WAVESTATE NEUROMONITOR

K092625 · Wavestate, Inc. · ORT · Jun 29, 2010 · Neurology

Device Facts

Record IDK092625
Device NameWAVESTATE NEUROMONITOR
ApplicantWavestate, Inc.
Product CodeORT · Neurology
Decision DateJun 29, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.1400
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

The Wavestate Neuromonitor System is intended to collect, record, and store up to 24 channels of adult EEG data for up to 24 hours. The System also can perform a post review of adult EEG data and identify burst suppression pattern in the stored EEG. The device displays the mean interburst interval reviewed up to that time point and the probability that the displayed value is within +/- 2 seconds of the mean of the interburst intervals for the entire dataset for that patient. The Wavestate Neuromonitor System does not provide any diagnostic conclusion about the patient's condition to the user. The Wavestate Neuromonitor is to be used under the guidance and interpretation of a licensed medical practitioner. Caution: (USA law) Restricts this device to sale by or on the order of a physician.

Device Story

Wavestate Neuromonitor is a software application running on a touch-screen tablet that processes EEG data recorded by the FDA-cleared Lifelines TrackIt-2 ambulatory hardware. The system collects up to 24 channels of adult EEG data; the software performs post-hoc analysis to identify burst suppression patterns. The algorithm calculates the mean interburst interval and provides a statistical confidence metric regarding the accuracy of this interval. The output is displayed to the clinician, who interprets the data to assist in clinical decision-making. The device does not provide diagnostic conclusions. It is intended for use by licensed medical practitioners in clinical settings.

Clinical Evidence

Bench testing only. No clinical data provided. Software verification and validation confirmed the accuracy of the burst suppression algorithm, including burst detection, channel logic, suppression duration, interburst interval calculation, and statistical confidence computation. Testing used digitized EEG files with inserted spikes of varying amplitude (10-10.5 microvolts) and duration to validate detection thresholds and interval calculations.

Technological Characteristics

Software-based EEG analysis system. Hardware: Lifelines TrackIt-2 (24 monopolar touchproof inputs). Input impedance >20 Mohms (differential), >100 Mohms (common mode). CMRR >110dB (0.16-70Hz). Gain 1000 ± 2%. Connectivity: Touch-screen tablet (Windows XP). Software: Proprietary burst suppression algorithm. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-26.

Indications for Use

Indicated for adult patients requiring collection, recording, and storage of up to 24 channels of EEG data for up to 24 hours, and post-review identification of burst suppression patterns. To be used under the guidance and interpretation of a licensed medical practitioner.

Regulatory Classification

Identification

An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ # 510(k) Summary K092625 807.92(c) JUN 2 9 2010 SPONSOR | Company Name: | Wavestate, Inc. | |-----------------|------------------------------------------------------| | Company Address | 333 Washington Blvd. #15<br>Marina del Ray, CA 90292 | | Telephone: | 310-591-8844 | | Fax: | 310-591-8191 | | Contact Person: | Michael J. Bier Ph.D., President and CEO | 807.92(a)(1)Summary Preparation Date: June 29, 2010 #### DEVICE NAME Trade Name: Common/Usual Name: Classification Name: Regulation Number: Product Code: Device Class: ## PREDICATE DEVICE 807.92(a)(2) Wavestate Neuromonitor EEG Electroencephalograph 882.1400 ORT / GWQ I I 807.92(a)(3) 510(k) # K021185 K061908 K010460 | Legally Marketed Equivalent Device | | |------------------------------------|-----------------------| | Company | Product | | Taugagreining HF | Nervus Monitor | | Viasys NeuroCare, Inc. | NicoletOne System V32 | | Lifelines Ltd. | Lifelines Trackit | ## DEVICE DESCRIPTION Wavestate, Inc. has created a new application for the TrackIt-2, an FDA-approved ambulatory EEG hardware unit manufactured by Lifelines, Ltd (UK). Our proprietary software analyzes EEG data files recorded with the TrackIt-2. Data are displayed on an Xplore touch-screen tablet computer using Microsoft Windows XP. Our application is used to quantify the inter-burst interval with 95% statistical confidence the duration of the interval within +/- 2 seconds. The Trackit-2 system is FDA approved. FDA-approved EEG electrodes will be bought separately by the end user. ## 807.92(a)(4) {2}------------------------------------------------ ## DEVICE INTENDED USE The Wavestate Neuromonitor System is intended to collect, record, and store up to 24 channels of adult EEG data for up to 24 hours. The System also can perform a post review of adult EEG data and identify burst suppression pattern in the stored EEG. The device displays the mean interburst interval reviewed up to that time point and the probability that the displayed value is within +/- 2 seconds of the mean of the interburst intervals for the entire dataset for that patient. The Wavestate Neuromonitor System does not provide any diagnostic conclusion about the patient's condition to the user. The Wavestate Neuromonitor is to be used under the guidance and interpretation of a licensed medical practitioner. Caution: (USA law) Restricts this device to sale by or on the order of a physician. Predicate Product Comparison NONCLINICAL AND CLINICAL TEST 807.92(b) SAFETY and EFFECTIVENESS Testing of the Wavestate Neuromonitor was performed. Testing included: - 1. Software verification and validation - 2. Electrical and EMC safety compliance to: - IEC 60601-1 IEC 60601-1-2 IEC 60601-2-26 ## Software Verification and Validation Summary ## V&V SUMMARY The validation testing demonstrates how each feature of the burst suppression algorithm is accurately implemented. The algorithm consists of five features: (1) burst detection, (2) channel logic, (3) suppression duration, (4) interburst interval, and (5) statistical confidence. First, the accuracy of burst detection is tested in a single EEG channel. 40-ms-duration spikes of varving amplitude are inserted into digitized EEG files consisting of background activity (< 5 microvolts). The results demonstrate the accuracy of detection as only spikes of 10 microvolts or higher are identified. Second, the accuracy of burst detection is tested in multiple channels. Test results demonstrate accurate detection of 10.5 microvolt spikes identified independently and simultaneously in each EEG of the 19 channels tested. Third, the accuracy of EEG suppression detection is tested. EEG suppression is defined as 500 ms of activity below 10 microvolts. Spikes of 10.5 microvolt amplitude are inserted, at increasing interval length, into an EEG file consisting of baseline background ## 807.92(a)(5) {3}------------------------------------------------ activity. Test results demonstrate detection of suppression only when spikes are separated by 500 ms or longer. Fourth, the accuracy of calculating and displaying the statistical mean of the interburst interval is tested. Spikes of 10.5 microvolt amplitude are inserted at increasing intervals into an EEG file consisting of background activity. Test results demonstrate accurate calculation and display of the mean interburst interval as an integer. Fifth, the accuracy of the statistical confidence computation is tested. A series of interburst intervals are constructed with 10.5 microvolt spikes. The series converge on a 10-second mean with variance decreasing initially, followed by increasing variance. This results in an initial mean of low statistical confidence and with increasing stability interburst interval stability the confidence interval reaches 95%. Test results demonstrate that the mean interburst interval is displayed once statistical confidence attains 95% and is not displayed when confidence is below 95%. {4}------------------------------------------------ #### COMPARISON OF TECHNICAL CHARACTERISTICS` 807.92(a)(6) ## Caution: (USA law) Restricts this device to sale by or on the order of a physician. | Parameters | New Device | Predicate Device | Predicate Device | Predicate Device | |--------------------------------------------------|---------------------------------------------------------|---------------------------------------------------------|-----------------------------------------|---------------------------------------------------------| | Proprietary Name | Neuromonitor | Trakit | Nicolet OneViasys | Nervus | | Manufacturer | Wavestate Inc. | Lifelines | NeuroCare, Inc | Taugagreining Hf | | Intended User | Licensed medical<br>professional | Licensed medical<br>professional | Licensed medical<br>professional | Licensed medical<br>professional | | Target population | Adult | Adult | Adult | Adult and pediatric<br>patients | | No of EEG<br>channels | 24 monopolar<br>touchproof inputs | 24 monopolar touchproof<br>inputs | 16 to 32 monopolar<br>touchproof inputs | 16 to 32 monopolar<br>touchproof inputs | | Max. common<br>mode input<br>voltage | 2V pk-pk | 2V pk-pk | | 2V pk-pk | | Input bias current | <25nA @25° C | <25nA @25° C | | <20nA @25° C | | Differential Input<br>Impedance | >20 Mohms | >20 Mohms | | >20 Mohms | | Common mode<br>input impedance | >100 Mohms | >100 Mohms | >100 Mohms | >100 Mohms | | Common mode<br>rejection ratio | >110d @ 0.16Hz to<br>70Hz w/ active ground<br>connected | >110d @ 0.16Hz to 70Hz<br>w/ active ground<br>connected | >110d @ 50/60 Hz | >110d @ 0.16Hz to<br>70Hz w/ active ground<br>connected | | Channel hardware<br>gain | 1000 ± 2% | 1000 ± 2% | | 500 | | Max. differential<br>AC input before<br>clipping | 10 mV pk-pk | 10 mV pk-pk | | 10 mV pk-pk | | DC Input<br>tolerance | ±500mV | ±500mV | ±350mV | | | Burst Suppression | yes | no | yes | yes | Predicate Product Comparison {5}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Wavestate, Inc. c/o Ned Devine Underwriters Laboratories, Inc. 333 Pfingsten Road Northbrook, IL 60062 JUN 2 9 2010 Re: K092625 Trade/Device Name: Wavestate Neuromonitor Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: ORT, GWO Dated: April 28, 2010 Received: April 30, 2010 Dear Mr. Devine: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {6}------------------------------------------------ device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resources.forYou/Industry/default.htm. Sincerely yours, Eithim im to Malvina B. Eydelman, M Director Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ # Indications for Use 510(k) Number (if known): K092625 Device Name: Wavestate Neuromonitor Indications for Use: The Wavestate Neuromonitor System is intended to collect, record, and store up to 24 channels of adult EEG data for up to 24 hours. The System also can perform a post review of adult EEG data and identify burst suppression pattern in the stored EEG. The device displays the mean interburst interval reviewed up to that time point and the probability that the displayed value is within +/- 2 seconds of the mean of the interburst intervals for the entire dataset for that patient. The Wavestate Neuromonitor System does not provide any diagnostic conclusion about the patient's condition to the user. The Wavestate Neuromonitor is to be used under the guidance and interpretation of a licensed medical practitioner. Caution: (USA law) Restricts this device to sale by or on the order of a physician. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Rick Rahn Page 1 of 1 (Division Sign-Off). Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices 510(k) Number K092628
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...