K091787 · Ivy Biomedical Systems, Inc. · DQA · Oct 9, 2009 · Cardiovascular
Device Facts
Record ID
K091787
Device Name
MODEL 6000 TWO PARAMETER BEDSIDE MONITOR
Applicant
Ivy Biomedical Systems, Inc.
Product Code
DQA · Cardiovascular
Decision Date
Oct 9, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Model 6000 is intended to provide continuous nonitoring of SpO2 and EtCO2 indications. The parameters are derived from both the measurements of arterial oxygen hemoglobin saturation and pulse rates as well as relative expired and inspired CO2 levels. The recorded data may be displayed locally and also transmitted wirelessly to any device supporting the 802.11 communication protocol. This device is intended for use on adult, geriatric, pediatric, and neonatal patients in bedside or mobile applications within health care institutions and/or health care environments. Caution: Federal (USA) law restricts this device to sale by or on the order of a physician.
Device Story
Model 6000 is a two-parameter bedside monitor for continuous noninvasive SpO2 and EtCO2 monitoring. Device utilizes internally integrated commercially available modules to process signals from patient-connected accessories (pulse oximetry sensor, nasal cannula). It displays waveforms and numerical values on a color TFT LCD; transmits data wirelessly via 802.11a-g protocol to network-based computing devices. Operated by clinicians in healthcare environments for bedside or mobile monitoring. Output allows clinicians to track patient respiratory and oxygenation status; supports clinical decision-making through real-time monitoring and trend storage. Benefits include continuous patient surveillance and remote data availability.
Clinical Evidence
Bench testing only. Verification and validation performed according to product specifications, including environmental and mechanical stress testing, electromagnetic immunity/emissions testing, and medical device safety testing. Software verified and validated for operational functionality. No clinical data presented.
Technological Characteristics
Two-parameter monitor; integrates commercially available SpO2 and EtCO2 modules. Power: AC Mains or internal Li-polymer battery. Display: VGA color TFT LCD. Connectivity: 802.11a-g wireless. Protection: IPX1 water ingress rating; Class II electrical shock protection. Standards: UL 60601-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-1-6, IEC 60601-1-8, IEC 60601-2-49, ISO 13485.
Indications for Use
Indicated for continuous noninvasive monitoring of SpO2, pulse rate, and EtCO2 (expired/inspired CO2 levels) in adult, geriatric, pediatric, and neonatal patients within healthcare institutions. Use restricted to physician or qualified clinician order.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Oridion Capnography, Inc. Capnostream20 Two Parameter Bedside Monitor (K060065)
Submission Summary (Full Text)
{0}------------------------------------------------
K091787
## 510(k) Summary
June 8, 2009
# Name and Address of Applicant
Ivy Biomedical Systems, Inc. 11 Business Park Drive Branford, CT, 06405
(203) 481-4183 Phone: (203) 481-8734 Fax :
#### Primary Contact
Jennifer Laine Electrical Engineer (203) 481-4183 x185
#### Proprietary Name of Device
Model 6000 Two Parameter Bedside Monitor
## Common Names of the Device
SpO2 and EtCO2 Monitor
#### Device Classification
The device is classified as a Class II device by the both the Cardiovascular and Anesthesiology Panels under 21 CFR Part 870.2700 "Oximeter" per DQA and 21 CFR Part 868.1400 "Carbon Dioxide Gas Analyzer" per CCK respectively.
## Establishment Registration Number of Manufacturer
1221108
#### Legally Marketed Predicate
Oridion Capnography, Inc. Capnostream20 Two Parameter Bedside Monitor, cleared under 510(k), K060065
OCT - 9 2009
{1}------------------------------------------------
### Summary of Technological Characteristics of Device Compared to Predicate
The device has similar indications for use and intended use as the predicate device. Both systems monitor EtCO2 or the relative level of carbon dioxide in exhaled breath, respiratory rate, oxygen saturation in arterial hemoglobin (SpO2), and pulse rate. In both devices, the measurements are collected noninvasively and displayed on a VGA color TFT LCD display. The trend of each measured indication may be stored and represented at multiple resolutions. They are both non-ambulatory devices and applicable to patients of all ages. Unlike the predicate, the Model 6000 device is not intended to be used in home environments.
Both devices deliver alarms meeting the voluntary harmonized standard IEC 60601-1-8.
The energy source for both the device and predicate is equivalent. They receive power and ground either through AC Mains or a battery pack. Neither device makes physical contact with the patient; the patient-side electronics of both are electrically isolated from Mains power.
## Description
The device is a two parameter patient monitor. Both EtCO2 and SpO2 parameters are derived from commercially available modules that are incorporated internally. The device displays the waveforms and numerical values associated with each parameter on a color LCD display, and transmits the information to a network based computing device over the 802.11a - g IEEE wireless protocol. The device is intended for use on adult, geriatric, pediatric, and neonatal patients only under the direction of a physician or qualified clinician.
#### Intended Use
The Model 6000 is intended to provide continuous nonitoring of SpO2 and EtCO2 indications. The parameters are derived from both the measurements of arterial oxygen hemoglobin saturation and pulse rates as well as relative expired and inspired CO2 levels. The recorded data may be displayed locally and also transmitted wirelessly to any device supporting the 802.11 communication protocol. This device is intended for use on adult, geriatric, pediatric, and neonatal patients in bedside or mobile applications within health care institutions and/or health care environments.
Caution: Federal (USA) law restricts this device to sale by or on the order of a physician.
{2}------------------------------------------------
## Performance and Safety Testing
The device complies voluntarily with the following industrial standards:
UL 60601-1. 2nd Edition IEC 60601-1-2:2001 IEC 60601-1-4:2004 IEC 60601-1-6, 2nd Edition IEC 60601-1-8, 2nd Edition IEC 60601-2-49:2001 ISO 13485:2003 CAN/CSA 22.2 No 601.1 M90 CAN/CSA-ISO13485:2003 21 CFR Part 870 MDD: 93/42/EEC Annex II, IIa (per manufacturer) CISPR11 Group 1, Class B
These standards guarantee that the device was designed by a company possessing a full quality system and meets the safety and performance criteria required for an Electrical Medical Device in markets within the European Union. The device was verified and validated according to the product specifications. The test criteria consist of standardized levels and internal product requirements. Tests performed on the device include environmental and mechanical stress testing, electromagnetic immunity and emissions testing, and medical device safety testing. Software on the device was verified and validated according to the functionality of the operations of the device. The test results confirm that the device is in accordance with its specifications.
## Classification Criteria
- The device provides Class II protection against electric shock. .
- The device does not contact the patient directly but plugs into a pulse oximetry sensor . and/or EtCO2 cable with nasal cannula which do contact the skin of the patient. These accessories are legally marketed and sold in the United States.
- The degree of protection against the ingress of water is IPX1. .
- . The device is not sterile.
- . The use of the device is not suitable in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.
- During normal operating conditions, the device is in a continuous mode of operation. .
- The device is externally supplied by AC Mains power, but may also be powered by an . internal Li-polymer battery pack.
- The device is portable. It may be moved between uses. .
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the seal of the Department of Health & Human Services USA. The seal features an eagle with its wings spread, clutching a staff with a snake wrapped around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. The seal is in black and white.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-0609 Silver Spring, MD 20993-0002
Ms. Jennifer Laine Electrical Engineer Ivy Biomedical Systems. Incorporated 11 Business Park Drive Branford, Connecticut 06405
OCT - 9 2009
Re: K091787
Trade/Device Name: Model 6000 Two Parameter Bedside Monitor Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DOA, CCK Dated: October 6, 2009 Received: October 7, 2009
Dear Ms. Laine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Ms. Laine
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.goy/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
for
Susan Runner, D.D.S., M.A. Acting Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use Statement
#### 510(k) Number (if known): K091787
Device Name: Model 6000 Two Parameter Bedside Monitor
#### Indications for Use:
The Model 6000 is intended to provide continuous noninvasive monitoring of SpO2 and EtCO2 indications. The parameters are derived from both the measurements of arterial oxygen hemoglobin saturation and pulse rates as well as relative expired and inspired CO2 levels. The recorded data may be displayed locally and also transmitted wirelessly to any device supporting the 802.11 communication protocol
This device is intended for use on adult, geriatric, and neonatal patients in bedside or mobile applications within health care institutions and/or health care environments only under the direction of a physician or qualified clinician.
Prescription Use x (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE
IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
and the same of the same of the same of the same of the same of
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K091787
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.