BOVIE ICON GS ELECTROSURGICAL GENERATOR (GS100), HANDPIECE (GS-OP-R), REUSABLE HANDPIECE CABLE (GS807-RC)

K090586 · Bovie Medical · GEI · Aug 7, 2009 · General, Plastic Surgery

Device Facts

Record IDK090586
Device NameBOVIE ICON GS ELECTROSURGICAL GENERATOR (GS100), HANDPIECE (GS-OP-R), REUSABLE HANDPIECE CABLE (GS807-RC)
ApplicantBovie Medical
Product CodeGEI · General, Plastic Surgery
Decision DateAug 7, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Bovie ICON GS system is an electrosurgical device that utilizes Helium gas for the coagulation of soft tissues during open soft tissue surgery.

Device Story

Bovie ICON GS is an electrosurgical generator system using high-frequency RF energy and Helium gas to create an ionized plasma flame for soft tissue coagulation. The system comprises a generator, accessory cable, and a single-use handpiece. The handpiece features a finger-activation button and an integrated programmable memory chip requiring a unique code for generator operation. During use, the generator ionizes Helium gas flowing through the handpiece tip to produce a plasma flame. The generator allows adjustment of RF output intensity, duty cycle, and gas flow. Used in open surgery settings by clinicians, the device provides controlled coagulation. The user interface screen displays settings for clinician monitoring. The system is designed to comply with medical electrical equipment safety and electromagnetic compatibility standards.

Clinical Evidence

No clinical data provided; substantial equivalence is based on technological characteristics and bench testing of the plasma output and safety compliance.

Technological Characteristics

Electrosurgical generator utilizing high-frequency RF energy and Helium gas. Features a single-use handpiece with a programmable memory chip for device identification/compatibility. User interface screen for setting selection. Complies with medical electrical equipment safety and electromagnetic compatibility standards. Sterilization method not specified.

Indications for Use

Indicated for coagulation of soft tissues during open soft tissue surgery.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K090586 # 510(k) SUMMARY (As Required per 21 CFR 807.92(c)) ### GENERAL INFORMATION: | 510k Owner's Name | Bovie Medical | |-------------------|-----------------------------------------------------------------------------------------------------------------------------------------| | Address | 3200 Tyrone Boulevard, Suite A<br>St. Petersburg, Florida 33710-2902 | | Contact Person | Richard A. Kozloff<br>Vice-President; Quality Assurance/Regulatory Affairs<br>Telephone #: (727) 803-8513<br>FAX Number: (727) 347-9144 | | Date Prepared: | July 1, 2009 | ## DEVICE DESCRIPTION: #### Trade Name: Bovie ICON GS General Purpose Electrosurgical Generator, Accessory Cable and Handpiece ## Common Name: Electrosurgical Generator ## Classification Name: Electrosurgical Cutting and Coagulation Devices and Accessories (21CFR 878.4400; Class II; Product Code GEI) ## Predicate Devices: Tenerco, Inc. Helica TC Thermo Coagulator (K972267) Plasma Surgical Ltd. PlasmaJet System (K030819) {2}------------------------------------------------ ## 510(k) SUMMARY (As Required per 21 CFR 807.92(c)) ### DEVICE OPERATION: The Bovie ICON GS (the Generator) operates by utilizing high frequency radiofrequency (RF) energy to create an ionized gas plasma channel which, when used in conjunction with the handpiece accessory, is used coagulate tissues. The plasma output is comparable between the generator and predicate device. The Generator has an RF mode of operation that can vary the output intensity, duty cycle and Helium gas flow to control the coagulation intensity. The generator is designed to comply with applicable Medical Electrical Equipment safety standards, including electromagnetic compatibility and other safety standards. The Generator uses technology substantially equivalent to the predicate devices. The generator incorporates an ergonomically designed user interface screen for the selection of device settings. Although different from the user interface of the predicate devices, the difference does not affect the safety and effectiveness and may provide improved visualization of the device settings. The Handpiece is a single use device designed to only be compatible with the ICON GS generator and cable. This handpiece contains a programmable memory device that requires a unique code for the generator to operate. The handpiece has a button for finger activation of the generator. Once activated, helium gas flows out the tip of the device and is excited by the high electric field resulting in a plasma flame. There are no new hazards presented with the use of the Bovie ICON GS generator and handpiece as compared with the named predicate device. #### INTENDED USE: "The Bovie ICON GS system is an electrosurgical device that utilizes Helium gas for the coagulation of soft tissues during open soft tissue surgery." {3}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with three stripes forming its body and wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. #### Public Health Service AUG 0-7 2009 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Bovie Medical % Mr. Richard Kozloff Vice-President. Ouality Assurance and Regulatory Affairs 7100 30th Avenue North Saint Petersburg, Florida 33710-2902 Re: K090586 Trade/Device Name: Bovie ICON GS Generator and Handpiece Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: July 1, 2009 Received: July 2, 2009 Dear Mr. Kozloff: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ ## Page 2 - Mr. Richard Kozloff If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/McdicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, urs, Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use 510(k) Number (if known): Bovie ICON GS Generator and Handpiece Device Name: Indications for Use: "The Bovie ICON GS system is an electrosurgical device that utilizes Helium gas for the coagulation of soft tissues during open soft tissue surgery." Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Hhmnil Gayer (Division Sign-Off) (Division Sign=On), Orthopedic, nd Restorative Devices 510(k) Number K090586
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...