POGS, MODEL 65/130

K092636 · On Site Gas Systems, Inc. · CAW · Jan 7, 2010 · Anesthesiology

Device Facts

Record IDK092636
Device NamePOGS, MODEL 65/130
ApplicantOn Site Gas Systems, Inc.
Product CodeCAW · Anesthesiology
Decision DateJan 7, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5440
Device ClassClass 2
AttributesTherapeutic, Real-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K092636 · Jan 7, 2010POGS, MODEL 65/130On Site Gas Systems, Inc.Military operational field experience/feedbackReal-world operational data from the predicate device (POGS 33C) used in military settings globally informed the design improvements and functional upgrades for the POGS 65/130.Military deployment; Operational feedback; Predicate device performance

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Military operational use of POGS 33C; Retrospective observational/operational feedback; Study Period: Prior to 2009Military personnel in global deployment settings; Number of Sites: Multiple (global military settings)Not applicable for this studySystem reliability, oxygen output requirements, maintenance needs

Indications for Use

The POGS 65/130 is intended to generate and deliver USP 93% supplemental oxygen. This device is intended to be used only by trained personnel in disaster relief and emergency preparedness situations, military settings or where bottled oxygen is not readily available.

Device Story

POGS 65/130 is a modular, portable oxygen generation system using Pressure Swing Absorption (PSA) with molecular sieves to separate oxygen from ambient air. Input: filtered ambient air; Output: USP 93% oxygen at 65-130 LPM (50-60 PSI). System comprises feed air compressors, oxygen generators, and a boost/distribution center. Used in disaster, emergency, or military settings by trained personnel. Output connects to hospital distribution systems, ventilators, anesthesia machines, or cylinder filling accessories. Hospital distribution has priority. PLC-controlled for maintenance and troubleshooting. Benefits include reliable oxygen supply in austere environments where liquid oxygen is unavailable.

Clinical Evidence

Bench testing only. No clinical data provided. Performance validated using ASTM F-1464-93 standards, focusing on oxygen purity and system output capabilities.

Technological Characteristics

Pressure Swing Absorption (PSA) technology; MDX molecular sieve; PLC-based control (Allen-Bradley); modular design (compressors, generators, boost/distribution center); output 65-130 LPM at 50-60 PSI; USP 93% oxygen purity. Complies with ASTM F-1464-93.

Indications for Use

Indicated for patients requiring supplemental oxygen in disaster relief, emergency, or military settings where bottled oxygen is unavailable. Intended for use by trained personnel only.

Regulatory Classification

Identification

A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ | | <i>O<sub>2</sub>N<sub>2</sub></i> <b>SITE</b> | | |--|------------------------------------------------|----------------------| | | <b>On Site Gas Systems, Inc.</b> | | | | Manufacturers / Designers of Oxygen & Nitrogen | Generating Equipment | | 5. 510(k) Summary | JAN - 7 2010 | |-------------------|-----------------------------------------------| | | POGS 65/130 Portable Oxygen Generation System | This 510(k) summary information is being submitted in accordance with the requirements of 21 CFR 807.92. | Submitter: | | |------------------------|-----------------------------------------------------| | | On-Site Gas Systems, Inc. | | | 35 Budney Road | | | Newington, CT 06111 | | Contact Person: | | | | Guy T. Hatch, Chief Executive Officer | | | Phone: 860-667-8888, extension 267 | | | Fax: 860-667-2222 | | | e-mail: ghatch@onsitegas.com | | Date Prepared: | September 1, 2009 | | Trade Name: | POGS 65/130, Portable Oxygen Generation System | | Common Name: | POGS 65 | | Classification Names: | Oxygen Concentrator | | Device Classification: | | | Regulatory Class: | Class II | | Product Code: | CAW | | Classification Panel: | Anesthesiology | | Regulation Number: | 21 CFR 868.5440 | | Predicate Devices: | | | | POGS 33C, 510(k) K063454, On-Site Gas Systems, Inc. | | Description of Device: | | The Portable Oxygen Generation System 65/130 (POGS 65) has been designed to accommodate medical personnel that require a portable, reliable source of supplemental oxygen in a setting where liquid oxygen may be unavailable. | | <b>On Site Gas Systems, Inc.</b> | |--|--------------------------------------------------------------------| | | 35 Budney Road, Budney Industrial Park, Newington, CT 06111 U.S.A. | | | Telephone: 860.667.8888 • Fax: 860.667.2222 | | | Website: www.onsitegas.com • Email: info@onsitegas.com | | | A BUSINESS INCORPORATED IN THE STATE OF CONNECTICUT, U.S.A | . {2}------------------------------------------------ # e Gas Systems. Inc. #### Manufacturers / Designers of Oxygen & Nitrogen Generating Equipment Requirements are based on the deployment oxygen system- medium (DOGS-M), operational requirements documents that have been issued by the United States Air Force and the United States Army. These two services are preparing a joint requirement document. This system is based on the Pressure Swing Absorption principle and uses a molecular sieve to separate gasses from the filtered ambient air. The oxygen is stored and delivered to the patient(s) through the oxygen boost and distribution system. The POGS 65/130 is a system that generates USP 93% oxygen at a flow rate of up to 65 LPM at a pressure of 50-60 PSI. Included with the generator are medical-grade oxygen hose for both low and high pressure, spare parts and tools that are required, but may be not be easily accessible to the user. The POGS 65/130 is compatible with commercial oxygen consuming equipment and accessories, including D. E, H, K, and M cylinder filling accessories, ventilators, nasal cannulas, Draeger Narkomed Anesthesia machines and Impact Univent/Eagle 754/754M Ventilators. It also connects with the high pressure Hospital Oxygen Backup System and the low-pressure Patient Oxygen Distribution System. A complete POGS 65/130 system consists of two (2) feed air compressors, two (2) oxygen generators, one (1) boost and distribution center, and one (1) oxygen back-up system. Each feed air compressor and oxygen generator pair is designed to provide up to 65 LPM independently or can operate in parallel to achieve up to 130 LPM to provide the user flexibility in the deployment of their resources. The approach is modular and enables the user to add or subtract systems to meet the requirements of the situation. Each POGS 65 pair responds to the oxygen demand that is being used and when operated in parallel, will load share between the pair. The boost and distribution system receives the oxygen from up to two oxygen generators. The boost and distribution system can feed the oxygen to 1) the hospital distribution system, 2) fill the back-up system, 3) fill cylinders or 4) fill the hospital back-up system depending upon the requirements of the situation. To ensure patient care, the hospital distribution system has priority over all other options. One or more type of cylinder or back-up system may be filled in conjunction with supplying the hospital distribution system. The System includes the capability to evacuate high pressure oxygen cylinders to 25 inches Hg vacuum prior to filling. #### Indications for Use: The POGS 65/130 is intended to generate and deliver USP 93% supplemental oxygen. This device is intended to be used only by trained personnel in disaster relief and emergency preparedness situations, military settings or where bottled oxygen is not readily available. #### Difference with Predicate Device: The POGS 65/130 has included several improvements identified through the operation of the POGS 33C by the military around the world. To roughly double the oxygen output of the system in the same footprint, a new molecular sieve material, MDX made by UOP was used in place of the Oxysieve 5 made by UOP used in the POGS 33C. An input and exhaust valve block has replaced individual valves. The ### On Site Gas Systems. Inc. {3}------------------------------------------------ # e Gas Systems, Inc. ### Manufacturers / Designers of Oxygen & Nitrogen Generating Equipment valve block is designed by the same company, MAC, that produces the individual valves, thus we are confident of the lifetime, materials and quality of the components. The system control has been upgraded from a mechanical timer to a PLC. We utilize PLC's throughout the rest of our product line and the manufacturer, Allen-Bradley, is a well known manufacturer of PLC's. The functionality improvement is mainly in the areas of maintenance and troubleshooting and the ease of operation is improved as even those with limited experience with the system can quickly place it into operation and ensure continuous oxygen generation. The POGS 65/130 is not required to produce medical air, therefore there are no provisions for dew point or CO monitoring of the air. #### Improvements from Previously Cleared Device: Several improvements have been made to the predicate device. Improvements to the feed air compression packaging, increased oxygen production, enhanced user interface, and improved operator manuals are provided with the new device. #### Non-Clinical Performance: Performance Standard ASTM F-1464-93 was used. Non-clinical bench testing conducted by On-Site Gas Systems, Inc. is sufficient in establishing substantial equivalence on the POGS 33C to the predicate device on which substantial equivalence is claimed, including oxygen purity evaluations. ## On Site Gas Systems, Inc. 35 Budney Road, Budney Industrial Park, Newington, CT 06111 U.S.A. Telephone: 860.667.8888 · Fax: 860.667.2222 Website: www.onsitegas.com · Email: info@onsitegas.com A BUSINESS INCORPORATED IN THE STATE OF CONNECTICUT, U.S.A. {4}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in a bold, sans-serif font and is centered horizontally. The text is likely part of a document or presentation from the U.S. Department of Health and Human Services. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002 Mr. Guy Hatch Chief Executive Officer On-Site Gas Systems, Incorporated 35 Budney Road Newington, Connecticut 06111 JAN - 7 2010 Re: K092636 Trade/Device Name: POGS 65/130 Portable Oxygen Generation System Regulation Number: 21 CFR 868.5440 Regulation Name: Portable Oxygen Generator Regulatory Class: II Product Code: CAW Dated: December 31, 2009 Received: January 4, 2010 Dear Mr. Hatch: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ #### Page 2 - Mr. Hatch Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. for Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health #### Enclosure {6}------------------------------------------------ #### Indications for Use 4. 510(k) Number: K092636 Device Name: POGS 65/130 Portable Oxygen Generation System Indications for Use: The POGS 65/130 is intended to generate and deliver USP 93% supplemental oxygen. This device is intended to be used only by trained personnel in disaster relief and emergency preparedness situations, military settings or where bottled oxygen is not readily available Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) L. Schultheis (Division Sign-Uff) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K092636
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...