K090636 · Tso3, Inc. · FLF · Dec 2, 2009 · General Hospital
Device Facts
Record ID
K090636
Device Name
TSO3 OZONE STERILIZER, MODEL 125L
Applicant
Tso3, Inc.
Product Code
FLF · General Hospital
Decision Date
Dec 2, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
TSO3 125L Ozone Sterilizer is intended for use in the sterilization processing of reusable medical devices in health care facilities. TSO 125L Ozone Sterilizer is designed for sterilization of both metal and non-metal medical devices at low temperatures. TSO3 125L Ozone Sterilizer has the ability to sterilize successfully packaged rigid and flexible lumen medical devices, including single and multi channel flexible endoscopes such as fiberoptic and video endoscopes.
Device Story
TSO3 125L Ozone Sterilizer is a low-temperature sterilization system for previously cleaned, packaged reusable medical devices. Device uses ozone gas as the sterilant; requires USP grade oxygen, water, and electricity. Operates via a factory-programmed control system; no aeration required post-cycle. Items are cool to touch and ready for immediate use or storage. Used in healthcare facilities; operated by trained personnel. Sterilization process involves exposure to 85 mg/L ozone for 15 minutes at 30.8°C–36.1°C. Benefits include rapid turnaround of heat-sensitive instruments like flexible endoscopes without toxic gas aeration.
Clinical Evidence
Bench testing only. Validation performed using overkill approach, half-cycle, and simulated use conditions to demonstrate sterilization efficacy for flexible endoscopes. Testing achieved a sterility assurance level (SAL) of 10^-6. Load configuration included multi-channel and single-channel endoscopes (total 14 channels) plus 14 additional medical devices (49 lbs total). Material effects, functional compatibility, and biocompatibility were evaluated to confirm safety.
Technological Characteristics
125-liter sterilization chamber; ozone gas sterilant; requires USP grade oxygen, water, and electricity. Compatible with non-woven wrapping, pouches, and anodized aluminum containers. Operates at 30.8°C–36.1°C. Factory-programmed control system. No exhaust ventilation required if room is adequately ventilated. By-products are oxygen and water vapor.
Indications for Use
Indicated for sterilization of reusable metal and non-metal medical devices, including rigid and flexible lumen devices and single/multi-channel flexible endoscopes (e.g., colonoscopes, gastroscopes, bronchovideoscopes, choledofiberscopes, ureteroscopes) in healthcare facilities.
Regulatory Classification
Identification
An ethylene gas sterilizer is a nonportable device intended for use by a health care provider that uses ethylene oxide (ETO) to sterilize medical products.
{0}------------------------------------------------
TSO
### RILIZATION OF FLEXIBLE ENDOSCOPES INTO HE TSO2 OZONE STERILIZER. MODEL 125L
JC090636
## 510(k) Summary
DEC - 2 2009
## Applicant's Name and Address
TSO3 Inc. 2505 avenue Dalton Québec, Qc Canada G1P 3S5
## Contact Person, Telephone, FAX
Marc Chaunet, Director of Quality Assurance and Regulatory Affairs Tel : (418) 651-0003 FAX : (418) 653-5726 E-mail : mchaunet@tso3.com
# U.S. Agent
Charles O. Hancock Inc. Contact person: Charles O. Hancock, RAC Tel : (585) 223-1850 FAX : (585) 223-6855 E-mail : chancock@rochester.rr.com
# Submission Date
March 4, 2009
## Trade Name
TSO3 STERIZONE® 125L Sterilizer
# Common Name
TSO3 125L Ozone Sterilizer
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a chemical structure diagram. The diagram depicts the letters "TSO" with a circle above the "O". The letters are in a simple, sans-serif font and are arranged horizontally. The diagram is simple and lacks any additional details.
### STERILIZATION OF FLEXIBLE EN E TSO3 OZONE STERILIZER. MODEL 125/
## Classification Name
Sterilizer, Chemical
Class II (as per 21CFR, part 880.6860 equivalent device)
# Legally Marketed Equivalent Device Name(s)
TSO3 Ozone Sterilizer, model 125L (K020875) (K051595)
# Description of Device
TSO3 Ozone Sterilizer, model 125L is intended to sterilize medical devices that has been previously cleaned.
The sterilization chamber has a capacity of 125 liters (4.3 cu. ft.).
It requires USP grade oxygen, water and electricity. TSO3 Sterilizer could be installed as a free standing unit or recessed behind the wall. No exhaust gas ventilation duct is required as long as the room is adequately ventilated. By-products are oxygen and low humidity water vapor.
Model 125L is equipped with a unique factory-programmed control system.
Non-woven wrapping material or pouches and anodized aluminum containers are used as packaging for medical devices to be sterilized.
OZO-TEST® self-contained Biological Indicators (B. stearothermophilus) are recommended for use in evaluating cycle performance.
TSO3 Chemical Indicators are available for this process.
No aeration is needed following the sterilization cycle. The sterilized items are cool to the touch and can be removed immediately and used, or stored for future use.
### Effectiveness
Ozone Sterilizer validation testing for sterilization of flexible endoscopes was performed using the « overkill » approach to demonstrate the effectiveness of the process.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a chemical structure diagram. The diagram includes the letters "TS" on the left side. To the right of the letters, there is a chemical structure with circles and lines, representing atoms and bonds. The structure appears to be a simplified representation of a molecule, possibly an organic compound.
## STERILIZATION OF FLEXIBLE ENDOSCOPES INTO THE TSO3 OZONE STERILIZER, MODEL 125L
The effectiveness of the ozone sterilization process in sterilization of flexible endoscopes is demonstrated by the full cycle validation testing in simulated use conditions, half-cycle validation testing and in-use testing.
# Safety
Safety of the TSO3 Ozone Sterilizer, model 125L was demonstrated into the original 510(k) submission (K020875). The content of the actual submittal does not compromise the safety of the device.
The safety of the ozone sterilization process in sterilization of flexible endoscopes is demonstrated by the processed device/material qualification testing which includes material effects, functional compatibility and biocompatibility evaluation.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized depiction of an eagle or bird-like figure with outstretched wings, symbolizing protection and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
TSO3 Incorporated C/O Mr. Charles O. Hancock Regulatory Affairs Consultant Charles O. Hancock Associates, Incorporated 33 Black Watch Trail Fairport. New York 14450
DEC - 2 2009
Re: K090636
Trade/Device Name: TSO3 Ozone Sterilizer, Model 125L Regulation Number: 21 CFR 880.6860 Regulation Name: Ethylene Oxide Gas Sterilizer Regulatory Class: II Product Code: FLF Dated: November 12, 2009 Received: November 12, 2009
Dear Mr. Hancock:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
### Page 2- Mr. Hancock
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Antman Q. muster for
Susan Runner, D.D.S., M.A. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
### Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K090636
Device Name: TSO3 Ozone Sterilizer, model 125L
### Indications for Use:
TSO3 125L Ozone Sterilizer is intended for use in the sterilization processing of reusable medical devices in health care facilities. TSO 125L Ozone Sterilizer is designed for sterilization of both metal and non-metal medical devices at low temperatures.
TSO3 125L Ozone Sterilizer has the ability to sterilize successfully packaged rigid and flexible lumen medical devices, including single and multi channel flexible endoscopes such as fiberoptic and video endoscopes.
Notes:
- · Testing on flexible endoscopes was conducted employing half cycle to demonstrate achievement of a sterility assurance level (SAL) of 10°.
- · During the sterilization cycle, items are exposed to ozone at a concentration of 85 milligrams per liter for 15 minutes at a temperature of 30.8℃ to 36.1℃ (87.4 F to 97 F).
- · Sterilization efficacy was demonstrated for a load comprising packaged multi channel (2) and single channel (1) flexible endoscopes having a total of 14 channels in the presence of other packaged medical devices. The load configuration used for qualification testing contained 2 multi channel and one single channel flexible endoscope, plus 14 medical devices packaged in TSO3 Sterliization Pouches and a Process Challenge Device. Total weight of the load was 49 lbs.
· For the purpose of this submission:
- · Single channel flexible endoscopes refer to medical devices having only one channel.
- · Multi channel flexible endoscopes refer to medical devices that may have two or more channels such as: Working channel (biopsy); Water jet; Air feeding; Water feeding; Suction;
- · Endoscopes evaluated include: Colonoscope, Gastroscope, Broncho videoscope, Choledofiberscope and Ureteroscope.
AND/OR
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
(Division Sign-Off)
510(k) Number: 19063600
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.