K090532 · Vision Quest Industries, Inc. · IPF · Nov 13, 2009 · Physical Medicine
Device Facts
Record ID
K090532
Device Name
T.E.A.R. TECH3
Applicant
Vision Quest Industries, Inc.
Product Code
IPF · Physical Medicine
Decision Date
Nov 13, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The T.E.A.R. Tech3 IF/Muscle Stimulator is intended for the following applications: The High Volt Pulsed Current Stimulation and Neuromuscular Electrical Stimulation can be used in the following applications: - Relaxation of muscle spasms - . Prevention or retardation of disuse atrophy - . Increases local blood circulation - . Muscle re-education - . Maintaining or increasing range of motion - Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis Interferential Stimulation can be used in the following applications: - Symptomatic relief of post traumatic acute pain - Symptomatic relief of chronic pain intractable pain - . Relaxation of muscle spasms - Maintaining or increasing range of motion - Increases local blood circulation Pulsed Direct Current stimulation can be used in the following applications: - Reduction of edema (under negative electrode) - . Relaxation of muscle spasm - Increasing local blood circulation - Retardation or prevention of disuse atrophy - Muscle re-education - Maintaining or increasing of range of motion
Device Story
T.E.A.R. Tech 3 is a multi-modal electrical stimulator combining Interferential (IF), Neuromuscular (NM), High Voltage (HV), and Pulsed Direct Current (PDC) stimulation. Device delivers low electrical current via lead wires to electrodes placed on patient skin by a clinician. System features LCD display and keypad for user interface. Clinicians adjust operating parameters; end-users limited to protocol selection and amplitude. Used in clinical settings to manage pain, improve circulation, and support muscle rehabilitation. Output affects clinical decision-making by providing therapeutic electrical stimulation to target tissues, potentially benefiting patients through pain relief, edema reduction, and muscle function maintenance.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics and intended use comparisons to predicate devices.
Technological Characteristics
Multi-modal electrical stimulator (IF, NM, HV, PDC). Features LCD display and keypad interface. Delivers electrical current via lead wires and electrodes. Powered muscle stimulator (21 CFR 890.5850).
Indications for Use
Indicated for patients with chronic or acute pain syndromes and/or recovery from orthopedic surgery. Used for muscle spasm relaxation, disuse atrophy prevention, local blood circulation increase, muscle re-education, range of motion maintenance, post-surgical calf stimulation to prevent venous thrombosis, and edema reduction.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
K090532
## 510(k) SUMMARY
| 510(k) Owner: | Vision Quest Industries, Inc.<br>18011 Mitchell South,<br>Irvine, CA, 92614<br>Ph. (800) 266-6969<br>Fax (800) 821-8012 | NOV 13 2009 |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------|-------------|
| Contact: | Doug Humphrey<br>Dir. QA and Regulatory Affairs | |
| Date Summary Prepared: | May 22, 2009 | |
| Proprietary Name: | T.E.A.R. Tech 3 | |
| Device Type: | Interferential Stimulator<br>High Volt Pulsed Current Stimulator<br>Neuromuscular Electrical Stimulator | |
| Substantial Equivalence: | This device is similar in design, composition<br>and function to the following devices: | |
| Manufacturer | Product Name | 510(k) Number |
|-------------------------------------------------------------------------------------------------|----------------|---------------|
| Vision Quest Industries, Inc.<br>18011 Mitchell South<br>Irvine, CA 92614 | T.E.A.R. Tech3 | K030507 |
| Staodyn, Inc.<br>1225 Florida Avenue<br>P. O. Box 1379<br>Longmont, Colorado 80502-1379 | Tuwave | K921668 |
| Compex Technologies, Inc.<br>(formerly Staodyn)<br>1811 Old Highway 8<br>New Brighton, MN 55112 | IF3 Wave | K050046 |
Device Description:
The T.E.A.R. Tech 3 IF/Muscle Stimulator is a device which combines the functionality of Interferential (IF), Neuromuscular (NM), High Voltage (HV) and Pulsed Direct Current (PDC) in one device. The device produces a low electrical current that is
{1}------------------------------------------------
transmitted via lead wires to electrodes placed on the area predetermined by a clinician. Operating parameters can be adjusted throughout their range by a trained clinician but the end-user is limited to protocol selection and amplitude. The user interface consists of an LCD display and keypad.
Indications for Use:
The T.E.A.R. Tech3 IF/Muscle Stimulator is intended for the following applications:
The High Volt Pulsed Current Stimulation and Neuromuscular Electrical Stimulation can be used in the following applications:
- Relaxation of muscle spasms
- . Prevention or retardation of disuse atrophy
- . Increases local blood circulation
- . Muscle re-education
- . Maintaining or increasing range of motion
- Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
Interferential Stimulation can be used in the following applications:
- Symptomatic relief of post traumatic acute pain
- Symptomatic relief of chronic pain intractable pain
- Relaxation of muscle spasms
- Maintaining or increasing range of motion
- Increases local blood circulation
Pulsed Direct Current stimulation can be used in the following applications:
- Reduction of edema (under negative electrode)
- . Relaxation of muscle spasm
- Increasing local blood circulation
- Retardation or prevention of disuse atrophy
- Muscle re-education
- Maintaining or increasing of range of motion
The target population for this device is patients with chronic, acute pain syndrome, and/or recovering from orthopedic related surgery.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features the department's name arranged in a circular pattern around the perimeter. In the center of the seal is a stylized graphic of an eagle, which is a common symbol of the United States.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609
Silver Spring MD 20000 Silver Spring MD 20000 Silver Spring, Control Room W-C
Silver Spring, MD 20993-00022
Vision Quest Industries, Inc. % Mr. Wallace Fischer 18011 Mitchell South Irvine, California 92614
Re: K090532
NOV 1:3 2009
Trade/Device Name: T.E.A.R. Tech3 Regulation Number: 21 CFR 890.5850 Regulation Name: Powered muscle stimulator
Regulation Name: Powered muscle stimulator Regulatory Class: II Product Code: IPF, LIH Dated: October 13, 2009 Received: October 16, 2009
Dear Mr. Fischer:
We have reviewed your Section 510(k) premar
We have reviewed your Section 510(k) premarket notification of intent to market the device i
referenced above and have determined the device is substantially accuracy of the referenced your Section 5 IQ(x) premarket notification of intent to market the device
for use stated in the enclosure is substantially equivalent (for the indications for use as associated in the device is substantial of market the device the device on the world in the chooder to legally marked produced double smarketings
devices that have been reclicate devices mexical devices markets marketstand
devices that have bee devices that have been redastiled in accordine and of the American in interstate
and Cosmetic Act (at in accordance we of the Porculical Derhe Pood, Drug, to
r ou may thet (A and Commit Act (Act (Act the device appre with the provisions of the Federal Food, your and Food, your and Food, your and Food, your and Food, your and your and You may, therefore, make the device, subject to all of a permaker approval approval operiod.
general controls provide approval of a premaket approval of the Act. The a more of the cole, thanks the device, subject to the general controls of the optisation of the optisation of the optisions of the of the of the of the of the of f
edulterat the Actuations provisions of the Act include requirements for annual registration, little Artistian
devices, good manufacturing practice, labeling, and prohibitions against m
If your device is classified (see above)
If your device is classified (see above) into either class II (Special Controls) or class III (PMA),
it may be subject to additional controls. Existing major regulations a ing and of too is eassified isee above) into either class II (Special Controls) or class III (PMA),
it may be subject o additional controls. Existing major regulations affe and in the Code of Federal Regulations, Title 21, 2014 of the Solars (11 (PM) 2017) of class III (PM)
found in the Code of Federal Regulations, Title 21, Dr. addition, FDA ma publish further announcements concerning your device in the Reduble Sept. In additions
Plansa has device in the Federal Regulations, Title 21, Parts 800 to 898. In addiser.
Please be advised that FDA is issuance of a substantial equivalence detemination does not mean
that FDA has made a determination that your device complies with over con that Free and FDA s issuance of a substanial equivalence detemination dos not mean
or arrean statues and regulation that your device complies with other requirements of the A on any Federal station that your device completed with are regulters and regulters not one notener more nonement met nom more nomen more monement met monther more . You must on the real outlines and reguliements, including the Federal agences. You more
Comply with all the Act's requirements, including the more of the Actives (2) CFR Print and the Act Srequirements, including, but not be ited to: registration registration must
device-related adverse events, including, but not be registration of medica device of overse events) (21 CFR 801); medical device reportion and listing (2
device related adverse events) (21 CFR 803); good not meditor regulentes as st സംഭവങ്ങൾ പ്രസ്സ് (21 CFR 803); good manufactupsnag ractive of medical
forth in the quality systems (21 CFR 803); and franceller emergenere electronic the was quality systems (QS) regulation (21 CFR Part 820); and if applicable, the product radiation
product radiation control provisions (Sections 531-542 of the Act); 21 CFR
{3}------------------------------------------------
Page 2 - Mr. Wallace Fischer
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
[signature]
EDK Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
K090532 510(k) Number (if known):
Device Name: T.E.A.R. Tech3
Indications For Use:
The T.E.A.R. Tech3 IF/Muscle Stimulator is intended for the following applications:
The High Volt Pulsed Current Stimulation and Neuromuscular Electrical Stimulation can be used in the following applications:
- Relaxation of muscle spasms
- Prevention or retardation of disuse atrophy .
- Increases local blood circulation .
- Muscle re-education •
- Maintaining or increasing range of motion
- Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
Interferential Stimulation can be used in the following applications:
- Symptomatic relief of post traumatic acute pain
- . Symptomatic relief of chronic pain intractable pain
- . Relaxation of muscle spasms
- Maintaining or increasing range of motion
- Increases local blood circulation
Pulsed Direct Current stimulation can be used in the following applications:
- Reduction of edema (under negative electrode)
- Relaxation of muscle spasm
- . Increasing local blood circulation
- Retardation or prevention of disuse atrophy
- Muscle re-education
- Maintaining or increasing of range of motion
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (C
FOR M.NELSON
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
Page 1 of 1
510(k) Number
Number K090532
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.