← Product Code [DSB](/productcode/DSB) · K090602

# CARDIODYNAMICS BIOZ RX, MODEL 7101-SYS (K090602)

_Cardiodynamics International Corp. · DSB · May 28, 2009 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K090602

## Device Facts

- **Applicant:** Cardiodynamics International Corp.
- **Product Code:** [DSB](/productcode/DSB.md)
- **Decision Date:** May 28, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2770
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The BioZ Rx device has the same intended uses as the predicate BioZ Dx device which was intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. The devices are intended to monitor hemodynamic parameters within hospitals and other healthcare facilities providing patient care. The Optional 12-Lead ECG Function of the BioX Rx device is indicated for use where the physician decides to evaluate the clectrocardiogram of adult and pediatric patients as part of decisions regarding possible diagnosis, potential treatment, effectiveness of treatment or to rule-out causes for symptoms.

## Device Story

Noninvasive impedance cardiography (ICG) monitor; measures thoracic electrical bioimpedance via neck and mid-axillary sensors; injects high-frequency, low-amplitude alternating current. Calculates hemodynamic parameters (e.g., stroke volume, cardiac output, cardiac index) based on impedance changes over time. Includes optional 12-lead ECG module. Used in hospitals/healthcare facilities by trained personnel/clinicians. Updated platform features Advantech PC/touch-screen interface and off-the-shelf printer. ICG signal processing remains identical to predicate. Output displayed on screen for clinician review to support hemodynamic assessment and diagnostic decision-making. Benefits include noninvasive monitoring of cardiac function and fluid status.

## Clinical Evidence

Bench testing only. No clinical data. Verification and validation included electrical safety, electromagnetic emissions, and immunity testing per EN60601-1, EN60601-1-1, and EN60601-1-25. System-level software and firmware verification performed.

## Technological Characteristics

Noninvasive ICG using thoracic bioimpedance sensing. PC-based platform with touch-screen interface. Connectivity via USB. ICG Patient Interface Module (PIM) unchanged from predicate. ECG module uses Welch/Allyn USB interface. Standards: EN60601-1, EN60601-1-1, EN60601-1-25. ISO 13485 certified manufacturing.

## Regulatory Identification

An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- CardioDynamics BioZ® Dx Hemodynamic Monitor with 12-Lead ECG ([K070156](/device/K070156.md))
- CardioDynamics BioZ® Dx Hemodynamic Monitor and Philips 12-Lead ECG ([K051228](/device/K051228.md))
- Welch/Allyn Cardioperfect Software with Accessories ([K051228](/device/K051228.md))
- Welch/Allyn Cardioperfect Portable ([K962854](/device/K962854.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## K090602 pg. 1 of 3

## Section 5 - Premarket Notification 510(k) Summary

#### GENERAL INFORMATION:

Device Name and Classification

| Product Name:         | BioZ Rx Hemodynamic Monitor with 12-Lead ECG            |
|-----------------------|---------------------------------------------------------|
| Classification Name:  | Plethysmography, Impedance                              |
| Common Name:          | Impedance Cardiograph (ICG)                             |
| Classification Panel: | Cardiovascular                                          |
| Regulation Number:    | 21 CFR 870.2770 (for ICG) and 21 CFR 870.2340 (for ECG) |
| Device Class:         | II                                                      |
| Product Code:         | 74 DSB (for ICG) and 74 DPS (for ECG)                   |

Manufacturer and Contact Person

CardioDynamics International Corporation 6175 Nancy Ridge Drive, # 300 San Diego, CA 92121 USA Phone: 858-535-0202 x1028 Fax: 866-581-9123

Donald J. Brooks Chief Technology Officer

Date of Summary Preparation: February 27, 2009

#### SUBSTANTIAL EQUIVALENCE:

Predicate Devices:

K070156 - CardioDynamics BioZ® Dx Hemodynamic Monitor with 12-Lead ECG K051228 - CardioDynamics BioZ® Dx Hemodynamic Monitor and Philips 12-Lead ECG K052158 - Welch/Allyn Cardioperfect Software with Accessories K962854 - Welch/Allyn Cardioperfect Portable

#### Device Description:

This submission for the BioZ Rx covers a PC platform/printer/cart change and user software modifications compared to the predicate BioZ Dx System (K070156 and K051228), which is a noninvasive impedance cardiography (ICG) device that provides hemodynamic parameters based on the measurement of thoracic electrical bioimpedance. The BioZ Rx measures this change in impedance by injecting a high frequency, low amplitude alternating electrical current through the thorax between a pair of sensors placed on the neck and another pair placed on the mid-axillary line at the xiphoid process level. By detecting and measuring the change in thoracic impedance as a function of time, the BioZ Rx is able to calculate stroke volume, cardiac output and many other hemodynamic parameters.

The device additionally includes the capability of performing a standard 12-Lead ECG test using previously cleared Welch/Allyn software and USB patient interface accessories (K052158 and K962854).

All ICG signal processing and parameter measurements and calculations are identical to predicate device. Those tasks are performed within the ICG Patient Interface Module (PIM) which is essentially unchanged from the predicate device.

{1}------------------------------------------------

### Intended Use:

The BioZ Rx device has the same intended uses as the predicate BioZ Dx device which was intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. The devices are intended to monitor hemodynamic parameters within hospitals and other healthcare facilities providing patient care.

The Optional 12-Lead ECG Function of the BioX Rx device is indicated for use where the physician decides to evaluate the clectrocardiogram of adult and pediatric patients as part of decisions regarding possible diagnosis, potential treatment, effectiveness of treatment or to rule-out causes for symptoms.

#### Technology:

The BioZ Rx device has the identical ICG Patient Interface Module as the predicate BioZ Dx device, with the exception of:

- A small change made to the USB communication processor in the PIM to enable and . service the watchdog timer to facilitate automatic reconnection and restoration of monitoring in the event of a communication interruption between the PIM and the CPU.
- Some other additional small changes made to facilitate communication with the faster . CPU platform.

The communication processor is not involved in the acquisition and processing of patient hemodynamic data.

The computer platform for the user interface has been changed from the ARM-PC-based Philips Page Writer Trim platform to a self-contained PC and touch-screen monitor manufactured by Advantech, and an off the shelf inkjet printer. Numerous human factors improvements has been made to the user software, facilitated by the significantly larger touch-screen interface of the new platform. The ECG Patient Interface Module has been replaced by an equivalent device (also USB-based) and software from Welch/Allyn, Inc., previously cleared under K052158 and K962854.

#### General Safety & Effectiveness Concerns:

The instructions for use for this device contain the necessary cautions and warnings to provide for safe and effective use for the device.

Risk management is an essential element of the design and development process, and hazard analysis assessments performed on the BioZ Rx device at the System and Software levels revealed that the enhancements present no additional risks from the predicate devices. The Software and System Hazard Analyses for the modified device are included in Section 18.

#### Manufacturing:

CardioDynamics International Corporation is presently ccrtified to ISO13485 2003 as of 2/17/09 by NEMKO AS. No significant audit findings have been presented since initial ISO 9004 (now 13485) certification in May of 1998. All products are designed and manufactured under CDIC's Quality Management System, which includes Design Control and Good Manufacturing Practices.

{2}------------------------------------------------

### Performance Testing:

There have been no hardware, cable, or signal processing modifications in the patient interface portion of the BioZ Rx compared to the predicate BioZ Dx monitor device. The ICG changes are limited to the computer platform for the user interface and the use of an off-the-shelf printer.

The Optional 12-lead ECG function uses a USB-based ECG Patient Interface Module like the predicate device. The supplier has been changed from Philips to Welch/Allyn. Both the predicate device and the ECG device and software of the BioZ Rx are tested and certified to comply with the relevant recognized consensus standards, including EN60601-1, EN60601-1-1, and EN60601-1-25. The issued Certificates of Conformity to these standards are included in Section 9. We are awaiting final agency processing on the CB Scheme report at the system level. Testing of the Welch/Allyn patient interface and software at the system level is included in the testing documented in Scction 18. The new user interface platform, including the cart, printer, ICG and ECG interface modules and all cables have been tested by NEMKO and TUV for electrical safety, electromagnetic emissions, and electromagnetic immunity to the appropriate International consensus standards. The test reports are included in Section 18 and the Certificates of Conformity are included in Section 9.

#### Design Controls

Design Review is required, at a minimum, at the completion of cach step in the design control process (i.e. Design Input, Design Output, Design Verification, Design Validation and Design Transfer). The design review process identifies the required deliverables for cach step in the process and verifies the content and accuracy of each deliverable.

The following quality assurance measures were conducted for the modified BioZ Rx device and are included in this submission:

- . Risk analysis
- . Design requirements and traceability
- Unit and system level software and firmware verifications .
- System level validations .

The results of verification and validation tests concluded that the functionality and performance characteristics of the modified BioZ Dx are comparable to the currently marketed predicate devices.

#### Conclusion:

The results of all testing demonstrate that the BioZ Rx Hemodynamic Monitor with Optional 12-Lead ECG does not raise any new significant issues of safety, effectiveness or performance of the device when compare to the existing predicate devices.

{3}------------------------------------------------

Public Health Service

Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular pattern around the symbol.

MAY 28 2009

ood and Drug Administration 200 Corporate Boulevard Rockville MD 20850

Cardiodynamics International Corporation C/O Mr. Don Brooks 6175 Nancy Ridge Drive, #300 San Diego, CA 92121

Re: K090602

Trade/Device Name: BioZ Rx HemoDynamic Monitor with Optional 12-lead ECG Regulation Number: 21 CFR 870.2770 Regulation Name: Impedance Plethysmograph Regulatory Class: Class II Product Codes: DSB Dated: February 27, 2009 Received: March 5, 2009

Dear Mr. Brooks:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 - Mr. Don Brooks

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Bram D. Zuckerman, M.D.

Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

K090602 510(k) Number (if known):

Device Name: BioZ Rx HemoDynamic Monitor with Optional 12-lead ECG Indications For Use:

#### For the ICG function:

The BioZ Rx Hemodynamic Monitor is intended to monitor and display a patient's hemodynamic parameters. These parameters include:

Heart Rate (HR) Systolic Blood Pressure (SBP) Diastolic Blood Pressure (DBP) Mean Arterial Blood Pressure (MAP) Stroke Index (SI) Stroke Volume (SV) Cardiac Index (CI) Cardiac Output (CO) Systemic Vascular Resistance (SVR) Systemic Vascular Resistance Index (SVRI) Left Cardiac Work (LCW) Acceleration Index (ACI) Velocity Index/Index of Contractility (VI, IC) Thoracic Fluid Content (TFC)

Thoracic Fluid Content Index (TFCI) Systolic Time Ratio (STR) Systolic Time Ratio Index (STRI) Pre-Ejection Period (PEP) Left Ventricular Ejection Time (LVET) Total Arterial Compliance (TAC) Total Arterial Compliance Index (TACI) Left Stroke Work Index (LSWI) Heather Index (HI) Q-C Interval (QC) Left Cardiac Work Index (LCWI) Systemic Stroke Resistance Index (SSRI/SSVRI) Base Impedance (TFI or Za) Electrocardiogram (ECG)

For the optional 12-lead ECG function:

Where the clinician decides to evaluate the electrocardiogram of adult and pediatric patients as part of decisions regarding possible diagnosis, potential treatment, effectiveness of treatment or to ruleout causes for symptoms.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

h

Division Sign-Off) n of Cardiovascular Devices K090602 510(k) Number

---

**Source:** [https://fda.innolitics.com/device/K090602](https://fda.innolitics.com/device/K090602)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
