← Product Code [FLL](/productcode/FLL) · K091676

# MC-245, MC-246, MC-247, MC-341, MC-343, MC-344 (K091676)

_Omron Healthcare, Inc. · FLL · Sep 1, 2009 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K091676

## Device Facts

- **Applicant:** Omron Healthcare, Inc.
- **Product Code:** [FLL](/productcode/FLL.md)
- **Decision Date:** Sep 1, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.2910
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Pediatric

## Indications for Use

The Omron electronic thermometer Models MC-245, MC-341, MC-341, MC-343, and MC-344 are intended to measure the body temperature either oral, axillaries (under arm) and rectal and Model - MC-247 measure the body temperature oral (basal) and to be used by medical professionals in clinical and hospital environments and consumers in a home environment. It is intended for use on people of all ages.

## Device Story

Electronic clinical thermometers; thermistor-based temperature sensing. Input: thermal energy from patient body site (oral, axillary, rectal, or basal). Transformation: thermistor detects temperature; internal circuitry calculates reading; buzzer signals peak temperature. Output: digital temperature reading on LCD. Used in clinics, hospitals, and home environments by medical professionals or patients. Provides objective temperature data to assist healthcare providers in clinical decision-making; enables patient monitoring of body temperature.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing and comparison of technological characteristics and specifications to the predicate device.

## Technological Characteristics

Thermistor-based temperature sensing; battery-operated (LR41); LCD digital display; 4-digit resolution (0.1 increments); accuracy ±0.2°F/±0.1°C. Materials tested per ISO 10993. Models feature rigid or flexible tips. Connectivity: None. Sterilization: Not specified.

## Regulatory Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

## Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; 
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and 
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

## Predicate Devices

- MC-3B ([K881446](/device/K881446.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Premarket Notification 510(k) Section 5 - 510(k) Summary

Omron Thermometers

KC916676

| Omron Healthcare, Inc.<br>1200 Lakeside Dr.<br>Bannockburn, IL 60015 | Tel - 847-247-5678<br>Fax - 847-680-6269                   | SEP - 1 2009 |
|----------------------------------------------------------------------|------------------------------------------------------------|--------------|
| Official Contact:                                                    | Ronald Dudek - Director of Quality                         |              |
| Proprietary or Trade Name:                                           | Models – MC-245, MC-246, MC-247, MC-341, MC-343,<br>MC-344 |              |
| Common/Usual Name:                                                   | Clinical electronic thermometer                            |              |
| Classification Name/Code:                                            | FLL – clinical electronic thermometer<br>21 CFR 880.2910   |              |
| Device:                                                              | Models – MC-245, MC-246, MC-247, MC-341, MC-343,<br>MC-344 |              |

#### Device Description:

These are simple electronic thermometer which is a thermistor to measure temperature.

- . All the functions, except posture symbol, are the same as the predicate device, MC-3B, K881446.
- Power source battery operated .
- Calculation method (system) is the same and identical to the predicate, K881449 �

The differences between the various models are:

- Some models have a flexible or rigid or rigid / flexible tip .
- Offered with a probe cover .
- Temperature scale offered in C° & F° or F° only .
- Readout time is different .

#### Indications for Use:

The Omron electronic thermometer Models MC-245, MC-341, MC-341, MC-343, and MC-344 are intended to measure the body temperature either oral, axillaries (under arm) and rectal and Model - MC-247 measure the body temperature oral (basal) and to be used by medical professionals in clinical and hospital environments and consumers in a home environment. It is intended for use on people of all ages.

Environment of Use: Clinics, hospital and home environments

Contraindications: None

{1}------------------------------------------------

# 510(k) Summary Page 2 of 3 5-Jun-09

#### Summary of substantial equivalence .

|                       | MC-3B<br>K881446                                                    | MC-245/246/247<br>MC-341/343/344                                    |
|-----------------------|---------------------------------------------------------------------|---------------------------------------------------------------------|
| Measurement<br>Method | Human temperature is detected by<br>thermistor and calculated       | Human temperature is detected by<br>thermistor and calculated       |
| Display               | Reading value is displayed on LCD<br>with 4 digits.(0.1 increments) | Reading value is displayed on LCD<br>with 4 digits.(0.1 increments) |
| Buzzer                | Beeper sound at peak temperature                                    | Beeper sound at peak temperature                                    |
| Power Source          | Battery(LR41 type)                                                  | Battery(LR41 type)                                                  |

| SPECIFICATION                         | MC-3B<br>K881446                       | MC-245, MC-246, MC-247<br>MC-341, MC-343, MC-344                                                      |
|---------------------------------------|----------------------------------------|-------------------------------------------------------------------------------------------------------|
| Display                               | LCD Digital Display                    | LCD Digital Display                                                                                   |
| Measurement<br>Range                  | 89.6 °F to 107.6 °F                    | 89.6 °F to 107.6 °F, 32.0 °C to 42.0 °C<br>These models switch the display unit (°F, °C) alternately. |
| Measurement area                      | Oral, rectum, under arm                | Oral, rectum, under arm, except Model MC-247 is for oral (basal) only                                 |
| Accuracy                              | ±0.2 °F                                | ±0.2 °F, ±0.1 °C                                                                                      |
| Display Resolution                    | 4 Digits<br>(0.1 °F increments)        | 4 Digits<br>(0.1 °F / 0.1 °C increments)                                                              |
| Weight                                | Approx. 1/3 oz /(11g)                  | Approx. 1/3 oz / (11g)                                                                                |
| Length                                | Approx. 130mm                          | Approx. 130mm                                                                                         |
| Battery                               | 1 each LR41 Alkali<br>1.5V D.C         | 1 each LR41 Alkali 1.5V D.C                                                                           |
| Battery Life                          | Approx. 300 hours.<br>(Continuous use) | Approx. 2 years (3 times measurements / day )                                                         |
| Ambient<br>Temperature for<br>use     | 50 °F to 104 °F<br>(10 °C to 40 °C)    | 50 °F to 104 °F<br>(10 °C to 40 °C)                                                                   |
| Ambient<br>Temperature for<br>use     | (10 ° C to 40 ° C)                     | (10 ° C to 40 ° C)                                                                                    |
| Ambient<br>Temperature for<br>storage | 4 °F to 140 °F<br>(-20 °C to 60 °C)    | 4 °F to 140 °F<br>(-20 °C to 60 °C)                                                                   |
| Ambient<br>Temperature for<br>storage | (-20 ° C to 60 ° C)                    | (-20 ° C to 60 ° C)                                                                                   |
| Ambient Humidity<br>for use           | 30 to 85%RH                            | 30 to 85%RH                                                                                           |
| Ambient Humidity<br>for storage       | 30 to 95%RH                            | 10 to 95%RH                                                                                           |

.

{2}------------------------------------------------

#### 510(k) Summary Page 3 of 3

5-Jun-09

#### Differences Between Other Legally Marketed Predicate Devices

The Omron thermometer models MC-245, MC-247, MC-341, MC-341, MC-343, and MC-344 are viewed as substantially equivalent to the predicate device because:

# Indications -

- . Identical to predicate - K881446
### Technology -

- . Identical algorithms and means of measuring temperature to predicate - K881446
# Operating specifications -

- Identical to predicate K881446 .
# Materials -

The materials have been tested in accordance to ISO 10993.

# Environment of Use -

- Identical to predicate K881446 �
#### Patient Population -

- Identical to predicate K881446 .

{3}------------------------------------------------

Image /page/3/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol consists of four abstract shapes that resemble a stylized human figure or a flame-like design. The logo is in black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

Omron Healthcare, Incorporated C/O Mr. Paul Dryden President ProMedic, Incorporated 24301 Woodsage Drive Bonita Springs, Florida 34134

Re: K091676

Trade/Device Name: Omron Thermometer Models MC-245, MC-246, MC-247,

MC-341, MD-343, MC-344 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: July 25, 2009 Received: July 29, 2009

Dear Mr. Dryden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

# Page 2- Mr. Dryden

Enclosure

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH /CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance,

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Anthony L., Amato Jr.

Susan Runner, D.D.S., M.A. Acting Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{5}------------------------------------------------

# Indications for Use Statement

Page 1 of 1

### 510(k) Number:

1. 0 91676 (To be assigned)

Device Name:

Omron Thermometers Models MC-245, MC-246, MC-247, MC-341, MC-343, MC-344

# Indications for Use:

The Omron electronic thermometer Models MC-245, MC-341, MC-341, MC-343, and MC-344 are intended to measure the body temperature either oral, axillaries (under arm) and rectal and Model - MC-247 measure the body temperature oral (basal) and to be used by medical professionals in clinical and hospital environments and consumers in a home environment. It is intended for use on people of all ages.

| Prescription Use            | or | Over-the-counter use XX |
|-----------------------------|----|-------------------------|
| (Part 21 CFR 801 Subpart D) |    | (21 CFR 807 Subpart C)  |

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Y. Schacter

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K091676

---

**Source:** [https://fda.innolitics.com/device/K091676](https://fda.innolitics.com/device/K091676)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
