POWER PROCESSOR SAMPLE PROCESSING SYSTEM WITH GENERIC CONNECTION MODULE, MODEL: 4210
K090588 · Beckman Coulter, Inc. · CDD · May 27, 2009 · Clinical Chemistry
Device Facts
Record ID
K090588
Device Name
POWER PROCESSOR SAMPLE PROCESSING SYSTEM WITH GENERIC CONNECTION MODULE, MODEL: 4210
Applicant
Beckman Coulter, Inc.
Product Code
CDD · Clinical Chemistry
Decision Date
May 27, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1810
Device Class
Class 2
Indications for Use
The basic Power Processor is an automated sample handling system which processes sample tubes from the pre-centrifugation, pre-sorting step to presentation of centrifuged and decapped samples into Generic or Personality Racks for specific instruments. The Power Processor can be configured with optional software and hardware to allow processing of sample tubes on Generic Connection Instruments. The Power Processor performs all pre-analytical sample tube preparation, and then sorts the sample tubes directly to Generic Connection Modules where the samples are pipetted by the Generic Connection instrument for testing. After the samples are pipetted, the tubes can route to other instruments for additional testing or to Outlet Racks. The UniCel Dxl 800 Access Immunoassay System with laboratory automation connection is a microcomputer-controlled, random and continuous access analyzer that includes an external computer. This computer stores the system user interface (UI) software and allows the operator to interface with and direct the instrument software. The UniCel Dxl 800 System uses enzyme immunoassays (utilizing paramagnetic particle solid phase and chemiluminescent detection) for determination of various analytes, such as Vitamin B12, Ferritin, Folate and hTSH along with other various enzyme immunoassays assays that may be adaptable to the analyzer depending on the reagent used to induce the enzyme immunoassay reaction. The UniCel Dxl 800 System is an in vitro diagnostic device for use in the clinical laboratory. The Access Ferritin assay is a paramagnetic particle, chemiluminescent assay for the quantitative determination of ferritin levels in human serum and plasma (heparin) using the Access Immunoassay Systems. Measurements of ferritin aid in the diagnosis of diseases affecting iron metabolism. The Access Folate assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of folic acid levels in human serum, plasma (heparin) and red blood cells using the Access Immunoassay Systems. Folic acid measurements are used in the diagnosis and treatment of megaloblastic anemia. The Access HYPERsensitive hTSH assay is a paramagnetic particle, chemiluminescent assay for the quantitative determination of human thyroid-stimulating hormone (thyrotropin, hTSH) levels in human serum using the Access Immunoassay Systems. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders. The Access Vitamin B12 assay is a paramagnetic particle, chemiluminescent assay for the quantitative determination of vitamin B12 in human serum and plasma (heparin) using Access Immunoassay Systems. Measurements obtained by this device are used in the diagnosis and treatment of anemias of gastrointestinal malabsorption.
Device Story
Power Processor is a scalable laboratory automation system (LAS) for pre-analytical sample preparation; automates centrifugation, decapping, and sorting of patient sample tubes. Integrates with UniCel DxI 800 Immunoassay System via Generic Connection Module using outboard sampling; system reads barcodes to identify samples and route them based on reagent/calibration status. Employs point-in-space pipetting technology for direct aspiration from tubes on the track. System software (PrepLink, Line Control) manages sample tracking and routing logic. Benefits include reduced biohazard exposure for lab personnel and streamlined workflow. Healthcare providers use the resulting immunoassay data for diagnosis of iron metabolism, anemia, and thyroid disorders.
Clinical Evidence
Bench testing only. Methods comparison studies evaluated a representative connected system using on-line sample preparation against an unconnected analyzer using off-line centrifugation. Results demonstrated good correlation between the Power Processor with GC module/UniCel Dxl 800 connection and the stand-alone UniCel Dxl 800 System.
Technological Characteristics
Scalable LAS with open architecture; modules include inlet, centrifugation, decapper, and outlet. Employs outboard sampling via point-in-space pipetting technology per CLSI guidelines. Connectivity via track system and LAS interface. Software-controlled routing and sample identification. System utilizes paramagnetic particle solid phase and chemiluminescent detection for immunoassays.
Indications for Use
Indicated for clinical laboratory use to automate pre-analytical sample processing (centrifugation, decapping, sorting) and facilitate automated immunoassay testing (Ferritin, Folate, hTSH, Vitamin B12) in human serum, plasma, and red blood cells for patients requiring diagnosis of iron metabolism disorders, megaloblastic anemia, thyroid/pituitary disorders, or anemias of gastrointestinal malabsorption.
Regulatory Classification
Identification
A vitamin B12 test system is a device intended to measure vitamin B12 in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of anemias of gastrointestinal malabsorption.
Predicate Devices
Siemens (Dade Behring) StreamLAB® Analytical WorkCell/Sample Transfer Module (k043546)
Submission Summary (Full Text)
{1}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k090588
B. Purpose for Submission:
This is a new 510(k) to support the integration of a Power Processor Sample Processing System to the UniCel DxI 600/800 Access Immunoassay System (k023764) using cleared Vitamin B12, Ferritin, Folate and TSH Access Immunoassays to show acceptable performance.
C. Measurand:
Vitamin B12, Ferritin, Folate and TSH
D. Type of Test:
Quantitative Chemiluminescent Immunoassay
E. Applicant:
Beckman Coulter, Inc.
F. Proprietary and Established Names:
Power Processor Sample Processing System with Generic Connection Module, Model: 4210
G. Regulatory Information:
1. Regulation section:
21CFR Sec.-862.1810 - Vitamin B12 test system.
21CFR Sec.-866.5340 - Ferritin immunological test system.
21CFR Sec.-862.1295 - Folic acid test system.
21CFR Sec.-862.1690 - Thyroid stimulating hormone test system.
21CFR Sec.-862.2160 - Discrete photometric chemistry analyzer for clinical use.
2. Classification:
II, II, II, II, I respectively
3. Product code:
CDD - Radioassay, Vitamin B12
DBF - Ferritin, Antigen, Antiserum, Control
CGN - Acid, Folic, Radioimmunoassay
JLW - Radioimmunoassay, Thyroid-Stimulating Hormone
JJE - Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
4. Panel:
Chemistry (75) Immunology (91)
{2}
H. Intended Use:
1. Intended use(s):
See indication(s) for use below.
2. Indication(s) for use:
The basic Power Processor is an automated sample handling system which processes sample tubes from the pre-centrifugation, pre-sorting step to presentation of centrifuged and decapped samples into Generic or Personality Racks for specific instruments. The Power Processor can be configured with optional software and hardware to allow processing of sample tubes on Generic Connection Instruments. The Power Processor performs all pre-analytical sample tube preparation, and then sorts the sample tubes directly to Generic Connection Modules where the samples are pipetted by the Generic Connection instrument for testing. After the samples are pipetted, the tubes can route to other instruments for additional testing or to Outlet Racks.
The UniCel DxI 800 Access Immunoassay System with laboratory automation connection is a microcomputer-controlled, random and continuous access analyzer that includes an external computer. This computer stores the system user interface (UI) software and allows the operator to interface with and direct the instrument software. The UniCel DxI 800 System uses enzyme immunoassays (utilizing paramagnetic particle solid phase and chemiluminescent detection) for determination of various analytes, such as Vitamin B12, Ferritin, Folate and hTSH along with other various enzyme immunoassays assays that may be adaptable to the analyzer depending on the reagent used to induce the enzyme immunoassay reaction. The UniCel DxI 800 System is an in vitro diagnostic device for use in the clinical laboratory.
The Access Ferritin assay is a paramagnetic particle, chemiluminescent assay for the quantitative determination of ferritin levels in human serum and plasma (heparin) using the Access Immunoassay Systems. Measurements of ferritin aid in the diagnosis of diseases affecting iron metabolism.
The Access Folate assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of folic acid levels in human serum, plasma (heparin) and red blood cells using the Access Immunoassay Systems. Folic acid measurements are used in the diagnosis and treatment of megaloblastic anemia.
The Access HYPER sensitive hTSH assay is a paramagnetic particle, chemiluminescent assay for the quantitative determination of human thyroid-stimulating hormone (thyrotropin, hTSH) levels in human serum using the Access Immunoassay Systems. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders.
{3}
The Access Vitamin B12 assay is a paramagnetic particle, chemiluminescent assay for the quantitative determination of vitamin B12 in human serum and plasma (heparin) using Access Immunoassay Systems. Measurements obtained by this device are used in the diagnosis and treatment of anemias of gastrointestinal malabsorption.
3. Special conditions for use statement(s): Prescription use
4. Special instrument requirements: UniCel DxI 800 Access Immunoassay System
I. Device Description:
The Power Processor is a scalable laboratory automation system (LAS) designed to streamline pre-analytical processes in the clinical laboratory. A basic Power Processor System is comprised of a Line Control Computer, Preplink™ Computer, Inlet Module, Hematology Module, Centrifugation Module, Decapper Module, and Outlet Module. In the basic configuration, patient sample tubes are loaded onto the Power Processor system to be sorted to a Hematology Module, or to be centrifuged, decapped, and sorted to Personality Racks for further processing on other instruments. Additional modules may be added for aliquot capability, sample capping, and refrigerated storage.
The Power Processor is an open architecture system that can connect to a variety of clinical analyzers. Connection modules are extensions of the Power Processor track system that link with an analyzer's existing LAS interface. Connection modules support one of two types of sample transfer methods: onboard or outboard sampling. Onboard sampling physically transfers the sample tube/rack from the automation track to the analyzer's sample load and identification area. With outboard sampling, the connection unit performs the sample bar code read function, presents the sample ID to the connected analyzer, and then signals for direct sampling of the open tube by the connected instrument at an aspiration point on the automation track. The Power Processor Generic Connection Module is specifically designed to support the outboard sampling method based on point-in space pipetting technology aligned with the CSLI guidelines. This method is used to establish connection with Beckman Coulter's UniCel DxI 800 Immunoassay System. Power Processor software version 3.5 establishes a dynamic or "smart" connection with the UniCel DxI 800 System to enable sample routing based on reagent and calibration status.
J. Substantial Equivalence Information:
1. Predicate device name(s):
| Predicate Device | 510(k) Number |
| --- | --- |
| Siemens (Dade Behring) StreamLAB® Analytical WorkCell/Sample Transfer Module | k043546 |
{4}
2. Predicate 510(k) number(s):
See Predicate Device table above.
3. Comparison with predicate:
| Aspect/ Characteristic | Device | Predicate |
| --- | --- | --- |
| Intended Use | The basic Power Processor is an automated sample handling system which processes sample tubes from the pre-centrifugation, pre-sorting step to presentation of centrifuged and decapped samples into Generic or Personality Racks for specific instruments. The system is designed to free laboratory personnel from biohazard exposure and routine sample preparation. The Power Processor can be configured with optional software and hardware to allow processing of sample tubes. The Power Processor performs all preanalytical sample tube preparation, then sorts the sample tubes directly to Generic Connection Modules where the samples are pipetted for testing. After the samples are pipetted, the tubes can route to other instruments for additional testing or to Outlet Racks. | Same: Automated sample handling, routing, and management system designed to increase productivity and improve operator safety. |
| System Design | Open LAS architecture enables multiple analyzer connections. Scalable, modular configuration. | Same |
| Sample Transfer Method | Outboard sampling capability using the analyzer's existing LAS interface and Generic Connection Module. | Same, using STM device. |
| Fundamental Technology | Outboard analyzer connections are based on "Point-in-space" pipetting technology aligned with CSLI guidelines. | Same |
| Operating Environment | Operating Software, Computer Console with Single User Interface | Same |
| System | The core Power Processor System is | Control unit, operator |
{5}
| Aspect/ Characteristic | Device | Predicate |
| --- | --- | --- |
| Modules | comprised of a line controller computer, a system console with PrepLink™ software, inlet module, hematology module, centrifugation module, decapper module, track transport system, and output module. The Generic Connection module is an optional module to enable analyzer connectivity. | interface, sample input/output module, decapping module, track transport module. The centrifugation module is optional, and the STM device is an optional module to enable analyzer connectivity. |
| Sample Identification | Identification of patient tubes and sample programming using bar codes. Does not use Radio Frequency Identification Device (RFID) sample identification. | Positive sample identification via a barcode reader and RFID chip. |
| Sample Handling | The ability to interface with a LIS device to receive patient identification and test requests via a communications protocol to provide sample tracking via bar code labeling. | Same |
K. Standard/Guidance Document Referenced (if applicable):
- IEC 61326-2-6 IEC: Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 2-6: Particular requirements - In vitro diagnostic (IVD) medical equipment. 12/15/2005
- AUTO5-A CSLI: Laboratory Automation: Electromechanical Interfaces; Approved Standard. 3/1/2001
L. Test Principle:
Paramagnetic particle solid phase and chemiluminescent detection with an automated sample handling system
M. Performance Characteristics (if/when applicable):
Assay performance characteristics were established in Beckman Coulter Access® Ferritin Assay k926221, Beckman Coulter Access® Folate Assay k060774, Beckman Coulter Access® HYPERsensitive hTSH Assay k042281, Beckman Coulter Access® Vitamin B12 Assay k955436.
1. Analytical performance:
a. Precision/Reproducibility: Provided in above referenced assay 510(k)'s
b. Linearity/assay reportable range: Provided in above referenced assay 510(k)'s
c. Traceability, Stability, Expected values (controls, calibrators, or methods): Provided in above referenced assay 510(k)'s
{6}
d. Detection limit:
Provided in above referenced assay $510(\mathrm{k})$ 's
e. Analytical specificity:
Provided in above referenced assay $510(\mathrm{k})$ 's
f. Assay cut-off:
Provided in above referenced assay $510(\mathrm{k})$ 's
# 2. Comparison studies:
a. Method comparison with predicate device:
In order to establish equivalence to manually introduced samples to the analyzer verses automated sample introduction and identification to the analyzer a method comparison was performed using the above referenced assays. The method comparison study employed paired sample testing and Deming regression analysis to assess the data. The assay test menu was selected by the evaluator and specimen selection was based on the site test requests. In the study, two blood samples were drawn from each patient. One sample tube was processed through the laboratory protocol using the Power Processor System with Generic Connection Module interface to the DxI 800 analyzer, and the other tube was processed on the AutoMate system and manually loaded on the analyzer.
The Deming regression analysis demonstrates comparable performance across the range of sample concentrations tested for each representative assay.
| Dxl assay (manual sample processing) | Sample Range | N | Slope | Intercept | r | Comparative Method |
| --- | --- | --- | --- | --- | --- | --- |
| Vitamin B12 | 94 -671 | 101 | 0.912 | 23.88 | 0.983 | Dxl 800 Analyzer w/ automated sample processing (Power Processor w/ Generic Connection Module) |
| Ferritin | 3 - 993 | 101 | 1.042 | -4.37 | 0.995 | |
| Folate | 11 - 42 | 85 | 0.944 | 1.19 | 0.982 | |
| Hypersensitive TSH | 0.46 – 12.09 | 104 | 1.005 | 0.01 | 0.994 | |
The sponsor's evaluator determined the test results showed no clinically significant difference in assay performance between the two sample processing methods. This data demonstrates substantial equivalence between DxI 800 System with Power Processor connection and the predicate stand-alone analyzer.
{7}
b. Matrix comparison:
Provided in above referenced assay 510(k)'s
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable
b. Clinical specificity:
Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Provided in above referenced assay 510(k)'s
4. Clinical cut-off:
Not Applicable
5. Expected values/Reference range:
Provided in above referenced assay 510(k)'s
N. Instrument Name:
Power Processor Sample Processing System with Generic Connection Module
O. System Descriptions:
1. Modes of Operation:
The Power Processor system can consist of the following interfaces: Instrument Manager, PrepLink, and Line Control Computer. The LIS sends sample programming to the PrepLink through the optional Instrument Manager in some software setups. Sample programming information is used to determine where to sort sample tubes. The Instrument Manager consolidates multiple sample programming requests for each sample tube, allowing the Power Processor system to sort sample tubes to the appropriate racks.
The PrepLink receives sample programming information which is used to determine the routing information when queried by the Line Control Computer. PrepLink sorts sample tubes to racks based on sample programming and rack configurations specified by the operator during setup.
The Line Control software communicates with the PrepLink software to keep track of sample tubes on the Power Processor system. The Line Control software sends and receives information from the PrepLink software and the Power Processor track. This information is contained in the Systems Error Log and the Operations Error Log (both can be used to evaluate and correct processing problems). The Line Control Computer Main screen displays a graphical representation of the automation line.
{8}
The Power Processor can be configured with optional hardware to automatically load centrifuged and decapped sample tubes onto Beckman Coulter's 800 DxI System. The Connection Modules operate identically and are designed to support an onboard sampling mechanism, where the sample tubes are transferred directly to the connected analyzer sample load area and read by the system barcode reader.
The Power Processor can be configured with optional software and hardware to allow processing of sample tubes on Connection Instrument. The Power Processor performs all pre-analytical sample tube preparation, and then sorts the sample tubes directly to Connection Module where the samples are pipetted for testing. After the samples are pipetted, the tubes can route to other instruments for additional testing or to Outlet Racks. Routing decisions are made by the PrepLink, and are based on the sample programming of each sample tube, the availability of the Connection instrument, and configuration parameters set by the operator.
2. Software:
FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:
Yes ☐ X or No ☐
The applicant provided software documentation that supports the device was designed and developed under good software LifeCycle processes.
3. Specimen Identification:
Bar code
4. Specimen Sampling and Handling:
On the DxI 800 connection, the Connection Module reads the sample bar code read prior to releasing the sample from the aspiration point, resulting in the following sequence from the DxI perspective:
- Read sample data
- Verify sample data matches software queue
- Aspirate sample from tube
- Read sample data again
- Does sample data read before and after aspiration match? If Yes, keep sample and perform analysis; if No, discard sample and do not perform analysis
- Release Sample from Aspiration point
5. Calibration:
Provided in k023764
{9}
6. Quality Control:
Provided in k023764
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.