SPIN-LOC SUTURE ANCHOR SYSTEM FG-0302-01,FG-0305-01
K090530 · Ross Creek Medical, Inc. · MBI · May 20, 2009 · Orthopedic
Device Facts
Record ID
K090530
Device Name
SPIN-LOC SUTURE ANCHOR SYSTEM FG-0302-01,FG-0305-01
Applicant
Ross Creek Medical, Inc.
Product Code
MBI · Orthopedic
Decision Date
May 20, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Spin-Loc Suture Anchor System is intended to be used for soft tissue fixation to bone in the shoulder, foot and ankle, knee, and elbow. The anchor is intended for use in the following procedures: Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift/Capsulolabral Reconstruction. Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Illiotibial Band Tenodesis. Foot and Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair. Elbow: Biceps Tendon Reattachment. The Ross Creek Spin-Loc Suture Anchor System is intended for single-use only.
Device Story
The Spin-Loc Suture Anchor System is a sterile, single-use orthopedic fixation device consisting of a PEEK polymer and stainless steel or cobalt-chromium alloy anchor loaded on an insertion handle. Used by surgeons in clinical settings (OR) to facilitate soft tissue-to-bone fixation. The device allows surgeons to adjust suture tension on the tissue after the anchor is inserted into the bone. It functions as a mechanical fastener to secure soft tissue during various reconstructive procedures. The device is provided sterile and is intended for single-use only.
Clinical Evidence
Bench testing only. In vitro studies evaluated anchor pull-out strength in accordance with the FDA Guidance Document for Testing Bone Anchor Devices (April 20, 1996). Results demonstrated pull-out strength comparable to predicate devices with identical failure modes.
Technological Characteristics
Materials: PEEK polymer, stainless steel, or cobalt-chromium alloy. Form factor: Suture anchor loaded on an insertion handle. Sterilization: Ethylene oxide (SAL 10^-6). Residuals tested per ISO 10993-7.
Indications for Use
Indicated for patients requiring soft tissue-to-bone fixation in the shoulder, foot, ankle, knee, or elbow during orthopedic repair procedures. Intended for single-use only.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the words "ross CREEK medical" in a bold, sans-serif font. The words "ross CREEK" are stacked on top of the word "medical". The word "medical" is italicized and slightly smaller than the words above it. The text is black and the background is white.
#### 510(k) for the Spin-Loc Suture Anchor System K090530 April 16, 2009
# 510(k) Summary (per 21CFR807.92)
| General Company Information | | | |
|-------------------------------------------------------------------------------------------------|--------------------------------------------------------|--|---------------|
| Name: | Ross Creek Medical, Inc. | | |
| Contact: | Michael Kolber | | |
| | Vice President, Regulatory Affairs | | |
| Address: | 16417 Peacock Lane | | |
| | Los Gatos, CA 95032 | | |
| Telephone: | 408-505-6626 | | |
| Fax: | 408-358-2381 | | |
| Date Prepared: | April 16, 2009 | | |
| General Device Information | | | |
| Product Name: | Spin-Loc Suture Anchor System | | |
| Common Name: | Suture Anchor | | |
| Classification: | Screw, fastener, fixation, non-degradable, soft tissue | | |
| | 21 CFR 888.3040 | | |
| Device Class: | Product Code: MBI | | |
| Product Code: | Class: II | | |
| Predicate Devices | | | |
| Manufacturer | Device Name | | 510(k) Number |
| Arthrex | PushLock | | K063479 |
| Mitek | Versalok | | K063478 |
| Teleflex Medical | Force Fiber | | K063778 |
| Description | | | |
| The Spin-Loc Suture Anchor System consists of a PEEK polymer and stainless steel or cobalt | | | |
| chromium alloy suture anchor loaded on an insertion handle. The Spin-Loc anchor is | | | |
| designed to facilitate fixation of soft tissue to bone. The design allows the surgeon to adjust | | | |
| the tension on the tissue after insertion of the anchor into bone. The Spin-Loc Suture Anchor | | | |
| System will be validated to a SAL of 10-6 using ethylene oxide. The ethylene oxide residuals | | | |
| will be tested according to ISO10993-7. | | | |
| Intended Use (Indications) | | | |
| The Spin-Loc Suture Anchor System is intended to be used for soft tissue fixation to bone in | | | |
| the shoulder, foot and ankle, knee, and elbow. The anchor is intended for use in the following | | | |
| procedures: | | | |
| Shoulder | Knee | | |
| ▪ Rotator Cuff Repair | ▪ Medial Collateral Ligament Repair | | |
- Bankart Repair 미
- SLAP Lesion Repair
- Biceps Tenodesis
- 제 Acromio-Clavicular Separation Repair
- 요 Deltoid Repair
- Capsular Shift/Capsulolabral Reconstruction. I
- M Lateral Collateral Ligament Repair
- 트 Patellar Tendon Repair
- . Posterior Oblique Ligament Repair
- . Illiotibial Band Tenodesis
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the words "ross creek" in large, bold, black letters on the top line. Below that, the word "medical" is written in a smaller, italicized font, also in black. The words appear to be a logo or heading for a medical practice or facility.
510(k) for the Spin-Loc Suture Anchor System K090530 April 16, 2009
- Foot and Ankle Elbow · Lateral Stabilization · Biceps Tendon Reattachment Medial Stabilization Achilles Tendon Repair · Hallux Valgus Reconstruction Midfoot Reconstruction . . Metatarsal Ligament Repair Substantial Equivalence
This submission supports the position that the Ross Creek Spin-Loc Suture Anchor System is substantially equivalent to a number of previously cleared devices, including the Arthrex PushLock [510(k) Number K063479], the Mitek Versalok [(510(k) K063478) and the Teleflex Medical Force Fiber [K063778].
The 510(k) notice contains summaries of in vitro studies that were conducted to evaluate the anchor pull-out strength as specified in the FDA Guidance Document for Testing Bone Anchor Devices (dated April 20, 1996). The data presented demonstrate that the anchor pullout for the Ross Creek Spin-Loc Suture Anchor System compares favorably with the predicate devices. The failure modes observed for the Spin-Loc Suture Anchor were the same as those of the predicate devices.
The single patient use components of the bone anchor system are provided sterile.
#### Conclusions
Ross Creek Medical. Inc. believes that the information provided demonstrates that the proposed device is substantially equivalent to the predicate devices and do not raise any new issues of safety or efficacy. Based on the indications for use, technological characteristics. and comparison to predicated devices, the Spin-Loc Suture Anchor System has been shown to be substantially equivalent to predicate devices as described under the Federal Food, Drug and Cosmetic Act.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle with its wings spread, with three horizontal lines representing the feathers.
MAY 20 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ross Creek Medical Inc. c/o Mr. Michael Kolber 14734 LA Rinconada Drive Los Gatos, California 95032
Re: K090530
Trade/Device Name: Spin-Loc Suture Anchor System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Code: MBI Dated: April 16 2009 Received: April 21, 2009
Dear Mr. Kolber:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{3}------------------------------------------------
### Page 2 – Mr. Michael Kolber
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
for Pete
Des O.f
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the words "ross creek" in large, bold, black letters on the top line. Below that, the word "medical" is written in a smaller, italicized font, also in black. The text appears to have a slightly distressed or textured effect, giving it a somewhat vintage or rugged look.
510(k) for the Spin-Loc Suture Anchor System K090530 April 16, 2009
# Indications for Use .
510(k) Number (if known): K090530
Device Name: Spin-Loc Suture Anchor System
Indications for Use: The Ross Creek Spin-Loc Suture Anchor System is intended to be used for soft tissue fixation to bone in the shoulder, foot and ankle, knee, and elbow. The anchor is intended for use in the following procedures:
#### Shoulder
- Rotator Cuff Repair .
- Bankart Repair .
- I SLAP Lesion Repair
- . Biceps Tenodesis
- Acromio-Clavicular Separation Repair I
- Deltoid Repair 제
- Capsular Shift/Capsulolabral Reconstruction ■
#### Foot and Ankle
- " Lateral Stabilization
- . Medial Stabilization
- · Achilles Tendon Repair
- · Hallux Valgus Reconstruction
- " Midfoot Reconstruction
- · Metatarsal Ligament Repair
#### Knee
- Medial Collateral Ligament Repair .
- · Lateral Collateral Ligament Repair
- · Patellar Tendon Repair
- . Posterior Oblique Ligament Repair
- Illiotibial Band Tenodesis
#### Elbow
- Biceps Tendon Reattachment li
The Ross Creek Spin-Loc Suture Anchor System is intended for single-use only.
Prescription Use *X*
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Anita Le PHS
(Division Sign-off)
loss of Suries Device of CDRH, Office of Device Evaluation (ODE) and Restorative Devices
510(k) Number K090530
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.