The Tornier PITON™ Fixation Implant is intended for fixation of soft tissue to bone. The PITON™ Implant is intended for use in the following applications: Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulabral Reconstruction Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair. Illiotibial Band Tenodesis Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction Elbow: Biceps Tendon Reattachment, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction Pelvis: Bladder Neck Suspension for female urinary incontinence due to urethral hypermobility or intrinsic sphincter deficiency Hip: Capsular Repair Acetabular Labral Repair
Device Story
PITON™ Fixation Implant System comprises 2.8 mm diameter bone anchor and disposable pre-loaded inserter; secures soft tissue to bone using USP #2 high-strength non-absorbable UHMWPE braided suture. Anchor design features curved titanium base and flared nitinol clip; allows insertion directly into bone without predrilled, punched, or tapped pilot hole. Used by surgeons in clinical settings for various orthopedic repairs (shoulder, foot/ankle, knee, hand/wrist, elbow, hip) and pelvic floor procedures. Device provides mechanical fixation; surgeon manually deploys anchor via inserter to secure tissue; facilitates healing by maintaining tissue-to-bone contact.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing in accordance with applicable FDA guidelines and established history of use for device materials.
Technological Characteristics
2.8 mm diameter bone anchor; materials: titanium (base) and nitinol (clip); suture: USP #2 non-absorbable UHMWPE. Mechanical fixation principle; no energy source. Designed for direct insertion without pilot hole. Non-software device.
Indications for Use
Indicated for soft tissue to bone fixation in shoulder, foot/ankle, knee, hand/wrist, elbow, and hip orthopedic repairs, and for female urinary incontinence (bladder neck suspension) due to urethral hypermobility or intrinsic sphincter deficiency.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
TwinFix Ti 2.8mm, 3.5mm and BioRaptor 2.9mm Suture Anchors (K053344)
Submission Summary (Full Text)
{0}------------------------------------------------
SEP 2 1 2009
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS (Per 21 CFR 807.92)
6691870
# General Company Information
Name: Contact:
Tornier, Inc. Howard Schrayer Regulatory Affairs Consultant
Address:
100 Cummings Center Suite 444C Beverly, MA 01915
(978) 232 - 9997 Telephone: (978) 232 – 9998 Fax:
Date Prepared June 22, 2009
## General Device Information
PITON™ Fixation Implant System Product Name:
"Non-degradable soft tissue fixation fastener" Classification: Product code: MBI - Class II
## Predicate Device
| Tornier, Inc. | CINCH™ Fixation Implant System.<br>[510(k) Number K080335] |
|----------------------|-----------------------------------------------------------------------------------|
| Smith & Nephew, Inc. | TwinFix Ti 2.8mm, 3.5mm and BioRaptor 2.9mm<br>Suture Anchors<br>[510(k) K053344] |
{1}------------------------------------------------
## Description
The PITON™ Fixation Implant System includes a 2.8 mm diameter bone anchor and a disposable pre-loaded anchor Inserter. The PITON™ Fixation Implant is designed to secure soft tissue to bone using USP #2 high strength non-absorbable UHMWPE braided suture. The device is designed for inserting the implant directly into bone without a predrilled, punched, or tapped pilot hole.
The PITON™ Anchor configuration is comprised of a curved titanium base and a flared nitinol clip.
## Intended Use (Indications)
The Tornier PITON™ Fixation Implant is intended for fixation of soft tissue to bone.
The PITON™ Implant is intended for use in the following applications:
- Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps . Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulabral Reconstruction
- Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair
- Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair. Illiotibial Band Tenodesis
- Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction
- Elbow: Biceps Tendon Reattachment, Ulnar Collateral Ligament Reconstruction, . Radial Collateral Ligament Reconstruction
- Pelvis: Bladder Neck Suspension for female urinary incontinence due to urethral ◆ hypermobility or intrinsic sphincter deficiency
- Hip: Capsular Repair Acetabular Labral Repair ●
## Substantial Equivalence
This submission supports the position that the Tornier PITON™ Fixation Implant System is substantially equivalent to previously cleared devices. including those listed above. A number of predicate devices list the same range of clinical uses.
# Conclusions
Tornier. Inc. believes that the information provided establishes that similar legally marketed devices have been used for the same clinical applications as the Tornier PITON™ Fixation Implant System. The materials from which the Tornier device is fabricated have an established history of use, and the devices have been tested in accordance with applicable FDA guidelines.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem of an eagle with outstretched wings.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Tornier, Inc. % Mr. Howard Schrayer Regulatory Affairs Consultant 100 Cummings Center, Suite 444C Beverly, Massachusetts 01915
Re: K091870
Trade/Device Name: Tornier, PITON™ Fixation Implant System Regulation Number: 21 CFR 888.3040. Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Code: MBI Dated: June 22, 2009 Received: June 23, 2009
Dear Mr. Schrayer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
### Page 2 - Mr. Howard Schrayer
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
http://www.fda.gov/MedicalDevices/ResourceforYou/Industry/default.htm.
Sincerely yours,
Mark M Millerson
Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### INDICATIONS FOR USE
510(k) Number (if known):
Device Name: Tornier, PITON™ Fixation Implant System
Indications For Use:
The Tornier PITON™ Fixation Implant is intended for fixation of soft tissue to bone.
The PITON™ Implant is intended for use in the following applications:
Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulabral Reconstruction
Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valqus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar \ Tendon Repair, Posterior Oblique Ligament Repair. Illiotibial Band Tenodesis Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament
Reconstruction, Radial Collateral Ligament Reconstruction
Elbow: Biceps Tendon Reattachment, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction
Pelvis: Bladder Neck Suspension for female urinary incontinence due to urethral hypermobility or intrinsic sphincter deficiency
Hip: Capsular Repair - Acetabular Labral Repair
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Quta for mxn
(Division Sign of)
Division of Surgic Orthopedic, and Restorative Devices
510(k) Number K091870
4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.