tel: (901) 399-6487 Fax: (901) 398-5146              |              |
| Proprietary Name:                   | SURESHOT™ TAN Nails and Accessories                                                                      |              |
| Common Name:                        | Intramedullary Nails and Accessories                                                                     |              |
| Classification Name and Reference:  | 21 CFR 888.3030, Single/multiple component metall<br>bone fixation appliances and accessories - Class II |              |
| Device Product Code and Panel Code: | JDS / Orthopedics / 87                                                                                   |              |

#### Device Description:

The subject devices are comprised of intramedullary nails and accessories such as lag screws, compression screws and set screws for use in the femur. All described components are manufactured from titanium material. The subject devices are available in the following size ranges:

| Device Type        | Available Range of Diameters | Available Length Range |
|--------------------|------------------------------|------------------------|
| Nails              | 9mm - 13mm                   | 32cm - 42cm            |
| Lag Screws         | 8.64mm (one diameter only)   | 70mm - 125mm           |
| Compression Screws | 6.22mm (one diameter only)   | 30mm - 120mm           |
| Set Screws         | 6.6mm (one diameter only)    | 25.4mm - 39.3mm        |

#### Intended Use:

SURESHOT™ TAN Nails are indicated for fractures of the femur including simple long bone fractures, severely comminuted, spiral, large oblique and segmental fractures; nonunions and malunions; polytrauma and multiple fractures; prophylactic nailing of impending pathologic fractures; reconstruction, following tumor resection and grafting; supracondylar fractures; bone lengthening and shortening; and for fixation of fractures that occur in and between the proximal third and distal fourth of the femur.

In addition, SURESHOT™ TAN Nails contain holes/slots proximally to accept screws that thread into the femoral head for compression and rotational stability and are indicated for the following: subtrochanteric fractures; intertrochanteric fractures; ipsilateral femoral shaft/neck fractures; and intracapsular fractures.

#### Technological Characteristics:

Components comprising SURESHOT™ TAN Nails and Accessories are very similar to legally marketed devices listed below in that they share very similar indications for use (compared to devices cleared under K981529, K040212, K040462, K061019, K050226 and K010801), are manufactured from identical material, and incorporate very similar technological design characteristics.

p. 1 of 2

{1}------------------------------------------------

### Substantial Equivalence Information:

When compared to the predicate devices listed below, substantial equivalence is based on similarities with regard to overall indications for use, material composition, and technological design characteristics.

- Titanium Nail System (TriGen™ Titanium Intramedullary Nail System) K981529
- · TriGen™ InterTAN Nail K040212
- · TriGen™ · Trochanteric Antegrade Nail K040462
- · TriGen™ Meta-Nails Retrograde Femoral and Tibia Nails K061019
- · Asian Intramedullary Hip Screw (IMHS) K050226
- · T2 Femoral Nail (Stryker/Howmedica/Osteonics Corp.) K010801

To further support a determination of substantial equivalence, various types of pre-clinical testing were conducted on the subject devices in comparison against one or more of the previously cleared predicate devices described above. The types of pre-clinical testing included: finite element analysis, construct fatigue testing of a nail construct with appropriate accessory devices and axial pullout testing of the lag screw from appropriate test media.

p. 2 of 2

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter of the circle. In the center of the circle is an abstract symbol that resembles an eagle or other bird in flight. The symbol is composed of three curved lines that form the body and wings of the bird, and a small, curved line that forms the head.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

## APR 2 3 2010

Smith and Nephew Inc. % Mr. David Henley Regulatory Affairs Project Manger 1450 Brooks Road Memphis, Tennessee 38116

Re: K092748

Trade/Device Name: SURESHOT™ TAN Nails and Accessories Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: II Product Code: HSB, HWC Dated: February 24, 2010 Received: February 25, 2010

Dear Mr. Henley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

· ·

{3}------------------------------------------------

Page 2 - Mr. David Henley

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFl)A/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Chubur Budup

Mark N. Melkers VDirector Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Premarket Notification Indications for Use Statement

K092748 510(k) Number (if known): __

Device Name: SURESHOT™ TAN Nails and Accessories

Indications for Use:

SURESHOT™ TAN Nails are indicated for fractures of the femur including simple long bone fractures, severely comminuted, spiral, large oblique and segmental fractures; bono naotaroo, covery connytrauma and multiple fractures; prophylactic nailing of impending pathologic fractures; reconstruction, following tumor resection and grafting; imponding patifores; bone lengthening and shortening; and for fixation of fractures that occur in and between the proximal third and distal fourth of the femur.

In addition, SURESHOT™ TAN Nails contain holes/slots proximally to accept screws that in addition, OOT(2017) head for compression and rotational stability and are indicated for the following: subtrochanteric fractures; intertrochanteric fractures; ipsilateral femoral shaft/neck fractures; and intracapsular fractures.

SURESHOT™ TAN Nails and Accessories are for single use only.

AND/OR Over-the-Counter Use Prescription Use XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX (Part 21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D)

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Onetu for mxn
(Division Sign 20)

Division Sign-Q Division of Surgical. Orthopedic. and Restorative Devices

510(k) Number K092748

---

**Source:** [https://fda.innolitics.com/device/K092748](https://fda.innolitics.com/device/K092748)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
