K092674 · Pan Global Implant · DZE · Dec 28, 2009 · Dental
Device Facts
Record ID
K092674
Device Name
SIMPLER TOADSTOOL MINI IMPLANT
Applicant
Pan Global Implant
Product Code
DZE · Dental
Decision Date
Dec 28, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Simpler Toadstool Mini Implant is intended to be used as an artificial root to provide immediate long term or transitional support for overdentures.
Device Story
Simpler Toadstool Mini Implants are narrow-diameter (2.5mm) endosseous dental implants used as artificial roots for overdenture support. Manufactured from 6/4 titanium alloy, they feature a one-piece design with a flattened, toadstool-shaped abutment and hexagonal top. Implants are grit-blasted and acid-etched; some models include a hydroxylapatite (HA) coating on the threaded portion. During clinical use, the hex ball abutment fits into a keeper with a rubber o-ring within the prosthesis to reduce loading and improve retention. Used by dental professionals in surgical settings, these implants provide a minimally invasive alternative to conventional implants for stabilizing soft tissue-supported overdentures in edentulous or partially edentulous patients. They support immediate loading when primary stability is achieved, benefiting patients by providing stable, removable prosthetic support.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing and comparison to predicate devices regarding materials, surface treatment, and fatigue performance.
Technological Characteristics
Narrow diameter (2.5mm) endosseous dental implants. Material: 6/4 Titanium alloy (90% Ti, 6% Al, 4% V). Surface: Grit-blasted and acid-etched; optional hydroxylapatite (HA) coating on threads. Design: One-piece with flattened toadstool-shaped abutment and hexagonal top. Lengths: 10mm, 13mm, 15mm, 18mm. Mechanical retention via rubber o-ring in prosthesis.
Indications for Use
Indicated for long-term intra-bony support of soft tissue-supported overdentures in partial or fully edentulous patients (mandible/maxilla). Suitable for immediate loading given good primary stability and appropriate occlusal load. Also indicated for temporary support. Contraindications: local infection, vascular impairment, uncontrolled diabetes, chronic high-dose steroids, clotting disorders, anticoagulant therapy, metabolic bone disease, or systemic disorders affecting bone/wound healing.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
172
3. SUMMARY
K092674
510(K) 092674
Dr. Harold Bergman, President Pan Global Implant Corp. Simpler Implants #404 1023 Wolfe Ave. DEC 2 8 2009 Vancouver, BC V6H 1V6 Phone: 604 736 9890 Fax: 604-736-9747 Contact: Karen Bergman. QA Assurance Officer or Dr. Bergman, President
Trade Name: Simpler Toadstool Mini Implant & Simpler Toadstool HA Mini Implant Common Name: Simpler Toadsfool Mini Implants & Simpler Toadstool Mini Implant Classification: Endosseous Dental Implants
Leaally marketed devices to which S.E. is claimed
K083886: Simpler Mini Implants K031106 : Imtec Sendax MDI and MDI plus
K050586: Leone Implant System
K70601: Mini Drive-Lock Implant
Description The Simpler Toadstool Minis and the Simpler HA Toadstool Minis are Narrow Diameter implants. They are manufactured using 6/4 Titanium alloy 90% titanium 6% aluminum and 4% vanadium for strength. All implants are arit blasted and also acid etched and are available uncoated or coated with hydroxylapatite only on the threaded part of the implant. The diameter is 2.5mm and the lenaths are 10mm. 13mm, 15 mm and 18 mm in both the coated and uncoated implants. The implants are designed as one piece with a flattened toadstool shaped abutment with a hexagonal outline at the top of the implant. The flattened, hex ball fits into a keeper with a rubber o-ring placed in the prosthesis to reduce loading and to help with retention.
Intended Use: The Simpler Toadstool Mini Implant is intended to be used as an artificial root to provide immediate long term or transitional support for overdentures.
Indications for Use: The Simpler Toadstool Mini Implants and the Simpler HA Toadstool Mini Implants are intended to provide long term intra-bony applications for Soft Tissue Supported Over Dentures only. They are designed for immediate loading when there is good primary stability and an appropriate occlusal load. These implants may also be used for
{1}------------------------------------------------
K092674
temporary soft tissue overdenture support for partial and fully edentulous restoration in the mandible and maxilla. They may be used as a minimally invasive surgical technique option for full or partial edentuloism as an alternative to more invasive surgery involving conventional implants
Contraindications: Contraindications customary to the placement of any dental implants may be observed. These include, but are not limited to current local infection, vascular impairment, uncontrolled diabetes, chronic high doses of steroids, clotting disorders, current anticoagulant therapy, metabolic bone disease, and other metabolic or systemic disorders which will affect bone or wound healing. Excessive loading or placement of implants in inadequate bone may result in fracture.
Complications: Possible complications with any oral reconstructive surgery include, infection, closure perforation, abscess formation, bone loss, pain soft tissue irregularities and additional complications associated with anesthesia and dental surgery.
Summary of Testing: Simpler Toadstool Mini Implants do not introduce new issues for materials, surface treatment, fatigue testing and risk management that have not been addressed in all other approved Simpler Implants.
{2}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
Dr. Harold Bergman Chief Executive Officer Pan Global Implant Corporation #404 1023 wolfe Avenue Vancouver, British Columbia CANADA V6H 1V6
DEC 2 8 2009
Re: K092674
Trade/Device Name: Simpler Toadstool Mini Implant, Simpler HA Toadstool Mini Implant Regulation Number: 21CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE Dated: December 17, 2009 Received: December 22, 2009
Dear Dr. Bergman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2- Dr. Bergman
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free. number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Cinchin 0.2m
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K092674
( 1)
## 510(k) K092674
Device Name: Simpler Toadstool Mini Implant Simpler HA Toadstool Mini Implant
## Indications for Use:
The Simpler Toadstool Mini Implants and the Simpler HA Toadstool Mini Implants are intended to provide long term intra-bony applications for Soft Tissue Supported Over Dentures only. They are designed for immediate loading when there is good primary stability and an minimal occlusal load. These implants may also be used for temporary soft tissue overdenture support for partial and fully edentulous restoration in the mandible and maxilla. They may be used as a minimally invasive surgical technique option for full or partial edentulism as an alternative to more invasive surgery involving conventional implants.
Prescription Use X AND/OR Part 21 CFR 801 Subpart D)
Over the Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Cunten her
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
4992674 510(k) Number:
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.