← Product Code [KPS](/productcode/KPS) · K091269

# NEUROPET (K091269)

_Photodetection Systems, Inc. · KPS · May 8, 2009 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K091269

## Device Facts

- **Applicant:** Photodetection Systems, Inc.
- **Product Code:** [KPS](/productcode/KPS.md)
- **Decision Date:** May 8, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1200
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

This device is a Positron Emission Tomography (PET) system used to detect and display the distribution of positron emitting radionuclides in parts of the human body that will fit in the patient aperture. This device is to be used by trained healthcare professionals. This information can assist in research, diagnosis, therapeutic planning and therapeutic outcome assessment.

## Device Story

NeuroPET is a small-aperture PET scanner; detects coincident gamma pairs from positron annihilation. System comprises ring gantry/acquisition unit, patient table, and console/PC. Used in clinical/research settings by trained healthcare professionals. Console/PC manages scan control, image reconstruction, data archiving, viewing, and analysis. Provides visual distribution maps of radionuclides to assist clinicians in diagnosis, treatment planning, and outcome assessment.

## Clinical Evidence

Bench testing and image data provided consistent with FDA guidance for PET systems and software-containing devices. No clinical trial data reported.

## Technological Characteristics

Small aperture PET scanner; ring gantry design; 31 cm bore size. Uses coincident gamma pair detection. Includes patient table and console/PC workstation. Connectivity via network. Software-based image reconstruction and analysis.

## Regulatory Identification

An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.

## Predicate Devices

- Toshiba Medical Systems PCA-8000A, Mark II, PET / CT System

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Traditional 510(k) Section 5 - 510(k) Summary PhotoDetection Systems- NeuroPET KO 91269

# 510(k) Summary

Date prepared

March 23, 2009

Company PhotoDetection Systems, Inc. 85 Swanson Rd Boxborough, MA 01719-1300

**Official Contact**

Paul Domigan, President Phone: (978) 266-0420 Fax: (978) 266-0426

## Trade name

NeuroPET

Common name PET (positron emission tomography) system

Classification name System, tomography, computed, emission

Regulation: 21 CFR 892.1200

Product code: KPS

Predicate device: Toshiba Medical Systems PCA-8000A, Mark II, PET / CT System

PhotoDetection Systems, Inc. 85 Swanson Rd Boxborough, MA 01719-1300 MAY ัง 8 2009

{1}------------------------------------------------

#### Device description:

The PhotoDetection Systems' NeuroPET is a small aperture positron emission tomography PET scanner that produces images based on the distribution of positron emitting isotopes in parts of the human body that will fit in the patient aperture for medical diagnostic and research purposes.

The NeuroPET system consists of three subsystems (shown below); the NeuroPET Ring Gantry/Acquisition System, Patient Table and the Console/PC. The NeuroPET Ring detects coincident gamma pairs emitted from positron annihilation and contains all the data acquisition electronics and reconstruction computers. The Patient Table is used to

Image /page/1/Picture/5 description: The image shows a medical examination room with a table, a computer, and a medical imaging machine. The table is in the center of the room and has wheels on the bottom. The computer is on a rolling stand to the left of the table. The medical imaging machine is on the right side of the room and has a large opening in the front.

comfortably position the patient relative to the scanner. The Console/PC provides the operator interface for the system including such functions as system and scan control, reconstruction, data archiving, image viewing, analysis, and network connectivity.

-For a more detailed description see Section 11 of this submission.

PhotoDetection Systems, Inc. 85 Swanson Rd Boxborough, MA 01719-1300 Page 2 of 3

{2}------------------------------------------------

#### Indications for Use:

This device is a Positron Emission Tomography (PET) system used to detect and display the distribution of positron emitting radionuclides in parts of the human body that will fit in the patient aperture. This device is to be used by trained healthcare professionals. This information can assist in research, diagnosis, therapeutic planning and therapeutic outcome assessment.

#### Differences Between Other Legally Marketed Predicate Devices

The NeuroPET is viewed as substantially equivalent to the PET portion of the Toshiba PCA-8000A/Mark II.

The primary differences between the two PET systems are the:

- Standalone PET vs. PET integrated into PET/CT
- 후 Method of attenuation correction: CAC vs. CTAC
- t Bore size: 31 cm vs. 90 cm
- . Operator Workstation: NeuroPET GUI vs. Toshiba PET/CT GUI

A detailed comparison of the two systems is found in section 12.

The two PET systems are substantially equivalent in function, safety, and effectiveness to the PET component of the Toshiba Mark II CT.

#### Conclusion:

Based on submitted non-clinical and image data consistent with FDA's guidance for PET systems and for devices that include software, the NeuroPET is substantially equivalent in technology, intended use, safety, and effectiveness to the PET imaging component of the Toshiba PCA-8000A/Mark II.

PhotoDetection Systems, Inc. 85 Swanson Rd Boxborough, MA 01719-1300 Page 3 of 3

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the graphic.

### DEPARTMENT OF HEALTH & HUMAN SERVICES

MAY - 8 2009

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

PhotoDetection Systems, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313

Re: K091269

Trade/Device Name: NeuroPET Positron Emission Tomography (PET) System Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II Product Code: KPS Dated: April 29, 2009 Received: April 30, 2009

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 |
|----------------|----------------------------------|----------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology)          | (240) 276-0115 |
| 21 CFR 892.xxx | (Radiology)                      | (240) 276-0120 |
| Other          |                                  | (240) 276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

[signature]

Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Traditional 510(k) Section 4 - Indications for Use

### Indications for Use

510(k) Number (if known): K071267

Device Name: NeuroPET Positron Emission Tomography (PET) System Indications for Use:

This device is a Positron Emission Tomography (PET) system used to detect and display the distribution of positron emitting radionuclides in parts of the human body that will fit in the patient aperture. This device is to be used by trained healthcare professionals. This information can assist in research, diagnosis, therapeutic planning and therapeutic outcome assessment.

Over-The-Counter Use Prescription Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Colin Mc Pollard

(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number

PhotoDetection Systems, Inc. 85 Swanson Rd Boxborough, MA 01719-1300 Page 1 of 1

---

**Source:** [https://fda.innolitics.com/device/K091269](https://fda.innolitics.com/device/K091269)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
