← Product Code [IKT](/productcode/IKT) · K091056

# RHYTHMLINK DISPOSABLE MONOPOLAR EMG NEEDLE, MODEL TFDN26032252R (K091056)

_Rhythmlink International, LLC · IKT · Jun 23, 2009 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K091056

## Device Facts

- **Applicant:** Rhythmlink International, LLC
- **Product Code:** [IKT](/productcode/IKT.md)
- **Decision Date:** Jun 23, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.1385
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine

## Indications for Use

Recording muscle activity for Electromyograph (EMG) applications. For single patient use only.

## Device Story

Disposable monopolar needle electrode for recording muscle activity during EMG procedures; used by clinicians in clinical settings. Device consists of medical-grade stainless steel needle coated with Teflon (PTFE) for insulation and friction reduction; needle hub made of black polyethylene; solderless crimped connection using gold-plated brass tube. Available in various diameters (0.32-0.57mm) and lengths (15-75mm). User holds needle hub to secure needle at recording site; crimped connection attaches to leadwire. Device captures electrical muscle signals for display on EMG equipment; aids clinicians in diagnostic assessment of neuromuscular activity. Single-patient use design minimizes cross-contamination risk.

## Clinical Evidence

No clinical data. Bench testing only.

## Technological Characteristics

Monopolar needle electrode; medical-grade stainless steel needle; Teflon (PTFE) coating; black polyethylene needle hub; gold-plated brass crimped connector. Available in 0.32-0.57mm diameters and 15-75mm lengths. Passive device; no energy source or software. Single-patient use.

## Regulatory Identification

A diagnostic electromyograph needle electrode is a monopolar or bipolar needle intended to be inserted into muscle or nerve tissue to sense bioelectrical signals. The device is intended for medical purposes for use in connection with electromyography (recording the intrinsic electrical properties of skeletal muscle).

## Predicate Devices

- Neuroline Disposable Monopolar Needle Electrode ([K071185](/device/K071185.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 5. 510(k) Summary

ﺔ، ﻓ

一

:4

## 510(k) Summary of Safety and Effectiveness

JUN 23 2009

·

1091056

| Company Name:            | Rhythmlink International, LLC<br>1256 First Street South Extension<br>Columbia, SC 29209<br>Phone: 803-252-1222<br>FDA Registration #: 1067162<br>Owner Operator #: 9052354                                                                                                                                                                                                                          |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact Person: | James M. Mewborne<br>Director of Engineering and Regulatory Affairs<br>Rhythmlink International, LLC<br>1256 First Street South Extension<br>Columbia, SC 29209<br>Phone: 803-252-1222 ext. 12<br>Email: jmewborne@rhythmlink.com                                                                                                                                                                    |
| Summary Date:            | April 9, 2009                                                                                                                                                                                                                                                                                                                                                                                        |
| Device Identification:   | Proprietary Device Name:<br>Rhythmlink Disposable Monopolar EMG Needle (Trade names have not been finalized at this time)<br><br>Generic Device Name:<br>Diagnostic Electromyograph Needle Electrode<br><br>Regulatory Class:<br>Class II<br><br>Classification Name: 21 CFR §890.1385, Diagnostic Electromyograph Needle Electrode<br><br>This device has not been previously submitted to the FDA. |
| Predicate Device(s):     | 510(k) Number: K071185<br>Manufacturer: Ambu®<br>Trade Name: Neuroline Disposable Monopolar Needle Electrode<br>Product Code: IKT                                                                                                                                                                                                                                                                    |

{1}------------------------------------------------

# Rhythmlink Disposable Monopolar EMG Needle is a device Device Description: intended for recording muscle activity during electromyography (EMG) procedures. The device consists of an insulated monopolar needle, needle hub, connector (solderless crimped connection), and plastic needle cover. The monopolar needle is made of medical grade stainless steel and is available in three lengths. The entire needle is coated with Teflon (PTFE), except for the recording tip, to insulate the needle and reduce friction upon insertion. The solderless crimped connection consists of a gold-plated brass tube 1mm in thickness, crimped onto the isolated needle. The connection is covered with a needle hub made of black polyethylene molding. The plastic needle cover is made of polyethylene. The device is available in seven different needle sizes. The needles are available in 0.32 to 0.57mm in diameter and are available in 15 to 75mm lengths. The needle is intended to record muscle activity during EMG procedures, and the needle hub is designed to be held by the user to secure the needle in the recording site. The solderless crimped connection secures the needle to the end of the leadwire. With the exception of the needle length, needle diameter, and leadwire length, the components and their dimensions are consistent throughout the product line, and these material and dimensional specifications are outlined in the product drawing under Device Description. Recording muscle activity for Electromyography (EMG) Indications for Use: applications. For single patient use only,

This concludes the 510(k) summary.

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract image of an eagle or other bird-like figure with three stylized wing segments.

Public Health Service

JUN 23 2009

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Rhythmlink® International, LLC c/o James M. Mewborne Director of Engineering and Regulatory Affairs 1256 First Street South Extension Columbia, SC 29209

Re: K091056

Trade/Device Name: Rhythmlink Disposable Monopolar EMG Needle Regulation Number: 21 CFR 890.1385 Regulation Name: Diagnostic Electromyograph Needle Electrode Regulatory Class: II Product Code: IKT Dated: April 9, 2009 Received: April 13, 2009

Dear Mr. Mewborne:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrl/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html,

Sincerely vours.

Kesia Alexander for

Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

### 4. Indications for Use Statement

# Indications for Use

510(k) Number (if known): Pending

K091056 Device Name Rhythmlink Disposable Monopolar EMG Needle

Indications ForUse: Recording muscle activity for Electromyograph (RMG) applications . For single patient use only.

scription Use

AND/OR

Over-The-Counter Use
(21 CFR 801 Subpart C)

LEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF

Concurrence of CDRH, Office of Device Evaluation (ODE)

JOE HUTTER

(Division Sign-Off) Division of Ophthalmic and Ear, Nose and Throat Devices

e 1 of .. 1

Kof 1056 510(k) Number

9 of 52

---

**Source:** [https://fda.innolitics.com/device/K091056](https://fda.innolitics.com/device/K091056)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
