← Product Code [MQV](/productcode/MQV) · K091218

# DBX DEMINERALIZED BONE MATRIX PUTTY (K091218)

_Musculoskeletal Transplant Foundation · MQV · Sep 28, 2009 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K091218

## Device Facts

- **Applicant:** Musculoskeletal Transplant Foundation
- **Product Code:** [MQV](/productcode/MQV.md)
- **Decision Date:** Sep 28, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3045
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

DBX® Inject is intended for use as a Demineralized Bone Matrix for voids or gaps that are not intrinsic to the stability of the bony structure. It can be used as follows: Extremities (Putty, Paste), Posterolateral spine (Putty), Pelvis (Putty, Paste). DBX® Inject is indicated for treatment of surgically created osseous defects or osseous defects created from traumatic injury. DBX® Inject Putty can be used as an extender in the spine with autograft. DBX® Inject can be used with bone marrow. DBX® Inject is for single patient use only.

## Device Story

DBX® Inject is a resorbable bone void filler composed of donated cortical and cancellous bone granules mixed with sodium hyaluronate. It is supplied as a putty or paste with a sterile delivery syringe. Used by surgeons in clinical settings to fill non-structural osseous defects in extremities, pelvis, or posterolateral spine. The device acts as an osteoconductive scaffold and possesses osteoinductive potential. It is intended to support bone healing in surgically created or traumatic defects. The device is single-donor processed and screened for viral safety. Every lot is assayed for osteoinductive potential using an athymic mouse model or alkaline phosphatase assay prior to release.

## Clinical Evidence

No clinical data provided. Evidence consists of bench testing and in vivo biological assessment. Osteoinductive potential is verified for every lot using an athymic mouse model or alkaline phosphatase assay. Viral inactivation potential was validated using a panel of model human viruses.

## Technological Characteristics

Resorbable bone void filler composed of human cortical and cancellous bone granules and sodium hyaluronate. Delivered via sterile plastic syringe. Osteoconductive and osteoinductive. Single-donor processed. Complies with 21 CFR Part 1270 and 1271 for human tissue.

## Regulatory Identification

A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.

## Predicate Devices

- DBX® Demineralized Bone Matrix ([K040262](/device/K040262.md))
- Sygnal™ DBM ([K080405](/device/K080405.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K091218

SEP 2 8 2009

# 510(k) SUMMARY OF SAFETY & EFFECTIVENESS

# General Company Information

Musculoskeletal Transplant Foundation Name: Nancy Bennewitz Joy Contact: Regulatory Affairs Submission Specialist

| Address: |  |  |  |
|----------|--|--|--|
|          |  |  |  |

125 May Street Edison, NJ 08837 USA

Telephone: Fax: Email:

(732) 661-2381 (732) 661-2189 Nancy Joy@mtf.org

#### Date Prepared

August 5, 2009

### General Device Information

Product Name:

DBX® Inject

Classification:

Bone Void Filler Containing Human Demineralized Bone Matrix (DBM) 21 CFR §888.3045 - Product code: MBP, MQV Class II

#### Predicate Devices

DBX® Demineralized Bone Matrix Musculoskeletal Transplant Foundation 510(k) K040262

Sygnal™ DBM Musculoskeletal Transplant Foundation 510(k) K080405

{1}------------------------------------------------

## Description

DBX® Inject is completely resorbable and is composed of donated cortical and cancellous bone. The bone granules are mixed with sodium hyaluronate (Hy). DBX® Inject consists of DBX Putty or Paste with a separate, sterile plastic syringe for delivery directly into the operative site.

# Intended Use (Indications)

DBX® Inject is intended for use as a Demineralized Bone Matrix for voids or gaps that are not intrinsic to the stability of the bony structure. It can be used as follows:

| Indications for Use  | Putty | Paste |
|----------------------|-------|-------|
| Extremities          | √     | √     |
| Posterolateral spine | √     |       |
| Pelvis               | √     | √     |

DBX® Inject is indicated for treatment of surgically created osseous defects or DBX - Info. In the matic injury. DBX® Inject Putty can be used as an extender in the spine with autograft. DBX® Inject can be used with bone marrow. DBX® Inject is for single patient use only.

#### Substantial Equivalence

This submission supports the position that DBX® Inject is substantially equivalent to a number of previously cleared devices, including:

DBX® Demineralized Bone Matrix - Musculoskeletal Transplant Foundation [K040262] Sygnal™ DBM-Musculoskeletal Transplant Foundation [K080405]

When comparing DBX® Inject to its predicate devices, there are no new types of safety and effectiveness questions. DBX® Inject has been demonstrated to be substantially equivalent to its predicate devices.

{2}------------------------------------------------

#### Osteoinductive Potential:

DBX® Inject is osteoconductive, and has been shown to have osteoinductive potential in an athymic mouse model. Every lot of final DBX® Inject Paste will be assayed in vivo for osteoinductive potential. Every lot of final DBX® Inject Putty product will be tested in an athymic mouse model or in an alkaline phosphatase assay, which has been shown to have a positive correlation with the athymic mouse model, to ensure the osteoinductive potential of the final product. Standard testing performed in an athymic mouse or alkaline phosphatase assay must prove positive for lot release. It is unknown how the osteoinductive potential, measured in the athymic mouse model or the alkaline phosphatase assay, will correlate with clinical performance in human subjects.

#### Viral Clearance and Inactivation:

The method for processing the demineralized bone matrix contained in DBX® Inject was evaluated for its viral inactivation potential. A panel of model potential human viruses representing various virus types, sizes, shapes and genomes were evaluated. The DBM processing methods were determined to provide significant viral inactivation potential for a wide range of potential viruses.

# Safety and Effectiveness Information

DBX® Inject is single-donor processed. The donor suitability criteria used to screen this donor are in compliance with the FDA regulations published in 21 CFR Part 1270 and Part 1271 Human Tissue Intended for Transplantation.

#### Conclusion

Musculoskeletal Transplant Foundation believes that the information provided in this 510(k) submission establishes that similar legally marketed devices have been used for the same clinical applications as the DBX® Inject. The materials from which DBX® Inject is fabricated have an established history of use, and the device has been tested in accordance with applicable FDA guidelines.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the three branches of government. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES" are arranged in a circular pattern around the eagle.

SEP % 8 2009

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

Musculoskeletal Transplant Foundation % Ms. Nancy Bennewitz Joy Regulatory Affairs Submission Specialist 125 May Street Edison, NJ 08837

Re: K091218

Trade/Device Name: DBX® Inject Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: II Product Code: MOV, MBP Dated: August 11, 2009 Received: August 12, 2009

Dear Ms. Bennewitz Joy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the individe for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your case in to rely found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Ms. Nancy Bennewitz Joy

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Sincerely yours,

Mark A. Millhiser

Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# INDICATIONS FOR USE

510(k) Number (if known): K091218

Device Name: DBX® Inject

Indications for Use:

DBX® Inject is intended for use as a Demineralized Bone Matrix for voids or gaps that are not intrinsic to the stability of the bony structure. It can be used as follows:

| Indications for Use  | Putty | Paste |
|----------------------|-------|-------|
| Extremities          | √     | √     |
| Posterolateral spine | √     |       |
| Pelvis               | √     | √     |

DBX® Inject is indicated for treatment of surgically created osseous defects or osseous defects created from traumatic injury. DBX® Inject Putty can be used as an extender in the spine with autograft. DBX® Inject can be used with bone marrow. DBX® Inject is for single patient use only.

Prescription Use X (Per 21 CFR 801 Subpart D) Over-The-Counter Use (Per 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE------------------------------------------------------------------------------------------------------------------------------------------IF NEEDED.)

OR

CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)

FOR M.MELKERSON

(Division Sign-Off) Division of Surgical, Orthopedic, ad Restorative Devices

.(k) Number K091218

MTF®

---

**Source:** [https://fda.innolitics.com/device/K091218](https://fda.innolitics.com/device/K091218)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
