HOSPIRA LATEX-FREE CRITICAL CARE AND ADVANCED SENSOR CATHETERS
Applicant
Hospira, Inc.
Product Code
DQE · Cardiovascular
Decision Date
Sep 18, 2009
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1230
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Hospira Advanced Sensor Catheters: Indicated for the assessment of the hemodynamic status of a patient, including but not restricted to the following: Venous Pressures, Cardiac Output. Oxyhemoglobin Saturation, and Venous Blood Sampling. A secondary indication is for the therapeutic infusion of solutions. Hospira Critical Care Catheters: Indicated for the assessment of hemodynamic status through right atrial, right ventricular, and pulmonary artery and/or wedge pressure monitoring for patients including the following : acute heart failure; differentiating ruptured ventricular septum form mitral regurgitation; diagnosis of tamponate; severe hypovolemia; complex circulatory situations (e.g., fluid management with acute burn patients); medical emergencies; adult respiratory distress syndrome; gram negative sepsis; drug intoxication; acute renal failure; hemorrhagic pancreatitis; intra- and postoperative management of high risk patients; history of pulmonary or cardiac disease; fluid shifts (such as extensive intra-abdominal operations); management of high risk obstetrical patients; known cardiac disease; toxemia; premature separation of the placenta; cardiac output determination by thermodilution method; and blood sampling. Additional indication for Hospira Critical Care Pacing Lead Catheters: Indicated for temporary transluminal ventricular pacing using a temporary ventricular lead.
Device Story
Multi-lumen catheters for pulmonary artery placement; distal balloon for flow-directed positioning; thermistor for core temperature/cardiac output; fiber optics for mixed venous oxygen saturation (SvO2); access ports for drug delivery/blood sampling; pacing lead port on specific models. Used in clinical settings (OR, ICU) by physicians for hemodynamic assessment. Provides real-time physiological data to guide fluid management, drug therapy, and cardiac support. Benefits include improved diagnostic accuracy in complex circulatory states and critical care management.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included biocompatibility, expansion symmetry, over-inflation, multiple inflations, deflation time, and burst strength; all passed acceptance criteria.
Indicated for patients requiring hemodynamic monitoring (e.g., acute heart failure, hypovolemia, sepsis, post-op management, cardiac/pulmonary disease) or temporary transluminal ventricular pacing. Applicable to patients needing venous pressure, cardiac output, SvO2 monitoring, or therapeutic infusion.
Regulatory Classification
Identification
A fiberoptic oximeter catheter is a device used to estimate the oxygen saturation of the blood. It consists of two fiberoptic bundles that conduct light at a desired wavelength through blood and detect the reflected and scattered light at the distal end of the catheter.
Predicate Devices
Hospira Critical Care and Advanced Sensor Catheter (K061450)
Submission Summary (Full Text)
{1}------------------------------------------------
K0911268
April 2009
Hospira, Inc. Confidential
Special 510(k) Hospira Critical Care and Advanced Sensor Catheters
# SEP 1 8 2009
#### 1 510(k) SUMMARY
#### 1. Submitted by:
Hospira, Inc. D-389 Bldg. H2 275 N. Field Drive Lake Forest, IL 60045 Phone: (224)212-5316 Fax: (224) 212-5401
Contact: Daniela Weksler
#### Manufacturer and Establishment Registration Number: 2.
#### Manufacturer:
ICU Medical (UTAH), Inc. 4455 Atherton Dr. Salt Lake City, UT 84123
Establishment Registration #: 1713468
| 3. Date Prepared: | April 24, 2009 |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 4. Name/Classification<br>of Device: | Class II<br>74 DYG - 21 CFR 870.1240 - Flow-Directed Catheter<br>74 KRB - 21 CFR 870.1915 - Thermodilution Probe<br>74 DQE - 21 CFR 870.1230 - Fiberoptic Oximeter<br>Catheter |
| 5. Trade Name of<br>Proposed Device: | Hospira Critical Care and<br>Advanced Sensor Catheter |
- 6. Predicate Devices: Hospira Critical Care and Advanced Sensor Catheter (K061450)
### 7. Proposed Device Description:
The Hospira Critical Care and Advanced Sensor Catheters have multi-lumens that incorporate some or all of the following components and features: a distal balloon for positioning the catheter tip via blood flow within the pulmonary artery, a heater coil for determining continuous cardiac output, a thermistor for monitoring core temperature and cardiac output, fiber optics for monitoring mixed venous oxygen saturation (SvQ2), and access ports for drug delivery or blood sampling. In addition, the Hospira Critical Care Pacing Lead Catheters include an access port for use with Transluminal Right Ventricular Pacing Leads. The catheters also incorporate insertion distance markings and are provided with a syringe for inflating the balloon.
{2}------------------------------------------------
#### 8. Statement of Intended Use:
#### Hospira Advanced Sensor Catheters:
Indicated for the assessment of the hemodynamic status of a patient, including but not restricted to the following: Venous Pressures, Cardiac Output, Oxyhemoglobin Saturation, and Venous Blood Sampling. A secondary indication is for the therapeutic infusion of solutions.
#### Hospira Critical Care Catheters:
Indicated for the assessment of hemodynamic status through right atrial, right ventricular, and pulmonary artery and/or wedge pressure monitoring for patients including the following : acute heart failure; differentiating ruptured ventricular septum form mitral regurgitation; diagnosis of tamponate; severe hypovolemia; complex circulatory situations (e.g., fluid management with acute burn patients); medical emergencies; adult respiratory distress syndrome; gram negative sepsis; drug intoxication; acute renal failure; hemorrhagic pancreatitis; intra- and postoperative management of high risk patients; history of pulmonary or cardiac disease; fluid shifts (such as extensive intra-abdominal operations); management of high risk obstetrical patients; known cardiac disease; toxemia; premature separation of the placenta; cardiac output determination by thermodilution method; and blood sampling.
Additional indication for Hospira Critical Care Pacing Lead Catheters:
Indicated for temporary transluminal ventricular pacing using a temporary ventricular lead.
#### 9. Summary of Substantial Equivalence
The Hospira Critical Care and Advanced Sensor Catheters as described in this submission are substantially equivalent to the predicate Hospira Critical Care and Advanced Sensor Catheters (K061450) with respect to manufacturing methods and materials with respect to the following characteristics:
#### Similarities:
- 1) The catheters have the same intended use and indications for use.
- 2) The catheters contain the same type of components.
- 3) The method of sterilization is the same.
- 4) The manufacturing process is the same.
- 5) The resin blend used for the balloon is the same.
- 6) The resin used for the bump and extrusion tubing of the Advanced Sensor Catheters is the same.
76
{3}------------------------------------------------
#### Differences:
- 1) The material used to manufacture the balloon in both the Critical Care and Advance Sensor Catheters will be a slightly different formulation.
- 2) The resin dye and stiffness specification for the extrusion used in the Advance Sensor Catheters.
- 3) The bond coat insulation on the heater lead wire for the Advance Sensor Catheters
### 10. Statement of Safety and Effectiveness
The Hospira Critical Care and Advanced Sensor Catheters have been tested for biocompatibility and for expansion symmetry, over inflation, multiple inflations, deflation time, and burst strength and have passed all of the acceptance criteria. The Hospira Critical Care and Advanced Sensor Catheters meet the functional claims and intended use as described in the product labeling, and are as safe and effective in terms of substantial equivalence as the predicate catheters described in this document.
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the circle is an abstract symbol that resembles a stylized eagle or bird with three wing-like shapes. The symbol is black, and the text is also black.
#### Public Health Service
SEP 1 8 2009
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Hospira, Inc. c/o Mr. Keith R. Dunn-Manager, Global Regulatory Affairs 275 N. Field Dr. Lake Forest, IL 60045
Re: K091268
Trade/Device Name: Hospira Critical Care and Advanced Sensor Catheters (see enclosure for listing of 29 models) Regulatory Number: 21 CFR 870.1230 Regulation Name: Fiberoptic Oximeter Catheter Regulatory Class: Class II (Two) Product Code: DQE, DYG and KRB
Dated: August 17, 2009
Received: August 18, 2009
### Dear Mr. Dunn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
## Page 2 - Mr. Keith R. Dunn
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set. forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
W.M.S.T.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{6}------------------------------------------------
# Indications for Use
510(k) Number: K091268
Device Name:
Hospira Critical Care and Advanced Sensor Catheters
### Indications for Use:
Hospira Advanced Sensor Catheters:
Indicated for the assessment of the hemodynamic status of a patient, including but not restricted to the following: Venous Pressures, Cardiac Output. Oxyhemoglobin Saturation, and Venous Blood Sampling. A secondary indication is for the therapeutic infusion of solutions.
#### Hospira Critical Care Catheters;
Indicated for the assessment of hemodynamic status through right atrial, right ventricular, and pulmonary artery and/or wedge pressure monitoring for patients including the following : acute heart failure; differentiating ruptured ventricular septum form mitral regurgitation; diagnosis of tamponate; severe hypovolemia; complex circulatory situations (e.g., fluid management with acute burn patients); medical emergencies; adult respiratory distress syndrome; gram negative sepsis; drug intoxication; acute renal failure; hemorrhagic pancreatitis; intra- and postoperative management of high risk patients; history of pulmonary or cardiac disease; fluid shifts (such as extensive intra-abdominal operations); management of high risk obstetrical patients; known cardiac disease; toxemia; premature separation of the placenta; cardiac output determination by thermodilution method; and blood sampling.
W.M.P.
(Division Sign-Off) Division of Cardiovascular Devices
510(k) Number_
Page 1 of 2
{7}------------------------------------------------
Additional indication for Hospira Critical Care Pacing Lead Catheters:
Indicated for temporary transluminal ventricular pacing using a temporary ventricular lead.
Prescription Use __ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (Part 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
1.
(Division Sign-Off) Division of Cardiovascular Devices
510(k) Number K081268
{8}------------------------------------------------
K091268 Hospira Critical Care and Advanced Sensor Cathters (29 models)
and the same of the same
.
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
and the same of the same
.
:
.
.
. . . . .
.
:
1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 -
:
. .
:
:
:
| Model # | Model Title |
|-----------|------------------------------------------------------------|
| 412400406 | Pentalumen™ Torque-Line® Catheter with Extra Infusion Port |
| 412370406 | Pentalumen™ Torque-Line® Catheter with Extra Infusion Port |
| 412390406 | Multiflex Torque-Line™ Catheter |
| 412480406 | Torque-Line® Catheter |
| 412500406 | Torque-Line® Catheter |
| 412510406 | Multiflex Torque-Line® S-Tip Catheter |
| 412520406 | Pentalumen™ Torque-Line® Catheter with RV Pacing Port |
| 413000402 | Pentalumen™ Catheter with RV Pacing Port |
| 412160402 | Pentalumen™ Catheter with Extra Infusion Port |
| 412170402 | Pentalumen™ Catheter with Extra Infusion Port |
| 412230402 | Multiflex Cather |
| 412270402 | Multiflex S-Tip Catheter |
| 412290402 | Multiflex Cather |
| 412310406 | Multiflex Torque-Line™ Catheter |
| 412320402 | Pentalumen™ Catheter with Extra Infusion Port |
| 412330406 | Pentalumen™ Catheter with Extra Infusion Port |
| 503240406 | Opticath® PA Catheter |
| 503280406 | Opticath ® PA Catheter with Extra Port |
| 503370406 | Opticath® PA Catheter with Extra Port |
| 503270406 | Opticath ® PA Catheter with Extra Port |
| 503250406 | Opticath ® PA Catheter |
| 503440406 | Opticath® PA Catheter with RV Pacing Port |
| 525090413 | Opti Q® SvO2/CCO Fiberoptic Catheter (with Q-Tip) |
| 525100413 | Opti Q® SvO2/CCO Fiberoptic Catheter (with J-Tip) |
| 525110413 | Opti Q® SvO2/CCO Fiberoptic Catheter (with Q-Tip) |
| 525160413 | TDQ™ CCO Catheter |
| 525150413 | TDQ™ CCO Catheter |
| 503540406 | Opticath ® PA Catheter |
| 503550406 | Opticath ® PA Catheter |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.