GAS TRANSFILL

K091191 · Respironics Inc., Sleep & Home Respiratory Group · CAW · Oct 23, 2009 · Anesthesiology

Device Facts

Record IDK091191
Device NameGAS TRANSFILL
ApplicantRespironics Inc., Sleep & Home Respiratory Group
Product CodeCAW · Anesthesiology
Decision DateOct 23, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5440
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The intended use of the Gas Transfill System is to provide supplemental oxygen to patients and to supply pressurized oxygen to fill gas cylinders for the patient's personal ambulatory use. The device is not intended to be life supporting nor life sustaining.

Device Story

Gas Transfill system consists of high-pressure oxygen compressor paired with external oxygen concentrator; concentrator provides gaseous oxygen to compressor for cylinder filling and to patient for breathing; compressor fills medical oxygen cylinders for ambulatory use; patient cannot breathe directly from compressor; used in home setting; operated by patient; provides supplemental oxygen and portable supply; benefits patient by enabling mobility.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

High-pressure oxygen compressor; utilizes external oxygen concentrator (up to 2LPM for filling, 3LPM for breathing); standalone system; non-life-supporting.

Indications for Use

Indicated for patients requiring supplemental oxygen and pressurized oxygen for personal ambulatory use via filled gas cylinders. Not for life-supporting or life-sustaining use.

Regulatory Classification

Identification

A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Traditional 510(k) Tab 5 - 510(k) summary KC091191 Gas Transfill OCT 2:3 2009 TAB 5 510(K) SUMMARY Date of Submission April 20, 2008 Official Contact Zita A. Yurko Director, Regulatory Affairs Respironics, Inc. 1001 Murry Ridge Lane Murrysville, PA 15668 Zita.yurko@respironics.com 724-387-4120 t 724-882-4120 c 724-387-7490 f 21 CFR 868.5440 Classification Reference Product Code CAW - Portable oxygen generator Portable Oxygen generator Common/Usual Name Proprietary Name Gas Transfill Predicate Device(s) Invacare Corp. Home Fill II (K003939) Chad Therapeutics Total O2 (K013472) DelVilbiss Healthcare iFill (K053240) Reason for submission new device ## Substantial Equivalence The Gas Transfill is similar to the previously cleared predicate devices in: - Intended use to produce and fill medical oxygen cylinders with high pressure gaseous oxygen - Operating principle produce high pressure oxygen from gaseous oxygen provided by an ට external Oxygen Concentrator. {2}------------------------------------------------ # Intended Use The intended use of the Gas Transfill System is to provide supplemental oxygen to patients and to supply pressurized oxygen to fill gas cylinders for the patient's personal ambulatory use. The device is not intended to be life supporting nor life sustaining. # Device Description The Respironics Gas Transfill System is comprised of a high pressure oxygen compressor and an external oxygen concentrator. The oxygen concentrator provides up to 2LPM of gaseous oxygen to the high pressure oxygen compressor for filling medical oxygen cylinders. In addition, the external oxygen concentrator provides up to 3LPM of gaseous oxygen for patient breathing. The external oxygen concentrator can be one of the Respironics series of FDA cleared devices or other manufacturers FDA cleared oxygen concentrators. The patient cannot breathe directly from the high pressure oxygen compressor. The purpose of the high pressure oxygen compressor is to fill medical oxygen cylinders only. (End of Tab.) {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002 Ms. Zita A. Yurko Director of Regulatory Affairs Respironics, Incorporated Sleep & Home Respiratory Group 1001 Murry Ridge Lane Murrysville, Pennsylvania 15668 Re: K091191 Trade/Device Name: Gas Transfill Regulation Number: 21 CFR 868.5440 Regulation Name: Portable Oxygen Generator Regulatory Class: II Product Code: CAW Dated: August 24, 2009 Received: August 25, 2009 Dear Ms. Yurko: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. OCT 2 3 2009 If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ #### Page 2 - Ms. Yurko Please be advised that FDA's issuance of a substantial equivalence determination docs not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Foc Susan Runner, D.D.S., M.A. Acting Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {5}------------------------------------------------ ### Indications for Use 510(k) Number (if known): Device Name: ___Gas Transfill The intended use of the Gas Transfill System is to provide supplemental oxygen to patients and to supply pressurized oxygen to fill gas cylinders for the patient's personal ambulatory use. The device is not intended to be life supporting nor life sustaining. Prescription Use > Over-The-Counter Use 1 rescription Ose ____________________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) L. Schuttheis (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K091191 Page 000015
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...