LOCKBOX FOR USE WITH MEDFUSION 3000 SERIES PUMPS

K091181 · Smiths Medical MD, Inc. · MRZ · Jun 12, 2009 · General Hospital

Device Facts

Record IDK091181
Device NameLOCKBOX FOR USE WITH MEDFUSION 3000 SERIES PUMPS
ApplicantSmiths Medical MD, Inc.
Product CodeMRZ · General Hospital
Decision DateJun 12, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5725
Device ClassClass 2

Indications for Use

The Lockbox is intended to hold a Medfusion™ 3000 Series Pump and provide reasonably secure access to the medication syringe contained within.

Device Story

Lockbox is a lockable, plastic enclosure accessory for Medfusion™ 3000 Series Syringe Pumps. Device holds pump and medication syringe; provides physical security to prevent unauthorized access to medication. Attaches to IV poles or horizontal square rails via mounting brackets. Used in clinical settings; operated by healthcare providers. Benefits patient by securing medication delivery system against tampering or accidental adjustment.

Clinical Evidence

No clinical data. Bench testing only, including physical property assessment and usability studies. Biocompatibility established via material similarity to existing marketed devices.

Technological Characteristics

Lockable plastic enclosure. Mounts to IV poles or horizontal square rails. Mechanical design; no energy source. Biocompatible materials consistent with existing Smiths Medical devices.

Indications for Use

Indicated for use as a protective, lockable enclosure for Medfusion™ 3000 Series Syringe Pumps to secure medication syringes during clinical use.

Regulatory Classification

Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### 510(k) Summary (As required by 21 CFR 807.92(c)) #### K091181 510(k) Number: Page 1 of 2 #### Date Prepared 20 April 2009, Updated 11 June, 2009 #### Submitter Information # JUN 1 2 2009 | Submitter's Name: | Smiths Medical MD, Inc. | |-------------------|-------------------------| | Address: | 1265 Grey Fox Road | | | St. Paul, MN 55112 | Establishment Registration: 2183502 Contact Person: James Chapman Sr. Regulatory Affairs Associate Phone: (651) 628-7611 Fax: (651) 628-7457 #### Device Information | Trade Name: | Lockbox | |----------------------|---------------------------------------------------| | Common Name: | Lockbox for use with Medfusion™ 3000 Series Pumps | | Classification Name: | | | Product Code: | MRZ | | Regulation: | 21 CFR §880.5725 | ## Predicate Device(s) The predicate devices are the currently marketed LockBox for use with the CADD®-Solis Ambulatory Infusion Pump (K080743) and LockBox with Syringe Holder (K942161). #### Device Description The Lockbox is an accessory to the Medfusion™ 3000 Series Syringe Pump. It is a lockable, plastic enclosure for the pump when the pump is loaded with a commercially available syringe containing medication. The Lockbox may be attached to an IV pole with the Pole Mount Bracket or with modification to a horizontal square rail. #### Intended Use/Indications for Use The Lockbox is intended to hold a Medfusion™ 3000 Series Syringe Pump and provide reasonably secure access to the medication syringe contained within. Lockbox for use with Medfusion™ 3000 Series Pumps Smiths Medical MD, Inc. Smiths Medical MD, Inc. Submission Page 15 {1}------------------------------------------------ ## 510(k) Summary Page 2 of 2 ### Summary of Non-Clinical Testing The non-clinical testing included assessment of the physical properties of the Lockbox and its ability to achieve its intended use. Bench testing of the Lockbox confirmed the suitability of the device for its intended use. The following physical tests were performed: > Bench Testing Usability Studies A biocompatibility assessment of the device was also performed. The purpose of the biocompatibility assessment was to ensure that biocompatibility had been established for the device. The device is biocompatible based on the similarity of the materials of construction to other devices currently marketed by Smiths Medical MD, Inc. #### Summary of Clinical Testing Human clinical studies were deemed not necessary to evaluate the safety or effectiveness of the Lockbox. #### Statement of Equivalence The Lockbox for Medfusion™ 3000 Series Pumps is substantially equivalent to the currently marketed LockBox for use with the CADD® -Solis Ambulatory Infusion Pump and LockBox with Syringe Holder based on a comparison of the indications for use and the technological characteristics of the device. #### Conclusion The Lockbox for use with Medfusion™ 3000 Series Pumps is substantially equivalent to the currently marketed LockBox for use with the CADD -Solis Ambulatory Infusion Pump and LockBox with Syringe Holder based on the technological characteristics of the devices. Bench tests confirmed the suitability of the device for its intended use. Lockbox for use with Medfusion™ 3000 Series Pumps Smiths Medical MD, Inc. Smiths Medical MD, Inc. Submission Page 16 {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle faces to the right. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA". #### Public Health Service JUN 1 2 2009 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. James Chapman Senior Regulatory Affairs Associate Smiths Medical MD, Incorporated 1265 Grey Fox Road Sait Paul, Minnesota 55112 Re: K091181 Trade/Device Name: Lockbox for Medfusion™ 3000 Series Pumps Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: MRZ Dated: April 20, 2009 Received: April 23, 2009 Dear Mr. Chapman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ ## Page 2- Mr. Chapman Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/Centers Offices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Susan Runner Susan Runner, D.D.S., M.A. Acting Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {4}------------------------------------------------ ## SMITHS MEDICAL MD, INC. 510(k) Premarket Notification ## Indications for Use Statement Koqußi 510(k) Number: __ Device Name: Lockbox for Medfusion™ 3000 Series Pumps Indications for Use: "The Lockbox is intended to hold a Medfusion™ 3000 Series Pump and provide reasonably secure access to the medication syringe contained within." Prescription Use ਮ (Per 21 CFR 801 . 109) AND/OR Over-The Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Sef. R.R. (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K091181 Lockbox for use with Medfusion™ 3000 Series Pumps Submission Page 14 Tab 5 - 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...