← Product Code [KCT](/productcode/KCT) · K092551

# SMITH & NEPHEW INSTRUMENT TRAYS (K092551)

_Smith & Nephew Inc., Endoscopy Div. · KCT · Sep 14, 2010 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K092551

## Device Facts

- **Applicant:** Smith & Nephew Inc., Endoscopy Div.
- **Product Code:** [KCT](/productcode/KCT.md)
- **Decision Date:** Sep 14, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6850
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

Smith & Nephew ACUFEX™ GRAFTMASTER™ III SYSTEM TRAYS are intended to contain Smith & Nephew reusable surgical instruments for convenient organized storage, sterilization and transport between usages. The instrument trays are suitable for use in a prevacuum steam method. They are intended to be used in conjunction with a validated sterilization wrap in order to maintain sterility of the enclosed devices.

## Device Story

The Smith & Nephew ACUFEX™ GRAFTMASTER™ III SYSTEM TRAY is a reusable stainless steel sterilization container designed to hold and protect surgical instruments. It features a lid, silicone pin mat, and internal instrument holders. The tray is perforated to allow sterilant exposure during the prevacuum steam sterilization process (132°C for 4 minutes with 60 minutes drying time). Used in clinical settings, the tray facilitates the organized storage and transport of instruments between uses. It must be used with a validated sterilization wrap to maintain sterility. The device benefits patients and healthcare providers by ensuring instruments are properly organized, protected from damage, and effectively sterilized before surgical procedures.

## Clinical Evidence

Bench testing only. Validation testing for sterilization efficacy and functional strength was conducted in accordance with AAMI ST77:2006 Containment Devices for reusable medical device sterilization. Results demonstrated that the device meets pre-defined acceptance criteria and performs equivalently to the predicate devices.

## Technological Characteristics

Reusable stainless steel sterilization tray and lid. Features silicone pin mats and instrument holders. Perforated design for steam penetration. Validated for prevacuum steam sterilization (132°C/270°F, 4-minute exposure, 60-minute dry time). Complies with AAMI ST77:2006 standards for containment devices.

## Regulatory Identification

A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.

## Predicate Devices

- Smith & Nephew Instrument Tray ([K073551](/device/K073551.md))
- Smith & Nephew Instrument Trays ([K090562](/device/K090562.md))
- Smith & Nephew Instrument Trays ([K091627](/device/K091627.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# SECTION IV

# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION

as required by the Safe Medical Devices Act of 1990 and codified in 21 CFR 807.92 upon which the substantial equivalence is based.

## Smith & Nephew ACUFEX™ GRAFTMASTER™ III SYSTEM TRAY

#### Date Prepared: August 19, 2009

#### A. Submitter's Name:

Smith & Nephew, Inc., Endoscopy Division 150 Minuteman Road Andover, MA 01810

### B. Company Contact

Christina Flores Regulatory Affairs Specialist II T: 508-261-3705 F: 508-261-3620 Christina.flores@smith-nephew.com

#### C. Device Name

| Trade Name:            | Smith & Nephew ACUFEX™ GRAFTMASTER™ III SYSTEM TRAY                   |
|------------------------|-----------------------------------------------------------------------|
| Common Name:           | Sterilization Tray                                                    |
| Classification Name:   | Sterilization Wrap Containers, Trays, Cassettes and Other Accessories |
| Class:                 | II                                                                    |
| Product Code:          | KCT                                                                   |
| Classification Number: | 21 CFR §880.6850                                                      |

Traditional 510(k) Section IV Updated September 14, 2010

Page 1 of 3

# 0103

(092551

SEP 1 4 2010

{1}------------------------------------------------

K092551
p²δ3

### D. Predicate Devices

The subject Smith & Nephew ACUFEX™ GRAFTMASTER™ III SYSTEM TRAY is substantially equivalent in Intended Use and Fundamental Scientific Technology to the following legally marketed device in commercial distribution: Smith & Nephew Instrument Tray cleared under K073551, Smith & Nephew Instrument Trays cleared under K090562, and Smith & Nephew Instrument Trays cleared under K091627.

#### E. Description of Device

The Smith & Nephew ACUFEX™ GRAFTMASTER™ III SYSTEM TRAY is a reusable stainless steel sterilization tray and lid fitted with a silicone pin mat and instrument holders designed to contain and protect reusable surgical instruments during transport, sterilization, and storage. Perforations in the tray and lid allow for optimal. exposure of the tray's contents to sterilant during the sterilization process.

#### F. Intended Use

Smith & Nephew ACUFEX™ GRAFTMASTER™ III SYSTEM TRAYS are intended to contain Smith & Nephew reusable surgical instruments for convenient organized storage, sterilization and transport between usages. The instrument trays are suitable for use in a prevacuum steam method. They are intended to be used in conjunction with a validated sterilization wrap in order to maintain sterility of the enclosed devices.

Validated Sterilization Parameters:

| Method           | Temperature      | Exposure Time | Drying Time |
|------------------|------------------|---------------|-------------|
| Pre-vacuum steam | 132 C<br>(270 F) | 4 minutes     | 60 minutes  |

Device model that is subject of this pre-market notification:

| REF      | Description                  |
|----------|------------------------------|
|          | Smith & Nephew ACUFEX™       |
| 72202441 | GRAFTMASTER™ III SYSTEM TRAY |

#### G. Comparison of Technological Characteristics

The technological characteristics of the proposed Smith & Nephew ACUFEX™ GRAFTMASTER™ III SYSTEM TRAY are identical to the predicate devices. The material of construction, the type of instruments contained, and the intended use are unchanged from the predicate trays. The internal configuration and number of instrument holders, pin mats, and organizing racks differ to accommodate each trays designated instrument set. These differences do not represent new technological

Traditional 510(k) Section IV Updated September 14, 2010

Page 2 of 3

{2}------------------------------------------------

characteristics for the proposed device and therefore do not raise new questions of safety or efficacy.

#### Summary Performance Data H.

Non clinical validation testing was conducted for sterilization and functional strength in order to demonstrate that the subject device is safe and effective, and whose performance meets the requirements of its pre-defined acceptance criteria and intended uses.

Performance testing was conducted in accordance with AAMI ST77:2006 Containment Devices for reusable medical device sterilization and demonstrates substantial equivalence to the predicates.

Traditional 510(k) Section IV Updated September 14, 2010

Page 3 of 3

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the department's name around the perimeter. In the center of the logo is an abstract symbol that resembles an eagle or bird-like figure.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Ms. Christina Flores Regulatory Affairs Specialist II Smith & Nephew, Incorporated 150 Minuteman Road Andover, Massachusetts 01810

# SEP 1 4 2010

Re: K092551

Trade/Device Name: Smith & Nephew Instrument Trays Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: KCT Dated: September 3, 2010 Received: September 7, 2010

Dear Ms. Flores:

· We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

#### Page 2- Ms. Flores

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance,

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

h for

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## INDICATIONS FOR USE STATEMENT

#### 510(K) Number:

Device Name:

Smith & Nephew ACUFEX™ GRAFTMASTER™ III SYSTEM TRAY ·

Indications for Use: Smith & Nephew ACUFEX™ GRAFTMASTER™ III SYSTEM TRAYS are intended to contain Smith & Nephew reusable surgical instruments for convenient organized storage, sterilization and transport between usages. The instrument trays are suitable for use in a prevacuum steam method. They are intended to be used in m a previloudin storalidated sterilization wrap in order to maintain sterility of the enclosed devices.

Validated Sterilization Parameter:

| Method              | Temperature      | Exposure Time | Drying Time |
|---------------------|------------------|---------------|-------------|
| Pre-vacuum<br>steam | 132 C<br>(270 F) | 4 minutes     | 60 minutes  |

Device models that are the subject of this pre-market notification:

| REF      | Description                                          |
|----------|------------------------------------------------------|
| 72202441 | Smith & Nephew Acufex Graftmaster III<br>System Tray |

(Division Sign-Off)

510(k) Number: Ko

Division of Anesthesiology, General Hospital

Infection Control, Dental Devices

Prescription Use __

## AND/OR

Over-The-Counter Use _X (21 CFR 807 Subpart C)

25 5

(Per 21 CFR 801 Subpart D)

# (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

---

**Source:** [https://fda.innolitics.com/device/K092551](https://fda.innolitics.com/device/K092551)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
