← Product Code [GEI](/productcode/GEI) · K092634

# OBS DISPOSABLE ELECTROSURGICAL PENCILS, MODELS: OBS-DB, OBS-DR, OBS-DF (K092634)

_Jiangmen City Xinhui Baisheng Medical Equipment CO · GEI · Dec 1, 2009 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K092634

## Device Facts

- **Applicant:** Jiangmen City Xinhui Baisheng Medical Equipment CO
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Dec 1, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Indications for Use

The OBS Disposable Electrosurgical Pencils, models: OBS-Db, OBS-Dr, OBS-Df are used for cutting and coagulation to remove tissue and control bleeding by using high frequency current during electrosurgical surgery with a 510k-clearance ESU generator. The device is disposable and supplied sterile with an electrode tip.

## Device Story

Disposable electrosurgical pencils (models OBS-Db, OBS-Dr, OBS-Df) used for soft tissue cutting and coagulation; connects to 510k-cleared monopolar ESU generators via 3M conductive cable. Hand-controlled models (OBS-Db, OBS-Dr) feature plastic housing with two-button or rocker switch interface for CUT/COAG mode selection; foot-controlled model (OBS-Df) activated via external generator-connected footswitch. Used by clinicians in surgical settings to manage bleeding and remove tissue. Device provides high-frequency current to the surgical site via an electrode tip. Benefits include sterile, single-use convenience for electrosurgical procedures.

## Clinical Evidence

Bench testing only. No clinical data. Testing included electrical safety (IEC 60601-1, IEC 60601-2-2), biocompatibility (ANSI/AAMI/ISO 10993-5, -10, -7), sterilization/packaging (ANSI/AAMI/ISO 11135, 11737-1/2, BS EN ISO 11607-1), and shelf-life (ASTM F1980-07).

## Technological Characteristics

Plastic handpiece; monopolar electrode tip; 3M conductive cable. Standards: IEC 60601-1:1988, IEC 60601-2-2:2006, ANSI/AAMI/ISO 10993 series, ANSI/AAMI/ISO 11135-1:2007, ANSI/AAMI/ISO 11737 series, USP 31:2008 <71>, BS EN ISO 11607-1:2006, ISO 15223, ASTM F1980-07. Sterilization: Ethylene Oxide (EO).

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- SHINMED various models of electro-surgical pencils (SW12200, SW12202, SW12300) ([K033027](/device/K033027.md))
- Disposable Hand and Foot-Switching Pencil ([K993647](/device/K993647.md))
- Tecno Disposable Fingerswitch Pencil (150-100) ([K002257](/device/K002257.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

DEC - 1 2009 FDA-Electrosurgical-D-05

# Chapter 05

## PREMARKET NOTIFICATION 510(K) Summary

OBS Disposable Electrosurgical Pencils, models: OBS-Db, OBS-Dr, OBS-Df

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR §807.92

### 1.0 Submitter's Name:

JIANGMEN CITY XINHUI BAISHENG MEDICAL EQUIPMENT CO.,LTD

| Address: | Building 1 No.1,Huimin East Road, Huicheng Town,<br>XinHui District JiangMen City, GuangDong, People's<br>Republic of China |
|----------|-----------------------------------------------------------------------------------------------------------------------------|
| Phone:   | +86-750-6628113                                                                                                             |
| Fax:     | +86-750-6616122                                                                                                             |
| Contact: | Mr. Chen Xi                                                                                                                 |
| E-mail:  | renzheng@bs0750.com                                                                                                         |
| Date     | June 1, 2008                                                                                                                |

- OBS Disposable Electrosurgical Pencils Proprietary Name: 2.0 Model: OBS-Db, OBS-Dr, OBS-Df
Electrosurgical Pencil Common Name :

Classification Name: Electrosurgical, cutting & coagulation & accessories Class II Classification: Product Code: GEI

### 3.0 Predicate Device Information:

1) SHINING WORLD HELTH CARE CO.,LTD

Company Name: SHINING WORLD HELTH CARE CO.,LTD

- Address: 6F, No.8, Lane 7, Wu-Chun Road, Wu-Ku Industrial Park, Taipei, China(Taiwan)
SHINMED various models of electro-surgical pencils model : Proprietary Name: SW12200,SW12202, SW12300

lt's 510(K) number is K033027

Code: FDA-005

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows a logo with the letters "OBS" in a stylized font. Below the letters, there are two Chinese characters, which may represent the name of the company or brand associated with the logo. The logo is black and white, with the letters and characters appearing in black against a white background. The overall design is simple and modern, with a focus on readability and visual impact.

2) Unimed Surgical Products, Inc.

Company Name: Unimed Surgical Products, Inc.

Address: 10401 Belcher Road Largo, Florida 33777

Proprietary Name: Disposable Hand and Foot-Switching Pencil

lt's 510(K) number is K993647

3) Tecno Instruments(PVT) Ltd

Company Name: Tecno Instruments(PVT) Ltd

Address: 316-C Small Industrial Estate Sialkot-51340 Pakistan

Proprietary Name: Tecno Disposable Fingerswitch Pencil

Model: 150-100

lt's 510(K) number is K002257

### 4.0 Device Description

The OBS Disposable Electrosurgical Pencils, models: OBS-Db, OBS-Dr, OBS-Df are used for the cutting and coagulation of soft tissue and have a connector attached to a 3M conductive cable and are designed for use with 510K-clearance generator. The connector or pluq fits into the mono-polar side of a 510K-clearance generator. The handpieces (OBS-Db or OBS-Dr) are made of plastic with two buttons or a rocker switch toward the distal part of the pencil in the case of the hand-controlled pencils, while the foot-controlled pencils(OBS-Df) that is activated by a monopolar footswitch connected to the generator (the footswitch is an accessory of the generator). One button or switch is to control the CUT mode of the ESU while the other controls the COAG mode.

#### Indications for Use: 5.0

The OBS Disposable Electrosurgical Pencils, models: OBS-Db, OBS-Dr, OBS-Df are used for cutting and coagulation to remove tissue and control bleeding by using high frequency current during electrosurgical surgery with a 510k-clearance ESU generator. The device is disposable and supplied sterile with an electrode tip.

#### Comparison to Predicate Devices 6.0

OBS Disposable Electrosurgical Pencils, models: OBS-Db, OBS-Dr, OBS-Df, have been carefully compared to legally marketed devices with respect to intended use, appearance, essential components, materials and performance specifications. They are similar in intended use, appearance, essential components, materials and performance specifications. Although they may differ from the predicate devices in color and size, it won't affect safety and effectiveness of subject devices. In addition, performance and safety have been done to validate the performance and safety of the device. testing

| Code: FDA-005 | Version: A/0 | Page 2 of 3 | Issuing Date: 10.2009 |
|---------------|--------------|-------------|-----------------------|
|---------------|--------------|-------------|-----------------------|

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows a logo with the letters "OBS" stacked above the Chinese characters "百生". The letters "OBS" are in a bold, sans-serif font, with the "O" resembling a water droplet. The Chinese characters are in a traditional font and are stacked vertically below the letters.

FDA-Electrosurgical-D-05

- Discussion of Non-Clinical Tests Performed for Determination of Substantial 7.0 Equivalence are as below:
Electrical safety test, Mechanical performance test, Biocompatibility test and EO Residue Test have been done to demonstrate the safety and performance of subject devices. Tests was conducted in accordance with the "510(k) Guidance Document for General Surgical Electrosurgical Devices", which outlines safety and performance requirements.

Tests were conducted on the OBS Disposable Electrosurgical Pencils, models: OBS-Db, OBS-Dr. OBS-Df as per the requirements of IEC60601-1:1988 . IEC60601-2-2:2006, ANSVAAMMSO10993-5:1999, ANSVAAMIISO10993-10:2002 ,ANSVAAMMSO10993-7:1995, ANSVAAMIISO11135-12007 and ANSVAAMIISO11737-1:2006, ANSVAAMIISO11737-2: 1998., USP31:2008, <71>, BSEN ISO11607-1:2006, ISO 15223, ASTM F1980-07. None of the test demonstrated and design characteristics that violated the requirements of the above mentioned standards or resulted in any safety hazards.

### 8.0 Conclusions:

The comparison and validation results presented in this 510k notification to the FDA show that the OBS Disposable Electrosurgical Pencils, models: OBS-Db, OBS-Dr, OBS-Df are substantially equivalent to predicated devices and is safe and effective in their intended use.

| Code: | FDA-005 |
|-------|---------|
|-------|---------|

| Version: | A/0 |
|----------|-----|
|----------|-----|

| Page | 3 of 3 |
|------|--------|
|------|--------|

| Issuing Date: | 10.2009 |
|---------------|---------|
|---------------|---------|

{3}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

Jiangmen City Xinhui BaiSheng Medical Equipment Co., Ltd. % Intertek Testing Services NA, Inc. Mr. Daniel W. Lehtonen 2307 East Aurora Road, Unit B7 Twinsburg, Ohio 44087 DEC - 1 2009

Re: K092634

Trade/Device Name: OBS Disposable Electrosurgical Pencils, models: OBS-Db, OBS-Dr, OBS-Df

Regulation Number: 21 CFR 878.4400

Regulation Name: Electrosurgical cutting and coagulation device and accessories, Regulatory Class: Class II Product Code: GEI Dated: November 13, 2009 Received: November 16, 2009

Dear Mr. Lehtonen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2- Mr. Daniel W. Lehtonen

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

FDA-Electrosurgical-D-04

OBS

百生

# Chapter 04 PREMARKET NOTIFICATION Indications for Use

510(k) Number (if known): Device Name: OBS Disposable Electrosurgical Pencils, models: OBS-Db, OBS-Dr, OBS-Df Indications For Use:

The OBS Disposable Electrosurgical Pencils, models: OBS-Db, OBS-Dr, OBS-Df are used for cutting and coagulation to remove tissue and control bleeding by using high frequency current during electrosurgical surgery with a 510k-clearance ESU generator. The device is disposable and supplied sterile with an electrode tip.

Prescription Use (Part 21 CFR 801 Subpart D)

## AND/OR Over-The-Counter Use

(21 CFR 801 Subpart C)

· (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Yauvence N. Ceying FOR M. MELKERSON

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K092634

---

**Source:** [https://fda.innolitics.com/device/K092634](https://fda.innolitics.com/device/K092634)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
