Retrospective clinical samples were used to optimize the assay's lot-specific threshold factor, which was subsequently validated by ROC analysis and applied to the clinical trial lots.
The RAMP RSV Assay is a qualitative immunochromatographic test for the detection of Respiratory Syncytial Virus (RSV) F-protein antigens in nasal wash/aspirate, nasopharyngeal aspirate and nasopharyngeal swab samples. It is an in vitro diagnostic assay that aids in the rapid diagnosis of RSV infections in symptomatic patients 21 years of age and younger. A negative test is presumptive and it is recommended that all negative results be confirmed by cell culture or direct specimen fluorescence assay (DSFA). Negative results do not preclude RSV infection and should not be used as the sole basis for treatment or other management decisions. The test is intended for professional use.
Device Story
RAMP RSV Assay is a qualitative immunochromatographic test for detecting RSV F-protein antigens. Input: nasal wash/aspirate, nasopharyngeal aspirate, or nasopharyngeal swab samples mixed with buffer. Operation: sample migrates along a test strip; fluorescent-dyed latex particles coated with anti-RSV antibodies bind to viral antigens; RSV-bound particles are captured at a detection zone. Output: qualitative result based on numerical RAMP Ratio values. Used in central labs, stat-labs, and point-of-care settings by professional operators. Results aid rapid diagnosis of RSV infections; negative results require confirmation by cell culture or DSFA. Benefits include faster turnaround compared to traditional culture methods.
Clinical Evidence
Prospective multi-center study (n=1140) compared RAMP RSV Assay to cell culture and DSFA. Overall sensitivity 87.3%, specificity 95.6%. Performance varied by sample type: NP Swab (Sens 88.2%, Spec 97.4%), NP Aspirate (Sens 85.3%, Spec 95.1%), and Nasal Wash/Aspirate (Sens 89.9%, Spec 93.4%). Additional PCR analysis of discordant samples suggested some RAMP false positives were PCR positive.
Technological Characteristics
Immunochromatographic assay using fluorescent-dyed latex particles coated with anti-RSV antibodies. Qualitative detection based on numerical RAMP Ratio values. Compatible with various transport media and swab materials (foam, polyester, rayon, nylon).
Indications for Use
Indicated for symptomatic patients 21 years of age and younger for the rapid diagnosis of RSV infections using nasal wash/aspirate, nasopharyngeal aspirate, or nasopharyngeal swab samples. For professional, prescription use only.
Regulatory Classification
Identification
Respiratory syncytial virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to respiratory syncytial virus in serum. Additionally, some of these reagents consist of respiratory syncytial virus antisera conjugated with a fluorescent dye (immunofluorescent reagents) and used to identify respiratory syncytial viruses from clinical specimens or from tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of respiratory syncytial virus infections and provides epidemiological information on diseases caused by these viruses. Respiratory syncytial viruses cause a number of respiratory tract infections, including the common cold, pharyngitis, and infantile bronchopneumonia.
Predicate Devices
Quidel QuickVue RSV Assay
Submission Summary (Full Text)
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
K091235
B. Purpose for Submission:
New device.
C. Measurand:
Respiratory Syncytial Virus (RSV) F-protein antigen
D. Type of Test:
Qualitative immunochromatographic test
E. Applicant:
Response Biomedical Corporation
F. Proprietary and Established Names:
RAMP® RSV Assay
G. Regulatory Information:
1. Regulation section:
21 CFR 866.3480 – Respiratory Syncytial Virus Serological Reagents
2. Classification:
Class I
3. Product code:
GQG
4. Panel:
Microbiology (83)
H. Intended Use:
{2}
1. Intended use(s):
The RAMP RSV Assay is a qualitative immunochromatographic test for the detection of Respiratory Syncytial Virus (RSV) F-protein antigens in nasal wash/aspirate, nasopharyngeal aspirate and nasopharyngeal swab samples. It is an *in vitro* diagnostic assay that aids in the rapid diagnosis of RSV infections in symptomatic patients 21 years of age and younger. A negative test is presumptive and it is recommended that all negative results be confirmed by cell culture or direct specimen fluorescence assay (DSFA). Negative results do not preclude RSV infection and should not be used as the sole basis for treatment or other management decisions. The test is intended for professional use.
2. Indication(s) for use:
Same as Intended Use
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
To be used only with the RAMP® 200 Reader
I. Device Description:
The RAMP RSV Assay consists of the following components:
RAMP RSV test cartridges, RAMP RSV Assay tips, RAMP RSV sample buffer vials, disposable droppers, transfer device, RSV positive control swab, RSV negative control swab, lot card, package insert, and procedure card.
The device requires the following components which are not provided as part of the assay:
RAMP 200 reader, Transport Media and Specimen collection materials
The RAMP Assay Tip contains test-specific reagents. The supplied fixed volume transfer device and the RAMP Assay Tip are used to mix the test-specific reagents with the respiratory sample. Using the same transfer device and tip, an aliquot of the prepared sample is introduced to the sample well of the RAMP Test Cartridge. The RAMP Test Cartridge is a single-use; disposable, plastic housing that encloses an analyte-specific immunochromatographic assay strip. Sample is metered onto the assay strip by a pad at the bottom of the sample well. The assay strip is a Mylar-backed nitrocellulose membrane, near the terminal end of which are two closely spaced capture antibody zones, one of which is analyte-specific and the other of
{3}
which is an internal standard. An absorbent pad at the terminal end of the assay strip soaks up excess sample. A section of the bottom of the Test Cartridge is open so the Reader can scan the test strip

## J. Substantial Equivalence Information:
1. Predicate device name(s):
Quidel QuickVue RSV Assay
2. Predicate K number(s):
K070747
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | The RAMP RSV Assay is a qualitative immunochromatographic test for the detection of Respiratory Syncytial Virus (RSV) F-protein antigens in nasal wash/aspirate, nasopharyngeal aspirate and nasopharyngeal swab samples. It is an in vitro diagnostic assay that aids in the rapid diagnosis of RSV infections in symptomatic patients 21 years of age and younger. A negative test is presumptive and it is recommended that all negative results be confirmed by cell culture or direct specimen fluorescence assay (DSFA). Negative results do not preclude RSV infection and should not be used as the sole basis for treatment or other management decisions. The test is intended for professional use. | The QuickVue RSV test is a dipstick immunoassay, which allows for the rapid, qualitative detection of Respiratory Syncytial virus (RSV) antigen (viral fusion protein) directly from nasopharyngeal swab, nasopharyngeal aspirate, or nasal/nasopharyngeal wash specimens for symptomatic pediatric patients (Eighteen years of age and younger). The test is intended for use as an aid in the diagnosis of acute Respiratory Syncytial viral infections. It is recommended that negative test results be confirmed by cell culture. Negative results do not preclude RSV infection and it is recommended that they not be used as the sole basis for treatment or other management decisions. The test is intended for professional and laboratory use. |
{4}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Specimen Type | Nasal wash, nasopharyngeal aspirate, and nasopharyngeal swab specimens | Nasal/ Nasopharyngeal wash, nasopharyngeal aspirate, and nasopharyngeal swab specimens |
| Target Antigen | RSV-F antigen | RSV-F antigen |
| Antibodies | Two mouse monoclonal antibodies recognizing the F- protein of RSV | Mouse monoclonal antibodies recognizing the F- protein |
| Test Time | 15 minutes | 15 minutes |
| Reagent Stability | In sealed pouch: Up to the stated expiration date stored at 15-30°C. Do not remove from pouch until ready to use. Do not freeze. | Store the kit at room temperature, 15-30°C, out of direct sunlight. Kit contents are stable until the expiration date printed on the outer box. Do not freeze |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Instrument | RAMP 200 Reader | None |
| Automated Processing | Instrument transport of test cartridge within reader is the only moving step. No internal liquid handling. | Visual results appear on the test strip, no automation |
| Automated Analysis | Yes | No |
| Quality Controls | Provided in every Test Cartridge. Built in performance controls for routine QC requirements. A comparison of the internal standard and the assay result indicates that sufficient properly mixed sample was applied to the test device and that unbound fluorescent label washed sufficiently from the detection zone, and the device was inserted and read properly by the instrument. Background fluorescence measurement serves as a negative control. Reader controls also prevent a used or expired cartridge from being read by the reader. Antibody quality, system function and assay timing are checked on each assay run. An unacceptable result from the control displays a warning message without a test result on the instrument indicating that the test should be repeated. | Provided on every test strip. The two-color result format provides a simple interpretation for positive and negative results. The appearance of a blue procedural Control Line provides several forms of positive control by demonstrating sufficient flow has occurred and the functional integrity of the Test Strip was maintained. If the blue procedural Control Line does not develop within 15 minutes, the test result is considered invalid. A built-in negative control is provided by the clearing of red background color, verifying that the test has been performed correctly. Within 15 minutes, the result area should be white to light pink and allow the clear interpretation of the test result. If background color remains and interferes with interpretation of the test result, the result is considered invalid |
| System Quality Controls | A built-in Internal Quality Control (IQC) function performs an automated test of all Reader analytical systems on a scheduled or operator determined basis to provide a record of proper Reader functionality. For complete quality control of the system refer to the Operator's Manual | NA |
| Calibrator | Lot card contains calibration and expiration information and is supplied with every kit. System will not process | NA |
{5}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | cartridge if lot card information is not stored in instrument. | |
| Method Comparison | vs. Viral culture and DSFA | vs. Viral Culture |
## K. Standard/Guidance Document referenced (if applicable):
Not Applicable
## L. Test Principle:
The RAMP® RSV Assay is a qualitative immunochromatographic test that utilizes the RAMP® 200 Reader for the identification of the presence of RSV F-protein antigens in nasal wash/aspirate, nasopharyngeal aspirate (NP aspirate) and nasopharyngeal swab (NP swab) samples.
A wash/aspirate or swab sample is added to the Sample Buffer. This sample is then mixed with the Assay Tip containing fluorescent-dyed particles conjugated to RSV specific antibodies and applied into the sample well of the Test Cartridge. The sample migrates along the strip.
Fluorescent-dyed particles coated with anti-RSV antibodies bind to RSV F-protein antigens, if present in the sample. As the sample migrates along the strip, RSV-bound particles are captured at the detection zone which is coated with monoclonal antibodies anti RSV-F. Excess fluorescent-dyed particles are captured at the internal standard zone which is coated with goat anti-mouse IgG.
The Reader then measures the amount of fluorescence emitted by the complexes at the detection zone and at the internal standard zone. The instrument calculates a ratio (RAMP® Ratio) of the detection zone fluorescence reading to the internal standard zone fluorescence reading. The Reader then compares these ratios to pre-defined threshold limits to determine a positive or negative result for RSV in the tested sample. The RAMP® 200 is a general use fluorometer that analyzes results produced by immunoassays that use a fluorophore having an excitation wavelength of 560 nm and an emission wavelength of 610 nm.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
The precision and reproducibility of the RAMP RSV Assay was evaluated using a panel consisting of a high negative RSV sample, a limit of detection (LoD) RSV sample (low positive), and a 2x LoD RSV sample (moderate positive). The RSV strain used to prepare the samples was RSV A (A-2)
{6}
(ATCC VR-1540) following viral titer determinations. To evaluate reproducibility, multiple operators at multiple sites tested each of the three precision samples on 5 different days. Testing at Site 2 was performed in a point-of-care (POC) setting (pediatric clinic, under CLIA umbrella of the hospital). At one site (Site 3), to evaluate precision, operators tested each of the three precision samples an additional 7 days for a total of 12 days. There was 99.2% agreement (393/396) with the expected test results for all specimens tested, with no significant differences within run (same operator on same day) between run, operators or sites. The RAMP RSV Assay is a qualitative assay based on numerical RAMP Ratio values.
| | Panel Member ID | RSV High Negative | RSV Low Positive | RSV Moderate Positive | Total Agreement All (%) |
| --- | --- | --- | --- | --- | --- |
| | Viral Titer (TCID50/mL) | 200 | 1200 | 2400 | |
| Site 1 | Agreement | 30/30 | 30/30 | 30/30 | 90/90 (100%) |
| | Mean RAMP Ratio* Value | 1176 | 3226 | 5460 | |
| | % CV | 11% | 11% | 9% | |
| Site 2 | Agreement | 30/30 | 29/30 | 30/30 | 89/90 (98.9%) |
| | Mean RAMP Ratio Value | 1209 | 3273 | 5600 | |
| | % CV | 13% | 14% | 18% | |
| Site 3 | Agreement | 71/72 | 71/72 | 72/72 | 214/216 (99.1%) |
| | Mean RAMP Ratio Value | 1330 | 3615 | 6193 | |
| | % CV | 17% | 12% | 11% | |
| | Total Agreement | 131/132 (99.2%) | 130/132 (98.5%) | 132/132 (100%) | 393/396 (99.2%) |
| | 95% CI | 95.8 – 99.9% | 94.6 – 99.6% | 97.1 – 100% | 97.8 – 99.7% |
| | Overall Mean RAMP Ratio | 1268 | 3449 | 5892 | |
| | Overall % CV | 16% | 13% | 14% | |
b. Linearity/assay reportable range:
Not Applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Not Applicable
d. Detection limit:
{7}
The RAMP RSV Assay was evaluated for analytical sensitivity and reactivity after viral titer was determined by testing 6 strains of RSV (RSV A-Long, RSV A (A-2), RSV B CH93 (18)-18, RSV B Wash/18537/62, RSV B WV/14617/85, RSV B9320) at the LoD concentration in either viral transport media (VTM) to simulate a swab sample type or saline solution to simulate a wash sample type. Although the specific RSV strains causing infection in humans can vary year to year, all contain the conserved membrane-bound F-protein antigens targeted by the RAMP RSV Assay. Analytical sensitivity (LOD) ranged from $3.5 \times 10^{2}$ to $>1.7 \times 10^{5} \mathrm{TCID}_{50} / \mathrm{mL}$ .
| RSV Sample | LoD Concentration (TCID50/mL) | RAMP Result |
| --- | --- | --- |
| A-Long in Saline | 6.5x103 | 100% RSV Positive |
| A-Long in VTM | 2.5x103 | 100% RSV Positive |
| A (A-2) in Saline | 1.0x103 | 100% RSV Positive |
| A (A-2) in VTM | 1.2x103 | 100% RSV Positive |
| B CH93(18)-18 in Saline | 3.5x102 | 100% RSV Positive |
| B CH93(18)-18 in VTM | 3.5x102 | 100% RSV Positive |
| B Wash/18537/62 in Saline | 7.0x103 | 95% RSV Positive |
| B Wash/18537/62 in VTM | 6.0x103 | 100% RSV Positive |
| B WV/14617/85 in Saline | 5.0x103 | 100% RSV Positive |
| B WV/14617/85 in VTM | 2.5x103 | 100% RSV Positive |
| B9320 in ZMC matrix | >1.7x105 | 90% RSV Positive |
# e. Analytical specificity:
# Cross Reactivity:
The analytical specificity of the RAMP RSV Assay was determined by testing a panel consisting of 16 viruses and 17 bacteria that may be present in the nasal cavity or nasopharynx. Bacterial and viral isolates were tested at the concentrations listed after titer determination. None of the organisms tested gave a positive result in the RAMP RSV Assay. Note: RAMP RSV Assay potential cross-reactivity with Chlamydophila pneumoniae has not been determined.
| Strain/Isolate | Concentration | RSV result |
| --- | --- | --- |
| Adenovirus, Type 1 | 103TCID50/mL | Negative |
| Adenovirus, Type 7a | 103TCID50/mL | Negative |
| Human coronavirus, Strain OC43 | 103TCID50/mL | Negative |
| Human coronavirus, Strain 229E | 103TCID50/mL | Negative |
{8}
| Strain/Isolate | Concentration | RSV result |
| --- | --- | --- |
| Cytomegalovirus | 10^5 TCID50/mL | Negative |
| Enterovirus, Type 71 | 10^5 TCID50/mL | Negative |
| Epstein Barr Virus | 10^5 TCID50/mL | Negative |
| Human Parainfluenza, Type 1 | 10^5 TCID50/mL | Negative |
| Human Parainfluenza, Type 2 | 10^5 TCID50/mL | Negative |
| Human Parainfluenza, Type 3 | 10^5 TCID50/mL | Negative |
| Influenza A, Brisbane/10/07 | 10^5 EID50/mL | Negative |
| Influenza B, Ohio/01/06 | 10^5 TCID50/mL | Negative |
| Measles | 10^5 TCID50/mL | Negative |
| Human metapneumovirus | 10^5 TCID50/mL | Negative |
| Mumps virus | 10^5 TCID50/mL | Negative |
| Human Rhinovirus, Strain 1A | 10^5 TCID50/mL | Negative |
| Bordetella pertussis | 10^6 cfu/mL | Negative |
| Corynebacterium Sp. | 10^6 cfu/mL | Negative |
| Escherichia coli | 10^6 cfu/mL | Negative |
| Haemophilus influenzae | 10^6 cfu/mL | Negative |
| Lactobacillus casei | 10^6 cfu/mL | Negative |
| Legionella pneumophila | 10^6 cfu/mL | Negative |
| Moraxella catarrhalis | 10^6 cfu/mL | Negative |
| Mycobacterium tuberculosis, Avirulent | 10^6 cfu/mL | Negative |
| Mycoplasma pneumoniae | 10^6 cfu/mL | Negative |
| Neisseria meningitidis | 10^6 cfu/mL | Negative |
| Neisseria sicca | 10^6 cfu/mL | Negative |
| Pseudomonas aeruginosa | 10^6 cfu/mL | Negative |
| Staphylococcus aureus, Strain 176 | 10^6 cfu/mL | Negative |
| Staphylococcus epidermidis, Strain 78 | 10^6 cfu/mL | Negative |
| Streptococcus pneumoniae | 10^6 cfu/mL | Negative |
| Streptococcus pyogenes | 10^6 cfu/mL | Negative |
| Streptococcus salivarius | 10^6 cfu/mL | Negative |
# Interfering Substances:
Whole blood and a number of other potentially interfering substances (medications and over the counter (OTC) products) that may be present naturally or artificially introduced in the nasal cavity or nasopharynx were evaluated in the RAMP RSV Assay. The substances were added to a simulated negative sample (0.85% saline), an RSV LoD positive sample and an RSV 2x LoD positive sample and tested in the RAMP RSV Assay. The RSV strain used to prepare the samples was RSV A (A-2), interference was not evaluated with an RSV B strain. The following substances were found to have no effect on the test results of the RAMP RSV Assay when present in positive and negative RSV simulated respiratory samples at the concentrations indicated:
{9}
| Substance Tested | Conc. Tested | Negative | | RSV LoD | | RSV 2x LoD | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | RAMP Results | Percent of Mean Control Ratio | RAMP Results | Percent of Mean Control Ratio | RAMP Results | Percent of Mean Control Ratio |
| Control Saline (0.85%) only (No interfering substance) | N/A | Neg 3/3 | 100% | Pos 3/3 | 100% | Pos 3/3 | 100% |
| Whole Blood | 2% v/v | Neg 3/3 | 118% | Pos 2/3 | 89% | Pos 3/3 | 86% |
| Mucin | 1% w/v | Neg 3/3 | 123% | Pos 3/3 | 96% | Pos 3/3 | 94% |
| Scope® Mouthwash | 40% v/v | Neg 3/3 | 117% | Pos 2/3 | 91% | Pos 3/3 | 85% |
| Good and Kind™ Mouthwash | 40% v/v | Neg 3/3 | 122% | Pos 3/3 | 93% | Pos 3/3 | 83% |
| Cepacol® Mouth Wash | 40% v/v | Neg 3/3 | 128% | Pos 2/3 | 90% | Pos 3/3 | 78% |
| Cepacol® Lozenge (Benzocaine) | 30% w/v | Neg 3/3 | 115% | Pos 3/3 | 93% | Pos 3/3 | 69% |
| Fisherman’s Friend® Throat Drop (Menthol) | 30% w/v | Neg 3/3 | 107% | Pos 3/3 | 94% | Pos 3/3 | 80% |
| Rhinocort® Nasal Spray (Budesonide) | 15% v/v | Neg 3/3 | 125% | Pos 3/3 | 102% | Pos 3/3 | 76% |
| Nasacort® Nasal Spray (Triamcinolone) | 15% v/v | Neg 3/3 | 141% | Pos 3/3 | 98% | Pos 3/3 | 85% |
| Flonase® Nasal Spray (Fluticasone Furoate) | 30% v/v | Neg 3/3 | 145% | Pos 3/3 | 105% | Pos 3/3 | 85% |
| Nasonex® Nasal Spray (Mometasone Furoate) | 15% v/v | Neg 3/3 | 134% | Pos 3/3 | 94% | Pos 3/3 | 79% |
| Tylenol® Tablets (4-Acetamidophenol) | 20 mg/mL | Neg 3/3 | 96% | Pos 3/3 | 90% | Pos 3/3 | 81% |
| Aspirin® Tablets (Acetylsalicylic Acid) | 30 mg/mL | Neg 3/3 | 127% | Pos 3/3 | 98% | Pos 3/3 | 80% |
| Chlor-Tripolon™ Tablets (Chloropheniramine) | 10 mg/mL | Neg 3/3 | 106% | Pos 3/3 | 110% | Pos 3/3 | 98% |
| Benadryl™ Allergy Tablet (Diphenhydramine) | 5 mg/mL | Neg 3/3 | 136% | Pos 3/3 | 107% | Pos 3/3 | 91% |
| Delsym® DM Cough Syrup (Dextromethorphan) | 2 mg/mL | Neg 3/3 | 135% | Pos 3/3 | 121% | Pos 3/3 | 105% |
| Robitussin® Syrup (Guaiacol Glycerol Ether) | 40 mg/mL | Neg 3/3 | 140% | Pos 3/3 | 89% | Pos 3/3 | 82% |
| Phenylpropanol-amine HCl (pure) | 40 mg/mL | Neg 3/3 | 142% | Pos 3/3 | 105% | Pos 3/3 | 96% |
{10}
| Substance Tested | Conc. Tested | Negative | | RSV LoD | | RSV 2x LoD | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | RAMP Results | Percent of Mean Control Ratio | RAMP Results | Percent of Mean Control Ratio | RAMP Results | Percent of Mean Control Ratio |
| Afrin® Nasal Spray (Oxymetazoline HCl) | 0.05% v/v | Neg 3/3 | 103% | Pos 3/3 | 86% | Pos 3/3 | 76% |
| Neo-Synephrine® Nasal Spray (Phenylephrine HCl) | 20 mg/mL | Neg 3/3 | 121% | Pos 3/3 | 87% | Pos 3/3 | 84% |
| Otrivin® Saline (NaCl w/preservatives) | 1.4% w/v | Neg 3/3 | 122% | Pos 3/3 | 91% | Pos 3/3 | 94% |
| Rebetol® (Ribavirin) | 100 ng/mL | Neg 3/3 | 116% | Pos 2/3 | 92% | Pos 3/3 | 82% |
| Relenza® (Zanamivir) | 20 mg/mL | Neg 3/3 | 105% | Pos 3/3 | 89% | Pos 3/3 | 76% |
| Rimantadine HCl | 500 ng/mL | Neg 3/3 | 118% | Pos 3/3 | 87% | Pos 3/3 | 85% |
| Tamiflu® (Oseltamivir Phosphate) | 50 mg/mL | Neg 3/3 | 146% | Pos 3/3 | 127% | Pos 3/3 | 108% |
f. Assay cut-off:
The kit lot threshold is determined by applying a standard factor to the difference in the mean RAMP Ratios between the negative and the QC standard. This estimated range of the factor was determined though pre-clinical testing with retrospective positive and negative clinical samples. The value of the factor was found to be optimized at 0.7 and validated by Receiver Operator Curve (ROC) analysis. This lot-specific threshold is then programmed into the lot cards and must be uploaded to the reader when a new kit lot is opened in order to run a test. This factor was then applied to the clinical trial lots to determine the clinical results and will be the standard factor used to generate the lot-specific Ratio Threshold for all manufacturing lots.
2. Comparison studies:
a. Method comparison with reference method:
Not Applicable
b. Matrix comparison:
Sample Collection Swab Compatibility:
Four swab materials (see below) were evaluated for compatibility in the RAMP RSV Assay by testing a negative sample (swab alone with no virus
{11}
present), an RSV LoD positive sample and an RSV 2x LoD positive sample in the RSV test. The RSV strain used to prepare the samples was RSV A (A-2) following titer determination. Each swab was dosed with the appropriate sample and extracted into Copan Universal Transport Media prior to testing in the RAMP RSV Assay. The RAMP RSV Assay is a qualitative assay based on numerical RAMP Ratio values. The $\% \mathrm{CV}$ s were calculated for the RAMP Ratios. The negative sample (swab alone) tested negative and all 2x LoD samples tested positive. None of the swabs tested interfered with the performance of the RAMP RSV Assay. Note: In general, calcium alginate swabs are not recommended because they may be cytotoxic to cells and cause viral culture assay inhibition.
| Swab Material Tested | RAMP Results | | |
| --- | --- | --- | --- |
| | Negative | RSV LoD | RSV 2x LoD |
| Foam | 100% Neg | 100% Pos | 100% Pos |
| Polyester | 100% Neg | 67% Pos | 100% Pos |
| Rayon | 100% Neg | 67% Pos | 100% Pos |
| Nylon | 100% Neg | 100% Pos | 100% Pos |
| Average RAMP Ratio % CV | 9% | 21% | 21% |
# Transport Media Compatibility :
Multiple lots of each of seven commercially available transport media (see below) were evaluated for compatibility in the RAMP RSV Assay by testing a negative sample (transport media only), an RSV LoD positive sample and an RSV 2x LoD positive sample. The RSV strain used to prepare the samples was RSV A (A-2) following titer determination. The RAMP RSV Assay is a qualitative assay based on numerical RAMP Ratio values. The $\% \mathrm{CV}$ s were calculated for the RAMP Ratios. None of the tested transport media interfered with the performance of the RAMP RSV Assay.
| Transport Media Tested | RAMP Results | | |
| --- | --- | --- | --- |
| | Negative | RSV LoD | RSV 2x LoD |
| Copan: Universal Transport Media (UTM) | 100% Neg | 100% Pos | 100% Pos |
| Remel: M4 Media | 100% Neg | 100% Pos | 100% Pos |
| RAMP: M4 Media | 100% Neg | 100% Pos | 100% Pos |
{12}
| Transport Media Tested | RAMP Results | | |
| --- | --- | --- | --- |
| | Negative | RSV LoD | RSV 2x LoD |
| Remel: M4-RT Media | 100% Neg | 100% Pos | 100% Pos |
| Remel: M5 Media | 100% Neg | 100% Pos | 100% Pos |
| Starplex Transport Media | 100% Neg | 100% Pos | 100% Pos |
| 0.85% Saline Solution | 100% Neg | 93% Pos | 100% Pos |
| Phosphate Buffered Saline (PBS) Solution | 100% Neg | 93% Pos | 100% Pos |
| Average RAMP Ratio: % CV | 12% | 16% | 14% |
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable
b. Clinical specificity:
Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Prospective Clinical Studies:
The performance of the RAMP RSV Assay was compared to cell culture and DSFA in a prospective study conducted as part of a multi-center trial during the 2008-2009 RSV season. Eight (8) independent laboratories located in distinct geographical regions across the United States (NY, MO(2), MD, OH(2), MA, AR) evaluated the RAMP RSV Assay in parallel with cell culture / DSFA. Testing staff included both laboratory and non-laboratory personnel, and two sites (MO, AR) also performed testing in point-of-care settings (near patient). The sites included an Emergency Department and Pediatric Testing Unit (both under CLIA umbrella of the individual hospital). One thousand, two hundred and seventy nine (1279) fresh specimens were collected from pediatric patients 21 years of age and younger. Of these specimens, valid results were obtained from 1140, with an approximately equal mix of male and female patients. During the clinical study, if either the cell culture or DSFA result was positive for a patient sample, it was determined to be a positive reference result. A reference result was deemed negative only if both, culture and DSFA, were negative
{13}
Results with Prospective Nasopharyngeal Swab Samples
| NP Swab | | | |
| --- | --- | --- | --- |
| Culture / DSFA | RAMP RSV | | |
| | Positive | Negative | Total |
| Positive | 112 | 15 | 127 |
| Negative | 7* | 257 | 264 |
| Total | 119 | 272 | 391 |
| | | | 95% CI |
| Sensitivity | 88.2% | | 81.4 – 92.7 |
| Specificity | | 97.4% | 94.6 – 98.7 |
* 1 was positive by the RAMP RSV Assay and by the Prodesse ProFlu RT-PCR analysis.
Results with Prospective Nasopharyngeal Aspirate Samples
| NP Aspirate | | | |
| --- | --- | --- | --- |
| Culture / DSFA | RAMP RSV | | |
| | Positive | Negative | Total |
| Positive | 168 | 29* | 197 |
| Negative | 15** | 291 | 306 |
| Total | 183 | 320 | 503 |
| | | | 95% CI |
| Sensitivity | 85.3% | | 79.7 – 89.6 |
| Specificity | | 95.1% | 92.1 – 97.0 |
* 8 were negative by the RAMP RSV Assay and by an investigational RT-PCR.
** 3 were positive by the RAMP RSV Assay and by an investigational RT-PCR.
Results with Prospective Nasal Wash / Aspirate Samples
| Nasal Wash / Aspirate | | | |
| --- | --- | --- | --- |
| Culture / DSFA | RAMP RSV | | |
| | Positive | Negative | Total |
| Positive | 98 | 11* | 109 |
| Negative | 9** | 128 | 137 |
| Total | 107 | 139 | 246 |
| | | | 95% CI |
| Sensitivity | 89.9% | | 82.8 – 94.3 |
| Specificity | | 93.4% | 88.0 – 96.5 |
* 1 was negative by the RAMP RSV Assay and by an investigational RT-PCR.
** 1 was positive by the RAMP RSV Assay and by an investigational RT-PCR
4. Clinical cut-off:
Not Applicable
5. Expected values/Reference range:
{14}
The prevalence of RSV varies each year with epidemics occurring during the fall and winter months in the US. Variables that affect the rate of positivity observed in respiratory testing include: the efficiency and timing of specimen collection, handling and transport of the specimen, the time of year, age of the patient, and local disease prevalence.
The prevalence observed with culture and DSFA during the clinical study (November 2008 – February 2009) was 33.9% (443/1306). During the previous U.S. respiratory season (October 2007 – March 2008), the prevalence of RSV was 15.9%8.
## N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
## O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
14
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.