The Cor12+ Holter Recorder is intended to acquire, record and store ECG data of patients that have been connected to the Cor12+ recorder and are undergoing Holter monitoring. The Cor12+ acquires, digitizes and stores data to be analyzed on a computer with special software. The Cor12+ is indicated for use: in a clinical setting, by qualified medical professionals only, for recording ECG data of symptomatic patients requiring ambulatory (Holter) monitoring of up to 24 hours. The Cor12+ performs no cardiac analysis by itself; it solely records the ECG signals. The patient population for this device is from 4 years on and a minimum weight of 10 kg.
Device Story
Cor12+ is a portable, battery-operated digital Holter ECG recorder; captures ECG signals via patient-connected leads; digitizes and stores data on an SD memory card. Device features a single-button interface for event recording and communication activation; dual-color LED provides visual feedback on electrode contact quality (green for good, red for insufficient). PC-based software connects via Bluetooth to program the device; measurement data is transferred to a computer via SD card for analysis. Used in clinical settings by medical professionals; performs no onboard cardiac analysis. Benefits include ambulatory monitoring for symptomatic patients to aid clinical diagnosis of cardiac conditions.
Clinical Evidence
Bench testing only. Testing included electrical safety per IEC 60601-1 and IEC 60601-2-47, and EMC testing per EN 60601-1-2 and EN 60601-2-47. Risk analysis performed via Failure Modes and Effects Analysis (FMEA). No clinical data provided.
Technological Characteristics
Portable ECG recorder; PC/ABS Cycoloy C1200 housing; 24-bit A/D conversion; 1000 Hz or 250 Hz sampling rate; 0.05-150 Hz frequency response. Powered by 2x AA Lithium batteries. Connectivity: Bluetooth (programming only), SD card (data storage). Electrode impedance measurement for contact quality assessment. Software-controlled via PC interface.
Indications for Use
Indicated for symptomatic patients aged 4 years and older, weighing at least 10 kg, requiring ambulatory (Holter) ECG monitoring for up to 24 hours in a clinical setting.
Regulatory Classification
Identification
A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.
{0}------------------------------------------------
K091505
P1/6
# JUN 1 0 2009
# 510(k) Summary
## GENERAL INFORMATION
#### Type of Submission 5.1
Traditional 510(k) Submission
Submission date: 04/21/2009
#### Submitter 5.2
Cardinal Health Germany 234 GmbH
Address:
Name:
Leibnizstrasse 7 D-97204 Hoechberg Germany
### Contact person in Germany:
| Address: | Thomas Rust<br>Cardinal Health Germany 234 GmbH<br>Leibnizstrasse 7, 97204 Hoechberg<br>Germany |
|----------|-------------------------------------------------------------------------------------------------|
| Phone: | +49 931 49 72 - 383 |
| FAX: | +49 931 49 72 - 62383 |
| E-mail | Thomas.Rust@cardinalhealth.com |
Contact person in the U.S .: Thomas Gutierrez (Official Correspondent) Cardinal Health 207 INC Address 22745 Savi Ranch Parkway Yorba Linda, CA 92887 Phone/Fax: 714-283-8472 Thomas. Gutierrez@cardinalhealth.com E-mail
{1}------------------------------------------------
- 5.3 Establishment Registration Number 9615102
- 5.4 Common Name or Classification Name Recorder, Magnetic Tape, Medical (CFR 870.2800, Product Code DSH)
- 5.5 Trade Name Cor12+
- 5.6 Device Classification This is a Class II device
#### 5.7 Classification Panel
74 Cardiovascular Part 870 Code DSH
- 5.8 Reason for Premarket Notification New Device
#### Legally predicate marketed device 5.9
| 1. H12+ Holter Recorder | K050896 | Code DSH |
|---------------------------|---------|----------------|
| 2. Cardiette Microtel ECG | K082124 | Code DPS |
| 3. MasterScope ECG | K082539 | Code BTY / DPS |
#### 5.10 Predicate Device Company
- 1. Mortara Instrument. INC.
- 2. Et Medical devices SpA
- 3. Cardinal Health Germany 234 GmbH
#### 5.11 Device Description
The digital Holter ECG recorder Cor12+ is a portable battery operated ECG amplifier with build in data storage on memory card (Mini SD size):
- Single dual color (green/yellow) LED .
- Button control (one Button) to activate communication interface, event . recording
- Two standard size AA batteries .
- Speaker for voice output for alarm messages .
- PC program to set-up recorder via blue-tooth communication interface 0
#### 5.12 Intended Use Statement
The Cor12+ Holter Recorder is intended to acquire, record and store ECG data of patients that have been connected to the Cor12+ recorder and are undergoing Holter monitoring. The Cor12+ acquires, digitizes and stores data to be analyzed on a computer with special software.
The Cor12+ is indicated for use:
{2}------------------------------------------------
K091505
p3/6
in a clinical setting, by qualified medical professionals only, for recording ECG data of symptomatic patients requiring ambulatory (Holter) monitoring of up to 24 hours.
The Cor12+ performs no cardiac analysis by itself; it solely records the ECG signals. The patient population for this device is from 4 years on and a minimum weight of 10 kg.
#### 5,13 Required Components
Cor12+ recorder Accessories User Manual
#### 5.14 Summary Table of Comparison
a) Comparison of Cor12+ with H12+ (complete device functions)
| | H12+ Holter Recorder<br>(K050896) | Cor12+<br>Holter Recorder |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | The H12+ Holter recorder is<br>intended to acquire, record and<br>store ECG data of patients that<br>have been connected to the<br>Mortara H12+ recorder and are<br>undergoing Holter monitoring. The<br>H12+ acquires, digitizes and stores<br>data to be analyzed by the H-Scribe<br>Holter system. | The Cor12+ Holter Recorder is<br>intended to acquire, record and<br>store ECG data of patients that<br>have been connected to the<br>Cor12+ recorder and are<br>undergoing Holter monitoring. The<br>Cor12+ acquires, digitizes and<br>stores data to be analyzed on a<br>computer with special software. |
| | The H12+ is indicated for use:<br>- in a clinical setting, by qualified<br>medical professionals only, for<br>recording ECG data of<br>symptomatic patients requiring<br>ambulatory (Holter) monitoring<br>of up to 48 hours.<br>- the H12+ performs no cardiac<br>analysis by itself and is<br>intended to be used with the<br>H-Scribe Holter analysis<br>system (K004017). ECG data<br>prerecorded by the H12+ is<br>acquired and analyzed by the<br>H-Scribe. | The Cor12+ is indicated for use:<br>- in a clinical setting, by qualified<br>medical professionals only, for<br>recording ECG data of patients<br>requiring ambulatory (Holter)<br>monitoring of up to 24 hours.<br><br>The Cor12+ performs no cardiac<br>analysis by itself; it solely records<br>the ECG signals.<br><br>The patient population for this<br>device is from 4 years on and a<br>minimum weight of 10 kg. |
| Patient<br>population | The H12+ Holter Recorder can be<br>used for patients from 4 years on<br>and older. | The Cor12+ Holter Recorder can<br>be used for patients from 4 years<br>on and older. Minimum weight of 10<br>kg. |
| Dimensions<br>(housing) | Length x Width x Height:<br>$112*82*34$ mm | Length x Width x Height:<br>$100*70*26$ mm |
| | Weight:<br>145 g (batteries included) | Weight:<br>155 g (batteries included) |
| Display | LCD module<br>Size: 46,0 x 18,4 mm<br>100 x 32 dots | Dual color LED |
| Key-panel | Foil Key-panel (4 keys):<br>- ESC (on/off)<br>- UP-ARROW<br>- DOWN-ARROW<br>- OK | Single button:<br>- Activate communication |
| Housing | Material:<br>Rotec ABS 1001FR V0 | Material:<br>PC/ABS<br>Cycoloy C1200 |
| Input<br>dynamic<br>range | +/- 300mV @ DC | +/- 1300mV @ DC |
| Frequency<br>response /<br>Bandwith | 0,05 -150 Hz<br>According to EC11 and<br>IEC 60601-1 | identical |
| A/D<br>conversion | 20 bit | 24 bit |
| Sampling<br>rate for data<br>storage | 1000 Hz or 180 Hz | 1000 Hz or 250 Hz |
| Leads | 12 Standard - US color coding | identical |
| Interface | Compact Flash Memory Card<br>(data storage and programming) | SD Memory Card (data storage)<br>Bluetooth (programming) |
| Energy type | 1 x 1,5 V (AA size) | 2 x 1,5 V (AA size, Lithium) |
| Operating<br>Require-<br>ments | Programming Software | Identical |
| Mode of<br>operation | Manual, Manual delayed and<br>Automatic recording | Automatic recording |
| | Cardiette Microtel<br>(K082124) | Cor12+<br>Holter Recorder |
| Interface | Bluetooth & USB | Bluetooth & SD-Card |
- . . ^ . . ^
{3}------------------------------------------------
K091505
P4/6
…
...
ד
·
ﺴﻪ ﺍ
ﻪ ﻣ
ページ ウェ
Pane 4 of 6
{4}------------------------------------------------
b) Comparison of Cor12+ with Cardiette Microtel ECG Recorder (communication interfaces)
### c) Comparison of Cor12+ with MasterScope ECG (electrode impedance)
| | MasterScope ECG<br>(K082539) | Cor12+<br>Holter Recorder |
|------------------------|----------------------------------------|---------------------------|
| Electrode<br>impedance | Impedance measurement for<br>electrode | Identical |
### Discussion to the device differences of the table a), b) and c) above.
The significant differences to the H12+ ECG Recorder are found as:
- No display The main purpose of the display on the H12+ recorder is the . programming of the device (set of time, subject ID, etc.). In the Cor12+ this programming of the device is performed by a PC program and direct bluetooth communication. Therefore no display is needed on the Cor12+ to allow its intended use
- Input Dynamic Range The Cor12+ has a better input range compared with . the predicate device. This allows the recording of the ECG even with increased baseline drift.
- Sampling Rate The lower sampling rate (250 Hz) for data storage is slightly . higher compared with the predicate device H12+ (160 Hz). This slightly higher frequency is beneficial for the accuracy and resolution of further data processing.
- Bluetooth is used for programming the Cor12+ instead programming by . keypad/display and memory-card as it is with the H12+ device. The Bluetooth is not for data transfer of measurement data, it is only for the purpose to program the Cor12+ for the patient. The measurement data are transferred to the computer by memory-card as it is provided also by the predicate device H12+.
{5}------------------------------------------------
The community to the Cardiette Microtel ECG Recorder is found as:
- Interface Bluetooth Common is that both devices work with a Bluetooth . interface, whereas the Corr2+ uses the Bluetooth interface only to program the patient device and not to transfer the measurement data.
The community to the MasterScope ECG is found as:
- Electrode impedance Common is that both devices work with electrod . impedance measurement. Electrode contact quality is determined by measuring the impediate at each electrode. If impedance is below a certain value, electrode contact is good and the green LED lights up. For impedance values above this value, the electrode contact is not sufficient to guarantee good ECG quality. The red LED on the corresponding electrode clip lights up. This feature enables easy and straightforward check of electrode contact quality.
#### 5.15 Summary of Device Testing
The following practices were followed and monitored for development of the Holter ECG recorder Cor12+:
- The Cor12+ was developed in accordance with the Cardinal Health . development standard operating procedures (000490 06 – Design Control).
- The risk analysis method used to assess the impact of Cor12+ was a � Failure Modes and Effects Analysis (FMEA).
- Safety test procedures demonstrate satisfaction of all safety . requirements and mitigation of all identified hazards.
- The EMC testing was performed nazards.
The EMC testing was performed according EN 60601-1-2 and . EN 60601-2-47.
- Electrical Safety.testing was performed according IEC 60601-1 and IEC
60601-2-47 . 60601-2-47.
#### · 5.16 Conclusions
Based on the above, Cardinal Health Germany 234 GmbH concludes that the Cor12+ is substantially equivalent to legally marketed predicate that the
Cor12+ is substantially equivalent to legally marketed predicate device "H12+" and the two options Bluetooth and Electrode impedance are equivalent to the fi
and the two options Bluetooth and Electrode impedance are equivalent to the devices "Cardiette Microtel" and "MasterScope ECG". The Cor12+ is safe and effective for its microlor and "MasterScope ECJ . The Cor12+ is safe a
ffective for its intended use, and performs at least as well as the predicate devices.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized eagle or bird-like symbol with outstretched wings, representing the department's mission related to health and human services.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 1 0 2009
Cardinal Health 207, Inc. c/o Mr. Thomas Gutierrez 22745 Savi Ranch Parkway Yorba Linda, CA 92887
K091505 Re:
> Cor12+ Regulation Number: 21 CFR 870. 2800 Regulation Name: Medical Magnetic Tape Recorder Regulatory Class: Class II (two) Product Code: DSH Dated: May 15, 2009 Received: May 21, 2009
Dear Mr. Gutierez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
### Page 2 - Mr. Thomas Gutierrez
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally. marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use
510(k) Number (if known): K
Device Name: Cor12+
Indications for Use:
The Cor12+ Holter Recorder is intended to acquire, record and store ECG data of patients that have been connected to the Cor12+ recorder and are undergoing Holter monitoring. The Cor12+ acquires, digitizes and stores data to be analyzed on a computer with special software.
The Cor12+ is indicated for use:
in a clinical setting, by qualified medical professionals only, for recording ECG data of symptomatic patients requiring ambulatory (Holter) monitoring of up to 24 hours.
The Cor12+ performs no cardiac analysis by itself; it solely records the ECG signals. The patient population for this device is from 4 years on and a minimum weight of 10 kg.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|------------------------------------|---------|
| Division of Cardiovascular Devices | |
| 510(k) Number | K091505 |
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.