K092744 · R & D Medical Products, Inc. · DRX · Dec 10, 2009 · Cardiovascular
Device Facts
Record ID
K092744
Device Name
NEONATAL ECG ELECTRODE, M203KEN
Applicant
R & D Medical Products, Inc.
Product Code
DRX · Cardiovascular
Decision Date
Dec 10, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2360
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Neonatal ECG Electrodes are intended for use in pediatric or neonatal electrocardiographic procedures where resting ECG monitoring is deemed necessary and is ordered by a physician. Such procedures include, in particular, patient ECG surveillance and ECG diagnosis recording. The Neonatal ECG Electrodes are applied to the surface of the body, single use only, and disposable. Neonatal ECG electrodes should be changed every 24 hours.
Device Story
Device consists of three prewired ECG electrodes on a single release card. Construction includes a surface layer (tricot/polyester fabric, polyethylene foam, or polypropylene), a metallic Ag/AgCl coated layer for leadwire connection, and a biocompatible conductive hydrogel layer. Electrodes are applied to the neonate's chest for resting ECG monitoring and diagnostic recording. Used in hospital neonatal/pediatric units; operated by clinical staff. Output is an electrical signal transmitted via leadwires to an ECG monitor for clinical surveillance and diagnosis. Benefits include non-invasive, disposable monitoring of infant cardiac activity.
Clinical Evidence
Bench testing only. No clinical data. Performance testing verified compliance with ANSI/AAMI EC12:2000 (R) 2005 and ANSI/AAMI EC53:1995 (R) 2008 standards, including AC impedance, DC offset voltage, defibrillation overload recovery, combined offset instability, internal noise, and bias current tolerance. Biocompatibility testing (cytotoxicity, sensitization, primary skin irritation) confirmed safety for skin contact.
Technological Characteristics
Multi-layer construction: surface layer (tricot/polyester, PE foam, or polypropylene), Ag/AgCl coated metallic layer, and conductive hydrogel. Complies with ANSI/AAMI EC12:2000/(R) 2005 and ANSI/AAMI EC53:1995/(R) 2008. Passive sensing device; no energy source or software.
Indications for Use
Indicated for resting ECG monitoring of newborn and premature infants in neonatal or pediatric hospital units. Single-use, disposable, applied to body surface, changed every 24 hours.
Regulatory Classification
Identification
An electrocardiograph electrode is the electrical conductor which is applied to the surface of the body to transmit the electrical signal at the body surface to a processor that produces an electrocardiogram or vectorcardiogram.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9. The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Electrocardiograph Electrodes.” See § 870.1(e) for availability information of guidance documents.
{1}------------------------------------------------
KD912744
1/2
# 510(k) Summary
.
.
| Submitter's Information: | R&D Medical Products, Inc.<br>20492 Crescent Bay Drive, Building 106<br>Lake Forest, CA 92630 |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | James Perrault<br>Phone: (949) 472-9346 x202<br>Fax: (949) 472-9347 |
| Date Prepared: | September 3, 2009, revised November 19, 2009 |
| | DEC 10 2009 |
| Proprietary Name: | Neonatal ECG Electrodes |
| Common Name: | Electrocardiograph (ECG) Electrodes |
| Classification Name: | Electrodes, Electrocardiograph |
| Regulation: | Electrocardiographic electrode, 21 C.F.R. § 870.2360 |
| Regulatory Class: | Class II |
| Product Code: | DRX |
| Predicate Devices: | K050443 - Ameritus Accu-Lead of Kentec Medical, Inc.<br>K011564 - Neolead of Neotech Products, Inc. |
| Description of Device: | The Neonatal ECG Electrodes are three prewired ECG electrodes<br>supplied on the same release card in a single package. The<br>electrodes are a multi-layer construction containing a first layer<br>surface (made of tricot/polyester fabric, polyethylene foam, or<br>polypropylene substrate), a second layer ( metallic with Ag/AgCl<br>coating ) prewired for connection to patient leadwires, and a third<br>layer (made of biocompatible conductive hydrogel coupling media)<br>The electrodes are placed on the neonate patient's chests. |
| Intended Use: | Neonatal ECG Electrodes are intended for use in pediatric or<br>neonatal electrocardiographic procedures where resting ECG<br>monitoring is deemed necessary and is ordered by a physician.<br>Such procedures include, in particular, patient ECG surveillance<br>and ECG diagnosis recording. The Neonatal ECG Electrodes are |
{2}------------------------------------------------
જેર
applied to the surface of the body, single use only, and disposable. Neonatal ECG electrodes should be changed every 24 hours.
Technological Comparison:
The Neonatal ECG Electrodes have technological characteristics that are substantially equivalent to those of the predicate devices as determined by testing, common materials, and common manufacturer. The following testing was conducted: AC impedance: DC offset voltage: defibrillation overload recovery; combined offset instability and internal noise; and bias current tolerance. The Neonatal ECG Electrodes and the predicate devices all meet the specifications as established in ANSI/AAMI EC12:2000/(R) 2005 and ANSI/AAMI EC53:1995/(R) 2008.
R&D Medical Products, Inc. contract manufactures the predicate Basis for Equivalence: devices and now seeks to make the same product under private label for distribution.
-Performance testing: Biocompatibility testing was performed, and the device passed the required skin sensitivity testing criteria. According to the performance data, the Neonatal ECG Electrodes met specifications as established in ANSI/AAMI EC12:2000 (R) 2005 for skin contact. The tests included cytotoxicity, sensitization and primary skin irritation tests. The predicate devices use the same materials and meet the same biocompatibility specifications.
> Bench testing demonstrated that the characteristics of the Neonatal ECG Electrodes are substantially equivalent to those of the predicate devices.
-Labeling:
Conclusions from Testing:
The labeling of the Neonatal ECG Electrodes is substantially equivalent to that of the predicate devices.
In all material respects, the Neonatal ECG Electrodes are substantially equivalent to the predicate devices. Testing was performed according to FDA recognized standards. Test results support the conclusion that the electrical output is substantially equivalent to the predicate devices, and there are no differences in construction and materials between the devices to pose new questions of safety or effectiveness.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image is a circular seal for the Department of Health & Human Services - USA. The seal features the department's logo, which is a stylized caduceus with three figures representing health, services, and humanity. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged around the top of the circle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
DEC 1 0 2009
R&D Medical Products, Inc. c/o Mr. James Perrault 20492 Crescent Bay Drive Building 106 Lake Forest, CA 92630
Re: K092744
Trade Name: Neonatal ECG Electrode, M203KEN Regulation Number: 21 CFR § 870.2360 Regulation Name: Electrocardiograph Electrode Regulatory Class: Class II (two) Product Code: 74 DRX Dated: November 23, 2009 Received: November 24, 2009
#### Dear Mr. Perrault:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
### Page 2 - Mr. James Perrault
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
W.M.P.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
.
### 510(k) Number: K092744
Device Name: Neonatal ECG Electrodes
Indications for Use:
These electrodes are to be used in the neonatal or pediatric unit of hospitals for the resting ECG monitoring of newborn and premature birth infants. They are applied to the surface of the body, single use only, and disposable. These electrodes should be changed every 24 hours.
| Prescription Use (Part 21 CFR 801 Subpart D) | X |
|----------------------------------------------|---|
| AND/OR | |
| Over-The-Counter Use (21 CFR 801 Subpart C) | |
(Division Sign-Off)
Division of Cardiovascular Devices
| 510(k) Number | K092744 |
|---------------|---------|
|---------------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.