K091287 · Common Sense , Ltd. · CEN · Aug 31, 2009 · Clinical Chemistry
Device Facts
Record ID
K091287
Device Name
VS-SENSE TEST
Applicant
Common Sense , Ltd.
Product Code
CEN · Clinical Chemistry
Decision Date
Aug 31, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1550
Device Class
Class 1
Indications for Use
The VS-SENSE™ TEST is a qualitative, visually-read swab for clinicians who wish to evaluate women with vaginal symptoms. The device is a vaginal pH indicator swab intended to be used in conjunction with other clinical examinations, such as the Amsel criteria or the Nugent Gram stain, to aid in determining conditions characterized by elevated vaginal pH, such as bacterial vaginosis.
Device Story
VS-Sense is a qualitative, colorimetric vaginal pH indicator swab. Device utilizes nitrazine yellow indicator to detect vaginal discharge pH levels. When vaginal discharge pH is ≥ 4.7, swab changes color from yellow to blue or green. Used in clinical settings by healthcare providers to evaluate symptomatic women. Device serves as an adjunct to standard clinical examinations (e.g., Amsel criteria, Nugent Gram stain) to aid in identifying conditions like bacterial vaginosis. Clinician visually interprets color change on swab; result informs clinical decision-making alongside other diagnostic tests. Benefits include rapid, point-of-care assessment of vaginal pH to assist in differential diagnosis of vaginal symptoms.
Clinical Evidence
Prospective clinical study of 254 pre-menopausal symptomatic women. Compared VS-Sense results against clinical diagnosis (defined by Trichomonas culture, Gram stain Nugent score > 7, or 3 Amsel criteria). VS-Sense demonstrated 86% sensitivity (95% CI: 78%-91%) and 94% specificity (95% CI: 88%-97%). Overall agreement with clinical diagnosis was 90%.
Technological Characteristics
Vaginal swab with tip coated in proprietary polymer matrix containing Nitrazine Yellow indicator. Qualitative colorimetric pH detection (yellow to blue/green) for pH >4.7. Standalone device, no energy source, no software.
Indications for Use
Indicated for pre-menopausal symptomatic women aged 18 or older presenting with acute vulvovaginal symptoms. Contraindicated in women who are menopausal, have blood in vaginal secretions, have had sexual relations or vaginal douching within 12 hours, have applied vaginal medication within 3 days, or have symptoms/signs of pelvic inflammatory disease.
Regulatory Classification
Identification
A urinary pH (nonquantitative) test system is a device intended to estimate the pH of urine. Estimations of pH are used to evaluate the acidity or alkalinity of urine as it relates to numerous renal and metabolic disorders and in the monitoring of patients with certain diets.
Predicate Devices
Phem Check™ (k960648)
Submission Summary (Full Text)
{1}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k091287
B. Purpose for Submission:
New device
C. Measurand:
pH (vaginal)
D. Type of Test:
Qualitative, colorimetric test for acidity read visually.
E. Applicant:
Common Sense Ltd.
F. Proprietary and Established Names:
VS-Sense™
G. Regulatory Information:
1. Regulation section:
21CFR862.1550, Urinary pH (non-quantitative) test system
2. Classification:
Class I, subject to limitation from exemption 862.9(6)
3. Product code:
CEN
4. Panel:
75, Chemistry
{2}
H. Intended Use:
1. Intended use(s):
See indications for use below.
2. Indication(s) for use:
The VS Sense Test is a qualitative, visually-read swab for clinicians who wish to evaluate women with vaginal symptoms. The device is a vaginal pH indicator swab intended to be used in conjunction with other clinical examinations, such as the Amsel criteria or Nugent Gram stain, to aid in determining conditions characterized by elevated vaginal pH, such as bacterial vaginosis.
3. Special conditions for use statement(s):
Prescription use
4. Special instrument requirements:
None
I. Device Description:
Vaginal swab containing the colorimetric indicator, nitrazine yellow
J. Substantial Equivalence Information:
1. Predicate device name(s):
Phem Check™
2. Predicate K number(s):
k960648
3. Comparison with predicate:
Both devices are vaginal pH indicators with similar indications. The VS Sense device shows a blue or green stain on a swab to indicate pH of 4.7 or above. The predicate device uses pH indicator paper mounted on a flat probe.
K. Standard/Guidance Document Referenced (if applicable):
No recognized standards were referenced.
{3}
L. Test Principle:
The test includes a colorimetric pH indicator incorporated into a vaginal swab. The swab turns blue or green when the vaginal discharge pH level is ≥ 4.7.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
The test swabs were tested with samples of 100mM citrate phosphate buffer, pH ranging from 3.0 to 7.0, and the color was recorded. Reproducibility was evaluated for three product lots, by three operators, over two days, using masked samples. The total number of samples tested per pH level was 54. The pH of the samples included 3.0, 4.0, 4.3, 4.5, 4.7, 5.0, 5.2, 5.5, 6.0, and 7.0. All samples at pH of 4.5 and below were negative (yellow); Samples at 4.7 and above were positive (colors ranging from light green near 4.7, to purple near pH=7.0).
Similar cutoff results were obtained when tests were shipped for transport/stability testing.
In addition, as supplemental support for repeatability, ten physicians each collected duplicate patient samples from patients they were examining. The same results were obtained for each repeat measurement.
b. Linearity/assay reportable range:
Not applicable. See precision and clinical study section for ranges tested.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The VS-Sense is traceable to standardized buffers by electrometric measurements (ANSI Standard ASTM D 1293-99).
Stability testing supported expiration date.
d. Detection limit:
See precision/reproducibility section above.
e. Analytical specificity:
Interference was tested using control solutions at pH 4.0, 5.0, and 5.5 (both 20 mM and 100 mM Phosphate-Citrate buffer). A variety of over-the counter
3
{4}
vaginal products were tested, including lubricants, antifungal treatments, antiseptic creams, and spermicides. Two hundred milligrams of each product tested was added to 7 uL of the buffered solutions. The VS Sense was exposed to each mixture and the result was recorded. No interference was observed for GLOVAN, micronazole nitrate, Liquibeads, Vagigard, Vagistat - 1, Monistat or Gyno Pevaryl. All tested products that showed some interference are listed in the package insert.
f. Assay cut-off:
See precision/reproducibility, section above.
2. Comparison studies:
a. Method comparison with predicate device:
See “Clinical Studies” section, below.
b. Matrix comparison:
Not applicable. The test uses only vaginal swabs.
3. Clinical studies:
a. Clinical Sensitivity:
A total of 254 women participated in a study to evaluate performance of the VS Sense. The study was conducted at three US sites. Subjects included were pre-menopausal symptomatic women aged 18 or above, with acute vulvovaginal symptoms, who were willing and able to sign the consent form. Women were excluded from the study if any of the following applied: menopausal, blood in vaginal secretions, had sexual relations or applied vaginal douche within previous 12 hours, applied vaginal medication within the last 3 days, or had symptoms and signs of pelvic inflammatory disease.
At the clinic, a nurse obtained a vaginal secretion specimen with the VS-Sense swab, and results were immediately recorded and were, unavailable to the clinician performing the subsequent exam and diagnosis. Following this, a clinician performed a speculum vaginal exam. Samples were obtained for microscopy, gram stain, and yeast and trichomonas culture (using Diamond’s medium or InPouch TV), KOH whiff (amine) test, and nitrazine paper test. Positive BV was defined by Gram stain Nugent score > 7 and positivity for 3 Amsel criteria. Nurse’s results with the VS-Sense were compared to the physician’s clinical diagnosis. Positive clinical diagnosis was defined as positive in one of the following tests: Trichomonas culture in Diamond’s medium or InPouch TMTV, BV as determined by Gram stain, or BV as
4
{5}
determined by 3 positive Amsel criteria (one of which had to include at least 20% Clue cells). Atrophic vaginitis and desquamative inflammatory vaginitis were diagnosed by standard clinical examination and by microscopic observation, respectively.
Results of this study in terms of VS Sense versus clinical diagnosis are shown in the table below:
| | Clinical Diagnosis | | |
| --- | --- | --- | --- |
| VS Sense | | Positive | Negative |
| | Positive | 101 | 9 |
| | Negative | 16 | 128 |
Sensitivity of 86%, with 95% CI: [78%, 91%]
Specificity of 94%, with 95% CI: [88%, 97%]
Overall agreement: 90%
Results in terms of the comparator/Nitrazine paper were also evaluated versus clinical diagnosis. Results are shown in the table below:
| | | Clinical Diagnosis | |
| --- | --- | --- | --- |
| Nitrazine paper | | Positive | Negative |
| | Positive | 91 | 24 |
| | Negative | 9 | 128 |
Sensitivity of 91%, with 95% CI: [84%, 96%]
Specificity of 84%, with 95% CI: [77%, 90%]
Overall agreement: 87%
b. Clinical specificity:
See clinical sensitivity, above.
c. Other clinical supportive data (when a. and b. are not applicable):
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
A pH level in the range of 3.8 to 4.2 is considered normal.
{6}
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.