The Pioneer Posterior Cervico-Thoracic System is indicated to promote fusion of the cervico- thoracic regions of the spine (C1-T3). The intended indications are as follows: - Degenerative Disc Disease (as defined by neck or back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies - Spondylolisthesis - Spinal Stenosis - Fracture/ Dislocation - Deformities or Curvature - Tumors - Pseudoarthrosis - Revision of previous cervical and upper thoracic spine surgery The use of the screws is limited to placement in the upper thoracic spine (T1-T3) in treating thoracic conditions only. The screws are not intended for use in the cervical spine. The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/ dislocation or trauma in the cervical/ upper thoracic (C1-T3) spine.
Device Story
Pioneer Posterior Cervico-Thoracic System is a spinal fixation construct consisting of rods, hooks, polyaxial screws, favored angle screws, locking caps, and connecting components. Used by surgeons in clinical settings to stabilize the cervico-thoracic spine (C1-T3) to promote fusion. Components are implanted to provide structural support following trauma, deformity correction, or degenerative disease treatment. System can integrate with Pioneer Quantum Spinal Rod System via parallel connectors. Device provides mechanical stabilization; clinical benefit is achieved through spinal fusion. No software or electronic components involved.
Clinical Evidence
Bench testing only. Performance data per recognized ASTM standards was presented to demonstrate mechanical integrity and substantial equivalence.
Technological Characteristics
Construct components manufactured from medical grade titanium alloy and cobalt chromium. Materials comply with ASTM standard specifications. System includes rods, hooks, polyaxial/favored angle screws, and locking caps. Mechanical fixation device; no energy source, software, or connectivity.
Indications for Use
Indicated for patients requiring spinal fusion in the cervico-thoracic region (C1-T3) due to degenerative disc disease, spondylolisthesis, spinal stenosis, fracture/dislocation, deformities/curvature, tumors, pseudoarthrosis, or revision of previous surgery. Screws are restricted to T1-T3 thoracic placement; not for cervical spine use.
Regulatory Classification
Identification
A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.
{1}------------------------------------------------
### Pioneer Posterior Cervico-Thoracic System
# 510(k) Summary
FEB 1 9 2010
| Sponsor: | Pioneer Surgical Technology<br>375 River Park Circle<br>Marquette, MI 49855<br>(906) 226-4812<br>Contact: Emily M. Downs | | |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----|-------------------|
| Device Name: | Pioneer Posterior Cervico-Thoracic System | | |
| Classification Name: | Spinal Interlaminal Fixation Orthosis<br>Pedicle Screw System | | |
| Classification Number: | Product Code/ Regulation Number:<br>KWP/ 888.3050 - Spinal Interlaminal Fixation Orthosis<br>MNI/ 888.3070(b)(1)- Pedicle Screw Spinal System<br>MNH/ 888.3070(b)(1)- Pedicle Screw Spinal System<br>Panel Code: 87 | | |
| Description: | The Pioneer Posterior Cervico-Thoracic System consists of a variety of<br>rods, hooks, polyaxial screws, favored angle screws, locking caps, and<br>connecting components used to build a cervico-thoracic spinal<br>construct. | | |
| | The Pioneer Posterior Cervico-Thoracic System components are<br>manufactured from medical grade titanium alloy and medical grade<br>cobalt chromium. Medical grade titanium alloy and medical grade<br>cobalt chromium may be used together. | | |
| | The Pioneer Posterior Cervico-Thoracic System can be attached to<br>Pioneer Quantum® Spinal Rod System using parallel connectors. | | |
| | The system also contains Class 1 manual surgical instruments and cases<br>that are considered exempt from premarket notification. | | |
| Intended Use: | The Pioneer Posterior Cervico-Thoracic System is indicated to promote<br>fusion of the cervico- thoracic regions of the spine (C1-T3). The<br>intended indications are as follows: | | |
| | - Degenerative Disc Disease (as defined by neck or back pain of<br>discogenic origin with degeneration of the disc confirmed by history<br>and radiographic studies | | |
| | - Spondylolisthesis | | |
| Pioneer Surgical Technology | | 119 | February 12, 2010 |
Ploneer Surgi
K092295_Al
Rase
ನ್
**February 12, 2010**
.
{2}------------------------------------------------
| K0922 95 | |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Pioneer Posterior Cervico-Thoracic System | |
| | - Spinal Stenosis |
| | - Fracture/ Dislocation |
| | - Deformities or Curvature |
| | - Tumors |
| | - Pseudoarthrosis |
| | - Revision of previous cervical and upper thoracic spine surgery |
| | The use of the screws is limited to placement in the upper thoracic spine<br>(T1-T3) in treating thoracic conditions only. The screws are not<br>intended for use in the cervical spine. |
| | The hooks and rods are also intended to provide stabilization to promote<br>fusion following reduction of fracture/ dislocation or trauma in the<br>cervical/ upper thoracic (C1-T3) spine. |
| Material: | Materials used to manufacture the implants and instruments of this<br>system follow ASTM Standard Specifications. |
| Performance Data: | Testing per recognized ASTM standards was presented. |
| Performance and SE<br>Determination: | The characteristics of the Pioneer Posterior Cervico-Thoracic System<br>are similar to the following predicate devices: |
| | 1. Atoll Cervico-Thoracic System (K070638), Theken Spine, SE date<br>05/30/2007. |
| | 2. Summit OCT Spinal System (K002733), DePuy AcroMed, Inc., SE<br>date 12/15/2000. |
| | 3. VERTEXTM Reconstruction System (K003780), Medtronic Sofamor<br>Danek USA, Inc., SE date 09/28/2001. |
| | 4. Spine Oasys System (K032394), Stryker, SE date 2/20/2004 |
| | 5. Cervifix/ Starlock System (K990965) Synthes (USA), SE date<br>7/1/1999. |
| | Equivalence for Pioneer Posterior Cervico-Thoracic System is based on<br>similarities of intended use, design, and physical characteristics when<br>compared to predicate devices. Therefore, Pioneer Surgical Technology<br>believes that there is sufficient evidence to conclude that the Pioneer<br>Posterior Cervico-Thoracic System is substantially equivalent to<br>existing legally marketed devices. |
| | |
2002
) a
February 12, 2010
·
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. The bottom half of the circle contains the department's emblem, which is a stylized representation of a human figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
## FEB 1 9 2010
Pioneer Surgical Technology, Inc. % Ms. Emily M. Downs 375 River Park Circle Marquette, Michigan 49855
Re: K092295
Trade/Device Name: Pioneer Posterior Cervico-Thoracic System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: Class II Product Code: KWP, MNI, MNH Dated: February 12, 2010 Received: February 16, 2010
Dear Ms. Downs:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Ms. Emily M. Downs
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark A. Millman
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Pioneer Posterior Cervico-Thoracic System
#### 3.0 Indications for Use Statement
K09 2295 510(k) Number (if knówn):
Pioneer Posterior Cervico-Thoracic System Device Name:
Indications:
The Pioneer Posterior Cervico-Thoracic System is indicated to promote fusion of the cervico- thoracic regions of the spine (C1-T3). The intended indications are as follows:
- Degenerative Disc Disease (as defined by neck or back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies
- Spondylolisthesis -
- Spinal Stenosis -
- -Fracture/ Dislocation
- -Deformities or Curvature
- -Tumors
- Pseudoarthrosis -
- Revision of previous cervical and upper thoracic spine surgery -
The use of the screws is limited to placement in the upper thoracic spine (T1-T3) in treating thoracic conditions only. The screws are not intended for use in the cervical spine.
The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/ dislocation or trauma in the cervical/ upper thoracic (C1-T3) spine.
> Prescription Use (Per 21 CFR 801.109)
OR
Over-the-Counter Use _________________________________________________________________________________________________________________________________________________________
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | <div>(Division Sign-Off)</div> |
|--|-----------------------------------------------------------|
| | Division of Surgical, Orthopedic, and Restorative Devices |
| | Page 1 of 1 |
| Pioneer Surgical Technology Premarket Notification | 510(k) Number: K092295 | Confidential July 27, 2009 |
|----------------------------------------------------|------------------------|----------------------------|
|----------------------------------------------------|------------------------|----------------------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.