The intended use of the Philips NM3 monitor, Model 7900 is to provide: - cardiac output monitoring via the method of partial rebreathing in adult patients receiving mechanical ventilation during general anesthesia and in the intensive care unit (ICU). - spirometric, and carbon dioxide monitoring in neonatal, pediatric and adult patients during general anesthesia and in the intensive care unit (ICU) and the emergency department (ED). Separate combination CO2/flow sensors are provided for adult, pediatric and neonatal use. - continuous, non-invasive monitoring of functional arterial oxygen saturation and pulse rate in neonatal, pediatric and adult patients during both no motion conditions and for patients who are well or poorly perfused during general anesthesia and in the intensive care unit (ICU) and the emergency department (ED). The NM3 monitor Model 7900 and its sensors are intended to be used by trained operators when spirometric, capnographic, pulse oximetry, or cardiac output monitoring is indicated in the judgement of a physician. The use of the NM3 monitor Model 7900 for cardiac output monitoring is contraindicated in patients in which a small rise (3-5 mmHg) in their arterial partial pressure of CO2 level cannot be tolerated.
Device Story
Philips NM3 monitor provides multi-parameter physiological monitoring. Inputs include differential pressure signals from flow sensors, infrared CO2 absorption data from Capnostat 5 sensors, and red/infrared light absorption data from pulse oximetry sensors. Device uses microprocessor-based modules (Mercury, LoFlo, Masimo MX-1) to process signals. Cardiac output is calculated via non-invasive differential Fick partial rebreathing technique. Output displayed as numerical values and waveforms (plethysmogram) for clinician review. Used in ICU, anesthesia, and ED by trained operators. Assists clinical decision-making by providing real-time data on ventilation, oxygenation, and hemodynamics, potentially improving patient management during mechanical ventilation.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics and performance of previously cleared modules and sensors.
Technological Characteristics
Multi-parameter monitor integrating Mercury module (flow/CO2), LoFlo module (sidestream CO2), and Masimo MX-1 module (pulse oximetry). Uses fixed-orifice target flowmeters, infrared (IR) absorption for CO2, and dual-wavelength (red/IR) LED photoplethysmography for SpO2. Microprocessor-based data acquisition. Sensors include combination CO2/flow adapters and nasal/oral cannulas. Connectivity via front-panel sensor ports.
Indications for Use
Indicated for neonatal, pediatric, and adult patients in ICU, anesthesia, or ED settings requiring spirometry, capnography, or pulse oximetry. Cardiac output monitoring indicated for adult patients on mechanical ventilation. Contraindicated for cardiac output monitoring in patients unable to tolerate a 3-5 mmHg rise in arterial CO2.
Regulatory Classification
Identification
A monitoring spirometer is a device used to measure continuously a patient's tidal volume (volume of gas inhaled by the patient during each respiration cycle) or minute volume (the tidal volume multiplied by the rate of respiration for 1 minute) for the evaluation of the patient's ventilatory status.
{0}------------------------------------------------
K091459
# Section iv - 510 (k) Summary
[ Refer to 21 C.F.R § 807.92 ]
| Submitted by: | Respironics Novametrix, LLC<br>5 Technology Drive<br>Wallingford, CT 06492 | JUL 28 2009 |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Contact Person: | Kevin Mader<br>Q.A. and Regulatory Manager<br>Phone: 203-697-6466 | |
| Date Prepared: | 5/14/2009 | |
| Proprietary Name: | Philips NM3 monitor | |
| Common Name: | multi-parameter monitor (monitoring spirometer, CO2 monitor, pulse oximeter and<br>cardiac output monitor with partial rebreathing valve). | |
| Classification Name: | Class II, 21 CFR 868.1850, 868.1400, 868.5675 and 870.2700 | |
| Predicate Device: | NICO with MARS monitor, Model 7300 [510(k) K030886] | |
The Philips NM3 monitor is intended for non-invasive monitoring of the inspired Description of Device: and expired airflow and airway pressure of intensive care unit (ICU), anesthesia and emergency room (ER) patients, as well as capnography and pulse oximetry in all of these clinical settings. It is intended to serve the same purposes as the flow, carbon dioxide and pulse oximetry monitoring components of the predicate NICO with MARS monitor.
In the NM3 monitor as in the NCO system, combination CO2 adapter/flow sensors (neonatal, pediatric, adult), combination adult CO2 adapter/flow sensors with a partial rebreathing valve and flow sensors (infant/neonatal, pediatricladult) are connected with a male pneumatic connector to the NICO monitor. Sidestream airway adapters and nasal cannulas are available which are connected with a sample cell connector to a receptacle on the LoFlo Module which can be interfaced to the NM3 monitor. All of these sensors are already legally marketed as accessories of 510(k) cleared Respironics-Novametrix predicate devices. The pulse oximetry sensors are connected to the NM3 monitor via a connector on the front panel of the monitor. All of the pulse oximetry sensors are already legally marketed as accessories of 510(k) cleared Masimo predicate devices.
The principal function of the flow portion of combination sensors and flow sensors is to provide a differential pressure signal related to flow and airway pressure relative to atmospheric pressure. These sensors are often placed in the breathing circuit between the endotracheal tube and the ventilator circuit Y piece and may also be used in conjunction with a face mask or mouthpiece. The flow measurement portion of the NM3 monitor is contained in the Mercury module which consists of a microprocessor-based data acquisition system that measures flow, and pressure and interfaces with a Capnostat 5 CO2 sensor. The CO2 airway adapter portion of the combination sensors, allow the Respironics-Novametrix CO2 mainstream gas sensor, the Capnostat® 5, to attach to it and measure the concentration of CO2 in the airway using infrared technology. When CO2 measurements are combined with airway flow and volume measurements, other parameters such as CO2 production and dead space can be calculated. The Capnostat 5 sensor as a mainstream gas analyzer includes a sample cell positioned in the breathing circuit through which a patient's inspiratory and expiratory gases flow. The LoFlo module, a sidestream type of gas analyzer, samples gases at 50 ml/min from a sampling port in an adapter placed in a breathing circuit or from a nasal or oral cannula. The gas then passes through a sampling tube to the
## CONFIDENTIAL
{1}------------------------------------------------
sample cell, where the gas components are measured. The combination adult CO2 adapter/flow sensors with a partial rebreathing valve with periodic activation of the rebreathing valve allow pulmonary capillary blood flow and cardiac output to be calculated using the differential Fick method.
Intended Use of the Device: The intended use of the Philips NM3 monitor, Model 7900 is to provide:
- cardiac output monitoring via the method of partial rebreathing in adult patients receiving mechanical . ventilation during general anesthesia and in the intensive care unit (ICU).
- spirometric, and carbon dioxide monitoring in neonatal, pediatric and adult patients during general . anesthesia and in the intensive care unit (ICU) and the emergency department (ED). Separate combination CO2/flow sensors are provided for adult, pediatric and neonatal use.
- continuous, non-invasive monitoring of functional arterial oxygen saturation and pulse rate in . neonatal, pediatric and adult patients during both no motion conditions and for patients who are well or poorly perfused during general anesthesia and in the intensive care unit (ICU) and the emergency department (ED).
The NM3 monitor Model 7900 and its sensors are intended to be used by trained operators when spirometric, capnographic, pulse oximetry, or cardiac output monitoring is indicated in the judgement of a physician.
The use of the NM3 monitor Model 7900 for cardiac output monitoring is contraindicated in patients in which a small rise (3-5 mmHg) in their arterial partial pressure of CO2 level cannot be tolerated.
Technological Characteristics The NM3 monitor comprises the cleared Mercury module with the Capnostat 5 sensor for flow and carbon dioxide monitoring, the cleared LoFlo module for carbon dioxide monitoring and the cleared Masimo MX-1 module for pulse oximetry monitoring. It has been designed to include all of the functionality of the flow, carbon dioxide, cardiac output and pulse oximetry monitoring components of predicate NICO with MARS monitor.
The NM3 monitor uses flow sensors that are considered to be a fixed orifice, target flowmeters and as such the pressure drop is proportional to the square of the flow. Combination CO2/flow sensors are available in three flow ranges that are tailored for neonates, pediatric patients and adults.
The NM3 monitor uses an infrared absorption (IR) technique for monitoring CO2. The principle is based on the fact that CQ- molecules absorb infrared light energy of specific wavelengths, with the amount of energy absorbed being directly related to the CO2 concentration Solid state CO2 sensors (such as the Capnostat) use a beam splitter to simultaneously measure the IR light at two wavelengths: one which is absorbed by CO2 and one which is not. The wavelength which is not absorbed by CO2 is related to the intensity of the fR light source. Also, the IR light source is electronically pulsed in order to eliminate effects of changes in electronic components.
The NM3 monitor measures oxygen saturation and pulse rate with sensors that contain red and infrared light sources. Since oxygen saturated blood absorbs different amounts of light at each wavelength (red and infrared) as compared with unsaturated blood, the amount of light absorbed at each wavelength by the blood in each pulse can be used to calculate oxygen saturation. The light energy from red and infrared LEDs is beamed through a sample cell- a pulsating vascular bed, the patient's finger or toe for example. The remaining light energy not absorbed by the sample cell reaches a photodiode, on the opposing side of the sensor. The signal received by the photodiode is split into its red and infrared components, sampled, software filtered and displayed as a numerical value for oxygen saturation and as a waveform, the plethysmogram.
A variation on the traditional rebreathing methods, the non-invasive differential Fick partial re-breathing technique is used in the NM3 monitor. The change in VCO2 and the change in end-tidal CO2 in response to a change in ventilation is used to determine pulmonary capillary blood flow. This value is then corrected for the effect of shunt to determine cardiac output.
## CONFIDENTIAL
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the department's name around the perimeter. In the center of the logo is an abstract symbol that resembles a stylized caduceus, a symbol often associated with medicine and healthcare.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## JUL 288 2009
Mr. Kevin Mader Manager of Quality Assurance and Regulatory Affairs Respironics Novameterix, LLC Critical Care 5 Technology Drive Wallingford, Connecticut 06492
Re: K091459
Trade/Device Name: Philips NM3 Monitor Regulation Number: 21 CFR 868.1850 Regulation Name: Monitoring Spirometer Regulatory Class: II Product Code: BZK Dated: June 25, 2009 Received: June 30, 2009
### Dear Mr. Mader:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
## Page 2- Mader
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
For
Susan Runner, D.D.S., M.A. Acting Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
Respironics Novametrix LLC NM3 Monitor Special 510(k) - Device Modification
# Section ii Indications for Use
K 091459 510(k) Number (if known):
Device Name: Philips NM3 Monitor
Indications for Use:
Intended Use:
The intended use of the Philips NM3 monitor, Model 7900 is to provide:
- cardiac output monitoring via the method of partial rebreathing in adult patients receiving mechanical ventilation during general anesthesia and in the intensive care unit (ICU).
- spirometric, and carbon dioxide monitoring in neonatal, pediatric and adult patients during general anesthesia and in the intensive care unit (ICU) and the emergency department (ED). Separate combination CO2/flow sensors are provided for adult, pediatric and neonatal use.
- continuous, non-invasive monitoring of functional arterial oxygen saturation and pulse rate in neonatal, pediatric and adult patients during both no motion conditions and for patients who are well or poorly perfused during general anesthesia and in the intensive care unit (ICU) and the emergency department (ED).
The NM3 monitor Model 7900 and its sensors are intended to be used by trained operators when spirometric, capnographic, pulse oximetry, or cardiac output monitoring is indicated in the judgement of a physician.
The use of the NM3 monitor Model 7900 for cardiac output monitoring is contraindicated in patients in which a small rise (3-5 mmHg) in their arterial partial pressure of CO2 level cannot be tolerated.
| Prescription Use | X | AND/OR | Over-Th |
|----------------------------|---|---------|---------|
| (Per 21 CFR 801 Subpart D) | | (21 CF) | |
ne -Counter Use R 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|----------------------------------------------|--|
| Division of Anesthesiology, General Hospital | |
| Infection Control, Dental Devices | |
| 510(k) Number: | K91459 |
|----------------|--------|
|----------------|--------|
| Page 1 of | |
|----------------|--|
| Page 42 of 279 | |
CONFIDENTIAL
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.