The Phaser Lipoplasty System is intended to be used for liquefaction and aspiration of localized subcutaneous fatty deposits for the purpose of aesthetic body contouring. The Phaser Lipoplasty System is indicated for use in aesthetic body contouring.
Device Story
Phaser Lipoplasty System is a surgical liposuction device; utilizes pressurized, heated, and pulsed saline streams (Phaser stream) via specialized cannula to transition adipose tissue from solid to liquid/gel state for removal. Used by physicians in clinical settings for aesthetic body contouring. Device allows standard surgical liposuction techniques; low pressure/temperature operation prevents thermal damage to non-target tissues. Physician observes surgical field and controls device to facilitate fat removal; benefits patient through effective body contouring with minimized thermal risk.
Clinical Evidence
Bench testing and in-vivo porcine model study conducted. Porcine study assessed safety, heat generation, and thermal damage compared to suction assisted lipoplasty (SAL). Results showed no histological evidence of thermal damage for Phaser or SAL. Average skin temperatures: Phaser 31.4–35.8°C vs. SAL 26.2–31.1°C. Average cannula temperatures: Phaser 36.7–40.2°C vs. SAL 28.5–33.8°C. Simulated use study evaluated human factors, user interface, and environment, confirming device functioned as intended.
Technological Characteristics
Surgical liposuction system; utilizes pressurized, heated, and pulsed saline streams delivered through a specialized cannula. Operates at low pressure and temperature to liquefy adipose tissue. Class II device (21 CFR 878.5040).
Indications for Use
Indicated for aesthetic body contouring via liquefaction and aspiration of localized subcutaneous fatty deposits.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
Byron Medical, Liposuction Aspiration and Tumescent Infiltration Cannulae/Needles (K981172)
HK Surgical Inc., HK Liposuction Aspirator III (K032802)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.
Andrew Technologies, LLC Jonathan Kahan 555 13th St., NW Washington, District of Columbia 20004
June 8, 2021
Re: K092066
Trade/Device Name: Phaser Lipoplasty System (dur-01), Phaser Disposable Kit (dis-01), Phaser Waste Collection Kit (was-01), Phaser Cannulae Regulation Number: 21 CFR 878.5040 Regulation Name: Suction lipoplasty system Regulatory Class: Class II Product Code: QPB
Dear Jonathan Kahan:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated April 12, 2010. Specifically, FDA is updating this SE Letter because FDA has created a new product code to better categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Cindy Chowdhury, OHT4: Office of Surgical and Infection Control Devices, 240-402-6647, Cindy.Chowdhury@fda.hhs.gov.
Sincerely,
# Cindy Chowdhury -S
Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image is a circular logo for the U.S. Department of Health and Human Services. The logo features the department's symbol, which is a stylized caduceus with three lines representing the three branches of government. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the symbol in a circular fashion.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Andrew Technologies, LLC % Hogan & Hartson LLP Mr. Luis Davila 555 13th Street, Northwest Washington, District of Columbia 20004
APR 1 2 2010
Re: K092066
Trade Name: Phaser™ Lipoplasty System Regulation Number: 21 CFR 878.5040 Regulation Name: Suction lipoplasty system Regulatory Class: II Product Code: MUU Dated: April 9, 2010 Received: April 9, 2010
Dear Mr. Davila:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{2}------------------------------------------------
Page 2 - Mr. Luis Davila
forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
erely yours,
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
092066 y/
# Indications for Use Statement
510(k) Number (if known): K092066
Device Name: Phaser TM Lipoplasty System
Indications for Use:
The Phaser Lipoplasty System is intended to be used for liquefaction and aspiration of localized subcutaneous fatty deposits for the purpose of aesthetic body contouring.
The Phaser Lipoplasty System is indicated for use in aesthetic body contouring.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David Krone for MXM
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K092066
\\\DC - 030929/000002 - 3023709 v2
{4}------------------------------------------------
## 510(k) SUMMARY
## Andrew Technologies' Phaser Lipoplasty System (K092066)
Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Contact Person: Luis Davila Phone: 770-715-9258 Facsimile: 770-886-6585
APR 1 2 2010
Date Prepared: April 12, 2010
Name of Device and Name/Address of Sponsor:
Phaser™ Lipoplasty System
Andrew Technologies. LLC 3 South Haddon Avenue, Suite 3 Haddonfield, NJ 008033
Common or Usual Name: Suction Lipoplasty System
Classification Name: Suction Lipoplasty System, Class II, 21 C.F.R. § 878.5040 (MUU)
#### Predicate Devices
- 1. Alcon Laboratories, Cataract Liquefracture Device (K980292)
- 2. Misonix Inc., Lysonix 2000/3000 Ultrasonic Surgical Aspiration Systems (K041058)
- 3. Soring GmbH Medizintechnik, Sonoca-Lipo (K052183)
- 4. Byron Medical, PSI-Tec Liposuction Aspirator (K981215)
- 5. Byron Medical , PSI-Tec Syringe Infusion Pump (K040149)
- 6. Byron Medical, Liposuction Aspiration and Tumescent Infiltration Cannulae/Needles (K981172)
- 7. HK Surgical Inc., HK Liposuction Aspirator III (K032802)
## Intended Use / Indications for Use
The Phaser Lipoplasty System is intended to be used for liquefaction and aspiration of localized subcutaneous fatty deposits for the purpose of aesthetic body contouring.
The Phaser Lipoplasty System is indicated for use in aesthetic body contouring.
## Technological Characteristics
The Phaser Lipoplasty System is a surgical liposuction device that utilizes pressurized, heated and pulsed saline streams ("the Phaser stream") in conjunction with a specialized
{5}------------------------------------------------
cannula to transition adipose tissue from a solid state to a liquid or gel state so as to facilitate its removal from the body. The physician can employ standard surgical liposuction techniques when using the device, the low pressure and temperature allow adipose tissue removal without burning or otherwise damaging non-target tissues within the surgical field.
# Performance Data
The company conducted in-vivo testing in a porcine model to assess the safety and performance of the device and the heat generated at the entrance/puncture site, comparing these results to concurrent testing performed with suction assisted lipoplasty (SAL). Temperature readings were taken from worst case locations on the skin surface and on the highest temperature region of the cannula. In addition, tissue was examined histologically to assess potential thermal damage. The data gathered in this study confirms that there is no thermal damage association with the use of the Phaser, with the impact on the Phaser test animals' skin directly comparable to the effects seen with SAL. Specifically, average skin temperatures with the Phaser ranged from 31.4 to 35.8℃ as compared to 26.2 to 31.1℃ for SAL. Average Phaser cannula temperatures ranged from 36.7 to 40.2℃, with corresponding SAL cannula temperatures of 28.5 to 33.8℃. Importantly, no histological evidence of thermal damage was seen with either the Phaser or SAL.
A simulated use study to address the use environment and the device user interface was also performed. The study evaluated human factor considerations, including the device's use environment, user capabilities, and the product's user interface. These evaluations demonstrated that the device performed and functioned as intended.
# Substantial Equivalence
The Phaser Lipoplasty System has the same intended uses and similar indications, technological characteristics, and principles of operation as the following predicate devices: the Misonix Inc. Lysonix 2000/3000 Ultrasonic System (K041058); the Soring GmbH, Sonoca Lipo (K052183) ("Sonoca-Lipo"); the Alcon Cataract Liquefracture Device (K980292); the PSI-Tec Peristaltic Infiltration Pump, Byron Medical Inc. (K040149); the Byron Medical Inc. PS1-Tec Liposuction Aspirator (K981215); the Byron Medical Inc. Byron Liposuction Aspiration and Tumescent Infiltration Cannulae/Needles (K981172); and the HK Surgical Inc., HK Liposuction Aspirator III (K032802). The minor technological differences between the Phaser Lipoplasty System and its predicate devices raise no new issues of safety or effectiveness. Performance data demonstrate that the Phaser Lipoplasty System has comparable performance and safety as compared to the Lysonix 2000/3000, Sonoca-Lipo and the AquaLase. Thus, the Phaser Lipoplasty System is substantially equivalent to its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.