← Product Code [DJR](/productcode/DJR) · K092048

# ONE STEP EDDP (METHADONE METABOLITE) TEST STRIP (K092048)

_Biosite Incorporated · DJR · Apr 19, 2010 · Clinical Toxicology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K092048

## Device Facts

- **Applicant:** Biosite Incorporated
- **Product Code:** [DJR](/productcode/DJR.md)
- **Decision Date:** Apr 19, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.3620
- **Device Class:** Class 2
- **Review Panel:** Clinical Toxicology
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K092048 · Apr 19, 2010 | ONE STEP EDDP (METHADONE METABOLITE) TEST STRIP | Biosite Incorporated | Clinical urine specimens from three external clinical study sites | The device performance was evaluated by comparing results from clinical urine specimens against the reference method (GC/MS) to determine accuracy and agreement. | Clinical accuracy; Urine specimens; GC/MS comparison |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Clinical Study; Retrospective comparison of clinical urine specimens against reference method | Clinical urine specimens; Sample Size: 554; Number of Sites: 3 | GC/MS Analysis | Percent agreement with GC/MS analysis |

## Indications for Use

The One Step EDDP (methadone Metabolite) Test Strip is a rapid immunochromatographic assays, for the qualitative detection of 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP), an inactive metabolite of methadone, at a designates cutoff concentration of 300 ng/mL. This product is used to obtain a visual, qualitative result and is intended for professional use and professionals at point of care sites. This assay provides only preliminary results. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

## Device Story

Rapid immunochromatographic test strip; detects EDDP (methadone metabolite) in human urine. Principle: competitive binding immunoassay; urine sample migrates via capillary action; EDDP competes with drug conjugate for antibody-coated particle binding sites. Output: visual colored line in test region; line presence indicates EDDP below 300 ng/mL; absence indicates EDDP above 300 ng/mL. Used at point-of-care sites by professionals. Provides preliminary results; requires GC/MS confirmation for clinical decision-making. Benefits: rapid screening for methadone metabolite presence.

## Clinical Evidence

Clinical evaluation performed at three external sites comparing test strip results to GC/MS analysis. Total sample size of 554. Overall agreement with GC/MS was 98% (95% CI: 96% - 99%). Analytical sensitivity study demonstrated >99% accuracy at 50% above and below the 300 ng/mL cutoff.

## Technological Characteristics

Rapid immunochromatographic immunoassay; competitive binding principle; lateral flow test strip; visual readout; urine specimen input; no electronic components or software; no external energy source required.

## Regulatory Identification

A methadone test system is a device intended to measure methadone, an addictive narcotic pain-relieving drug, in serum and urine. Measurements obtained by this device are used in the diagnosis and treatment of methadone use or overdose and to determine compliance with regulations in methadone maintenance treatment.

## Special Controls

*Classification.* Class II (special controls). A methadone test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

## Predicate Devices

- ACON MTD One Step Methadone Test Strip (k012595)
- DRI Methadone Metabolite Enzyme Assay (k023617)
- Bionexia Single and Multi-Strip Cassette/Dipstick DOA Screen Panels (k081378)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE

A. 510(k) Number:
k092048

B. Purpose for Submission:
New device

C. Measurand:
Methadone (EDDP)

D. Type of Test:
Qualitative immunochromatographic assay

E. Applicant:
BioSite, Inc

F. Proprietary and Established Names:
One Step EDDP (Methadone Metabolite) Test Strip

G. Regulatory Information:

|  Product Code | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  DJR- Methadone Test System | II | 862.3620 | 91, Toxicology  |

H. Intended Use:

1. Intended use(s):
See Indications for Use below

2. Indication(s) for use:
The One Step EDDP (methadone Metabolite) Test Strip is a rapid immunochromatographic assays, for the qualitative detection of 2-ethylidene-1,5-

{1}

dimethyl-3,3-diphenylpyrrolidine (EDDP), an inactive metabolite of methadone, at a designates cutoff concentration of 300 ng/mL. This product is used to obtain a visual, qualitative result and is intended for professional use and professionals at point of care sites.

This assay provides only preliminary results. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method.

Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

3. Special conditions for use statement(s):

For Professional prescription use and point-of-care use

4. Special instrument requirements:

Not applicable, as the devices are visually-read single-use devices.

I. Device Description:

The One Step EDDP (Methadone Metabolite) Test Strip contains 25 test devices packaged in a desiccant sealed pouch and a package insert (directions for use).

J. Substantial Equivalence Information:

1. Predicate device name(s):

ACON MTD One Step Methadone Test Strip
DRI Methadone Metabolite Enzyme Assay, Microgenics Corporation
Bionexia Single and Multi-Strip Cassette/Dipstick DOA Screen Panels, Applied DNA Technologies Inc.

2. Predicate K number(s):

k012595, k023617 and k081378 respectively

{2}

3. Comparison with predicate:

|  Similarities/Differences  |   |   |   |   |
| --- | --- | --- | --- | --- |
|  Item | Device | ACON MTD strip and device | Bionexia Single and Multi-Strip Cassette/Dipstick | DRI Methadone Metabolite  |
|  Intended Use | for the qualitative detection of 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP), an inactive metabolite of methadone in human urine | Same | Same | Same  |
|  Test Principle | Competitive Binding immunoassay | Same | Same | Liquid Homogeneous Enzyme Assay  |
|  Sample | Urine | Same | Same | Same  |
|  Antibody | Monoclonal | Same | Same | Same  |
|  Cutoff | 300 ng/mL | Same | 100 ng/mL | Same  |
|  Tracer | Ab-Colloidal Gold complex | Same | Same | Glucos-6-Phosphate Dehydrogenase  |
|  Analyte | 2-ethylidene-1,5-dimethyl-3,3-diphenylpyr-rolidine (EDDP) | Methadone | Same | Same  |

K. Standard/Guidance Document Referenced (if applicable):

None were identified.

L. Test Principle:

The One Step EDDP (Methadone Metabolite) Test Strip is an immunoassay based on the principle of competitive binding. Drugs which may be present in the urine specimen compete against the drug conjugate for binding sites on the antibody. During testing, a urine sample migrates upward by capillary action. EDDP, if present in the urine sample below $300\mathrm{ng/mL}$, will not saturate the binding sites of antibody-coated particles in the test strip. The antibody-coated particles will then be captured by immobilized EDDP conjugate and a visible colored line will show up in the test region. The colored line will not form in the test region if the EDDP levels exceed $300\mathrm{ng/mL}$ because it will saturate all the binding sites of anti-EDDP antibodies.

{3}

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

Precision studies were performed at three Physician's office laboratories using drug free urine spiked to the following concentrations: 0, 150, 225, 375 and 450 ng/mL. The samples were aliquots, coded, randomized and blinded, than given to each site. A total of 45 determinations were made at each concentration. Testing was performed once a day for three days by three operators (1 per site). Three different lot numbers for each device were used for the study. Sample concentrations were confirmed by GC/MS. The results are displayed in the tables below:

|   | Concentration of sample ng/mL | Number of determinations | EDDP Test Strip Results #Neg/#Pos  |
| --- | --- | --- | --- |
|  Site A | 0 (negative) | 15 | 15/0  |
|   |  150 (-50%) | 15 | 15/0  |
|   |  225 (-25%) | 15 | 9/6  |
|   |  375 (+25%) | 15 | 0/15  |
|   |  450 (+50%) | 15 | 0/15  |
|  Site B | 0 (negative) | 15 | 15/0  |
|   |  150 (-50%) | 15 | 15/0  |
|   |  225 (-25%) | 15 | 8/7  |
|   |  375 (+25%) | 15 | 3/12  |
|   |  450 (+50%) | 15 | 0/15  |
|  Site C | 0 (negative) | 15 | 15/0  |
|   |  150 (-50%) | 15 | 14/1  |
|   |  225 (-25%) | 15 | 11/4  |
|   |  375 (+25%) | 15 | 1/14  |
|   |  450 (+50%) | 15 | 0/15  |
|  Combined | 0 (negative) | 45 | 45/0  |
|   |  150 (-50%) | 45 | 44/1  |
|   |  225 (-25%) | 45 | 28/17  |
|   |  375 (+25%) | 45 | 4/41  |
|   |  450 (+50%) | 45 | 0/45  |

Reading time:

The sponsor conducted a reading time study to validate the optimal reading time and stability of the test result for both devices. Drug-free urine specimens were spiked with commercially available EDDP to achieve three GC/MS confirmed concentration levels (+/-50% and a negative).

The samples were tested at 3, 4, 5, 6, 7, 10 minutes, 15 minutes, 20 minutes,

{4}

30 minutes, 60 minutes, 2 hours, 4 hours, 8 hours and 24 hours. The results support the sponsors reading time of 5 minutes and the stability of the results up to 8 hours for both formats.

b. Linearity/assay reportable range:

Not applicable, the device is intended for qualitative use.

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Procedural controls are included in the test strip and device. A colored line appearing in the control zone is considered as an internal procedural control. It confirms sufficient specimen volume and adequate membrane wicking. Users are informed not to interpret the test if no red line appears in the control zone.

Control standards are not supplied with these tests; however it is recommended that positive and negative controls be tested as a good laboratory practice to confirm the test procedure and to verify proper test performance. User should follow local, state and federal guidelines for testing QC material.

d. Detection limit:

To test the analytical sensitivity for EDDP seven different urine samples containing drug concentrations from negative to 3X the cutoff were tested; 30 devices, on three lots for each format were tested. Drug concentrations were confirmed by GC/MS:

|  EDDP | Strip  |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|   |   |  Lot 1 | Lot 2 | Lot 3 | Total  |
|  Conc (ng/mL) | % of Cutoff | Neg/Pos | Neg/Pos | Neg/Pos | Neg/Pos  |
|  0 | 0 | 30/0 | 30/0 | 30/0 | 90/0  |
|  150 | 50% cutoff | 30/0 | 30/0 | 30/0 | 90/0  |
|  225 | 75 % cutoff | 30/0 | 30/0 | 30/0 | 90/0  |
|  300 | Cutoff | 18/12 | 16/14 | 17/13 | 51/39  |
|  375 | 125% cutoff | 6/24 | 3/27 | 5/25 | 14/76  |
|  450 | 150% cutoff | 0/30 | 0/30 | 0/30 | 0/90  |
|  900 | 300% cutoff | 0/30 | 0/30 | 0/30 | 0/90  |

{5}

# e. Analytical specificity:

Cross-reactivity was established by spiking similarly structured drug compounds to a concentration of  $100\mu \mathrm{g / mL}$  into urine having concentrations of EDDP at  $-25\%$  of the cutoff and  $+25\%$  of the cutoff. These solutions were tested with both devices. The table below summarizes the results:

|  Compound | Concentration tested | Test Results  |   |
| --- | --- | --- | --- |
|   |  | EDDP Test Strip  |   |
|   |  | -25% CO | +25% CO  |
|  EDDP | 0 μg/mL | Negative | Positive  |
|  EDDP | 100 μg/mL | Positive | Positive  |
|  Doxylamine | 100 μg/mL | Negative | Positive  |
|  EMDP | 100 μg/mL | Negative | Positive  |
|  LAAM HCL | 100 μg/mL | Negative | Positive  |
|  Nor-LAAM HCL | 100 μg/mL | Negative | Positive  |
|  MDP | 100 μg/mL | Negative | Positive  |
|  (-) alpha methadol | 100 μg/mL | Negative | Positive  |
|  Methadone | 100 μg/mL | Negative | Positive  |

Interfering substances to a final concentration of  $100\mu \mathrm{g / mL}$  for each compound were spiked into a low negative urine and positive urine. The samples were tested with device and showed no interference at the concentration tested. The complete list of the compounds can be found in the package insert.

# pH and Specific Gravity:

To test for possible positive and/or negative interference drug free urine were adjusted to the following pH concentrations 4.0, 5.0, 6.0, 7.0, 8.0 and 9.0.

Each of these samples were divided into three aliquots: drug free urine, EDDP was spiked to  $-25\%$  of the cutoff and EDDP was spiked to  $25\%$  of the cutoff. Each sample was assayed in duplicate. No negative or positive interference due to pH was observed.

To test for possible positive and/or negative interference from specific gravity 15 drug free urine samples having specific gravity from 1.000-1.034 were used. Each of these samples were divided into three aliquots: drug free urine, EDDP was spiked to  $-25\%$  of the cutoff and EDDP was spiked to  $25\%$  of the cutoff. Each sample was assayed in duplicate. No negative or positive interference due to specific gravity was observed.

{6}

f. Assay cut-off:

Characterization of how the device performs analytically around the claimed cutoff concentration appears in the precision section, 1.d, above.

2. Comparison studies:

a. Method comparison with predicate device:

Performance for the EDDP One Step Methadone Metabolite Test Strip was evaluated at two Physician Office Laboratories and one clinical laboratory with a total of nine operators who are typical operators at these sites. Operators ran 183 unaltered clinical urine samples. The samples were blind labeled and compared to GC/MS results. The operators were only provided the labeling to perform the testing. The results are presented in the table below:

|  EDDP Test Strip |  | Negative | Negative (<50% cutoff concentration by GC/MS) | Near cutoff negative (-50% to the cutoff concentration) | Near cutoff positive (cutoff to 50%) | High Positive (>50% cutoff) | % Agreement  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  Site 1 | Positive | 0 | 0 | 3 | 8 | 70 | 98% (180/183)  |
|   |  Negative | 100 | 1 | 1 | 0 | 0  |   |
|  Site 2 | Positive | 0 | 0 | 4 | 8 | 70 | 97.8 (179/183)  |
|   |  Negative | 100 | 1 | 0 | 0 | 0  |   |
|  Site 3 | Positive | 0 | 0 | 4 | 8 | 70 | 97.8 (179/183)  |
|   |  Negative | 100 | 1 | 0 | 0 | 0  |   |

b. Matrix comparison:

Not applicable. The assay is intended for only one sample matrix, urine.

3. Clinical studies:

a. Clinical Sensitivity:

Not applicable.

b. Clinical specificity:

Not applicable.

c. Other clinical supportive data (when a. and b. are not applicable):

{7}

Not applicable.

4. Clinical cut-off:
Not applicable.

5. Expected values/Reference range:
Not applicable.

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

8

---

**Source:** [https://fda.innolitics.com/device/K092048](https://fda.innolitics.com/device/K092048)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
