DIMENSION VISTA IMMUNOGLOBULIN IGG SUBCLASSS 1FLEX REAGENT CARTRIDGE, DIMENSION VISTA IMMUNOGLOBULIN IGG

K092283 · Siemens Healthcare Diagnostics · DFZ · Oct 21, 2009 · Immunology

Device Facts

Record IDK092283
Device NameDIMENSION VISTA IMMUNOGLOBULIN IGG SUBCLASSS 1FLEX REAGENT CARTRIDGE, DIMENSION VISTA IMMUNOGLOBULIN IGG
ApplicantSiemens Healthcare Diagnostics
Product CodeDFZ · Immunology
Decision DateOct 21, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5510
Device ClassClass 2
AttributesPediatric

Indications for Use

The IGG1 method is an in vitro diagnostic test for the quantitative measurement of immunodobulin G subclass 1 in human serum, heparinized plasma and EDTA plasma on the Dimension Vista System. Measurements of immunoglobulin G subclass 1 aid in the diagnosis of plasma cell antibody forming abnormalities in conjunction with clinical and other laboratory findings. The IGG2 method is an in vitro diagnostic test for the quantitative measurement of immunodobulin G subclass 2 in human serum. heparinized plasma and EDTA plasma on the Dimension Vista" System. Measurements of immunoglobulin G subclass 2 aid in the diagnosis of plasma cell antibody forming abnormalities in conjunction with clinical and other laboratory findings. The IGG3 method is an in vitro diagnostic test for the quantitative measurement of immunodobulin G subclass 3 in human serum, heparinized plasma and EDTA plasma on the Dimension Vista System. Measurements of immunoglobulin G subclass 3 aid in the diagnosis of plasma cell antibody forming abnormalities in conjunction with clinical and other laboratory findings. The IGG4 method is an in vitro diagnostic test for the quantitative measurement of immunodobulin G subclass 4 in human serum, heparinized plasma and EDTA plasma on the Dimension Vista" System. Measurements of immunoglobulin G subclass 4 aid in the diagnosis of plasma cell antibody forming abnormalities in conjunction with clinical and other laboratory findinas.

Device Story

Reagent cartridges for Dimension Vista System; measure IgG subclasses (1-4) in human serum/plasma. Principle: immunochemical reaction forming immune complexes; IgG1/2 use direct nephelometry; IgG3/4 use particle-enhanced nephelometry. Scattered light intensity proportional to protein concentration. Used in clinical laboratories; operated by trained personnel. Output: quantitative concentration values. Results assist clinicians in diagnosing plasma cell antibody forming abnormalities. Benefits: automated, rapid quantitative assessment of IgG subclasses compared to manual methods.

Clinical Evidence

Bench testing only. Performance validated via precision (CLSI EP5-A2), linearity (EP6-A), and method comparison (EP9-A2) against predicate devices using 129-150 patient samples. Correlation coefficients (r2) ranged from 0.934 to 0.988. Interference testing (EP7-A2) confirmed bias <10% for common endogenous/exogenous substances. Reference intervals validated per CLSI C28-A2.

Technological Characteristics

Nephelometric assay; IgG1/2 use sheep polyclonal antisera; IgG3/4 use polystyrene particles coated with sheep polyclonal antisera. Reagents are liquid, ready-to-use in plastic cartridges. Energy source: light scattering measurement. Connectivity: integrated with Dimension Vista System. Software: automated processing on host system.

Indications for Use

Indicated for quantitative measurement of IgG subclasses 1, 2, 3, and 4 in human serum, heparinized plasma, and EDTA plasma to aid in the diagnosis of plasma cell antibody-forming abnormalities.

Regulatory Classification

Identification

An immunoglobulins A, G, M, D, and E immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the immunoglobulins A, G, M, D, an E (serum antibodies) in serum. Measurement of these immunoglobulins aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k092283 B. Purpose for Submission: New device C. Measurand: IgG subclasses, IgG1, IgG2, IgG3, IgG4 D. Type of Test: Quantitative assays using immunochemical reactions measured by nephelometry E. Applicant: Siemens Healthcare Diagnostics F. Proprietary and Established Names: Dimension Vista® IgG1 Flex® reagent cartridge Dimension Vista® IgG2 Flex® reagent cartridge Dimension Vista® IgG3 Flex® reagent cartridge Dimension Vista® IgG4 Flex® reagent cartridge G. Regulatory Information: 1. Regulation section: 21 CFR§866.5510 – Immunoglobulins A, G, M, D, and E Immunological Test System 2. Classification: Class II 3. Product code: DFZ: IgG (Gamma Chain Specific), Antigen, Antiserum, Control 4. Panel: Immunology (82) H. Intended Use: 1. Intended use(s): Dimension Vista® Immunoglobulin G Subclass 1 Flex® reagent cartridge: The IgG1 method is an *in vitro* diagnostic test for the quantitative measurement of immunoglobulin G subclass 1 in human serum, heparinized plasma and EDTA plasma on the Dimension Vista® System. Measurements of immunoglobulin G subclass 1 aid in the diagnosis of plasma cell antibody forming abnormalities in conjunction with clinical and other laboratory findings. Dimension Vista® Immunoglobulin G Subclass 2 Flex® reagent cartridge: The IgG2 method is an *in vitro* diagnostic test for the quantitative measurement of immunoglobulin G subclass 2 in human serum, heparinized plasma and EDTA plasma on the Dimension Vista® System. Measurements of immunoglobulin G subclass 2 aid in the diagnosis of plasma cell antibody forming abnormalities in conjunction with clinical and other laboratory findings. Dimension Vista® Immunoglobulin G Subclass 3 Flex® reagent cartridge: The IgG3 method is an *in vitro* diagnostic test for the quantitative measurement of {1} immunoglobulin G subclass 3 in human serum, heparinized plasma and EDTA plasma on the Dimension Vista® System. Measurements of immunoglobulin G subclass 3 aid in the diagnosis of plasma cell antibody forming abnormalities in conjunction with clinical and other laboratory findings. Dimension Vista® Immunoglobulin G Subclass 4 Flex® reagent cartridge: The IgG4 method is an in vitro diagnostic test for the quantitative measurement of immunoglobulin G subclass 4 in human serum, heparinized plasma and EDTA plasma on the Dimension Vista® System. Measurements of immunoglobulin G subclass 4 aid in the diagnosis of plasma cell antibody forming abnormalities in conjunction with clinical and other laboratory findings. 2. Indication(s) for use: Same as intended use 3. Special conditions for use statement(s): For prescription use only 4. Special instrument requirements: Dimension Vista System (k051087) I. Device Description: Dimension Vista Immunoglobulin G Subclass 1 Flex reagent Cartridge: Reagents are contained in segregated wells in a plastic cartridge. Wells 1 and 2 contain buffers and polyethylene glycol. Wells 3-10 are empty. Wells 11 and 12 contain liquid sheep polyclonal antisera to human IgG1. The antisera contain specific antibodies which form immune complexes with IgG1. There are 2 Flex cartridges per carton. Calibrators and controls are required, but not provided. Dimension Vista Immunoglobulin G Subclass 2 Flex reagent Cartridge: Reagents are contained in segregated wells in a plastic cartridge. Wells 1 and 2 contain buffers and polyethylene glycol. Wells 3-10 are empty. Wells 11 and 12 contain liquid sheep polyclonal antisera to human IgG2. The antisera contain specific antibodies which form immune complexes with IgG2. There are 2 Flex cartridges per carton. Calibrators and controls are required, but not provided. Dimension Vista Immunoglobulin G Subclass 3 Flex reagent Cartridge: Reagents are contained in segregated wells in a plastic cartridge. Wells 1 and 2 contain buffers and polyethylene glycol. Wells 3-10 are empty. Wells 11 and 12 contain liquid sheep polyclonal antisera to human IgG3. Polystyrene particles are coated with antibodies specific to human IgG3. There are 2 Flex cartridges per carton. Calibrators and controls are required, but not provided. Dimension Vista Immunoglobulin G Subclass 4 Flex reagent Cartridge: Reagents are contained in segregated wells in a plastic cartridge. Wells 1 and 2 contain buffers and polyethylene glycol. Wells 3-10 are empty. Wells 11 and 12 contain liquid sheep polyclonal antisera to human IgG4. Polystyrene particles are coated with antibodies specific to human IgG4. There are 2 Flex cartridges per carton. Calibrators and controls are required, but not provided. J. Substantial Equivalence Information: 1. Predicate device name(s): Siemens N Antisera to Human IgG Subclasses 1 and 2 assays Siemens N Latex IgG3 and IgG4 {2} 2. Predicate 510(k) number(s): N Antiserum to Human IgG Subclasses 1 and 2 -k860894 N Latex IG3 and IG4 - 510(k) Exempt and CLIA Categorized under classification 866.5530 code DAS in 3/2004. 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicates | | Intended Use | The IGG1-4 methods are in vitro diagnostic tests for the quantitative measurement of immunoglobulin G subclasses in human serum, heparinized plasma and EDTA plasma on the Dimension Vista® System. Measurements of immunoglobulin G subclasses aid in the diagnosis of plasma cell antibody forming abnormalities. | Same | | Technology | Nephelometric | Same | | Assay format | Quantitative | Same | | Reagents | Liquid, require no preparation | Same | | Antibody | Polyclonal from sheep | Same | | Matrix | Human serum, heparinized plasma and EDTA plasma | Same | | Storage | 2-8°C | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Instrument | Dimension Vista System | BN II and BN ProSpec System | K. Standard/Guidance Document Referenced (if applicable): Interference Testing in Clinical Chemistry; Approved Guideline (EP 7-A) Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline (EP09-A2) Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guideline (EP5-A) Evaluation of the Linearity of Quantitative Analytical Methods (EP6-P2) Protocols for Determination of Limits of Detection and Limits of Quantitation (NCCLS/CLSI EP17-A) Draft Guidance Document for 510(k) Submission of Immunoglobulins A, G, M, D and E Immunoglobulin System In Vitro Devices L. Test Principle: Dimension Vista® IGG 1 and 2 {3} Proteins contained in human body fluids form immune complexes in an immunochemical reaction with specific antibodies. These complexes scatter a beam of light passed through the sample. The intensity of the scattered light is proportional to the concentration of the respective protein in the sample. The result is evaluated by comparison with a standard of known concentration. ## Dimension Vista® IGG 3 and 4 Polystyrene particles coated with antibodies specific to human IGG3 or 4 are aggregated when mixed with samples containing IGG3 or 4. These aggregates scatter a beam of light passed through the sample. The intensity of the scattered light is proportional to the concentration of the respective protein in the sample. The result is evaluated by comparison with a standard of known concentration. ## M. Performance Characteristics (if/when applicable): ### 1. Analytical performance: #### a. Precision/Reproducibility: Precision testing for the IGG1-4 methods were performed over 20 days according to CLSI EP5-A2, at a single site. A single instrument, a single reagent lot and two operators were used. On each day of testing, each sample was run in duplicate, in two separate runs. The test samples consisted of 3 levels of Dimension Vista® Protein 1 Control, 2 serum samples and 2 plasma samples. Serum and plasma pools for all methods were established at levels that encompassed the analytical measuring range for each method. Analysis of variance (ANOVA) was used to evaluate the data consistent with the recommendations of EP5-A2. IgG1 | Material | Mean (g/L) | Repeatability | | Within Lab | | | --- | --- | --- | --- | --- | --- | | | | %CV | SD (g/L) | %CV | SD (g/L) | | PROT1CON L | 3.64 | 3.0 | 0.11 | 3.7 | 0.13 | | PROT1CON M | 4.85 | 3.5 | 0.17 | 3.8 | 0.18 | | PROT1CON H | 7.77 | 1.8 | 0.14 | 2.0 | 0.15 | | Serum pool | 3.38 | 2.9 | 0.1 | 3.6 | 0.12 | | Serum pool | 19.92 | 3.3 | 0.65 | 3.4 | 0.68 | | Plasma pool | 4.99 | 2.9 | 0.15 | 3.8 | 0.19 | | Plasma pool | 10.66 | 1.9 | 0.2 | 2.4 | 0.26 | IgG2 | Material | Mean (g/L) | Repeatability | | Within Lab | | | --- | --- | --- | --- | --- | --- | | | | %CV | SD (g/L) | %CV | SD (g/L) | | PROT1CON L | 1.76 | 2.0 | 0.04 | 5.3 | 0.09 | | PROT1CON M | 2.77 | 2.8 | 0.08 | 4.5 | 0.12 | | PROT1CON H | 4.24 | 3.0 | 0.13 | 3.5 | 0.15 | | Serum pool | 1.32 | 3.3 | 0.04 | 5.0 | 0.07 | | Serum pool | 7.20 | 2.8 | 0.20 | 4.1 | 0.3 | | Plasma pool | 1.50 | 2.3 | 0.03 | 5.7 | 0.09 | | Plasma pool | 7.29 | 2.9 | 0.21 | 3.8 | 0.28 | {4} IgG3 | Material | Mean (g/L) | Repeatability | | Within Lab | | | --- | --- | --- | --- | --- | --- | | | | %CV | SD (g/L) | %CV | SD (g/L) | | PROT1CON L | 0.27 | 2.7 | 0.007 | 3.6 | 0.010 | | PROT1CON M | 0.36 | 3.0 | 0.011 | 3.9 | 0.014 | | PROT1CON H | 0.52 | 3.7 | 0.019 | 5.2 | 0.027 | | Serum pool | 0.23 | 2.4 | 0.005 | 3.8 | 0.009 | | Serum pool | 1.28 | 4.6 | 0.059 | 5.7 | 0.074 | | Plasma pool | 0.20 | 1.9 | 0.004 | 3.5 | 0.007 | | Plasma pool | 0.52 | 3.5 | 0.018 | 3.9 | 0.020 | IgG4 | Material | Mean (g/L) | Repeatability | | Within Lab | | | --- | --- | --- | --- | --- | --- | | | | %CV | SD (g/L) | %CV | SD (g/L) | | PROT1CON L | 0.44 | 3.4 | 0.015 | 4.5 | 0.020 | | PROT1CON M | 0.62 | 3.8 | 0.024 | 4.4 | 0.027 | | PROT1CON H | 0.81 | 4.4 | 0.035 | 4.6 | 0.038 | | Serum pool | 0.18 | 2.4 | 0.005 | 3.3 | 0.006 | | Serum pool | 1.01 | 3.6 | 0.036 | 3.7 | 0.037 | | Plasma pool | 0.15 | 3.1 | 0.005 | 3.2 | 0.005 | | Plasma pool | 0.87 | 3.5 | 0.031 | 3.8 | 0.033 | b. Linearity/assay reportable range: Linearity The linearity for the IGG1-4 methods was determined according to the CLSI EP-6-A. One serum sample was serially diluted with System Diluent, and run in replicates of five. The observed value represents the mean of the five replicates. The bias was determined at each level. The acceptance criteria were defined as: the absolute value for the Difference (g/L) [G] must not exceed the Limit (g/L) [H] or the absolute value for the Relative Difference (%) [I] must not exceed the Limit (%) [J]. The Limit (g/L) [H] is defined as 15% of Expected value [D] at Dilution factor level 0.075. For all of the assays, the acceptance criteria were met. IgG1 Testing was performed using a human serum sample (26.29 g/L) that was serially diluted ten times. Expected concentrations vs. observed concentrations were plotted. Regression analysis showed a slope of 0.996 and y intercept of -0.027. The linear range was determined to be 2.34-26.29 g/L. The analytical measuring range was established from the linear range and the results of the Limit of Quantitation study and determined to be 2.20-23.40 g/L. IgG2 Testing was performed using a human serum sample (9.9 g/L) that was serially diluted eleven times. Measured concentrations vs. observed concentrations were plotted. Regression analysis showed a slope: 0.993 and y intercept of 0.0098. The linear range was determined to be 0.50-9.91 g/L. The analytical measuring range was {5} established from the linear range and the results of the Limit of Quantitation study and determined to be 0.44-8.30 g/L. ## IgG3 Testing was performed using a human serum sample (2.27 g/L) that was serially diluted eleven times. Measured concentrations vs. observed concentrations were plotted. Regression analysis showed a slope of 1.04 and y intercept of -0.0086. The linear range was determined to be 0.057-2.55 g/L. The analytical measuring range was established from the linear range and the results of the Limit of Quantitation study and determined to be 0.035-1.57 g/L. ## IgG4 Testing was performed using a human serum sample (2.833 g/L) that was serially diluted ten times. Measured concentrations vs. observed concentrations were plotted. Regression analysis showed a slope of 0.961 and y intercept of 0.017. The linear range was determined to be 0.045-2.833 g/L. The analytical measuring range was established from the linear range and the results of the Limit of Quantitation study and determined to be 0.057-2.55 g/L. ## Recovery A recovery study was performed by mixing equal volumes of two patient samples of known quantity for each of the IGG 1-4 assays. For IgG1, recovery ranged from 93-100% with a mean of 96% For IgG2, recovery ranged from 91.2-96.2%, with a mean of 93.2% For IgG3, recovery ranged from 91-105%, with a mean of 99%. For IgG4, recovery ranged from 99-105%, with a mean of 102%. c. Traceability, Stability, Expected values (controls, calibrators, or methods): There are no reference standards for IgG1 to IgG 4. IGG1-4 calibrator and control value assignment and stability data were provided in k071980. The calibrator master lot values for IgG 1-4 are assigned versus ERM-DA 470 (CR 470) using independently determined values for subclasses compared to total IgG value of the reference material. d. Detection limit: For each of the IgG subclasses, three serum samples with known concentrations of the IgG subclass were spiked with ERM®-DA 470 – a reference material certified for the mass concentration of 12 serum proteins (diluted with system diluent). For each IgG subclass, two concentrations were tested, both chosen to challenge the low end of the analytical measuring range and the extended low end. At each concentration, three samples were tested; three replicates per sample per run, for each of 5 runs. Testing was performed in one day with a single reagent lot, calibrator lot, instrument and operator and the mean and SD for the 15 replicates of each sample was calculated. The per sample bias (Bias = Mean - sample concentration) and the average bias, across the three samples, were calculated. The pooled SD across the three samples per the formula in Appendix B of CLSI EP17-A was calculated. The observed Total Analytical Error was calculated as Average Bias +2*Pooled SD. Based on the calculated Total Analytical Error, the LOQ was determined. The 6 {6} extended low range was calculated to measure low concentrations, such as those seen in children, for each of the IgG subclasses. ## IgG1 Concentrations tested were 2.2 g/L (initial assay range) and 0.44 g/L (extended low assay range). These levels were chosen to challenge the low end of the analytical measuring range (1:400 dilution) and the extended low range (1:20 dilution). The limit of quantitation (LoQ) was determined to be 2.2 g/L. The extended limit of quantitation for IgG1 was determined to be 0.44 g/L. ## IgG2 Concentrations tested were 0.44 g/L (initial assay range) and 0.11 g/L (extended low assay range). These levels were chosen to challenge the low end of the analytical measuring range (1:100 dilution) and the extended low range (1:5 dilution). The Limit of Quantitation (LoQ) for IgG2 was determined to be 0.44 g/L. The extended low range was determined to be 0.11 g/L. ## IgG3 Two concentrations were tested, 0.04 g/L (initial assay range) and 0.002 g/L (extended low assay range). These levels were chosen to challenge the low end of the analytical measuring range (1:2000 dilution) and the extended low range (1:20 dilution). The Limit of Quantitation (LoQ) for IgG3 was determined to be 0.04 g/L. The extended low range was determined to be 0.002 g/L. ## IgG4 Two concentrations were tested, 0.06 g/L (initial assay range) and 0.003 g/L (extended low assay range). These levels were chosen to challenge the low end of the analytical measuring range (1:200 dilution) and the extended low range (1:20 dilution). The Limit of Quantitation (LoQ) for IgG4 was determined to be 0.06 mg/L. The extended low range was determined to be 0.003 g/L. ## Hook effect The possibility of a hook effect occurring when using the IGG1-4 assays was evaluated. No hook effect was seen in concentrations up to: 78.6 g/L for IgG1; 51.9 g/L for IgG2; 35.7 g/L for IgG3; and 35.1 g/L for IgG4. ## e. Analytical specificity: Interference testing was performed according to CLSI/NCCLS EP7-A2 to determine the effect of various endogenous and exogenous substances on the Dimension Vista IGG1-4 assays. For all interferents except rheumatoid factors (RF), the percent bias was determined by testing a control sample without the interferent and comparing it to the value obtained from a test sample to which the potential interferent had been added. Test samples were prepared by spiking the potential interferent into serum. - IgG1 concentrations ranged from 4.79 - 12.8 g/L - IgG2 concentrations ranged from 1.48 - 7.31 g/L - IgG3 concentrations ranged from 0.225 - 0.781 g/L - IgG4 concentrations ranged from 0.471 - 1.892 g/L 7 {7} To evaluate interference from rheumatoid factors, samples with known levels of IgG subclasses 1-4, which had elevated RF concentrations and samples with no detectable RF concentration, were used to prepare samples for the study. One to one mixtures of samples of non-detectable levels of RF with samples with high concentrations of RF were prepared and the IGG1-4 assays concentrations determined in replicates of five on the Dimension® Vista System. For lipemic samples: special handling instructions include clarification by centrifugation (10 min at approximately $15,000\mathrm{~xg}$ ) prior to testing. Since specimens should be free of particulate matter, lipemic or turbid samples, which cannot be clarified by centrifugation, must not be used. Recovery for the IGG1-4 were tested and met the acceptance criteria of bias $&lt; 10\%$ . IgG1 | Substance Tested | Substance Concentration | IgG subclass 1 mg/dL [g/L] | Bias* % | | --- | --- | --- | --- | | Hemoglobin (hemolysate) | 1000 mg/dL [0.155 mmol/L] | 348 [3.48] | +1 | | | | 2151 [21.51] | +3 | | Bilirubin (unconjugated) | 60 mg/dL [1026 μmol/L] | 334 [3.34] | -2 | | | | 2138 [21.38] | -1 | | Bilirubin (conjugated) | 60 mg/dL [1026 μmol/L] | 340 [3.40] | -1 | | | | 2106 [21.06] | -3 | IgG2 | Substance Tested | Substance Concentration | IgG subclass 2 mg/dL [g/L] | Bias* % | | --- | --- | --- | --- | | Hemoglobin (hemolysate) | 1000 mg/dL [0.155 mmol/L] | 114 [1.14] | -1 | | | | 769 [7.69] | +2 | | Bilirubin (unconjugated) | 60 mg/dL [1026 μmol/L] | 118 [1.18] | +3 | | | | 747 [7.47] | +1 | | Bilirubin (conjugated) | 60 mg/dL [1026 μmol/L] | 115 [1.15] | 0 | | | | 747 [7.47] | +1 | IgG3 | Substance Tested | Substance Concentration | IgG subclass 3 mg/dL [g/L] | Bias* % | | --- | --- | --- | --- | | Hemoglobin (hemolysate) | 1000 mg/dL [0.155 mmol/L] | 23 [0.23] | +2 | | | | 119 [1.19] | +2 | | Bilirubin (unconjugated) | 60 mg/dL [1026 μmol/L] | 24 [0.24] | -3 | | | | 122 [1.22] | -4 | | Bilirubin (conjugated) | 60 mg/dL [1026 μmol/L] | 24 [0.24] | 0 | | | | 122 [1.22] | +2 | {8} IgG4 | Substance Tested | Substance Concentration | IgG subclass 4 mg/dL [g/L] | Bias* % | | --- | --- | --- | --- | | Hemoglobin (hemolysate) | 1000 mg/dL [0.155 mmol/L] | 8 [0.08] | -1 | | | | 351 [3.51] | -1 | | Bilirubin (unconjugated) | 60 mg/dL [1026 μmol/L] | 8 [0.08] | -3 | | | | 306 [3.06] | +9 | | Bilirubin (conjugated) | 60 mg/dL [1026 μmol/L] | 8 [0.08] | -3 | | | | 306 [3.06] | +9 | The following substances were determined not to interfere with the IGG1-4 method when present in serum and plasma at the concentrations indicated. Inaccuracies (biases) due to these substances are less than 10% at IgG 1 concentrations of 479 mg/dL to 1280 mg/dL [4.79 g/L to 12.8 g/L], IgG 2 concentrations of 148 mg/dL to 731 mg/dL [1.48 g/L to 7.31 g/L], IgG3 concentrations of 22.8 mg/dL to 72.0 mg/dL [0.228 g/L to 0.720 g/L], and IgG4 concentrations. of 42.9 mg/dL to 194.2 mg/dL [0.429 g/L to 1.942 g/L]. | Substance | Test Concentration | SI Units | | --- | --- | --- | | Acetaminophen | 20 mg/dL | 1328 μmol/L | | Amikacin | 15 mg/dL | 256 μmol/L | | Ammonium heparin | 3 U/mL | 3000 U/L | | Ampicillin | 5.3 mg/dL | 152 μmol/L | | Ascorbic acid | 5 mg/dL | 284 μmol/L | | Caffeine | 6 mg/dL | 308 μmol/L | | Carbamazepine | 3 mg/dL | 127 μmol/L | | Chloramphenicol | 5 mg/dL | 155 μmol/L | | Chlordiazepoxide | 5 mg/dL | 167 μmol/L | | Chlorpromazine | 0.2 mg/dL | 6.27 μmol/L | | Cholesterol | 500 mg/dL | 12.9 mmol/L | | Cimetidine | 2 mg/dL | 79.2 μmol/L | | Creatinine | 30 mg/dL | 2652 μmol/L | | Dextran | 6000 mg/dL | 1500 μmol/L | | Diazepam | 0.5 mg/dL | 17.6 μmol/L | | Digoxin | 5 ng/mL | 6.15 nmol/L | | Erythromycin | 6 mg/dL | 81.6 μmol/L | | Ethanol | 400 mg/dL | 86.8 mmol/L | | Ethosuximide | 25 mg/dL | 1770 μmol/L | | Furosemide | 6 mg/dL | 181 μmol/L | | Gentamicin | 12 mg/dL | 151 μmol/L | | Ibuprofen | 50 mg/dL | 2425 μmol/L | | Lidocaine | 1.2 mg/dL | 51.2 μmol/L | | Lithium chloride | 2.3 mg/dL | 3.2 mmol/L | | Lithium heparin | 3 U/mL | 3000 U/L | {9} | Substance | Test Concentration | SI Units | | --- | --- | --- | | Nicotine | 0.1 mg/dL | 6.2 μmol/L | | Penicillin | 25 U/mL | 25000 U/L | | Pentobarbital | 8 mg/dL | 354 μmol/L | f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: The new assays (Dimension Vista IGG 1-4) were compared to the predicate assays (N Antiserum to Human IgG subclasses 1 and 2 and the N Latex IgG3 and IgG4 on the BN Prospec System). Single determinations from 129-150 serum and plasma (lithium and sodium heparin) samples were run according to EP9-A2 on both the new and predicate assays for both the reportable and extended range. Passing-Bablok regression analysis was used to analyze the data for the initial measuring range. | Subclass | N= | Range tested mg/dL (g/L) | Slope (95% CI) | Intercept (95% CI) | r2 | | --- | --- | --- | --- | --- | --- | | IgG1 | 129 | 222 – 2285 | 1.002 | -0.165 | 0.988 | | | | (2.22 – 22.85) | (0.983, 1.022) | (-0.321, -0.027) | | | IgG2 | 147 | 63 – 762 | 1.000 | -0.020 | 0.988 | | | | (0.63 – 7.62) | (0.987,1.018) | (-0.070, 0.020) | | | IgG3 | 142 | 4 – 155.2 | 1.047 | -0.002 | 0.987 | | | | (0.040 – 1.552) | (1.030, 1.061) | (-0.005, 0.002) | | | IgG4 | 150 | 6.2 – 250 | 1.020 | -0.009 | 0.988 | | | | (0.062 – 2.50) | (1.004, 1.035) | (-0.015, -0.004) | | The performance of the low and high samples run with a new dilution was demonstrated by running a method comparison study versus the BN Prospec in the extended range low and high. The low extended range was validated by evaluating serum samples at a lower dilution. The high extended range was validated by evaluating serum samples at a higher dilution. Extended Low Range | Subclass | N= | Range tested mg/dL (g/L) | Slope (95% CI) | Intercept (95% CI) | r2 | | --- | --- | --- | --- | --- | --- | | IgG1 | 24 | 45-188 (0.45-1.88) | 1.000 (0.931, 1.073) | + 0.010 (-0.061, 0.061) | 0.984 | | IgG2 | 28 | 11-44 (0.11-0.44) | 1.000 (0.923-1.083) | 0.0 (-0.023-0.019) | 0.968 | | IgG3 | 23 | 0.3-3.3 (0.003-0.033) | 1.00 (1.00-1.00) | 0.0 (0.0-0.0) | 0.975 | | IgG4 | 29 | 0.3-3.7 (0.003-0.037) | 1.00 (0.97-1.00) | 0.0 (0.0-0.0) | 0.997 | Extended High Range {10} | Subclass | N= | Range tested mg/dL (g/L) | Slope (95% CI) | Intercept (95% CI) | r2 | | --- | --- | --- | --- | --- | --- | | IgG1 | 24 | 2784-9262 | 1.025 | -4.019 | 0.969 | | | | (27.84-92.62) | (0.961, 1.101) | (-7.733, -0.638) | | | IgG2 | 28 | 845-4068 | 0.998 | -0.639 | 0.956 | | | | (8.45.-40.68) | (0.928-1.054) | (-1.843-0.903) | | | IgG3 | 23 | 231.4.-534.8 | 1.008 | 0.159 | 0.961 | | | | (2.314-5.348) | (0.924-1.124) | (-0.292-0.457) | | | IgG4 | 29 | 300.4-939.7 | 1.004 | -0.135 | 0.934 | | | | (3.004-9.397) | (0.878-1.131) | (-0.733-0.452) | | # b. Matrix comparison: A separate study was performed using matched serum and plasma samples on the Dimension Vista® System. In this study, matched samples of serum, lithium heparin, sodium heparin and EDTA were tested on the Dimension Vista® System. For each plasma type, the % recovery and regression analyses compared to serum were performed. The acceptance criteria: a correlation coefficient of $\geq 0.950$ and a mean of % differences in recovery versus serum of $\leq 7\%$ . All samples met the acceptance criterion. IgG1 (n= 24 for low and high range) | | Linear Regression vs. Serum | | | % Recovery Statistics | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | | Li Hep | Na Hep | EDTA | | Li Hep | Na Hep | EDTA | | Slope | 0.96 | 1.00 | 1.01 | Mean | -2.5 | -3.2 | -3.1 | | Y-int | 0.13 | -0.22 | -0.28 | Min | -6.2 | -6.9 | -9.3 | | R | 0.999 | 1.000 | 0.999 | Max | 3.9 | 0.6 | 2.3 | | Syx: | 0.23 | 0.22 | 0.39 | | | | | | Slope 95% CI low | 0.938 | 0.982 | 0.980 | | | | | | Slope 95% CI high | 0.97 | 1.01 | 1.04 | | | | | IgG2 (n=28 for low and high ranges) | | Linear Regression vs. Serum | | | % Recovery Statistics | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | | Li Hep | Na Hep | EDTA | | Li Hep | Na Hep | EDTA | | Slope | 0.96 | 0.95 | 0.95 | Mean | -2.1 | -1.4 | -4.0 | | Y-int | 0.05 | 0.08 | 0.03 | Min | -4.5 | -5.0 | -9.5 | | R | 0.999 | 1.000 | 0.997 | Max | 1.7 | 2.8 | 2.7 | | Syx: | 0.09 | 0.08 | 0.20 | | | | | | Slope 95% CI low | 0.94 | 0.936 | 0.912 | | | | | | Slope 95% CI high | 0.98 | 0.97 | 1.00 | | | | | {11} IgG3 (n=23 for low range, n=30 for high range) | | Linear Regression vs. Serum | | | % Recovery Statistics | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | | Li Hep | Na Hep | EDTA | | Li Hep | Na Hep | EDTA | | Slope | 0.99 | 0.98 | 0.99 | Mean | -1.4 | -0.9 | -2.2 | | Y-int | -0.01 | 0.01 | -0.01 | Min | -6.6 | -3.4 | -9.1 | | R | 0.999 | 0.999 | 0.999 | Max | 2.5 | 1.9 | 2.0 | | Syx: | 0.02 | 0.02 | 0.02 | | | | | | Slope 95% CI low | 0.965 | 0.966 | 0.962 | | | | | | Slope 95% CI high | 1.02 | 1.00 | 1.02 | | | | | IgG4 (n=29 for low range and n=28 for high range) | | Linear Regression vs. Serum | | | % Recovery Statistics | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | | Li Hep | Na Hep | EDTA | | Li Hep | Na Hep | EDTA | | Slope | 1.00 | 1.02 | 0.99 | Mean | 1.0 | 2.2 | -1.9 | | Y-int | 0.00 | 0.00 | -0.01 | Min | -2.8 | -1.5 | -8.2 | | R | 1.000 | 1.000 | 0.999 | Max | 8.9 | 8.8 | 2.4 | | Syx: | 0.02 | 0.03 | 0.04 | | | | | | Slope 95% CI low | 0.994 | 1.000 | 0.967 | | | | | | Slope 95% CI high | 1.01 | 1.04 | 1.01 | | | | | 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: A literature reference was used for the IgG1-4 expected values: The expected values for the IgG1 method are: Age Group IgG1 (mg/dL) IgG1 [g/L] $&gt;1$ to $\leq 3$ years 265-938 2.65-9.38 $&gt;3$ to $\leq 6$ years 362-1228 3.62-12.28 $&gt;6$ to $\leq 12$ years 377-1131 3.77-11.31 $&gt;12$ to $\leq 18$ years 362-1027 3.62-10.27 $&gt;18$ years 405-1011 4.05-10.11 {12} The expected values for the IgG2 method are: | Age Group | IgG2 (mg/dL) | IgG2 [g/L] | | --- | --- | --- | | >1 to ≤ 3 years | 27.58 – 215.8 | 0.28 – 2.16 | | >3 to ≤ 6 years | 57.1 – 290.4 | 0.57 – 2.90 | | >6 to ≤ 12 years | 67.8 – 388.2 | 0.68 – 3.88 | | >12 to ≤ 18 years | 81.1 – 471.9 | 0.81 – 4.72 | | >18 years | 169 - 786 | 1.69 – 7.86 | The expected values for the IgG3 method are: | Age Group | IgG3 (mg/dL) | IgG3 [g/L] | | --- | --- | --- | | >1 to ≤ 3 years | 8.7 – 86.4 | 0.087 – 0.864 | | >3 to ≤ 6 years | 12.9 – 78.9 | 0.129 – 0.789 | | >6 to ≤ 12 years | 15.8 – 89.0 | 0.158 – 0.890 | | >12 to ≤ 18 years | 13.8 – 105.8 | 0.138 – 1.058 | | >18 years | 11.00 – 85.0 | 0.11 – 0.85 | The expected values for the IgG4 method are: | Age Group | IgG4 (mg/dL) | IgG4 (g/L) | | --- | --- | --- | | >1 to ≤ 3 years | 0.89 – 74.21 | 0.009 – 0.742 | | >3 to ≤ 6 years | 1.28 – 144.60 | 0.013 – 1.446 | | >6 to ≤ 12 years | 1.17 – 169.88 | 0.012 – 1.699 | | >12 to ≤ 18 years | 4.91 – 198.47 | 0.049 – 1.985 | | >18 years | 3.00 - 201 | 0.03 – 2.01 | The expected values were confirmed by running a reference interval transference study following the NCCLS/CLSI Guideline C28-A2. Twenty normal clinical samples were run in triplicate, and the mean of the three runs was calculated. The acceptance criterion for this study was that not more than two of twenty (10%) values could be outside the published range. The acceptance criterion was met for all of the assays within the adult and pediatric reference ranges. # N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. # O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...