APLIO XG DIAGNOSTIC ULTRASOUND SYSTEM, MODEL SSA - 790A VERSION 4.0R001
K092179 · Toshiba America Medical Systems, In.C · IYN · Jul 29, 2009 · Radiology
Device Facts
Record ID
K092179
Device Name
APLIO XG DIAGNOSTIC ULTRASOUND SYSTEM, MODEL SSA - 790A VERSION 4.0R001
Applicant
Toshiba America Medical Systems, In.C
Product Code
IYN · Radiology
Decision Date
Jul 29, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The Aplio v4.0R001 is intended to be used for the following type of studies: fetal, abdominal, intraoperative, pediatric, small organs, neonatal cephalic, adult cephalic, cardiac, transrectail, transvaginal, transesophageal, peripheral vascular and musculo-skeletal (both conventional and superficial).
Device Story
The Aplio XG v4.0R001 is a mobile diagnostic ultrasound system used in clinical settings by healthcare professionals. It utilizes various probes (linear, convex, sector) with frequencies from 2 MHz to 12 MHz to acquire ultrasound signals. The system processes these inputs to generate diagnostic images and fluid flow data. A key feature is the Elastography function, which provides information on the relative elasticity of lesions. The system supports multiple imaging modes, including B-mode, M-mode, Pulsed Wave Doppler, Color Doppler, and others. Clinicians view the output on the system display to assess tissue characteristics and blood flow, aiding in clinical decision-making and diagnosis. The device benefits patients by providing non-invasive diagnostic imaging and tissue elasticity assessment.
Clinical Evidence
Bench testing only. The device was evaluated for safety and performance in accordance with recognized standards including IEC 60601-1, IEC 60601-2-37, and AIUM-NEMA UD2/UD3 output standards. No clinical trial data was required for this 510(k) submission.
Technological Characteristics
Mobile ultrasound system; frequency range 2-12 MHz; supports flat linear, convex, and sector array probes. Complies with IEC 60601-1, IEC 60601-2-37, IEC 62304, and AIUM-NEMA UD2/UD3 standards. Software-based image processing; connectivity via standard ultrasound interfaces.
Indications for Use
Indicated for determining the relative elasticity of a lesion in patients undergoing diagnostic ultrasound imaging or fluid flow analysis.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Toshiba Ultrasound Diagnostic System Aplio XG MODEL SSA-790A V4.0 (K091295)
Siemens Medical Solutions Acuson S2000 Diagnostic Ultrasound System (K090334)
Submission Summary (Full Text)
{0}------------------------------------------------
10092179
510(k) Premarket Notification Aplio™ XG (v4.0R001) SSA-790A Ultrasound System
# 510(k) Summary
JUL 2 9 2009 Submitter's Name: Toshiba America Medical Systems, Inc. Address: PO Box 2068, 2441 Michelle Drive Tustin, CA 92781-2068 Contact: Paul Biggins, Director Regulatory Affairs Telephone No .: (714) 730-5000
Device Proprietary Name: Common Name: .
APLIO XG MODEL SSA-790A Version 4.0R001 Diagnostic Ultrasound System
# Classification:
- . Regulatory Class: II
- . Review Category: Tier II
- Ultrasonic Pulsed Doppler Imaging System Product Code: 90-IYN . [Fed. Reg. No .: 892.1550]
- Ultrasonic Pulsed Echo Imaging System Product Code: 90-IYO . [Fed. Reg.No .: 892.1560]
- Diagnostic Ultrasonic Transducer Product Code: 90-ITX t [Fed. Reg. No.: 892.1570]
# Identification of Predicate Devices:
Toshiba America Medical Systems believes that this device is substantially equivalent to:
- Toshiba Ultrasound Diagnostic System Aplio XG MODEL SSA-790A V4.0 510(k) K091295 .
- Siemens Medical Solutions Acuson S2000 Diagnostic Ultrasound System 510(k) K090334 .
# Device Description:
The Aplio XG v4.0R001 is a mobile system. This system is a Track 3 device that employs a wide array of probes that include flat linear array, convex linear array, and sector array with a frequency range of approximately 2 MHz to 12 MHz
### Intended Use:
The Aplio v4.0R001 is intended to be used for the following type of studies: fetal, abdominal, intraoperative, pediatric, small organs, neonatal cephalic, adult cephalic, cardiac, transrectail, transvaginal, transesophageal, peripheral vascular and musculo-skeletal (both conventional and superficial).
# Safety Considerations:
This device is designed and manufactured in conjunction with the Quality System Regulation, IEC 60601-1 (applicable portions), IEC 60601-1-1 (applicable portion), IEC 60601-1-2 (applicable portion), IEC 60601-1-4 (applicable portion), IEC60601-2-37 (applicable portions), IEC 62304 (applicable portion) and the AIUM-NEMA UD2 Output Measurement Standard as applied to Track 3 Ultrasound systems and the AIUM-NEMA UD3 Output Display Standard.
{1}------------------------------------------------
# Declaration of Conformity:
This device is designed and manufactured in conjunction with the Quality System Regulation, IEC 60601-1 (applicable portions), IEC 60601-1-1 (applicable portion), IEC 60601-1-2 (applicable portion), IEC 60601-1-4 (applicable portion), IEC 60601-2-37 (applicable portions), IEC 62304 (applicable porton) and the AIUM-NEMA UD2 Output Measurement Standard as applied to Track 3 Ultrasound systems and the AIUM-NEMA UD3 Output Display Standard a
## Standards Forms:
The standards forms are the same as those submitted for the Aplio XG v4.0 K091295.
# Clinical Trials Form
Please see the attached Clinical Trials Form FDA-3674
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird-like figure with outstretched wings.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Toshiba America Medical Systems, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313
JUL 2 9 2009
Re: K092179
Trade/Device Name: Elastography for SSA-790A Aplio XG v4.0R001 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasound pulsed doppler imaging system Regulatory Class: II Product Code: TYN, IYO, and ITX Dated: July 17, 2009 Received: July 21, 2009
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Elastography for SSA-790A Aplio XG v4.0R001 Diagnostic Ultrasound System, as described in your premarket notification:
Transducer Model Number
PLT-604AT PLT-704AT PLT-704SBT PLT-705BTH PLT-705BTF PLT-805AT PLT-1202S PLT-1204AT
PLT-1204BT PLT-1204AX PLT-1204BX PLT-1204MV PET-510MB PC-20M PC-50M PST-25BT
{3}------------------------------------------------
| PST-30BT | PVT-575MV |
|-----------|------------|
| PST-50AT | PVT-674BT |
| PST-65AT | PVT-661VT |
| PVT-375BT | PVT-681MV |
| PVT-375MV | PVT-745BTV |
| PET-382BT | PVT-770RT |
| PVT-382MV | |
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
If you have any questions regarding the content of this letter, please contact Sunder Rajan at (240) 276-3666.
Sincerely yours,
Janine M. Morris
Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{4}------------------------------------------------
510(k) Number (if known):
510(k) Premarket Notification Aplio™ XG (v4.0R001) SSA-790A Ültrasound System
Ko92179
Device Name: Elastography for SSA-790A Aplio XG v4.0R001 Diagnostic Ultrasound System
Indications for Use:
Elastography is indicated for determining the relative elasticity of a lesion.
Prescription Use Only (Per 21 CRF801.109)
(PLEASE DO NOT WRITE BELOW THIS LINE · CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices 510(k) Number.
{5}------------------------------------------------
### 510(k) Premarket Notification Aplio™ XG (v4.0R001) SSA-790A Ultrasound System
### System: _ Aplio XG v4.0R001 SSA-790A Transducer: __________________________________________________________________________________________________________________________________________________________________
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | | | | |
|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|-----|-----------------|-------|-----------|----|-----------------|--|
| Specific<br>(Tracks 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | THI | Dynamic<br>Flow | Power | CHI<br>2D | 4D | Other<br>[Note] | |
| Ophthalmic | | | | | | | | | | | | | |
| Fetal | P | P | P | P | P | 2 | P | P | P | P | P | 5,7,8,9,10 | |
| Abdominal | P | P | P | P | P | 2,3 | P | P | P | P | P | 5,7,8,9,10 | |
| Intra-operative (Abdominal) | P | P | P | P | P | 2 | P | P | P | | | 4,5 | |
| Intra-operative (Neuro) | | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | | |
| Pediatric | P | P | P | P | P | 2,3 | P | P | P | P | P | 5,7,8,9,10 | |
| Small Organ (Note 1) | P | P | P | P | P | 2 | P | P | P | | | 4,5,6,7,9,1 | |
| Neonatal Cephalic | P | P | P | P | P | 3 | P | P | P | P | | | |
| Adult Cephalic | P | P | P | P | P | 3 | P | P | P | P | | | |
| Trans-rectal | P | P | P | P | P | 2 | P | P | P | P | P | 4,5,7,11 | |
| Trans-vaginal | P | P | P | | P | 2 | P | P | P | P | P | 4,5,7 | |
| Trans-urethral | | | | | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | P | P | P | | P | 2 | P | P | P | P | | 4,5,6,7,9,1 | |
| Musculo-skeletal (Superficial) | P | P | P | | P | 2 | P | P | P | | | 4,5,6,7,9,1 | |
| Intravascular | | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | | |
| Cardiac Adult | P | P | P | P | P | 3 | P | P | P | P | | 4 | |
| Cardiac Pediatric | P | P | P | P | P | 3 | P | P | P | P | | 4 | |
| Intravascular (Cardiac) | | | | | | | | | | | | | |
| Trans-esoph. (Cardiac) | P | P | P | P | P | 3 | P | | | | | 4 | |
| Intra-cardiac | | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | | |
| Peripheral vessel | P | P | P | P | P | 2 | P | P | P | P | | 4,5,6,7,9,1 | |
| Other (Specify) | | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K091295
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
Note 3 Combined mode includes Birl, BDTMDF, BDFMDFF, BDF/MDF/P WD
Note 3 Combined mode includes B/M; B/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD; BDF/CWD; BDF/CWD; BDF/CWD; Note 4 TDI
.. .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Note 5 ApliPure
Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D color
Note 10 STIC Color
Note 11 Elastography (New)
vision Sign-Off Division of Reproductive, Abdominal, and Radiological De 510(k) Number
{6}------------------------------------------------
### 510(k) Premarket Notification Aplio™ XG (v4.0R001) SSA-790A Ultrasound System
#### System: Aplio XG v4.0R001 SSA-790A Transducer: PLT-604AT
Intended Use: Diagnostic ultrasound imaging or Nuid flow analysis of the human body as follows:
| Clinical Application<br>Specific<br>(Tracks 3) | Mode of Operation<br>B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify)* | THI | Dynamic<br>Flow | Power | CHI<br>2D | 3D | Other<br>[Note] |
|------------------------------------------------|------------------------|---|-----|-----|------------------|------------------------|-----|-----------------|-------|-----------|----|-----------------|
| Ophthalmic | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | | |
| Intra-operative (Abdominal) | | | | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | | |
| Small Organ (Specify) (1) | P | P | P | | P | 2 | P | P | P | P | | 5 |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | P | P | P | | P | 2 | P | P | P | P | | 5 |
| Musculo-skeletal (Superficial) | P | P | P | | P | 2 | P | P | P | P | | 5 |
| Intravascular | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
| Cardiac Adult | | | | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | | | | |
| Intra-cardiac | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
| Peripheral vessel | P | P | P | | P | 2 | P | P | P | P | | 5 |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous SIO(k) of the transducer: K091295
Note { Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes BM; BPWD; BDF/MDF; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD; BDF/CWD; BDF/CWD; BDF/CWD; BDF/CWD; BDF/CWD; BDF/CWD; BDF/CWD; BDF/CWD; BDF/CWD; BDF/C Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D color Note 10 STIC Color Note 11 Elastography (New)
Prescription Use Only (Per 21 CRF801 109)
signature
(Division Sign-Off) Division of Reproductive, Abdomin and Radiological Devices 510(k) Number
{7}------------------------------------------------
#### 510(k) Premarket Notification Aplio™ XG (v4.0R001) SSA-790A Ultrasound System
System: Aplio XG v4.0R001 SSA-790A Transducer:___PLT-704AT
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | |
|---------------------------------|-------------------|---|---------|---------------|--------------------|-----|--------------|----------|--------|--------------|
| Specific (Tracks 3) | B | M | PWD/CWD | Color Doppler | Combined (Specify) | THI | Dynamic Flow | Power 2D | CHI 3D | Other [Note] |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intra-operative (Abdominal) | | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (Specify) (1) | P | P | P | P | 2 | P | P | P | P | 5 |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | | |
| Musculo-skeletal (Conventional) | P | P | P | P | 2 | P | P | P | P | 5 |
| Musculo-skeletal (Superficial) | P | P | P | P | 2 | P | P | P | P | 5 |
| Intravascular | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | |
| Cardiac Adult | | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | | |
| Intra-cardiac | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | |
| Peripheral vessel | P | P | P | P | 2 | P | P | P | P | 5 |
| Other (Specify) | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K091295
Note I Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D color Note 10 STIC Color Note 11 Elastography (New)
Prescription Use Only (Per 2) CRF801.109)
Hee Rem
(Division Šign-Off) Division of Reproductive, Abdominal. and Radiological Device
510(k) Number
{8}------------------------------------------------
### 510(k) Premarket Notification Aplio™ XG (v4.0R001) SSA-790A Ultrasound System
System: Aplio XG v4.0R001 SSA-790A Transducer:_ PLT-704SBT
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | | | |
|------------------------------------|-------------------|---|-----|-----|------------------|------------------------|-----|-----------------|-------|-----------|----|-----------------|
| Specific<br>(Tracks 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify)* | THI | Dynamic<br>Flow | Power | CHI<br>2D | 3D | Other<br>[Note] |
| Ophthalmic | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | | |
| Intra-operative (Abdominal) | | | | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | | |
| Small Organ (Specify) (1) | P | P | P | | P | 2 | P | P | P | P | | 5 |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | P | P | P | | P | 2 | P | P | P | P | | 5 |
| Musculo-skeletal (Superficial) | P | P | P | | P | 2 | P | P | P | P | | 5 |
| Intravascular | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
| Cardiac Adult | | | | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | | | | |
| Intra-cardiac | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
| Peripheral vessel | P | P | P | | P | 2 | P | P | P | P | | 5 |
| Other (Specify) | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K091295
Note 1 Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes BM, BPWD; WD; BDF/MDF; BDF/MDFf/PWD; 2D/CWD; BDF/CWD; BDF/CWD; BDF/CWD; BDF/CWD; BDF/CWD; BDF/CWD; BDF/CWD; BDF/CWD; BDF/CWD; BDF/CWD; BDF/CWD; Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D color Note 10 STIC Color . (Division Sign-Off) Note 11 Elastography (New)
Division of Reproductive, Abdominal, and Radiological De 510(k) Number
{9}------------------------------------------------
# 510(k) Premarket Notification Aplio™ XG (v4.0R001) SSA-790A Ultrasound System
#### System: _ Aplio XG v4.0R001 SSA-790A Transducer: PLT-705BTH
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | | | |
|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|-----|-----------------|-------|-----------|----|-----------------|
| Specific<br>(Tracks 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | THI | Dynamic<br>Flow | Power | CHI<br>2D | 3D | Other<br>[Note] |
| Ophthalmic | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | |
| Abdominal | P | P | P | | P | 2 | P | P | P | P | | 5 |
| Intra-operative (Abdominal) | P | P | P | | P | 2 | P | P | P | P | | 5 |
| Intra-operative (Neuro) | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
| Pediatric | P | P | P | | P | 2 | P | P | P | P | | 5 |
| Small Organ (Specify) (1) | P | P | P | | P | 2 | P | P | P | P | | 5 |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | | | | | | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
| Cardiac Adult | | | | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | | | | |
| Intra-cardiac | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K091295
Note I Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M, B/PWD; BDF/PWD; BDF/MDF; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D color Note 10 STIC Color Note 11 Elastography (New)
Prescription Use Only (Per 21 CRF801.109)
Heflen
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Device
510(k) Number
{10}------------------------------------------------
#### 510(k) Premarket Notification Aplio™ XG (v4.0R001) SSA-790A Ultrasound System
System: ______________________________________________________________________________________________________________________________________________________________________ Aplio XG v4.0R001 SSA-790A Transducer:_ PLT-705BTF
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | | |
|------------------------------------|-------------------|---|---------|------------------|------------------------|-----|-----------------|-------|-----------|----|-----------------|
| Specific<br>(Tracks 3) | B | M | PWD CWD | Color<br>Doppler | Combined<br>(Specify)* | THI | Dynamic<br>Flow | Power | CHI<br>2D | 3D | Other<br>[Note] |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | | | | | | | | | | |
| Abdominal | P | P | P | P | 2 | P | P | P | P | | 5 |
| Intra-operative (Abdominal) | P | P | P | P | 2 | P | P | P | P | | 5 |
| Intra-operative (Neuro) | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Pediatric | P | P | P | P | 2 | P | P | P | P | | 5 |
| Small Organ (Specify) (1) | P | P | P | P | 2 | P | P | P | P | | 5 |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | | | | | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | |
| Cardiac Adult | | | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | | | |
| Intra-cardiac | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K091295
. .
Note 1 Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D color Note 10 STIC Color Note 11 Elastography (New)
Prescription Use Only (Per 21 CRF801.109)
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devi 510(k) Number
{11}------------------------------------------------
# 510(k) Premarket Notification Aplio™ XG (v4.0R00!) SSA-790A Ultrasound System
System: Aplio XG v4.0R001 SSA-790A Transducer:_ PLT-805AT -----
lntended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | Other<br>(Note) | |
|------------------------------------|-------------------|---|---------|------------------|-----------------------|-----|-----------------|-------|-----------|-----------------|-------|
| Specific<br>(Tracks 3) | B | M | PWD CWD | Color<br>Doppler | Combined<br>(Specify) | THI | Dynamic<br>Flow | Power | CHI<br>2D | | 3D |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | |
| Intra-operative (Abdominal) | | | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ (Specify) (1) | P | P | P | P | 2 | P | P | P | P | | 5,6,7 |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | P | P | P | P | 2 | P | P | P | P | | 5,6,7 |
| Musculo-skeletal (Superficial) | P | P | P | P | 2 | P | P | P | P | | 5,6,7 |
| Intravascular | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | |
| Cardiac Adult | | | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | | | |
| Intra-cardiac | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | |
| Peripheral vessel | P | P | P | P | 2 | P | P | P | P | | 5,6,7 |
| Other (Specify) | | | | | | | | | | | |
.. ... ...
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K091295
….
i
Note 1 Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD; BDF/CWD; BDF/CWD; BDF/CWD; BDF/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D color Note 10 STIC Color Note II Elastography (New)
. .
Hui Rem
(Division Sign-Off)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K092179
{12}------------------------------------------------
.
### 510(k) Premarket Notification Aplio™ XG (v4.0R001) SSA-790A Ultrasound System
System: _ Aplio XG v4.0R001 SSA-790A Transducer:____PLT-1202S
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | | | |
|------------------------------------|-------------------|---|---------|------------------|------------------------|-----|-----------------|-------|-----------|----|-----------------|--|
| Specific<br>(Tracks 3) | B | M | PWD CWD | Color<br>Doppler | Combined*<br>(Specify) | THI | Dynamic<br>Flow | Power | CHI<br>2D | 3D | Other<br>[Note] | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | | |
| Intra-operative (Abdominal) | P | P | P | P | 2 | P | P | P | | | 4,5,11 | |
| Intra-operative (Neuro) | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | | |
| Small Organ (Specify) (1) | P | P | P | P | 2 | P | P | P | | | 4,5,11 | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | P | P | P | P | 2 | P | P | P | | | 4,5,1 | |
| Musculo-skeletal (Superficial) | P | P | P | P | 2 | P | P | P | | | 4,5,11 | |
| Intravascular | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
| Cardiac Adult | | | | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | | | | |
| Intra-cardiac | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
| Peripheral vessel | P. | P | P | P | 2 | P | P | P | | | 4,5,1 | |
| Other (Specify) | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K091295
Note I Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note & STIC Note 9 3D color Note 10 STIC Color Note 11 Elastography (New)
Prescription Use Only (Per 21 CRF801.109)
·
Hut Rumen
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devic
51.0(k) Number.
{13}------------------------------------------------
### 510(k) Premarket Notification Aplio™ XG (v4.0R001) SSA-790A Ultrasound System
#### System: _ Aplio XG v4.0R00I SSA-790A PLT-1204AT____ Transducer:
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | | |
|------------------------------------|-------------------|---|---------|------------------|-----------------------|-----|-----------------|-------|-----------|----|-----------------|
| Specific<br>(Tracks 3) | B | M | PWD CWD | Color<br>Doppler | Combined<br>(Specify) | THI | Dynamic<br>Flow | Power | CHI<br>2D | 3D | Other<br>[Note] |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | |
| Intra-operative (Abdominal) | | | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ (Specify) (1) | P | P | P | P | 2 | P | P | P | | | 4,5,6,1 |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | P | P | P | P | 2 | P | P | P | | | 4,5,6,1 |
| Musculo-skeletal (Superficial) | P | P | P | P | 2 | P | P | P | | | 4,5,6,1 |
| Intravascular | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | |
| Cardiac Adult | | | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | | | |
| Intra-cardiac | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | |
| Peripheral vessel | P | P | P | P | 2 | P | P | P | | | 4,5,6,1 |
| Other (Specify) | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K091295
.. . ...
Note I Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D color Note 10 STIC Color Note 11 Elastography (New)
. .
.......
Prescription Use Only (Per 21 CRF801.109)
Dision Sign-Off) (Division of Reproductive, Abdominal, and Radiological Devi 510(k) Number
{14}------------------------------------------------
#### System: Aplio XG v4.0R001 SSA-790A Transducer: _ PLT-1204BT
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | | | |
|------------------------------------|-------------------|---|-----|-----|------------------|----------------------------|-----|-----------------|-------|-----------|----|-----------------|
| Specific<br>(Tracks 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify)<br>* | THI | Dynamic<br>Flow | Power | CHI<br>2D | 3D | Other<br>[Note] |
| Ophthalmic | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | | |
| Intra-operative (Abdominal) | | | | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | | |
| Small Organ (Specify) (1) | P | P | P | | P | 2 | P | P | P | | | 4,5,6,1 |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | P | P | P | | P | 2 | P | P | P | | | 4,5,6,1 |
| Musculo-skeletal (Superficial). | P | P | P | | P | 2 | P | P | P | | | 4,5,6,1 |
| Intravascular | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
| Cardiac Adult | | | | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | | | | |
| Intra-cardiac | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
| Peripheral vessel | P | P | P | | P | 2 | P | P | P | | | 4,5,6, |
| Other (Specify) | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K091295
Note 1 Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D color Note 10 STIC Color Note 11 Elastography (New)
Prescription Use Only (Per 21 CRF801.109)
Hellem
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devic 510(k) Number.
) 10(k). Numbe
{15}------------------------------------------------
# 510(k) Premarket Notification
Aplio™ XG (v4.0R001) SSA-790A Ultrasound System
Aplio XG v4.0R001 SSA-790A System: Transducer: PLT-1204AX____________________________________________________________________________________________________________________________________________________________________
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | Other<br>[Note] | | | | |
|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|-------|-----------------|-----------------|-------|-----------|---------|
| Specific<br>(Tracks 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | THI * | | Dynamic<br>Flow | Power | CHI<br>2D | 3D |
| Ophthalmic | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | | |
| Intra-operative (Abdominal) | | | | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | | |
| Small Organ (Specify) (1) | P | P | P | | P | 2 | P | P | P | | | 4,5,6,1 |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | P | P | P | | P | 2 | P | P | P | | | 4,5,6,1 |
| Musculo-skeletal (Superficial) | P | P | P | | P | 2 | P | P | P | | | 4,5,6,1 |
| Intravascular | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
| Cardiac Adult | | | | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | | | | |
| Intra-cardiac | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
| Peripheral vessel | P | P | P | | P | 2 | P | P | P | | | 4,5,6,1 |
| Other (Specify) | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous SIO(k) of the transducer: K091295
Note 1 Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD. Note 3 Combined mode includes B/M; B/PWD; BDF/MDF; BDF/MDF; BDF/MDF/PWD; BDF/CWD; BDF/CWD; BDF/CWD; BDF/CWD; BDF/CWD; BDF/CWD; BDF/CWD; BDF/CWD; BDF/CWD; BDF/CWD; BDF/CWD; BD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D color Note 10 STIC Color Note 11 Elastography (New)
Prescription Use Only (Per 21 CRF801.109)
H.D. Loum
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devic 510(k) Number
{16}------------------------------------------------
# 510(k) Premarket Notification
Aplio™ XG (v4.0R001) SSA-790A Ultrasound System
#### System: Aplio XG v4.0R001 SSA-790A Transducer: PLT-1204BX
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
......
| Clinical Application<br>Specific<br>(Tracks 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | THI | Dynamic<br>Flow | Power | CHI<br>2D | 3D | Other<br>[Note] | |
|------------------------------------------------|---|---|-----|-----|------------------|-----------------------|-----|-----------------|-------|-----------|----|-----------------|-----------|
| Ophthalmic | | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | | | |
| Intra-operative (Abdominal) | | | | | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | | | | | |
| Laparoscopic…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.