K091220 · Healthanywhere, an Igeacare System Company · DRG · Sep 21, 2009 · Cardiovascular
Device Facts
Record ID
K091220
Device Name
HEALTHANYWHERE SYSTEM
Applicant
Healthanywhere, an Igeacare System Company
Product Code
DRG · Cardiovascular
Decision Date
Sep 21, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2910
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The intended use of the HA System is for the collection, transmission, and persistence of medical measurement data and patient care data for the sole purpose of encouraging and tracking a patient's progress against the goals set between the Healthcare Professional and the patient. The goals would be established independent of the HA System in the pursuit of providing care for a chronic or post acute condition and/or to promote patient wellness. The HA System would enable Healthcare Professionals to make use of the persisted information as a guide in delivering ongoing feedback to their patients, to set periodic reminders for patients to act on scheduled activities such as taking a vital measurement, and to receive notifications for events they configure such as an out of range vital measurement. The HA System can record physiological information such as: Blood pressure measurements . Pulse rate measurements Body weight measurements . Blood glucose levels Body temperature measurements . Blood oxygen saturation (%SpO2) levels The HA System can be used by Healthcare Professionals to create health related questionnaires, exercise plans and nutrition plans; to forward these items to the patients; and to receive/review patient responses to the related questionnaires, exercise activities, and nutrition activities. The HA System is not intended for diagnosis or as a substitute for medical care; it is also not intended to provide real-time data. The data is not used for time-critical care. The device is contraindicated for patients requiring direct medical supervision or emergency intervention.
Device Story
System collects/transmits/persists physiological data and patient care updates; inputs include blood pressure, pulse, weight, glucose, temperature, SpO2 from off-the-shelf Bluetooth peripherals; also supports manual entry and patient questionnaires/plans. Client app runs on BlackBerry/PC; transmits data via cellular/WiFi/wired internet to server. Healthcare Professional Portal (HCPP) web app allows clinicians to configure patient goals, reminders, and notifications; review patient responses/data. Used in home/clinic settings; operated by patients and clinicians. Output provides clinicians with longitudinal data to guide feedback and track progress. Not for real-time or time-critical care.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Client software on Windows/BlackBerry; server on Linux. Connectivity via Bluetooth (peripherals), HTTPS/SSL (internet). Supports AND UA-767PBT, AND UC-321PBT, Nonin 9560, Polymap PWR-08-03. Role-based access control.
Indications for Use
Indicated for patients with chronic or post-acute conditions or those seeking wellness promotion. Contraindicated for patients requiring direct medical supervision or emergency intervention.
Regulatory Classification
Identification
A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.
{1}------------------------------------------------
### ··· Healthanywhere
510(k) Summary
Section 5
SEP 2 1 2009
| Submitter: | Healthanywhere, Inc. |
|-------------------------------------|-------------------------------------------------------------|
| Address: | 515 Legget Drive Suite 700 |
| | Ottawa, ON K2K 3G4 |
| | Canada |
| Phone number: | 613-271-0283 |
| Fax number: | 613-271-9831 |
| Contact person: | Shawn Slade, Director of Product Development and Operations |
| Date prepared: | 8/28/2009 |
| Trade name: | Healthanywhere System |
| Device classification: | 870.2910, DRG, Physiological Transmitter and Receiver |
| Substantial equivalence claimed to: | The Hermes System, Palaistra Systems Inc, (K050929) |
K091220
#### Device Description:
The Healthanywhere (HA) System is a user-friendly software system which provides the ability to remotely monitor basic wellness parameters. The HA System is comprised of an HA Client Application, anHA Server and the Healthcare Professional Portal (HCPP).
The HA Client Application provides capabilities to perform data collection, data transmission, message display, reminders/notifications, and reporting. It is used to collect data including vital measurements from noninvasive, commercially available, off-the-shelf, wireless medical peripherals, and patient care updates. The user data is transmitted though the HA Client Application to the HA Server using off-the-shelf. FCCapproved wireless/cellular/wired connectivity (including but not limited to GPRS, CDMA, WiFi, or Wired Ethernet) which is provided through the FCC-approved host hardware platform or through an external FCC approved wireless adapter.
The HA Server hosts the Healthcare Professional Portal (HCPP), a web application which is used to administer the HA System and users. The HCPP application provides user registration, care management, us configuration set-up, and data management. Based on specific parameters set by Healthcare Professionals, reminders and messages are scheduled and sent from the HA Server to the HA Client Applications.
#### Intended use:
The intended use of the HA System is for the collection, transmission, and persistence of medical measurement data and patient care data for the sole purpose of encouraging and tracking a patient's progress against the goals set between the Healthcare Professional and the patient. The goals would be established independent of the HA System in the pursuit of providing care for a chronic or post acute condition and/or to promote patient wellness.
The HA System would enable Healthcare Professionals to make use of the persisted information as a guide in delivering ongoing feedback to their patients, to set periodic reminders for patients to act on scheduled activities such as taking a vital measurement, and to receive notifications for events they configure such as ar out of range vital measurement.
{2}------------------------------------------------
# *** Healthanywhere
Image /page/1/Picture/1 description: The image shows the text "Page" with the number "2" above and to the right of it. The word "of" is to the right of "Page", and the number "3" is above and to the right of "of". This appears to be indicating page 2 of 3.
The HA System can record physiological information such as:
- Blood pressure measurements .
- Pulse rate measurements
- Body weight measurements .
- . Blood glucose levels
- Body temperature measurements .
- � Blood oxygen saturation (%SpO2) levels
The HA System can be used by Healthcare Professionals to create health related questionnaires, exercise plans and nutrition plans; to forward these items to the patients; and to receive/review patient responses to the related questionnaires, exercise activities, and nutrition activities.
The HA System is not intended for diagnosis or as a substitute for medical care; it is alsonot intended to provide real-time data. The data is not used for time-critical care. The device is contraindicated for patients requiring direct medical supervision or emergency intervention.
#### Summary of technological characteristics:
There are two distinct subsystems in the HA System – the client software runs on a Microsoft Windows and BlackBerry platforms; and the server components run on a LINUX platform hosted in a secure data centre.
The HA Client Application is distributed on the following hardware platforms: BlackBerry (models: 8100, 8120, 8130, 8300, 8320, 8330, 8820, 8830); PC (models: Samsung NP-Q1U Tablet PC, Dell Inspirion Mini 10 Notebook, Dell Latitude XT Tablet PC, Dell Optiplex SX280, ELO TouchSystems 17A2 Touchcomputer
The HA Client Application is a standalone software application which collects information from off-the-shel Bluetooth-enabled medical peripherals, such as a Blood Pressure monitor, using a Bluetooth wireless connection. To transmit data from the medical peripherals to the Healthanywhere Client Application, the HA Client Application makes use of the enibedded Bluetooth software on all of the hardware platforms listed above except for the Dell Optiplex SX280 and ELO TouchSystems 17A2 which require an external D-Link DBT-122 Bluetooth Adaptor. The following wireless, commercially-available medical peripherals are supported by the HA Client Application:
- AND UA-767PBT Blood Pressure Monitor .
- . AND UC-321PBT Weight Scale
- Nonin 9560 Pulse Oximeter .
- Polymap PWR-08-03 Bluetooth Adaptor ●
Note: The application supports manually entered thermometer measurements; the choice of a thermomet is based on user preference and is independent from the HA Client Application
The HA Client Application uses the internet to route information to the HA Server via HTTPS or SSL over HTTP protocols and maintains the data stack until an acknowledgement is transmitted from the server indicating that the information is successfully persisted in the server-side database. Periodically, the client application connects to the server to resynchronize with the updates set by the Healthcare Professional including patient care settings, reminders, and notifications via a secure connection over HTTPS. All platforms noted above connect to the internet via a customer preferred option and carrier or internet service provider using off-the-shelf, FCC approved wireless/cellular/wired connectivity (including but not limited to
{3}------------------------------------------------
### · Healthanywhere
GPRS, CDMA, WiFi, or Wired Ethernet) which is provided through the FCC-approved host hardware platform or through an external FCC approved wireless adapter.
The HCPP web application runs in any standard web browser. HCPP enables Healthcare Professionals to perform program administration and vital signs monitoring. HCPP uses role-based controls and patient group to limit access to the patient care data; only an authorized Healthcare Professional can access patient data within a specific group of patients. HCPP users can configure patient care based on individual patient assessments.
#### Substantial equivalence:
The Healthanywhere System is substantially equivalent to The Hermes System (K050929) in that it allows patients to record data from equivalent devices using equivalent methods.
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized depiction of an eagle with three waves emanating from its body. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
### SEP & 1 2009
Healthanywhere, Inc. c/o Mr. Shawn Slade Director Product Development and Operations 515 Legget Drive, Suite 700 Ottawa, Ontario CANADA K2K 3G4
Re: K091220
Trade/Device Name: Healthanywhere (HA) System Regulation Number: 21 CFR 870.2910 Regulation Name: Radiofrequency physiological signal transmitter and receiver Regulatory Class: Class II (two) Product Code: DRG Dated: August 28, 2009 Received: September 11, 2009
Dear Mr. Slade:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
{5}------------------------------------------------
#### Page 2 - Mr. Shawn Slade
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## ealthanywhere
Page 8 of 21
Section 4 Indications for Use Statement
K091220 510(k) Number: >
Device Name: Healthanywhere System
#### Indications for Use:
The intended use of the HA System is for the collection, transmission, and persistence of medical measurement data and patient care data for the sole purpose of encouraging and tracking a patient's progress against the goals set between the Healthcare Professional and the patient. The goals would be established independent of the HA System in the pursuit of providing care for a chronic or post acute condition andor to promote patient wellness.
The HA System would enable Healthcare Professionals to make use of the persisted information as a guide i delivering ongoing feedback to their patients, to set periodic reminders for patients to act on scheduled activities such as taking a vital measurement, and to receive notifications for events they confenue such as ar out of range vital measurement.
The HA System can record physiological information such as:
- Blood pressure measurements
- Pulse rate measurements
- Body weight measurements
- Blood glucose levels
- Body temperature measurements
- Blood oxygen saturation (%SpO2) levels
The HA System can be used by Healthcare Professionals to create health related questionnaires, exercise plans and nutrition plans; to forward these items to the patients; and to receive the patient responses for the related questionnaires, exercise activities, and nutrition activities.
The HA System is not intended for diagnosis or as a substitute for medical care; it is also not intended to provide real-time data. The data is not used for time-critical care. The device is contraindicated for patients requiring direct medical supervision or emergency intervention.
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Offi Device Evaluation (ODE)
> (Division Sign-Off Division of Cardiovascular Dev
Healthanywhere System
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.