← Product Code LZU · K090938

# FLOW CONTROLLER (K090938)

_Odyssey Medical, Inc. · LZU · Jun 4, 2009 · OP · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K090938

## Device Facts

- **Applicant:** Odyssey Medical, Inc.
- **Product Code:** LZU
- **Decision Date:** Jun 4, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** OP
- **Attributes:** Therapeutic

## Intended Use

Punctal Occluders (plugs) are indicated for the treatment of "dry eye" syndrome.

## Device Story

Micro Flow Punctal Occluder is a small, plug-shaped device inserted into the punctal opening of the eye to treat dry eye syndrome. Unlike permanent or difficult-to-reverse occlusion methods, this device provides partial occlusion via a through-hole design, allowing for functional, reversible treatment without breaching intact tissue. The device is manufactured from medical-grade silicone with titanium dioxide colorant and a proprietary surface treatment. It features a smaller nose angle and a .002-inch ridge to support the through-hole compared to the predicate. The device is intended for clinical use by eye care professionals.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on design and material characteristics.

## Technological Characteristics

Material: Medical-grade silicone with titanium dioxide colorant. Design: Plug-shaped with a through-hole for partial occlusion, .002-inch ridge on the nose, and modified nose angle. Energy source: None (passive device). Connectivity: None. Sterilization: Not specified.

## Predicate Devices

- Odyssey Punctal Occluder ([K970631](/device/K970631.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K090938

Image /page/0/Picture/1 description: The image shows the word "ODYSSEY" in a stylized font. To the left of the word is a graphic of an arrow piercing through three concentric circles. The arrow is pointing towards the right, and the circles are arranged around the arrow's shaft. The word "ODYSSEY" is written in all capital letters, with each letter outlined in black.

Odyssey Medical, Ir 2975 Brother Blvd. Bartlett, TN 38133 Ph 901-383-777 Fax 901-382-2712

# JUN - 4 2009

## 510K Summary

Submitted: 03/12/2009 By: Terry Green Registration Number: 1060840 Trade Name: Punctal Occluder Common Name: Plug, Punctum Product Name: Micro Flow Classification Name: LZU Classification Panel: Ophthalmic Device Class: Unclassified Performance Standards: None Established

#### Summary:

This device, made of medical grade silicone, titanium dioxide as a colorant and utilizing a proprietary surface treatment made under current good manufacturing practices, is a small, generally "plug" shaped design. Punctal Occluders are intended for the treatment of "dry eve" conditions through partial occlusion of individual punctual openings. By material, design and intended use, Micro Flow is substantially equivalent to the previous Odyssey Punctal Occluder (K970631).

Differences between the two devices may be described as follows:

The original Odyssey plug under 510 K970631 has a larger angle nose, no through hole and shorter in length.

The new design "Micro-Flow" is similar to the original plug with the exceptions of the following:

- A through hole which allows for partial occlusion of individual punctual openings. .
- A .002 ridge on nose to help support the designed through hole. .
- Angle on nose (tip) is smaller. ●

Alternative methods of punctual occlusion are either permanent or difficult to reverse. In contrast, the submitted method is both functionally safe and fully reversible. Unlike some other methods, this device does not require the breach of intact tissue surfaces or the risk of damaging said surface should reversal become desirable.

Kim Finch for Jerry Green Kim Finch
(Signature) (Typed Name)

6/2/09

(Date)

(Premarket Notification [510(k)] Number)

{1}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract symbol that resembles three stylized human figures.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Oddyssey Medical. Inc. c/o Terry R. Green 2975 Brother Blvd. Bartlett, Tennessee 38133

Re: K090938

Trade/Device Name: Micro Flow Punctal Occluder Regulatory Class: Unclassified Product Code: LZU Dated: March 26, 2009 Received: April 3, 2009

Dear Mr. Green:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Malvina B. Eplehnus, m.d.

Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

510(k) K090938 Number:

Name: Micro Flow Device

Indications for Use:

Punctal Occluders (plugs) are indicated for the treatment of "dry eye" syndrome.

Prescription Use

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

AND/OR

(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

510(k) Number_109

×

---

(Division Sign-Off)

Division of Ophthalmic and Ear, Nose and Throat Devices

---

**Source:** [https://fda.innolitics.com/device/K090938](https://fda.innolitics.com/device/K090938)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
