← Product Code LFL · K090864

# CUSA  SHEAR TIP, AN ACCESSORY OF THE CUSA EXCEL ULTRASONIC SURGICAL ASPIRATOR SYSTEM (K090864)

_Integra Radionics, Inc. · LFL · May 8, 2009 · SU · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K090864

## Device Facts

- **Applicant:** Integra Radionics, Inc.
- **Product Code:** LFL
- **Decision Date:** May 8, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic

## Intended Use

The CUSA Shear Tip, an accessory of the CUSA EXcel Ultrasonic Surgical Aspirator System, is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft tissue is desirable, including Neurosurgery, Gastrointestinal and affiliated organ surgery, Urological surgery, Plastic and Reconstructive surgery, General surgery, Orthopedic surgery, Gynecological surgery, Thoracic surgery, Laparoscopic surgery and Thoracoscopic surgery.

## Device Story

CUSA Shear Tip is an accessory tip for the CUSA EXcel Ultrasonic Surgical Aspirator System; used in various surgical specialties for soft tissue fragmentation, emulsification, and aspiration. The distal end features opposing angled lands instead of a flat surface; this geometry promotes refracted longitudinal waves propagating in multiple directions at the tissue interface to enhance fibrous tissue fragmentation. The device attaches to the existing CUSA EXcel handpiece; no modifications to the console, handpiece, or suction/irrigation systems are required. Operated by surgeons in clinical settings (OR); the device assists in tissue removal, potentially improving surgical efficiency in fibrous tissue cases.

## Clinical Evidence

Bench testing only. Performance testing was conducted to demonstrate the device's ability to fragment fibrous tissue compared to standard tips.

## Technological Characteristics

Ultrasonic surgical aspirator accessory; distal tip features opposing angled lands for refracted longitudinal wave propagation. Compatible with existing CUSA EXcel console and handpiece. No changes to energy source or suction/irrigation systems.

## Predicate Devices

- CUSA EXCEL Ultrasonic Surgical Aspirator System ([K981262](/device/K981262.md))
- CUSA EXCEL Ultrasonic Surgical Aspirator System with Bone Tip ([K051947](/device/K051947.md))
- CUSA Selector NXT Ultrasonic Tissue Ablation System ([K081459](/device/K081459.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Ko90864

Page 1 of 2

## CUSA Shear Tip, an accessory of the CUSA EXcel Ultrasonic Surgical Aspirator System 510(k) Summary

#### Submitter's Name and Address:

Integra Radionics 22 Terry Avenue Burlington, MA 01803 781-565-1227 (Telephone) 781-238-0645 (Fax)

## Contact Person and Telephone Number:

MAY - 8 2009

Kevin J. O'Connell Director Regulatory Affairs Integra Radionics, Inc. Tel.: (781) 565-1227

#### Date Summary was Prepared: May 7, 2009.

#### Name of the Device:

Trade Name: CUSA Shear Tip

Common Name: Ultrasonic Surgical Aspirator

Classification Name: Instrument, Ultrasonic Surgical Product Code: LFL

Classification Panel: General and Plastic Surgery

#### Substantial Equivalence:

The CUSA Shear Tip, an accessory of the CUSA EXcel Ultrasonic Surgical Aspirator System, is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft tissue is desirable, including Neurosurgery, Gastrointestinal and affiliated organ surgery, Urological surgery, Plastic and Reconstructive surgery, General surgery, Orthopedic surgery, Gynecological surgery, Thoracic surgery, Laparoscopic surgery and Thoracoscopic surgery. Such functions are within the indications of use for the predicate devices. The technological characteristics are similar to those found in the following predicate devices: CUSA EXCEL Ultrasonic Surgical Aspirator System via 510(k) K981262 on July, 6, 1998; CUSA EXCEL Ultrasonic Surgical Aspirator System with Bone Tip cleared via 510(k) K051947 on August 22, 2005 and CUSA Selector NXT Ultrasonic Tissue Ablation System cleared via 510(k) K081459 on August 13, 2008.

{1}------------------------------------------------

Ko90864

Page 2 of 2

The CUSA Shear Tip is an additional tip that is used with CUSA EXcel system to fragment fibrous tissue. The distal end of the CUSA Shear Tip has a series of opposing angled lands, instead of a flat surface. This pattern promotes refracted longitudinal waves propagating in greatly different directions at the interface to coupled tissue. No changes to the console, handpiece or suction/irrigation system were needed.

Performance testing has been completed to demonstrate that the CUSA Shear Tip can better fragment fibrous tissue.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features an eagle with its wings spread, symbolizing the federal government. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

MAY - 8 2009

ood and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Public Health Service

Integra Radionics, Inc % Mr. Kevin J O'Connell Director Regulatory Affairs 22 Terry Avenue Burlington, Massachusetts 01803

Re: K090864

Trade/Device Name: CUSA Shear Tip, An Accessory of The CUSA Excel Ultrasonic Surgical Aspirator System

Regulation Number: 21 CFR 878.4400

Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: LFL Dated: May 4, 2009 Received: May 5, 2009

Dear Mr. O'Connell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

## Page 2 - Mr. Kevin J O'Connell

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at (240) 276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Daniel Knane

Mark N. Melkerson Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K090864

## Indications for Use Statement

510(k) Number (if known):

Device Name: CUSA Shear Tip, an accessory of the CUSA EXcel Ultrasonic Surgical Aspirator System

Indications For Use:

The CUSA Shear Tip, an accessory of the CUSA EXcel Ultrasonic Surgical Aspirator System, is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft tissue is desirable, including Neurosurgery, Gastrointestinal and affiliated organ surgery, Urological surgery, Plastic and Reconstructive surgery, General surgery, Orthopedic surgery, Gynecological surgery, Thoracic surgery, Laparoscopic surgery and Thoracoscopic surgery.

PRESCRIPTION USE __ × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use_ (21 CFR 801 Subpart C)

# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David Krone for MXM

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K090864

---

**Source:** [https://fda.innolitics.com/device/K090864](https://fda.innolitics.com/device/K090864)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
