K091686 · Jeil Medical Corporation · HWC · Nov 13, 2009 · Orthopedic
Device Facts
Record ID
K091686
Device Name
LEFORTE NEUROSYSTEM BONE SCREW
Applicant
Jeil Medical Corporation
Product Code
HWC · Orthopedic
Decision Date
Nov 13, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This device is intended for use in selective trauma of the mid-face and craniofacial skeleton; craniofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin.
Device Story
LeForte Neuro System Bone Screw consists of titanium alloy (ASTM F136) bone screws, including common, self-drilling (auto), and NS auto variants. Available in micro, mini, and maxi sizes. Used by surgeons in clinical settings for fixation of mid-face and craniofacial skeleton during trauma, reconstructive, or orthognathic procedures. Device provides mechanical stabilization of bone segments to facilitate healing. Potential benefits include structural support for bone reconstruction; risks include infection, nonunion, migration, or metal sensitivity.
Clinical Evidence
No clinical data. Substantial equivalence is based on design, material, and functional comparison to legally marketed predicate devices.
Technological Characteristics
Material: Titanium alloy (ASTM F136). Types: Common bone screws, self-drilling (auto) screws, NS auto screws. Sizes: Micro, mini, maxi. Sterilization: Not specified (non-sterile). Connectivity: None. Software: None.
Indications for Use
Indicated for patients requiring fixation in mid-face and craniofacial trauma, craniofacial surgery, reconstructive procedures, and orthognathic surgery of the maxilla and chin.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Synthes (USA) Neuro Plate and Screw System (K042365)
{0}------------------------------------------------
107 4
K091686
# Medical Corporatio
Extribit 5 510(k) Summary of Safety and Effectiveness
NOV 1 8 2009
Model: LeForte Neuro System Bone Screw (Various Models)
| Submitter and US Official Correspondent | | |
|-----------------------------------------|--------------------------------|--|
| Submitter: | Jeil Medical Corporation | |
| Address: | #702, Kolon Science Valley 2nd | |
| | 811, Guro-dong, Guro-gu | |
| | Seoul, 152-050, Korea | |
| Official Correspondent: | Shin Kuk Yoo, Consultant | |
| Telephone No.: | 714-313-7442 | |
| Fax No.: | 801-303-7455 | |
| Email: | skyone@LSKBioPartners.com | |
- 2. Establishment Registration Number 3004049923
রে, Device Information Proprietary/Trade Name: LeForte Neuro System Bone Screw Common/Usual Name: Bone Screw Classification Name: Screw, Fixation, Intraosseous Classification Product Code: DZL Device Class: Class II per regulation 21 CFR 872.4880
- 心.
【人】【 New or Modification
The application is intended for premarket notification of new medical device, LeForte Neuro System Bone Screw, for US market
- ్త. - 1.40 Tay Marketed Device
- Plate, Souc (Model: Synthes (USA) Neuro Plate and Screw System) manufactured by ・・
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Synth > (USA) - K042365, Decision Date: 11/18/2004, JEY (Class II per regulation 21CRF 372.4760)
{1}------------------------------------------------
2 of
K091686
- 2) Screw, Fixation, Intraosseous (Model: Synthes (USA) 1.3mm Craniofacial Screws) manufactured by Synthes (USA) - K021850, Decision Date: 06/10/2002, DZL (Class II per regulation 21CFR 872.4880)
-
#### Description of the Device 6.
The LeForte Neuro system bone screw is made of Titanium alloy (ASTM F136) and consists of common bone screw, self-drilling screw (auto screw & NS auto screw). The common bone screws are classified to 4 classes: Micro screw, Mini screw, Mini screw, Maxi screw. And the auto screws (self-drilling screws) are divided to 3 classes; Micro auto, Mini auto screws. Additionally, the NS auto screws are divided to 2 classes. This device is manufactured and intended for use in selective trauma of the mid-face and craniofacial skeleton; craniofacial surgery; reconstructive procedures; selective orthognathic surgery of the maxilla and chin. Jeil Medical Corporation utilizes the state of the art technology and apply the essential requirements of MDD(93/42/EEC) and ISO 14630:1997 from the device design to manufacturing and QC.
#### 7. Indications for Use (Intended Use)
This device is intended for use in selective trauma of the mid-face and craniofacial skeleton; craniofacial surgery; reconstructive procedures; selective orthognathic surgery of the maxilla and chin.
8.
### Potential Adverse Affects and Complications (Common to all devices of this type)
- 1) · Poor bone formulation, Osteoporosis, Osteomyelitis, inhibited revascularization, or
- infection can cause loosening, bending, cracking or fracture of the device or premature loss of .
- ﻧ fixation with the bone, leading to nonunion.
- 2) Nonision or delayed union which may lead to breakage of the implant.
- Migration, bending, fracture or loosening of the implant. 3)
- Metal sensitivity, or allergic reaction to foreign body 4)
- Decrease in bone density due to stress shielding 5)
- 6) Pain. discomfort, or abnormal sensation due to the presence of the device
- Increased fibrous tissue.response around the facture site and/or the implant 7)
- 8) Necrosis of bone
- ਰੇ) Inadequate healing
Apart from these adverse effects there are always possible complications of any surgical procedure
{2}------------------------------------------------
3 of 4
such as, but not limited to, infection, nerve damage and pain which may not be related to the ...." implant. . -- --
K091686
#### Safety and Effectiveness Comparison to Predicate Devices റ്റ.
い
r.
| Device Name | Leforte Neuro System<br>Bone Screw (New) | Synthes(USA) 1.3mm<br>Craniofacial Screws<br>(K021850) | Synthes(USA) Neuro<br>Plate & Screw System<br>(K042365) |
|----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant | Jeil Medical Corporation | Synthes(USA) | Synthes(USA) |
| Classification | Screw, Fixation,<br>Intraosseous<br>(Class II, 21 CFR 872.<br>4880) | Screw, Fixation,<br>Intraosseous (Class II<br>21 CFR 872.4880) | Intraosseous fixation<br>screw or wire (Class II,<br>21 CFR 872. 4880)<br>Bone Plate<br>(Class II, 21 CFR 872.<br>4760)<br>Burr hole cover<br>(Class II, 21 CFR 882.<br>5250) |
| Material | Titanium Alloy | Titanium Alloy | Titanium Alloy |
| Intended Use | This device is intended for<br>use in selective trauma of<br>the<br>mid-face and craniofacial<br>skeleton;<br>craniofacial surgery:<br>Reconstructive procedures;<br>selective orthognathic<br>surgery of the maxilla and<br>chin; | This device is intended for<br>use in selective trauma of<br>the mid-face and<br>craniofacial skeleton;<br>craniofacial surgery;<br>reconstructive procedures;<br>and selective orthognathic<br>surgery of the maxilla and<br>chin. | This device is intended<br>for use in selective<br>trauma of the mid-face<br>and craniofacial<br>Skeleton; craniofacial<br>surgery; reconstructive<br>procedures;<br>and<br>selective<br>orthognathic surgery of<br>the<br>maxilla and chin. |
| Single Use | Single Use | Single Use | Single Use |
- : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : .
·
: ジ
۰۰ ،
-
{3}------------------------------------------------
K091686
." ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Non-sterile Non-sterile Sterile Non-sterile
10.
·
17:28:
lusion
ાં
. .
Conclusion and in and System Bone Screws are the equivalent of currently marketed devices. They are made of the same materials and have similar dimensions and characteristics. Potential adverse effects are identical to those of the predicate device is manufactured from material of titanium alloy (ASTM F136) that is used generally in this kind of bone screw. Similar devices made from titanium alloy (ASTM F136) to this device are manufactured and sold around the world. This device, LeForte Neuro system bone screw is substantially equivalent in design, material, and function to the products on the table above. These devices are certificated by notified body for CE.
. No ﺗﻨ - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
ு ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ( ) •
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans and provide essential human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
Jeil Medical Corporation C/O Mr. Shin Kuk Yoo -Manager LSK BioPartners, Incoporated 215 South State Street, Suite 100B Salt Lake City, Utah 84111
NOV 1 3 2009
Re: K091686
Trade/Device Name: LeForte Neuro System Bone Screw (Various Models) Regulation Number: 21 CFR 872.4880 Regulation Name: Intraosseous Fixation Screw or Wire Regulatory Class: II Product Code: DZL Dated: October 19, 2009 Received: October 21, 2009
Dear Mr. Yoo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
## Page 2- Mr. Yoo
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
for
Susan Runner, D.D.S., M.A. Acting Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
# Enclosure
{6}------------------------------------------------
#### Exhibit 4 Indications for Use
510(k) number (if known): _
Device Name: LeForte Neuro System Bone Screw (Various Models)
Indications for Use:
This device is intended for use in selective trauma of the mid-face and craniofacial skeleton; craniofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ > AND/OR Over-The-Counter Use _ (Part 21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ken Mulin for M SR
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K091686
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.