← Product Code [GXI](/productcode/GXI) · K090608

# MINTA RYAN RF THERMOCOUPLE ELECTRODE (RYAN-50, 100, 150 AND 200), MINTA INTERMEDIATE CABLE (RYAN-C, -R, -B) (K090608)

_Minta Medical Limited · GXI · Jun 14, 2010 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K090608

## Device Facts

- **Applicant:** Minta Medical Limited
- **Product Code:** [GXI](/productcode/GXI.md)
- **Decision Date:** Jun 14, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4725
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The Minta Radiofrequency Thermocouple Electrodes are indicated for use in the lesioning of peripheral nerve tissue.

## Device Story

Minta Ryan RF Thermocouple Electrode is a temperature-sensing probe used to deliver radiofrequency (RF) energy to peripheral nerve tissue. The device consists of a stainless steel shaft, PET hub, silicone rubber insulated lead, and plug connector. It functions by transferring RF energy to heat surrounding tissue, creating a lesion, while simultaneously measuring tip temperature via an integrated thermocouple. Used in conjunction with compatible RF lesion generators and disposable cannulas, the device is operated by clinicians in a medical setting. The output (temperature data) allows the physician to monitor the lesioning process, facilitating controlled tissue ablation. The device is a re-processed single-use, non-sterile instrument.

## Clinical Evidence

No clinical data. Bench testing only. Functional performance and compatibility with Neurotherm RF lesion generators were validated. Safety and effectiveness are supported by over 15 years of clinical use in the UK and other countries, and in the USA since 2001, without incident.

## Technological Characteristics

Stainless steel shaft, PET hub, non-latex silicone rubber insulated lead, plug connector. Sensing principle: thermocouple temperature measurement. Energy source: external RF lesion generator. Form factor: various lengths and gauges. Sterilization: steam sterilization.

## Regulatory Identification

A radiofrequency lesion probe is a device connected to a radiofrequency (RF) lesion generator to deliver the RF energy to the site within the nervous system where a lesion is desired.

## Predicate Devices

- Neurotherm RF Lesioning System Probes ([K011387](/device/K011387.md))

## Reference Devices

- Pajunk RF disposable cannula ([K060397](/device/K060397.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

-

:

こ

.

K090608

、

.

Page 1 of 3

:

## 510(K) SUMMARY

This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92 (c).

| Owners information:   | Minta Medical Limited<br>Caddick Road<br>Knowsley Business Park<br>Knowsley<br>Prescot<br>Merseyside<br>United Kingdom<br>L34 9HP                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Name:         | Andrew Edwards                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Position:             | Managing Director                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Telephone:<br>E-Mail: | +44(0)151 548 6818<br>andy.edwards@mintamedical.co.uk                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Date prepared:        | 8th July 2009                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Trade Name:           | Minta Ryan RF Thermocouple Electrode                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Common Name:          | Radiofrequency Lesion Probe                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Classification:       | Radiofrequency Lesion Probe<br>CFR882. 4725<br>Product Code: GXI<br>Class 2 Neurology Devices                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Predicate Devices:    | (SAME DEVICE) Neurotherm RF Lesioning System Probes<br>(K011387)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Description:          | The Minta Ryan RF Thermocouple Electrodes have<br>been in use in this format for over fifteen years in the UK<br>and other countries around the world including the<br>U.S.A. previously labelled as Neurotherm (K011387).<br>(Minta Medical Registration No. 3004617090,<br>owner/operator Number 9064097).<br>The Minta Ryan RF Thermocouple Electrodes are used in<br>conjunction with the commercially available Neurotherm<br>(K011387), RF Lesion Generators and Pajunk RF<br>disposable cannula (K060397) to create radiofrequency<br>(RF) lesioning of peripheral nerve tissue. |

{1}------------------------------------------------

The Minta Ryan RF Thermocouple Electrode is a temperature sensor (thermocouple) which is used to deliver RF energy to the tissue.

The RF energy is then transferred from the electrode which heats the surrounding tissue to create a lesion. The thermocouple also measures the temperature at the tip of the electrode.

The Minta Ryan RF Thermocouple Electrodes are provided as non-sterile re-usable devices (re-processed single use).

The Minta Ryan RF Thermocouple Electrode will be available in a variety of lengths and gauges.

The Minta Ryan RF Thermocouple Electrode consists of a stainless steel shaft, a PET (polyethylene terephthalate) hub; a non-latex silicone rubber insulated electrical lead and a plug connector.

#### INTENDED USE

The Minta Ryan RF Thermocouple Electrodes are indicated for use in the lesioning of peripheral nerve tissue.

## TECHNOLOGICAL CHARACTERISTICS

The Minta Ryan RF Thermocouple Electrodes have the same technological characteristics and intended use as the original device which was cleared for marketing in the U.S in 2001 under K011387.

They are the same device (name change only).

#### COMPARISON TO PREDICATE

The Minta RYAN RF Thermocouple Electrode is the same device that was cleared for marketing under K011387.

#### NON-CLINICAL DATA

All non-clinical data for the Minta Ryan RF Thermocouple Electrode was submitted under K011387 to confirm the performance characteristics.

A risk analysis identifying potential hazards and documentary mitigations of the hazards was developed and applied for K011387.

Testing was performed to validate the functional performance of the Minta Ryan RF Thermocouple Electrode. In particular, specific performance and compatibility testing was performed with the Neurotherm RF Lesion Generators to show that the performance was met.

{2}------------------------------------------------

Page 3 of 3

#### STERILISATION

Additional steam sterilisation validation studies were carried out in February 2010 to validate each of the recommended sterilisation processes.

### CONCLUSION

The Minta Ryan RF Thermocouple Electrodes are safe and effective for the application for which they are intended and have been tested to che the 
safety and effectiveness in this formation and have been tested to confirm their safety and effectiveness in this format for over 15 years, without incident in their
UK and other countries, plus the USA since 2001 , without incident in the UK and other countries, plus the USA since 2001.

The fundamental scientific technology of the Minta Ryan RF Thermocouple
Electrode was submitted under K011207 Electrode was submitted under K011387.

SIGNED:

:
**ANDREW EDWARDS**

MANAGING DIRECTOR

**DATED:** 7th May 2010

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three lines representing its wings, body, and legs. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

## JUN 1 4 2010

Minta Medical Limited c/o Mr. Andrew Edwards Managing Director Caddick Road Knowsley Business Park Knowsley Prescot Merseyside United Kingdom L34 9HP

Re: K090608

Trade/Device Name: Minta Ryan RF Therocouple Electrode . Regulation Number: 21 CFR 882.4725 Regulation Name: Radiofrequency Lesion Probe Regulatory Class: Class II Product Code: GXI Dated: June 4, 2010 Received: June 7, 2010

Dear Mr. Edwards:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for are prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be may of saysor de a decessed Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

http://www.fda.gov/MedicalDevices/ResourcestoYou/Industry/default.htm.

Sincerely yours,

Kesia Alexander

Image /page/4/Picture/7 description: The image contains a handwritten word "for" in cursive. The letter "f" has a descending loop that extends below the baseline, and the "o" and "r" are connected with a smooth curve. The writing appears to be done with a pen or marker, and the stroke thickness varies slightly, giving it a natural, handwritten look.

Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Section 4

Sheet 1 of 1

# Indications for Use

### 510(K) Number (if known): K090608

Device Name: Minta Ryan Radiofrequency Thermocouple Electrode

Indications for Use: The Minta Radiofrequency Thermocouple Electrodes are indicated for use in the lesioning of peripheral nerve tissue.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Dani Kaufman

(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

090608 510(k) Number.

---

**Source:** [https://fda.innolitics.com/device/K090608](https://fda.innolitics.com/device/K090608)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
