← Product Code [GCJ](/productcode/GCJ) · K091869

# SILS CLINCHER, SILS DISSECTOR, SILS GRASPER, SILS L-HOOK, SILS SHEARS (K091869)

_Covidien Lp, Formerly Registered AS United States · GCJ · Sep 1, 2009 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K091869

## Device Facts

- **Applicant:** Covidien Lp, Formerly Registered AS United States
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Sep 1, 2009
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The SILS™ Clincher* single use articulating clincher has application in a variety of endoscopic, gynecological, and general laparoscopic procedures for temporarily clamping or grasping tissues and small tubular structures. The SILS™ Dissector* single use articulating dissector with monopolar cautery has application in a variety of endoscopic, gynecological, and general laparoscopic procedures to temporarily grasp or clamp tissues and small vessels or body structures, and for use in blunt dissection. When connected to an electrosurgical power source, the device may be used to apply monopolar cautery in accordance with the recommendations of the electrosurgical unit's manufacturer. The SILS™ Grasper* single use articulating grasper has application in a variety of endoscopic, gynecological, and general laparoscopic procedures to temporarily grasp or clamp tissues and small vessels or body structures, and for use in blunt dissection. The SILS™ L-Hook* single use articulating hook with monopolar cautery has application in endoscopic, gynecological, and general abdominal and thoracic laparoscopic procedures. When connected by a standard cable to an electrosurgical power source, the device may be utilized for monopolar cautery. The SILS™ Shears* single use articulating shears with monopolar cautery has application in a variety of endoscopic, gynecologic, and general laparoscopic procedures for transection and cutting of tissues. When connected to an electrosurgical power source, the device may be used to apply monopolar cautery in accordance with the recommendations of the electrosurgical unit's manufacturer.

## Device Story

Single-use endoscopic hand instruments; pistol grip handle; rigid shaft; articulating/rotating end effector. Includes clincher, dissector, grasper, L-hook, and shears. Dissector, L-hook, and shears connect to external electrosurgical power source for monopolar cautery. Operated by surgeons in OR/clinic settings. Handle deflection articulates end effector; rotation wheel rotates end effector. Used for tissue manipulation, blunt dissection, and cutting. Provides surgeons with precise control during laparoscopic procedures; facilitates minimally invasive access.

## Clinical Evidence

Bench testing only. In-vitro and in-vivo testing performed to support intended use. Patient contact materials evaluated per ISO 10993-1:2003.

## Technological Characteristics

Single-use endoscopic instruments; pistol grip handle; rigid shaft; articulating/rotating end effector. Materials evaluated per ISO 10993-1:2003. Monopolar electrosurgical capability via external power source connection. Mechanical articulation via handle deflection and rotation wheel.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Auto Suture™ Modified Endoscopic Hand Instruments ([K951589](/device/K951589.md))
- Auto Suture™ Articulating Endoscopic Clamp ([K914753](/device/K914753.md))
- Auto Suture™ Articulating Endoscopic Scissors ([K914752](/device/K914752.md))
- Auto Suture™ Modified Endo Stream Suction/Irrigation Device ([K905426](/device/K905426.md))
- Auto Suture™ Modified Endoscopic Clamp ([K904578](/device/K904578.md))

## Reference Devices

- SILS™ Stitch Endoscopic Suturing Device ([K090419](/device/K090419.md))
- SILS™ Port Multiple Instrument Access Port ([K082619](/device/K082619.md))
- Convenience Kit for Single-Incision Laparoscopic Surgery ([K072814](/device/K072814.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Surgical Devices, a global business unit of Tyco Healthcare Group

### 5. 510(K) SUMMARY:

510(k) Summary of Safety and Effectiveness:

# SEP = 1 2009

SUBMITTER: Surgical Devices, a global business unit of Tyco Healthcare Group LP (d/b/a Covidien) 60 Middletown Avenue North Haven, CT 06473

CONTACT PERSON:

Robert Zott Program Director, Regulatory Affairs Phone: (203) 492-6013 Fax: (203) 492-5029

DATE PREPARED:

Revised August 13, 2009

TRADE/PROPRIETARY NAME:

SILSTM Clincher* / SILSTM Dissector* SILSTM Grasper* / SILSTM L-Hook* SILS™ Shears*

COMMON/USUAL NAME:

Endoscopic Hand Instruments

Endoscopic Hand Instruments

CLASSIFICATION NAME:

PREDICATE DEVICE(S):

K951589: Auto Suture™ Modified Endoscopic Hand Instruments K914753: Auto Suture™ Articulating Endoscopic Clamp K914752: Auto Suture™ Articulating Endoscopic Scissors K905426: Auto Suture™ Modified Endo Stream Suction/Irrigation Device K904578: Auto Suture™ Modified Endoscopic Clamp

## PRIOR RELATED SUBMISSION(S):

For Single Incision Procedures:

K090419: SILSTM Stitch Endoscopic Suturing Device

K082619: SILS™ Port Multiple Instrument Access Port

K072814: Convenience Kit for "Single-Incision Laparoscopic Surgery and other advanced laparoscopic procedures." 

{1}------------------------------------------------

All predicate devices including those cleared in the prior related submissions are manufactured by Surgical Devices, a global business unit of Tyco Healthcare Group LP (d/b/a Covidien)

### · DEVICE DESCRIPTION:

# INDICATIONS:

Each single use SILSTM Hand Instrument contains a pistol grip, a rigid shaft with an articulating and rotating end effector, and opposing jaws or an electrocautery hook at the distal end.

The SILS™ Clincher* single use articulating clincher has application in a variety of endoscopic, gynecological, and general laparoscopic procedures for temporarily clamping or grasping tissues and small tubular structures.

The SILSTM Dissector* single use articulating dissector with monopolar cautery has application in a variety of endoscopic, gynecological, and general laparoscopic procedures to temporarily grasp or clamp tissues and small vessels or body structures, and for use in blunt dissection. When connected to an electrosurgical power source, the device may be used to apply monopolar cautery in accordance with the recommendations of the electrosurgical unit's manufacturer.

The SILSTM Grasper* single use articulating grasper has application in a variety of endoscopic, gynecological. and general laparoscopic procedures to temporarily grasp or clamp tissues and small vessels or body structures, and for use in blunt dissection.

The SILS™ L-Hook* single use articulating hook with monopolar cautery has application in endoscopic, gynecological, and general abdominal and thoracic laparoscopic procedures. When connected by a standard cable to an electrosurgical power source, the device may be utilized for monopolar cautery.

The SILSTM Shears* single use articulating shears with monopolar cautery has application in a variety of endoscopic, gynecologic, and general laparoscopic procedures for transection and cutting of tissues. When connected to an electrosurgical power source, the device may be used to apply monopolar cautery in accordance with the recommendations of the electrosurgical unit's manufacturer.

{2}------------------------------------------------

# TECHNOLOGICAL CHARACTERISTICS:

.
.
.
.

# MATERIALS:

# PERFORMANCE DATA:

The clincher, dissector, grasper, and shears contain opposing jaws to manipulate tissue, and the dissector, L-hook, and shears administer electrocautery. The end effector of each instrument articulates when the handle is deflected relative to the shaft and rotates when a rotation wheel is turned.

All patient contact materials in the SILSTM Hand Instruments have been evaluated in accordance with ISO 10993-1: 2003, Biological Evaluation of medical devices -- Part 1: Evaluation and Testing.

In-vitro and in-vivo testing has been performed in support of the intended use of these devices.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, a symbol often associated with healthcare.

#### : | 11:4

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

Covidien LP % Mr. Robert Zott Program Director, Regulatory Affairs 60 Middletown Avenue North Haven, Connecticut 06511

Re: K091869

Trade/Device Name: SILS™ Clincher*, SILS™ Dissector*, SILS™ Grasper*, SILS™ L-Hook*, SILS™ Shears*

Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: GCJ Dated: August 13, 2009 Received: August 14, 2009

Dear Mr. Zott:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 - Mr. Robert Zott

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrl/indr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

for

Michal N. Meller

Mark N. Melkerso. Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K091869

# Indications for Use

## 510(k) Number (if known): K091869

Device Name:

Indications For Use:

SILS™ Clincher*, SILS™ Dissector*, SILS™ Grasper*, SILS™ L-Hook*, SILS™ Shears*

The SILS™ Clincher* single use articulating clincher has application in a variety . of endoscopic, gynecological, and general laparoscopic procedures for temporarily clamping or grasping tissues and small tubular structures.

The SILS™ Dissector* single use articulating dissector with monopolar cautery has application in a variety of endoscopic, gynecological, and general laparoscopic procedures to temporarily grasp or clamp tissues and small vessels or body structures, and for use in blunt dissection. When connected to an electrosurgical power source, the device may be used to apply monopolar cautery in accordance with the recommendations of the electrosurgical unit's manufacturer.

The SILS™ Grasper* single use articulating grasper has application in a variety of endoscopic, gynecological, and general laparoscopic procedures to temporarily grasp or clamp tissues and small vessels or body structures, and for use in blunt dissection.

The SILS™ L-Hook* single use articulating hook with monopolar cautery has application in endoscopic, gynecological, and general abdominal and thoracic laparoscopic procedures. When connected by a standard cable to an electrosurgical power source, the device may be utilized for monopolar cautery.

The SILS™ Shears* single use articulating shears with monopolar cautery has application in a variety of endoscopic, gynecologic, and general laparoscopic procedures for transection and cutting of tissues. When connected to an electrosurgical power source, the device may be used to apply monopolar cautery in accordance with the recommendations of the electrosurgical unit's manufacturer.

* Specific trade names to be determined.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Surgical, Orthopedic,
and Restorative Devices

510(k) Number. 2091865

Page 1 of **__**

---

**Source:** [https://fda.innolitics.com/device/K091869](https://fda.innolitics.com/device/K091869)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).