The sponsor used prospective and retrospective clinical cohorts to establish the clinical performance (sensitivity and specificity) of the ProParaflu+ Assay compared to standard-of-care cell culture/DFA methods.
ProParaflu+ Prospective Study; Prospective clinical study; Follow-up/Duration: May 2008 - September 2009; Study Period: May 2008 - September 2009
Symptomatic individuals suspected of respiratory infection; Sample Size: 857; Number of Sites: 4
Cell culture (rapid or traditional) followed by direct fluorescent antibody (DFA) screening
Sensitivity and specificity for HPIV-1, HPIV-2, and HPIV-3
ProParaflu+ Retrospective Study; Retrospective study
Symptomatic individuals; Sample Size: 91
Direct fluorescent antibody (DFA)
Sensitivity and specificity for HPIV-1
Indications for Use
The ProParaflu+ Assay is a multiplex Real Time RT-PCR in vitro diagnostic test for the qualitative detection and discrimination of Parainfluenza 1 Virus, Parainfluenza 2 Virus and Parainfluenza 3 Virus (HPIV-1, HPIV-2 and HPIV-3) nucleic acids isolated and purified from nasopharyngeal (NP) swab specimens obtained from individuals exhibiting signs and symptoms of respiratory tract infections. This assay targets the conserved regions of the Hemagglutinin-Neuraminidase (HN) gene of HPIV-1, HPIV-2 and HPIV-3, respectively. The detection and discrimination of HPIV-1, HPIV-2 and HPIV-3 nucleic acids from symptomatic patients aid in the diagnosis of human respiratory tract parainfluenza infections if used in conjunction with other clinical and laboratory findings. This test is not intended to detect Parainfluenza 4a or Parainfluenza 4b Viruses. Negative test results are presumptive and should be confirmed by cell culture. Negative results do not preclude Parainfluenza 1, 2 or 3 virus infections and should not be used as the sole basis for treatment or other management decisions.
Device Story
ProParaflu+ Assay is a multiplex Real Time RT-PCR test for HPIV-1, HPIV-2, and HPIV-3. Input: nucleic acids extracted from nasopharyngeal swabs using bioMérieux NucliSENS easyMAG or Roche MagNA Pure LC. Process: Taqman-based RT-PCR amplification on Cepheid SmartCycler II instrument; primers target conserved HN gene regions; dual-labeled probes (FAM, Cal Orange 560, Cal Red 610) generate fluorescent signals upon cleavage by Taq polymerase. Output: qualitative detection/discrimination of HPIV types. Used in clinical laboratories by trained personnel. Results interpreted via instrument software to aid diagnosis alongside clinical findings. Benefits: rapid identification of specific HPIV types to support clinical management.
Clinical Evidence
Clinical performance established via prospective study (n=857) and retrospective study (n=91) comparing ProParaflu+ to cell culture/DFA. Prospective sensitivity: HPIV-1 88.9%, HPIV-2 96.3%, HPIV-3 97.3%. Specificity: HPIV-1 99.9%, HPIV-2 99.8%, HPIV-3 99.2%. Reproducibility study (3 sites, 12-member panel) showed 97.8% overall agreement.
Technological Characteristics
Multiplex Real Time RT-PCR assay. Targets: HPIV-1, HPIV-2, HPIV-3 HN genes. Chemistry: Taqman (5'-3' exonuclease activity). Instrumentation: Cepheid SmartCycler II. Detection: Fluorescence-based (FAM, Cal Orange 560, Cal Red 610, Quasar 670). Extraction: Automated (NucliSENS easyMAG or MagNA Pure LC).
Indications for Use
Indicated for symptomatic individuals of all ages exhibiting signs/symptoms of respiratory tract infection. Used for qualitative detection and discrimination of HPIV-1, HPIV-2, and HPIV-3 nucleic acids from nasopharyngeal swab specimens. Not intended for HPIV-4a or HPIV-4b detection.
Regulatory Classification
Identification
A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B; (2) Influenza A subtype H1 and Influenza A subtype H3; (3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B; (4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus; (5) Human Metapneumovirus; (6) Rhinovirus; and (7) Adenovirus.
Special Controls
*Classification.* Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.
Predicate Devices
ID Tag Respiratory Virus Panel, Luminex Molecular Diagnostics (K063765)
Submission Summary (Full Text)
{1}------------------------------------------------
K09/053
Page 1 of 7 Date: November 5, 2009
## Attachment D 510(k) SUMMARY
#### CONTACT
NOV 2 0 2009
Karen Harrington Gen-Probe Prodesse, Inc. W229 N1870 Westwood Dr. Waukesha, WI 53186
### NAME OF DEVICE
Trade Name: Regulation Number: Classification Name:
ProParaflu+TM Assay 21 CFR 866.3980 Respiratory viral panel multiplex nucleic acid assay
## PREDICATE DEVICE
K063765 - ID Tag Respiratory Virus Panel, Luminex Molecular Diagnostics
#### INTENDED USE
The ProParaflu+ Assay is a multiplex Real Time RT-PCR in vitro diagnostic test for the qualitative detection and discrimination of Parainfluenza 1 Virus, Parainfluenza 2 Virus and Parainfluenza 3 Virus (HPIV-1, HPIV-2 and HPIV-3) nucleic acids isolated and purified from nasopharyngeal (NP) swab specimens obtained from individuals exhibiting signs and symptoms of respiratory tract infections. This assay targets the conserved regions of the Hemagglutinin-Neuraminidase (HN) gene of HPIV-1, HPIV-2 and HPIV-3, respectively. The detection and discrimination of HPIV-1, HPIV-2 and HPIV-3 nucleic acids from symptomatic patients aid in the diagnosis of human respiratory tract parainfluenza infections if used in conjunction with other clinical and laboratory findings. This test is not intended to detect Parainfluenza 4a or Parainfluenza 4b Viruses.
Negative test results are presumptive and should be confirmed by cell culture. Negative results do not preclude Parainfluenza 1, 2 or 3 virus infections and should not be used as the sole basis for treatment or other management decisions.
#### PRODUCT DESCRIPTION
The ProParaflu+ Assay enables the detection and differentiation of Parainfluenza 1 Virus, Parainfluenza 2 Virus, Parainfluenza 3 Virus and an Internal Control (IC) nucleic acid. Nasopharyngeal swab specimens from symptomatic patients using a polyester, rayon or nylon tipped swab and place into viral transport medium. The IC is added to every sample prior to nucleic acid extraction to monitor for inhibitors present in the specimens.
Isolation and purification of nucleic acids is performed using the bioMérieux NucliSENS
{2}------------------------------------------------
easyMAG automated extractor and the Automated Magnetic Extraction Reagents or the Roche MagNA Pure LC Instrument and the MagNA Pure Total Nucleic Acid Isolation Kit.
The purified nucleic acids are added to the ProParaflu+ Supermix along with enzymes included in the ProParaflu+ Detection Kit. The ProParaflu+ Supermix contains oligonucleotide primers that are complementary to highly conserved regions of hemagglutinin neuraminidase gene for each human Parainfluenza type (1, 2 and 3). The probes are dual-labeled with a reporter dye attached to the 5'-end and a quencher dye attached to the 3'-end (see table below).
RT-PCR amplification is performed in a Cepheid SmartCycler® II instrument. During this process, the primers and probes anneal specifically to the template (if present) followed by primer extension and amplification. The ProParaflu+ Assay is based on Taqman chemistry, which utilizes the 5' – 3' exonuclease activity of the Taq polymerase to cleave the probe thus separating the reporter dye from the quencher. This generates an increase in fluorescent signal upon excitation from a light source. With each cycle, additional reporter dye molecules are cleaved from their respective probes, further increasing the fluorescent signal. The amount of fluorescence at any given cycle is dependent on the amount of amplification product present at that time. Fluorescent intensity is monitored during each PCR cycle by the real-time instrument. Results are analyzed and interpreted as presented by the software.
| Analyte | Gene Targeted | Probe<br>Fluorophore | Absorbance<br>Peak | Emission<br>Peak | Instrument<br>Channel |
|--------------------------|--------------------------------|----------------------|--------------------|------------------|-----------------------|
| Parainfluenza 1<br>Virus | Hemagglutinin<br>neuraminidase | FAM | 495 nm | 520 nm | FAM |
| Parainfluenza 3<br>Virus | Hemagglutinin<br>neuraminidase | Cal Orange 560 | 540 nm | 561 nm | TET |
| Parainfluenza 2<br>Virus | Hemagglutinin<br>neuraminidase | Cal Red 610 | 595 nm | 615 nm | Texas Red |
| Internal Control | NA | Quasar 670 | 647 nm | 667 nm | Cy5 |
{3}------------------------------------------------
### SUBSTANTIAL EQUIVALENCE
#### Clinical Performance
The clinical performance of the ProParaflu+ Assay was established during a prospective study at 4 U.S. clinical laboratories during May 2008 - September 2009. Specimens used in the study represented excess nasopharyngeal (NP) swab specimens that were prospectively collected from symptomatic individuals suspected of respiratory infection, and were submitted for routine analysis. Demographic details for this patient population are summarized in the following table.
| Sex | Number of Subjects |
|---------------|--------------------|
| Female | 407 (47.5%) |
| Male | 450 (52.5%) |
| Age (yrs) | |
| ≤ 5 years | 580 (67.7%) |
| 6 - 21 years | 168 (19.6%) |
| 22 - 59 years | 67 (7.8%) |
| ≥ 60 years | 42 (4.9%) |
Gender and Age Demographic Detail for ProParaflu+ Prospective Study
Performance of the ProParaflu+ Assay was compared to the reference method of cell culture (rapid or traditional) followed by direct fluorescent antibody (DFA) screening and HPIV type identification.
A total of 857 eligible NP swab samples were tested with the ProParaflu+ Assay and by culture across four clinical sites. Of the ProParaflu+ Assay run on all eligible specimens, 99.2% (852/857) of these specimens were successful on the first attempt. The remaining 5 gave "Unresolved" results on the first attempt. Unresolved results occur when the sample is negative for all three HPIVs and the Internal Control, indicating potentially PCR-inhibiting samples. Of the 5 "Unresolved" specimens on the first attempt, 60.0% (3/5) gave a valid result on the second attempt. The remaining 2 were "Unresolved" on the second attempt and are not included in the analysis below. Both samples were culture negative.
Discrepant analysis for samples where ProParaflu+ Assay and culture results were in disagreement was performed using RT-PCR with virus specific primers obtained from literature® 9 (and different from those used in ProParaflu+) followed by bi-directional sequencing.
{4}------------------------------------------------
# Prospective Study
| | | Culture/DFA | | | |
|----------------------|----------|-------------|----------|-------|-------------------------------------------|
| | | Positive | Negative | Total | |
| ProParaflu+<br>Assay | Positive | 16 | 1a | 17 | Sensitivity 88.9% (67.2% - 96.9%) 95% CI |
| | Negative | 2b | 838 | 840 | Specificity 99.9% (99.3% - 100.0%) 95% CI |
| | Total | 18 | 839 | 857 | |
# Parainfluenza 1 Comparison Results
4One (1) sample positive for HPIV-1 by bi-directional sequence analysis.
bTwo (2) samples negative for HPIV-1 by bi-directional sequence analysis. One sample positive for HPIV-3 by ProParaflu+ and bi-directional sequence analysis.
# Parainfluenza 2 Comparison Results
| | | Culture/DFA | | | |
|----------------------|----------|-------------|----------|-------|---------------------------------------------|
| | | Positive | Negative | Total | |
| ProParaflu+<br>Assay | Positive | 26 | 2a | 28 | Sensitivity 96.3% (81.7% - 99.3%)<br>95% CI |
| | Negative | 1b | 828 | 829 | Specificity 99.8% (99.1% - 99.9%)<br>95% CI |
| | Total | 27 | 830 | 857 | |
4Two (2) samples positive for HPIV-2 by bi-directional sequence analysis.
bOne (1) sample negative for HPIV-2 by bi-directional sequence analysis.
# Parainfluenza 3 Comparison Results
| | | Culture/DFA | | | |
|----------------------|----------|-------------|----------|-------|---------------------------------------------|
| | | Positive | Negative | Total | |
| ProParaflu+<br>Assay | Positive | 36 | 8a | 44 | Sensitivity 97.3% (86.2% - 99.5%)<br>95% CI |
| | Negative | 1b | 812 | 813 | Specificity 99.2% (98.1% - 99.5%)<br>95% CI |
| | Total | 37 | 820 | 857 | |
4Seven (7) samples positive for HPIV-3 and one (1) sample negative for HPIV-3 by bidirectional sequence analysis.
bone (1) sample negative for HPIV-3 by bi-directional sequence analysis.
{5}------------------------------------------------
## Retrospective Study
Due to a minimal number of HPIV-1 positive samples, a retrospective study was also conducted using a total of 91 frozen NP swab samples that had been previously tested by direct DFA. Demographic details for this patient population are summarized in the following table.
Gender and Age Demographic Detail for ProParaflu+ Retrospective Study
| Sex | Number of Subjects |
|---------------|--------------------|
| Female | 40 (44.4%) |
| Male | 50 (55.6%) |
| Age (yrs) | |
| ≤ 5 years | 81 (90.0%) |
| 6 - 21 years | 5 (5.6%) |
| 22 – 59 years | 2 (2.2%) |
| ≥ 60 years | 2 (2.2%) |
# Parainfluenza 1 Comparison Results
| | | <i>DFA</i> | | | |
|------------------------------|----------|------------|----------|-------|---------------------------------------------|
| | | Positive | Negative | Total | |
| <i>ProParaflu+<br/>Assay</i> | Positive | 24 | 0 | 24 | Sensitivity 82.8% (65.4% - 92.4%)<br>95% CI |
| | Negative | 5a | 62 | 67 | Specificity 100% (94.2% - 100%)<br>95% CI |
| | Total | 29 | 62 | 91 | |
4Five (5) samples negative for HPIV-1 by bi-directional sequence analysis.
{6}------------------------------------------------
# ProParaflu Assay 510(k) Submission
### Reproducibility
The reproducibility of the ProParaflu+ Assay was evaluated at 3 laboratory sites. Reproducibility was assessed using a panel of 12 simulated samples that included medium positive, low positive (near the assay limit of detection, ≥95% positive), intermediate (1 log below the assay limit of detection), and high negative (below the assay limit of detection, <5% positive) samples. Panels and controls were tested at each site by 2 operators for 5 days. The overall percent agreement with the expected result for the ProParaflu+ Assay was 97.8%.
| Panel Member ID | HPIV-1 high negative" | HPIV-1 low positive | HPIV-1 medium positive | HPIV-2 high negative" | HPIV-2 low positive | HPIV-2 medium positive | HPIV-3 high negative" | HPIV-3 low positive | HPIV-3 medium positive | Para Extraction Control | Para RNA Control HPIV-1 | HPIV-2 | HPIV-3 | Negative Control | Total % Agreement | |
|-----------------------|--------------------------------------|--------------------------------|------------------------|-----------------------|---------------------|------------------------|-----------------------|---------------------|------------------------|-------------------------|-------------------------|---------------|---------------|------------------|-------------------|------------------|
| Concentration | 0.001 X LoD | 2 X LoD | 10X LoD | 0.001 X LoD | 2 X LoD | 10X LoD | 0.01 X LoD | 2 X LoD | 10X LoD | N/A | N/A | N/A | N/A | N/A | | |
| Site 1 | Agreement with Expected Result | 10/10<br>100% | 8/10<br>80% | 9/9<br>100% | 10/10<br>100% | 9/9<br>100% | 9/10<br>90% | 10/10<br>100% | 9/10<br>90% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% | 114/118<br>96.6% |
| | Mean Ct Value | 27.73 | 28.31 | 26.33 | 27.90 | 28.62 | 26.27 | 27.76 | 31.21 | 29.47 | 27.33 | 27.37 | 29.37 | 28.61 | 27.69 | |
| | % CV | 2.87 | 1.55 | 1.60 | 2.84 | 0.85 | 1.25 | 2.43 | 3.21 | 1.91 | 1.60 | 1.05 | 0.43 | 0.81 | 1.67 | |
| | Site 2 | Agreement with Expected Result | 8/10<br>80% | 8/10<br>80% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% |
| Mean Ct Value | | 28.59 | 28.47 | 26.12 | 28.83 | 28.91 | 26.61 | 28.30 | 31.64 | 29.51 | 27.56 | 23.86 | 26.09 | 25.23 | 28.68 | |
| % CV | | 1.36 | 1.72 | 1.26 | 2.80 | 1.31 | 1.83 | 1.18 | 2.17 | 2.66 | 2.47 | 1.48 | 1.12 | 0.92 | 1.15 | |
| Site 3 | | Agreement with Expected Result | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% |
| | Mean Ct Value | 26.35 | 29.91 | 27.67 | 26.28 | 29.51 | 27.44 | 26.67 | 33.13 | 30.43 | 28.61 | 28.73 | 30.98 | 29.84 | 26.46 | |
| | % CV | 0.88 | 0.81 | 1.25 | 1.35 | 0.57 | 2.25 | 4.13 | 2.36 | 1.02 | 1.54 | 3.31 | 3.39 | 3.19 | 1.04 | |
| | Total Agreement with Expected Result | 28/30<br>93.3% | 26/30<br>86.7% | 29/29<br>100% | 30/30<br>100% | 29/29<br>100% | 29/30<br>96.7% | 30/30<br>100% | 29/30<br>96.7% | 30/30<br>100% | 30/30<br>100% | 30/30<br>100% | | | 30/30<br>100% | 350/358<br>97.8% |
| 95% CI | 78.7% -<br>98.2% | 70.3% -<br>94.7% | 88.3% -<br>100% | 88.6% -<br>100% | 88.3% -<br>100% | 83.3%-<br>99.4% | 88.6% -<br>100% | 83.3%-<br>99.4% | 88.6%-<br>100% | 88.6%-<br>100% | 88.6% - 100% | | | 88.6%-<br>100% | 95.6% -<br>98.9% | |
| Overall Mean Ct Value | 27.56 | 28.90 | 26.72 | 27.67 | 29.03 | 26.79 | 27.58 | 32.02 | 29.80 | 27.83 | 26.65 | 28.81 | 27.89 | 27.62 | | |
| Overall % CV | 3.88 | 2.87 | 2.95 | 4.55 | 1.57 | 2.54 | 3.68 | 3.61 | 2.42 | 2.75 | 8.13 | 7.49 | 7.39 | 3.54 | | |
*Average Ct value for the Internal Control (IC)
{7}------------------------------------------------
An additional reproducibility study was performed to assess samples that were at an intermediate concentration, below the assay's LoD but above the "high negatives" tested during the original reproducibility study. The percent positive for the intermediate member across all sites was 56.7% for HPIV-1 (mean Ct = 35.1), 86.7% for HPIV-2 (mean Ct = 33.0), and 30.0% for HPIV-3 (mean Ct = 37.1). This result was expected as the intermediate concentration should be positive in the range of 5 - 95% as the samples were lower concentration than the LoD concentration (≥ 95% positive) and higher than the "high negative" concentration (< 5% positive).
| | Panel Member ID | HPIV-1 intermediate | HPIV-2 intermediate | HPIV-3 intermediate | Para Extraction Control | Parainfluenza RNA Control | | | Negative Controla |
|--------|--------------------------------------|---------------------|---------------------|---------------------|-------------------------|---------------------------|---------------|---------------|-------------------|
| | Concentration | 0.1 X LoD | 0.1 X LoD | 0.1 X LoD | N/A | HPIV-1 | HPIV-2 | HPIV-3 | N/A |
| Site 1 | Agreement with Positive Result | 4/10<br>40% | 8/10<br>80% | 1/10<br>10% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% | 10/10*<br>100% |
| Site 1 | Average Ct Value | 35.5 | 33.3 | 36.9 | 27.9 | 28.8 | 30.4 | 29.5 | 28.4 |
| Site 1 | % CV | 6.19 | 2.78 | N/A | 3.80 | 1.13 | 0.88 | 0.89 | 3.47 |
| Site 2 | Agreement with Positive Result | 8/10<br>80% | 10/10<br>100% | 7/10<br>70% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% |
| Site 2 | Average Ct Value | 34.4 | 32.2 | 37.3 | 27.5 | 29.0 | 30.6 | 29.9 | 27.7 |
| Site 2 | % CV | 1.38 | 2.22 | 1.50 | 2.92 | 1.04 | 0.89 | 0.72 | 2.80 |
| Site 3 | Agreement with Positive Result | 5/10<br>50% | 8/10<br>80% | 1/10<br>10% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% | 10/10<br>100% |
| Site 3 | Average Ct Value | 35.9 | 33.7 | 35.8 | 28.6 | 29.7 | 31.5 | 30.3 | 27.9 |
| Site 3 | % CV | 2.44 | 2.90 | N/A | 2.75 | 1.92 | 1.69 | 1.46 | 1.94 |
| | Total Agreement with Positive Result | 17/30<br>56.7% | 26/30<br>86.7% | 9/30<br>30.0% | 30/30<br>100% | 30/30<br>100% | 30/30<br>100% | 30/30<br>100% | 30/30<br>100% |
| | 95% CI | 39.2% -<br>72.6% | 70.3% -<br>94.7% | 16.7% -<br>47.9% | 88.7 -<br>100% | 88.7 - 100% | | | 88.7 - 100% |
| | Overall Average Ct Value | 35.1 | 33.0 | 37.1 | 28.0 | 29.1 | 30.8 | 29.9 | 28.0 |
| | Overall % CV | 3.68 | 3.25 | 1.89 | 3.46 | 1.91 | 1.92 | 1.53 | 2.92 |
ª Average Ct value for the Internal Control (IC)
*Agreement with Negative result
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image contains a logo and some text. The logo is a circular emblem with a stylized image of an eagle or bird in the center. The text is partially visible and appears to be part of the circular emblem. The text is not fully legible, but it seems to be related to a government agency or organization.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Building 66 Silver Spring, MD 20993
Dr. Karen Harrington Manager, Clinical Affairs Prodesse Inc. W229 N1870 Westwood Drive Waukesha, WI 53186
NOV 2 0 2009
Re: K091053
> Trade/Device Name: ProParaflu+ "M Assay Regulation Number: 21 CFR 866.3980 Regulation Name: Respiratory viral panel multiplex nucleic acid assay Regulatory Class: Class II Product Code: OQU Dated: October 14, 2009 Received: October 15, 2009
Dear Dr. Harrington:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not
{9}------------------------------------------------
limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Fally weArb
Sally A. Hojvat. M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
# Indication for Use
510(k) Number (if known): k091053
Device Name: ProParaflu+TM Assay
Indication For Use:
The ProParaflu+ Assay is a multiplex Real Time RT-PCR in vitro diagnostic test for the qualitative detection and discrimination of Parainfluenza 1 Virus, Parainfluenza 2 Virus and Parainfluenza 3 Virus (HPIV-1, HPIV-2 and HPIV-3) nucleic acids isolated and purified from nasopharyngeal (NP) swab specimens obtained from individuals exhibiting signs and symptoms of respiratory tract infections. This assay targets the conserved regions of the Hemagglutinin-Neuraminidase (HN) gene of HPIV-1, HPIV-2 and HPIV-3. respectively. The detection and discrimination of HPIV-1, HPIV-2 and HPIV-3 nucleic acids from symptomatic patients aid in the diagnosis of human respiratory tract parainfluenza infections if used in conjunction with other clinical and laboratory findings. This test is not intended to detect Parainfluenza 4a or Parainfluenza 4b Viruses.
Negative test results are presumptive and should be confirmed by cell culture. Negative results do not preclude Parainfluenza 1, 2 or 3 virus infections and should not be used as the sole basis for treatment or other management decisions.
Prescription Use X (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
the Schuf
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) k091053
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.