The iVue is an optical coherence tomography system indicated for the in vivo imaging and measurement of the retinal nerve fiber layer, optic disk, cornea, and anterior chamber of the eye as an aid in the diagnosis and management of ocular diseases.
Device Story
iVue is a compact, laptop-based optical coherence tomography (OCT) system for ophthalmic imaging. It uses a low-coherence interferometer to scan the eye and measure retinal tissue reflectivity; cross-sectional structures are composed of sequences of A-scans. The device includes a cornea lens adapter for anterior segment imaging. Operated by a clinician in a clinical setting, the patient rests their head on a chin/forehead rest while the operator uses a joystick for alignment. The system provides a graphical user interface for image acquisition and analysis. Output consists of ocular tissue images and thickness measurements (e.g., RNFL, retinal thickness, corneal thickness). These outputs assist clinicians in diagnosing and managing ocular pathologies like glaucoma, AMD, DME, and ERM. The device is a more compact, affordable redesign of the RTVue system.
Clinical Evidence
Clinical study compared iVue to RTVue across 48 subjects (normal, glaucoma, retina, and cornea patients). Study evaluated precision (repeatability/reproducibility) and agreement (Limits of Agreement) for three scan patterns: Glaucoma (RNFL thickness), Retina (full/inner retinal thickness), and Cornea (thickness). Results demonstrated high precision and strong agreement with predicate device measurements across all pathologies. Bench testing included optical emission safety, electrical safety, and EMC.
Technological Characteristics
Fourier Domain OCT system; laptop-based; non-invasive; low-coherence interferometer; cornea lens adapter for anterior segment imaging; computer-controlled; joystick-aligned; software-based image acquisition and analysis.
Indications for Use
Indicated for in vivo imaging and measurement of retinal nerve fiber layer, optic disc, cornea, and anterior chamber of the eye to aid diagnosis and management of ocular diseases in patients requiring ophthalmic evaluation.
Regulatory Classification
Identification
An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.
Special Controls
*Classification.* Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the word "optovue" in a simple, sans-serif font. The "o" in "optovue" is stylized to resemble an eye. A curved line is drawn above the word, arching over the letters.
# 510(k) Summary
APR 2 2 2010
### Optovue, Incorporated iVue
This 510(k) summary for the iVue is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92
#### General Information
| Manufacturer: | Optovue, Inc.<br>45531 Northport Loop West,<br>Fremont, CA 94538<br>Phone: (510) 623-8868<br>Fax: (510) 623-8668<br>Registration No.: 3005950902 |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Azimun Jamal<br>Regulatory Manager<br>Optovue, Inc.<br>Phone: (510)623-8868 x188<br>e-mail: azimun_jamal@optovue.com |
**Device Information**
| Classification: | Class II |
|----------------------|------------------------------------------------|
| Trade Name: | iVue |
| Common Name: | Optical Coherence Tomography (OCT) |
| Classification Name: | Ophthalmoscope, a-c powered (21 CFR§ 886.1570) |
### Predicate Devices
(1) RTVue (K062552) (2) CA (K071250)
#### Intended Use
The iVue is an optical coherence tomography system indicated for the in vivo imaging and The I vac is an optical vehicles temself of the rayer, optic disc, cornea, and anterior chamber of the eye as an aid in the diagnosis and management of ocular diseases.
Optovue, Inc.
Page 1 of 8
{2}------------------------------------------------
#### Device Description
iVue is a modification of its predicate device RTVue (K062552). The intent of use, system performance, majority sub-assemblies, and key components of the iVue are all the same as RTVue. The intention of this redesign is to make the size of the iVue more compact and more affordable.
iVue also has a cornea lens adapter, like the predicate device CA (K071250), which can be attached to the front of the instrument to enable the iVue to image the cornea and anterior chamber of the eye.
i Vue, based on the same Optical Coherence Tomography (OCT) technology that is using in the predicate device RTVue (K062552) and CA (K071250), is a non-invasive diagnostic device for viewing the ocular tissue structure with micrometer range resolution. Both iVue and RTVue are designed and manufactured by Optovue Inc. A brochure of the RTVue and CA system is in appendices [1].
Like RTVue, iVue is a computer controlled ophthalmic imaging system. RTVue uses a desktop computer while i Vue uses a laptop computer. The device scans a patient's eye and uses a low coherence interferometer to measure the reflectivity of the retinal tissue. The cross sectional retinal tissue structure is composed of sequence of A-scans. It has a traditional patient and instrument interface like most ophthalmic devices. The patient will rest their head on the forehead and chin rest while the operator uses joystick to align the device to patient's eye. The computer has a graphic user interface for acquiring and analyzing the image.
i Vue also has similar scan patterns and analysis functions as used in RTVue.
#### Safety
The power of the scanning beam entering into patient's pupil is at same level as predicate devices. There is no change in the safety from previously submitted devices, the RTVue (K062552) and CA (K071250). A copy of the previous safety analysis for RTVue is attached in the Appendix [2]. The analysis was completed by an international recognized expert in the field of optical radiation hazards and safety, Dr. Dave Sliney.
#### Effectiveness
The predicate device has been routinely used in clinic. Numerous clinical papers and data have been published over last 2 years. Since there is no performance change, the effectiveness in clinical use is the same as predicate devices. A clinical comparison report and the IRB approved test protocol are attached in the premarket notification.
#### Substantial Equivalence
The iVue is substantially equivalent to the predicate devices identified previously. The iVue is substantially equivalent to the predicate devices with regard to intended use, operating principle, function, material, and energy source. The only difference from the predicate device is that the iVue is more compact in size.
#### Performance Data
(a) Non-clinic tests:
Optovue, Inc.
{3}------------------------------------------------
The iVue has had accuracy tests, optical emission safety analysis, electrical safety, electromagnetic compatibility test, and software validation tests.
#### (b) Clinical tests:
#### Description of Precision Study, Subject Selection Criteria and Pathologies
#### Instruments:
Two types of Fourier Domain OCT devices were used in the study. The iVue device is a new OCT device and is compared to the predicate device, the RTVue. For the study, three iVues were paired with three RTVues with three different operators and results were compared in normal eyes, and eyes with retina pathology, cornea pathology, and glaucoma.
#### Scans:
The iVue device has three scan patterns optimized to evaluate three ocular pathologies, namely, i) Cornea scan, ii) Retina scan, and iii) Glaucoma scan. These scan patterns were compared to three equivalent RTVue scans, i) Pachymetry scan, ii) EMM5 scan, and iii) ONH scan, respectively. The measurements and parameters calculated from these respective scans were the same. Both eyes were scanned in each subject.
#### Subjects:
Four types of subjects were enrolled in the study at two clinical sites, namely 1) normal healthy eyes with no ocular pathology, 2) glaucoma patients, 3) retina patients, and 4) cornea patients. Normal subjects were free from ocular pathology as determined by the Principal Investigator (PI) at each clinical site. Glaucoma patients were diagnosed as having glaucoma by the PI at each site. Retina patients included subjects with any type of retina pathology diagnosed by the PI. Retina pathologies included but were not limited to, AMD, DME, and ERM. Cornea patients included subjects with any type of cornea pathology as diagnosed by the PI, or cornea alteration including LASIK. There were 4 normal subjects, 4 cornea patients, 4 retina patients, and 4 glaucoma patients enrolled for each pair of iVue and RTVue devices. There were 3 pairs of iVue and RTVue devices at 2 clinical sites. At one site, there were 2 iVues systems, 2 RTVues systems and 2 operators at different times. In total, there were 16 subjects enrolled for each of the 3 pairs of iVue and RTVue devices; a total of 48 subjects were included in the study.
#### Selection Criteria:
All data was carefully reviewed for completeness and quality in two levels, namely, the subject level and the individual scan level. At the subject level, the CRF was carefully reviewed to qualify each subject against all inclusion and exclusion criteria by comparing the study protocol with the CRF. At the individual scan level, the data was reviewed for quality to ensure data meeting inclusion criteria was accepted.
In a clinical environment, only scans with acceptable quality should be used. In order to match our analysis with acceptable clinical results, we reviewed and excluded all scans in the study with poor image quality. Image quality is based on a number of factors, including overall signal strength, localized weak signals, eye blink, data out of boundary, and data off-center. Due to the selection criteria, the sample size for each scan type varies.
#### Precision Results
Optovue, Inc.
{4}------------------------------------------------
The following tables provide the precision results for the iVue scans. The data in the table includes the number of scans per subject group, the overall mean, the standard deviation, the repeatability standard deviation (Median value of subfields with the minimum subfield value and the maximum subfield value), and reproducibility standard deviation (Median value of subfields with the minimum subfield value and the maximum subfield value).
#### Glaucoma Scan Results (Normals and Glaucoma Patients) i)
The following table shows the overall precision results with the Glaucoma scan for average RNFL thickness for normals and glaucoma patients.
| Avg RNFL | Normal Patients | Glaucoma Patients |
|---------------------------------|-------------------|-------------------|
| # of Scans | 72 | 65 |
| Overall Mean (Overall SD) | 100.26 (8.32) | 85.94 (14.40) |
| Repeatability SD* (Min, Max) | 1.30 (1.30, 1.30) | 1.09 (1.09, 1.09) |
| Reproducibility SD** (Min, Max) | 1.30 (1.30, 1.30) | 7.03 (7.03, 7.03) |
*estimate of the standard deviation among measurements taken on the same subject using the same operator and device in the same testing session with repositioning.
** estimate of the standard deviation among measurements taken on different subjects using different operators and devices, including repeatability; in some cases the maximum reproducibility estimates were impacted by mean thickness difference between subjects in different sites.
The following table shows the overall precision results with the Glaucoma scan for the average of 16 sectors for normals and glaucoma patients.
| 16 RNFL Sectors | Normal Patients | Glaucoma Patients |
|---------------------------------|--------------------|--------------------|
| # of Scans | 72 | 65 |
| Overall Mean (Overall SD) | 100.26 (16.33) | 85.94 (18.68) |
| Repeatability SD* (Min, Max) | 3.74 (3.74, 5.82) | 3.58 (3.58, 6.13) |
| Reproducibility SD** (Min, Max) | 5.29 (4.27, 14.04) | 7.56 (3.58, 17.32) |
*estimate of the standard deviation among measurements taken on the same subject using the same operator and device in the same testing session with repositioning
** estimate of the standard deviation among measurements taken on different subjects using different operators and devices, including repeatability; in some cases the maximum reproducibility estimates were impacted by mean thickness difference between subjects in different sites.
ii) Retina Scan Results (Normals, Retina Patients and Glaucoma Patients)
The following table shows the overall precision results with the Retina scan for full retinal thickness in the fovea for normals, retina patients, and glaucoma patients (since inner retinal layer is affected by the disease of glaucoma, the precision results of retina scan on glaucoma patients are produced below).
| Full Retina Fovea | Normal Patients | Retina Patients | Glaucoma<br>Patients |
|---------------------------------|----------------------|----------------------|----------------------|
| # of Scans | 72 | 71 | 67 |
| Overall Mean (Overall SD) | 247.00 (26.03) | 269.28 (70.95) | 247.93 (21.20) |
| Repeatability SD* (Min, Max) | 2.54 (2.54, 2.54) | 4.41 (4.41, 4.41) | 3.33 (3.33, 3.33) |
| Reproducibility SD** (Min, Max) | 17.41 (17.41, 17.41) | 13.73 (13.73, 13.73) | 3.33 (3.33, 3.33) |
Optovue, Inc.
Page 4 of 8
{5}------------------------------------------------
*estimate of the standard deviation among measurements taken on the same operator and device in the same testing session with repositioning
** estimate of the standard deviation among measurements taken on different subjects using different operators and devices, including repeatability; in some cases the reproducibility estimates were impacted by mean thickness difference between subjects in different sites.
The following table shows the overall precision results with the Retina scan for full retinal thickness in the periphery (average over 8 sectors outside the fovea) for normals, retina patients, and glaucoma patients.
| Full Retina Peripheral | Normal Patients | Retina Patients | Glaucoma<br>Patients |
|---------------------------------|--------------------|-------------------|----------------------|
| # of Scans | 72 | 71 | 67 |
| Overall Mean (Overall SD) | 295.03 (16.95) | 298.35 (30.99) | 281.55 (24.90) |
| Repeatability SD* (Min, Max) | 1.89 (1.89, 3.04) | 2.60 (2.60, 4.11) | 2.24 (2.24, 3.33) |
| Reproducibility SD** (Min, Max) | 2.65 (1.89, 10.10) | 3.18 (2.60, 4.11) | 2.99 (2.24, 8.95) |
*estimate of the standard deviation among measurements taken on the same subject using the same operator and device in the same testing session with repositioning
** estimate of the standard deviation among measurements taken on different subjects using different operators and devices, including repeatability; in some cases the maximum reproducibility estimates were impacted by mean thickness difference between subjects in different sites.
The following table shows the overall precision results with the Retina scan for inner retinal thickness in the fovea for normals, retina patients, and glaucoma patients.
| Inner Retina Fovea | Normal Patients | Retina Patients | Glaucoma<br>Patients |
|---------------------------------|-------------------|-------------------|----------------------|
| # of Scans | 72 | 71 | 67 |
| Overall Mean (Overall SD) | 72.91 (14.36) | 77.88 (20.64) | 72.35 (10.28) |
| Repeatability SD* (Min, Max) | 2.94 (2.94, 2.94) | 3.14 (3.14, 3.14) | 1.99 (1.99, 1.99) |
| Reproducibility SD** (Min, Max) | 8.96 (8.96, 8.96) | 3.14 (3.14, 3.14) | 1.99 (1.99, 1.99) |
*estimate of the standard deviation among measurements taken on the same subject using the same operator and device in the same testing session with repositioning
** estimate of the standard deviation among measurements taken on different subjects using different operators and devices, including repeatability; in some cases the reproducibility estimates were impacted by mean thickness difference between subjects in different sites.
The following table shows the overall precision results with the Retina scan for inner retinal thickness in the periphery (average over 8 sectors outside the fovea) for normals, retina patients, and glaucoma patients.
| Inner Retina Peripheral | Normal Patients | Retina Patients | Glaucoma<br>Patients |
|---------------------------------|-------------------|-------------------|----------------------|
| # of Scans | 72 | 71 | 67 |
| Overall Mean (Overall SD) | 118.01 (9.37) | 117.37 (13.83) | 106.73 (12.05) |
| Repeatability SD* (Min, Max) | 1.45 (1.45, 2.82) | 1.91 (1.91, 3.41) | 1.88 (1.88, 2.83) |
| Reproducibility SD** (Min, Max) | 3.19 (2.22, 5.32) | 3.26 (1.91, 5.20) | 2.37 (1.88, 2.83) |
*estimate of the standard deviation among measurements taken on the same operator and device in the same testing session with repositioning
**estimate of the standard deviation among measurements taken on different subjects using different operators and devices, including repeatability; in some cases the maximum reproducibility estimates were impacted by mean thickness difference between subjects in different sites.
{6}------------------------------------------------
- iii) Cornea Scan Results (Normals and Cornea Patients)
The following table shows the overall precision results with the Cornea scan for central 0-2 mm cornea thickness for normals and cornea patients.
| Central Cornea 0-2 mm | Normal Patients | Cornea Patients |
|---------------------------------|-------------------|----------------------|
| # of Scans | 70 | 72 |
| Overall Mean (Overall SD) | 549.83 (20.85) | 531.80 (48.20) |
| Repeatability SD* (Min, Max) | 1.71 (1.71, 1.71) | 3.58 (3.58, 3.58) |
| Reproducibility SD** (Min, Max) | 1.71 (1.71, 1.71) | 14.52 (14.52, 14.52) |
*estimate of the standard deviation among measurements taken on the same subject using the same operator and device in the same testing session with repositioning
**estimate of the standard deviation among measurements taken on different subjects using different operators and devices, including repeatability; in some cases the maximum reproducibility estimates were impacted by mean thickness difference between subjects in different sites.
The following table shows the overall precision results with the Cornea scan for 2-5 mm cornea thickness measurements for normals and cornea patients.
| Cornea 2-5 mm | Normal Patients | Cornea Patients |
|---------------------------------|-------------------|----------------------|
| # of Scans | 70 | 72 |
| Overall Mean (Overall SD) | 569.23 (22.10) | 559.10 (49.50) |
| Repeatability SD* (Min, Max) | 2.42 (2.42, 5.44) | 5.27 (5.27, 7.79) |
| Reproducibility SD** (Min, Max) | 4.04 (2.42, 5.44) | 21.38 (14.29, 29.19) |
*estimate of the standard deviation among measurements taken on the same subject using the same operator and device in the same testing session with repositioning
** estimate of the standard deviation among measurements taken on different subjects using different operators and devices, including repeatability; in some cases the maximum reproducibility estimates were impacted by mean thickness difference between subjects in different sites.
The following table shows the overall precision results with the Cornea scan for 5-6 mm cornea thickness measurements for normals and cornea patients.
| Cornea 5-6 mm | Normal Patients | Cornea Patients |
|----------------------------------|-------------------|----------------------|
| # of Scans | 70 | 72 |
| Overall Mean (Overall SD) | 590.93 (24.55) | 589.81 (49.69) |
| Repeatability SD* (Min, Max) | 3.63 (3.63, 7.55) | 6.50 (6.50, 11.72) |
| Reproducibility SD ** (Min, Max) | 5.77 (3.63, 7.55) | 24.31 (13.97, 38.47) |
*estimate of the standard deviation among measurements taken on the same subject using the same operator and device in the same testing session with repositioning
** estimate of the standard deviation among measurements taken on different subjects using different operators and devices, including repeatability; in some cases the maximum reproducibility estimates were impacted by mean thickness difference between subjects in different sites.
#### Agreement Results
The following tables provide the overall agreement results for the iVue scans compared to the RTVue scans. The data in the table includes the overall mean differences averaged across all measurements for that scan pattern (with the minimum and maximum differences provided in parentheses) and the
{7}------------------------------------------------
standard deviation of the differences (with the minimum standard deviations provided in parentheses).
#### i) Glaucoma LOA Results
The following table shows the overall agreement results for average RNFL thickness measurement in the iVue Glaucoma scan and the RTVue ONH scan for normals and glaucoma patients.
| Avg RNFL | Normal Patients | Glaucoma Patients |
|-----------------------------|----------------------|----------------------|
| Mean Differences (Min, Max) | -0.55 (-0.61, -0.49) | -0.52 (-0.87, -0.18) |
| Estimated STDEV (Min, Max) | 2.25 (2.12, 2.38) | 1.74 (1.71, 1.76) |
The following table shows the overall agreement results for the 16 sector RNFL thickness measurement in the iVue Glaucoma scan and the RTVue ONH scan for normals and glaucoma patients.
| 16 RNFL Sector Averages | Normal Patients | Glaucoma Patients |
|-----------------------------|---------------------|---------------------|
| Mean Differences (Min, Max) | -0.55 (-8.00, 4.49) | -0.53 (-7.19, 4.25) |
| Estimated STDEV (Min, Max) | 8.16 (6.26, 11.01) | 7.61 (5.01, 10.39) |
#### ii) Retina LOA Results
The following table shows the overall agreement results for full retina fovea thickness measurement in the iVue Retina scan and the RTVue EMM5 scan for normals, glaucoma patients, and retina patients
| Full Retina Fovea | Normal Patients | Glaucoma Patients | Retina Patients |
|-----------------------------|-------------------|-------------------|-------------------|
| Mean Differences (Min, Max) | 2.64 (1.91, 3.37) | 1.99 (1.64, 2.35) | 2.14 (1.13, 3.14) |
| Estimated STDEV (Min, Max) | 3.79 (3.39, 4.20) | 4.88 (4.60, 5.16) | 6.00 (5.36, 6.64) |
The following table shows the overall agreement results for full retina peripheral thickness measurements in the iVue Retina scan and the RTVue EMM5 scan for normals, glaucoma patients, and retina patients.
| Full Retina Periphery | Normal Patients | Glaucoma Patients | Retina Patients |
|-----------------------------|---------------------|---------------------|---------------------|
| Mean Differences (Min, Max) | -0.27 (-3.48, 3.21) | -1.00 (-4.18, 2.40) | -0.41 (-2.83, 2.30) |
| Estimated STDEV (Min, Max) | 3.48 (2.28, 4.92) | 4.56 (3.66, 5.34) | 4.54 (3.70, 5.51) |
The following table shows the overall agreement results for inner retina fovea thickness measurement in the iVue Retina scan and the RTVue EMM5 scan for normals, glaucoma patients, and retina patients.
| Inner Retina Fovea | Normal Patients | Glaucoma Patients | Retina Patients |
|-----------------------------|-------------------|-------------------|-------------------|
| Mean Differences (Min, Max) | 2.64 (1.91, 3.37) | 1.70 (1.58, 1.82) | 2.39 (1.92, 2.86) |
| Estimated STDEV (Min, Max) | 3.79 (3.39, 4.20) | 4.15 (3.51, 4.79) | 4.73 (4.61, 4.86) |
{8}------------------------------------------------
The following table shows the overall agreement results for inner retina peripheral thickness measurements in the iVue Retina scan and the RTVue EMM5 scan for normals, glaucoma patients, and retina patients.
| Inner Retina Periphery | Normal Patients | Glaucoma Patients | Retina Patients |
|-----------------------------|---------------------|---------------------|---------------------|
| Mean Differences (Min, Max) | -0.27 (-3.48, 3.21) | -0.50 (-3.22, 2.17) | -0.26 (-2.10, 2.18) |
| Estimated STDEV (Min, Max) | 3.48 (2.28, 4.92) | 4.05 (2.70, 5.94) | 4.28 (2.93, 5.91) |
iii) Cornea LOA Results
The following table shows the overall agreement results for the central 0-2mm cornea thickness measurement in the iVue Cornea scan and the RTVue Pachymetry scan for normals and cornea patients.
| 0-2 mm Central Cornea | Normal Patients | Cornea Patients |
|-----------------------------|----------------------|----------------------|
| Mean Differences (Min, Max) | -0.44 (-0.64, -0.23) | -1.04 (-1.41, -0.67) |
| Estimated STDEV (Min, Max) | 5.25 (4.86, 5.63) | 5.70 (5.63, 5.78) |
The following table shows the overall agreement results for the central 2-5 mm cornea thickness measurement in the iVue Cornea scan and the RTVue Pachymetry scan for normals and cornea patients.
| 2-5 mm Cornea | Normal Patients | Cornea Patients |
|-----------------------------|--------------------|---------------------|
| Mean Differences (Min, Max) | 1.27 (-0.74, 2.85) | 0.07 (-4.01, 3.72) |
| Estimated STDEV (Min, Max) | 7.93 (4.14, 11.71) | 12.35 (7.81, 16.64) |
The following table shows the overall agreement results for the central 5-6 mm cornea thickness measurement in the iVue Cornea scan and the RTVue Pachymetry scan for normals and cornea patients.
| 5-6mm Cornea | Normal Patients | Cornea Patients |
|-----------------------------|---------------------|---------------------|
| Mean Differences (Min, Max) | 2.71 (-1.14, 5.05) | 1.79 (-3.15, 8.29) |
| Estimated STDEV (Min, Max) | 10.27 (5.39, 13.32) | 15.28 (8.99, 21.84) |
The evidence from the clinical performance data study demonstrates that the iVue is substantially equivalent to the predicate device, the RTVue.
#### Conclusion
As described in this 510(k) Summary, comprehensive testing and analysis was conducted on the iVue to ensure that the device is safe and effective for its intended use when used in accordance with its instructions for use.
{9}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wings, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The eagle is facing left, and the text is in all capital letters.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Optovue, Inc c/o Ms. Azimun Jamal Manager of Quality/Regulatory 45531 Northport Loop W. Fremont, CA. 94538
APR 2 2 2010
Re: K091404
Trade/Device Name: iVue Model iVue100 Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: Class II Product Code: HLI Dated: April 19, 2010 Received: April 20, 2010
Dear Ms. Jamal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's fequirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{10}------------------------------------------------
Page 2 – Ms. Azimun Jamal
Enclosure
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Ehturn for
Malvina B. Eydelman, M.D. Director
Division of Ophthalmic, Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
{11}------------------------------------------------
## Statement of Indications for Use
510(k) Number (if known): K091404
Device Name: iVue
Indications for Use:
The iVue is an optical coherence tomography system indicated for the in vivo imaging and measurement of the retinal nerve fiber layer, optic disk, cornea, and anterior chamber of the eye as an aid in the diagnosis and management of ocular diseases.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
teon.
(Division Sign-೧೧) Division of Ophthalmic Newrological and Ear, Nose and Throat De vices
510(k) Number K091404
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.