← Product Code [NST](/productcode/NST) · K091468

# IVD TECHNOLOGIES ACHRAB TEST KIT (K091468)

_Ivd Technologies · NST · Nov 20, 2009 · Immunology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K091468

## Device Facts

- **Applicant:** Ivd Technologies
- **Product Code:** [NST](/productcode/NST.md)
- **Decision Date:** Nov 20, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5660
- **Device Class:** Class 2
- **Review Panel:** Immunology

## Indications for Use

The AchRAb Assay is a radioimmunoassay for the in vitro diagnostic semi-quantitative determination of autoantibodies against the acetylcholine receptor in human serum. The AchRAb assay is useful as an aid in the differential diagnosis of Myasthenia Gravis (MG) in conjunction with other clinical and laboratory findings.

## Device Story

The AchRAb Assay is a radioimmunoassay (RIA) used in clinical laboratories to detect autoantibodies against the acetylcholine receptor in human serum. The device utilizes human acetylcholine receptors labeled with 125I-α-bungarotoxin. Patient serum is incubated with the labeled receptor; autoantibodies present in the sample bind to the receptor to form complexes. These complexes are immunoprecipitated using goat anti-human IgG and centrifuged. The resulting pellet is measured in a gamma counter. The radioactivity count is directly proportional to the concentration of autoantibodies in the sample. Results are calculated against a standard curve to provide a semi-quantitative value. The assay aids clinicians in the differential diagnosis of Myasthenia Gravis by identifying elevated antibody titers (≥ 0.5 nmol/L).

## Clinical Evidence

No clinical data provided in the document; substantial equivalence determination based on regulatory review of the 510(k) submission.

## Technological Characteristics

Radioimmunoassay (RIA) utilizing 125I-labeled α-bungarotoxin. Components include lyophilized human acetylcholine receptor, 125I-AchR diluent (Tris buffer, surfactant), 6-level standards, positive/negative controls, goat anti-human IgG, and wash buffer. Detection requires a gamma counter. Assay range: 0.1–7.5 nmol/L. Functional sensitivity: 0.1 nmol/L.

## Regulatory Identification

A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number:
k091468

B. Purpose for Submission:
New device

C. Analyte:
Acetylcholine receptor autoantibodies

D. Type of Test:
Semi-quantitative, Radioimmunoassay

E. Applicant:
IVD Technologies

F. Proprietary and Established Names:
Acetylcholine Receptor AB RIA Kit (AChRAb Assay)

G. Regulatory Information:
1. Regulation section:
21CFR §866.5660, Multiple autoantibodies immunological test system
2. Classification:
Class II
3. Product code:
NST, Autoantibodies, Acetylcholine Receptor, Acetylcholine Blocking and Non-Blocking
4. Panel:
Immunology (82)

H. Intended Use:
1. Intended use(s):
The AchRAb Assay is a radioimmunoassay for the in vitro diagnostic semi-quantitative determination of autoantibodies against the acetylcholine receptor in human serum. The AchRAb assay is useful as an aid in the differential diagnosis of Myasthenia Gravis (MG) in conjunction with other clinical and laboratory findings.
2. Indication(s) for use:
Same as intended use
3. Special conditions for use statement(s):
Prescription use only
4. Special instrument requirements:
Gamma counter for I¹²⁵

I. Device Description:
The Acetylcholine Receptor (AchRAb) kit consists of the following:
1. Human Acetylcholine Receptor, labeled with ¹²⁵I α-bungarotoxin, lyophilized
2. ¹²⁵I-AchR Diluent [contains 0.05M Tris buffer with 1.5% surfactant]
3. Standards (6 Levels): 0.00 (normal human serum), 0.2, 0.50, 1.25, 2.50, and 7.5 nmol/L human anti-acetylcholine receptor in human serum; &lt;0.01% NaN₃
4. Positive Control, contains human anti-acetylcholine receptor in human serum, with &lt;0.01% NaN₃

{1}

5. Negative Control Cut Off Point, contains human anti-acetylcholine receptor in human serum, with  $&lt; 0.01\% \mathrm{NaN}_3$
6. Goat anti-human IgG, contains  $&lt; 0.01\% \mathrm{NaN}_3$
7. Normal human serum (for diluting test sera), contains  $&lt; 0.01\% \mathrm{NaN}_3$
8. Wash buffer [contains 0.05M Tris buffer with  $1.5\%$  surfactant]

# J. Substantial Equivalence Information:

1. Predicate device name(s): IBL  $^{125}$ I Acetylcholine Receptor Antibody Kit
2. Predicate 510(k) number(s): k051144
3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Intended Use/Indications for Use | Determination of autoantibodies against human acetylcholine receptor, aid in the differential diagnosis of Myasthenia Gravis | Same  |
|  Test principle | Radioimmunoassay (radioreceptor assay) | Same  |
|  Test platform | Radiolabeled receptor binds to serum antibodies in liquid phase, antibody-antigen complexes are precipitated and radioactivity measured | Same  |
|  Detection instrument | Gamma counter | Same  |
|  Calculation | Semi-quantitative determination with standard curve | Same  |
|  Quality Control | Positive control and cut-off control | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Sample Types | Serum | Serum and plasma  |
|  Negative cutoff | ≤ 0.25 nmol/L | ≤ 0.25 nmol/L  |
|  Equivocal | 0.25 – 0.5 nmol/L | 0.25 – 0.4 nmol/L  |
|  Positive | ≥ 0.5 nmol/L | ≥ 0.4 nmol/L  |
|  Limit of quantitation | 0.1 nmol/L | 0.01 nmol/L  |
|  Assay range | 0.1 – 7.5 nmol/L | 0.25 – 1.5 nmol/L  |

# K. Standard/Guidance Document Referenced (if applicable):

None referenced.

{2}

# L. Test Principle:

Patient specimens, standards, and controls are incubated with acetylcholine receptor (AchR) which has been labeled with  $^{125}\mathrm{I}-\alpha$ -bungarotoxin. The resulting bound complexes of labeled AchR and autoantibodies are then immunoprecipitated with anti-human IgG. After centrifugation, the supernatant is aspirated and the pellet containing labeled AchR/autoantibody-bound complexes is counted in a gamma counter. Counts are directly proportional to the amount of autoantibodies present.

# M. Performance Characteristics (if/when applicable):

# 1. Analytical performance:

# a. Precision/Reproducibility:

# Intra-assay

Two different intra-assay precision studies were performed. In the first study, two different technicians ran 16 replicates of three different serum samples. Results are summarized in the following table:

|   | Tech 1 |   |   |   | Tech 2  |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Sample | N | Mean (nmol/L) | SD (nmol/L) | CV(%) | N | Mean (nmol/L) | SD (nmol/L) | CV(%)  |
|  1 | 16 | 0.37 | 0.033 | 9.0% | 16 | 0.46 | 0.028 | 6.1%  |
|  2 | 16 | 1.67 | 0.19 | 11.3% | 16 | 1.65 | 0.20 | 12.2%  |
|  3 | 16 | 6.45 | 0.19 | 2.9% | 16 | 5.91 | 0.20 | 3.3%  |

In the second study, one technician ran 20 replicates of 8 different serum pools. The following table summarizes the results:

|  Serum Sample | N | Mean (nmol/L) | SD (nmol/L) | CV(%)  |
| --- | --- | --- | --- | --- |
|  Pool 1 | 20 | 0.22 | 0.025 | 11.3%  |
|  Pool 2 | 20 | 0.44 | 0.027 | 6.2%  |
|  Pool 3 | 20 | 0.71 | 0.048 | 6.8%  |
|  Pool 4 | 20 | 0.92 | 0.037 | 4.1%  |
|  Pool 5 | 20 | 1.54 | 0.084 | 5.5%  |
|  Pool 6 | 20 | 2.02 | 0.088 | 4.4%  |
|  Pool 7 | 20 | 4.22 | 0.130 | 3.1%  |
|  Pool 8 | 20 | 6.66 | 0.175 | 2.6%  |

# Inter-assay

Inter-assay variation was assessed by analyzing 8 serum samples over 15 assay runs. The assays were performed in different labs at the manufacturer's facility with multiple technicians with one kit lot. The mean coefficient of variation was found to be  $8.1\%$  (range  $4.2 - 10.3\%$ ) Results are shown in the table below.

{3}

|  Serum Sample | N | Mean (nmol/L) | SD (nmol/L) | CV(%)  |
| --- | --- | --- | --- | --- |
|  Pool 1 | 15 | 0.24 | 0.024 | 10.1%  |
|  Pool 2 | 15 | 0.46 | 0.047 | 10.3%  |
|  Pool 3 | 15 | 0.78 | 0.058 | 7.5%  |
|  Pool 4 | 15 | 1.00 | 0.092 | 9.2%  |
|  Pool 5 | 15 | 1.50 | 0.14 | 9.1%  |
|  Pool 6 | 15 | 2.11 | 0.21 | 9.9%  |
|  Pool 7 | 15 | 4.30 | 0.20 | 4.8%  |
|  Pool 8 | 15 | 6.81 | 0.29 | 4.2%  |

Lot-to-lot

Lot-to-lot variation was assessed by testing 9 low value serum samples in duplicate on three different kit lots. The mean coefficient of variation was found to be 8.6% (range 3.0-16.6%). Results are summarized in the table below.

|  Serum Sample | Mean (nmol/L) | CV(%)  |
| --- | --- | --- |
|  A | 0.22 | 16.6%  |
|  B | 0.17 | 8.8%  |
|  C | 0.39 | 10.5%  |
|  D | 0.32 | 3.9%  |
|  E | 0.58 | 6.7%  |
|  F | 0.48 | 9.2%  |
|  G | 0.71 | 5.1%  |
|  H | 1.30 | 3.0%  |
|  I | 1.21 | 5.5%  |

b. Linearity/assay reportable range:

Two serum samples with different acetylcholine receptor levels were diluted with the zero calibrator, in ratios of 1:1.2, 1:1.4, 1:1.8, 1:2, 1:3, 1:4, and 1:8. The graph of measured vs. expected results had slope = 0.989, intercept = 0.032, and r² = 0.993. The results are summarized in the table below. The assay is linear up to 7.5 nmol/L, the level of the highest calibrator.

{4}

|  Serum Sample—Pool 8 |   |   | Serum Sample—Pool 2  |   |   |
| --- | --- | --- | --- | --- | --- |
|  Dilution | Measured (nmol/L) | Recovery (%) | Dilution | Measured (nmol/L) | Recovery (%)  |
|  Undiluted | >8.5 | -- | Undiluted | 0.72 | 106  |
|   | >8.5 | -- |  | 0.64 | 94  |
|  1:1.2 | 9.14 | 103 | 1:1.2 | 0.54 | 95  |
|   | 8.12 | 92 |  | 0.49 | 86  |
|  1:1.4 | 7.63 | 101 | 1:1.4 | 0.54 | 111  |
|   | 7.56 | 100 |  | 0.40 | 82  |
|  1:1.8 | 6.61 | 112 | 1:1.8 | 0.39 | 103  |
|   | 5.99 | 101 |  | 0.35 | 93  |
|  1:2 | 6.40 | 120 | 1:2 | 0.28 | 82  |
|   | 5.88 | 111 |  | 0.30 | 88  |
|  1:3 | 3.95 | 112 | 1:3 | 0.24 | 106  |
|   | 3.47 | 98 |  | 0.21 | 96  |
|  1:4 | 2.80 | 105 | 1:4 | 0.14 | 82  |
|   | 2.76 | 104 |  | 0.15 | 88  |
|  1:8 | 1.36 | 102 | 1:8 | 0.10 | 118  |
|   | 1.17 | 88 |  | 0.10 | 118  |

Recovery: Increasing amounts of acetylcholine receptor antibodies were added to three different serum samples with various initial acetylcholine receptor antibody levels, resulting in 9 different pools overall. Each sample was assayed in quadruplicate. The mean recovery ranged  $94\%$  to  $105\%$ .

|  Spiked Sample | Expected (nmol/L) | Recovered (nmol/L) | Recovery (%)  |
| --- | --- | --- | --- |
|  A | 0.86 | 0.81 | 94%  |
|  B | 1.13 | 1.09 | 97%  |
|  C | 1.29 | 1.21 | 94%  |
|  D | 1.55 | 1.49 | 96%  |
|  E | 2.08 | 2.13 | 102%  |
|  F | 2.42 | 2.36 | 97%  |
|  G | 3.69 | 3.89 | 105%  |
|  H | 4.15 | 4.27 | 103%  |
|  I | 4.64 | 4.85 | 105%  |

c. Traceability, Stability, Expected values (controls, calibrators, or methods): Traceability: No reference material is available.

Stability: Real-time stability testing of two lots of the AChRAb tracer (human acetylcholine receptor labeled with  $^{125}\mathrm{I}$ $\alpha$ -bungarotoxin) demonstrated that the unopened reagent is stable for three months at  $2 - 8^{\circ}\mathrm{C}$ . Other kit reagents are stable for at least 12 months at  $2 - 8^{\circ}\mathrm{C}$ . The reconstituted tracer is stable for at

{5}

least 30 days at 2-8°C.

Complete unopened kits were placed at different temperatures to evaluate the effect of extreme temperatures. The kits are stable for up to 30 days at 25°C, at least 9 days at -10°C and 37°C, and up to 6 days at 41°C.

d. Detection limit:

The Limit of Blank was determined to be 0.058 nmol/L. Functional sensitivity (Limit of Quantitation) was determined to be 0.1 nmol/L.

e. Analytical specificity:

Cross-reactivity with autoantibodies potentially interfering with the AchRAb Assay was assessed. Eight patient serum samples containing autoantibodies (anti-Sm/RNP, -Ro (SS-A), -La (SS-B), -dsDNA, -RF, and ANA antibodies) were tested in the assay. No cross-reactivities were found to AchRAb. In addition, hemoglobin (1.9 - 400 mg/dL), bilirubin (1.25 - 50 mg/dL), and triglycerides (25-3000 mg/dL) were evaluated for interference. All test results are available in the submission. Results showed no interference at the concentrations tested.

f. Assay cut-off:

Serum samples from 72 apparently healthy male and female adult blood donors were assayed. The range of values was between 0.0 and 0.22 nmol/L with a mean of 0.082 nmol/L and a standard deviation of 0.061 nmol/L. The cutoff was set at 0.25 nmol/L (approximately the mean + 3SD of the normal range). Antibody titers greater than 0.5 nmol/L are indicative of Myasthenia Gravis (positive). Values between 0.25 and 0.5 nmol/L are considered equivocal.

2. Comparison studies:

a. Method comparison with predicate device:

131 patient samples were tested with the AchRAb test and with the predicate assay. Equivocal results are not counted as positive or negative for the purpose of determining % Agreement.

|   | Predicate Assay (IBL ARAb)  |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|   |   |  + | Equivocal | - | Total  |
|  IVD AchRAb Assay | + | 96 | 1 | 0 | 97  |
|   |  Equivocal | 0 | 0 | 2 | 2  |
|   |  - | 1 | 3 | 28 | 32  |
|   |  Total | 97 | 4 | 30 | 131  |

Treating equivocal results as positive:

Positive agreement (97/101) = 96.0% (95% CI: 90.2 - 98.4)

Negative agreement (28/30) = 93.3% (95% CI: 78.7% to 98.2%)

Overall agreement (125/131) = 95.4% (95% CI: 90.4% to 97.9%)

Treating equivocal results as negative:

Positive agreement (96/97) = 99.0% (95% CI: 94.4% to 99.8%)

Negative agreement (31/34) = 97.1% (95% CI: 85.1% to 99.5%)

{6}

Overall agreement (127/131) = 98.5% (95% CI: 94.6% to 99.6%)

b. Matrix comparison:
Not applicable.

3. Clinical studies:

a. Clinical Sensitivity and Specificity:
40 samples with known clinical diagnosis of Myasthenia Gravis (MG) were tested using the AchRAb kit and the predicate kit. Both the AchRAb kit and the predicate kit correctly identified MG in all 40 samples. In addition, 16 samples from patients diagnosed with other neurological diseases were tested using the AchRAb kit and the predicate kit. All were identified as negative for MG.

|   | Myasthenia Gravis  |   |   |   |
| --- | --- | --- | --- | --- |
|   |   |  + | - | Total  |
|  IVD AchRAb Assay | + | 40 | 0 | 40  |
|   |  Equivocal | 0 | 0 | 0  |
|   |  - | 0 | 16 | 16  |
|   |  Total | 40 | 16 | 56  |

Sensitivity (40/40) = 100.0% (95% CI: 91.2% to 100%)
Specificity (16/16) = 100.0% 95% CI: 80.6% to 100%)
Overall agreement (56/56) = 100.0% (95% CI: 93.6 to 100%)

c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.

4. Clinical cut-off:
See Assay cut-off

5. Expected values/Reference range:
The expected value in normal population is &lt; 0.25 nmol/L.

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/device/K091468](https://fda.innolitics.com/device/K091468)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
